The words you are searching are inside this book. To get more targeted content, please make full-text search by clicking here.
Discover the best professional documents and content resources in AnyFlip Document Base.
Search
Published by prasanth181296, 2018-09-25 07:11:56

rosemount seal

rosemount seal

March 2017 Rosemount DP Level

Europe Special Conditions for Safe Use (X):

E8 ATEX Flameproof and Dust 1. The apparatus is not capable of withstanding the 500 V
Certificate: KEMA00ATEX2013X; Baseefa11ATEX0275X insulation test required by clause 6.3.12 of
Standards: EN60079-0:2012, +A11:2013, EN60079-11:2012. This must be taken into account when
installing the apparatus.
EN60079-1:2014,EN60079-26:2015,
2. The enclosure may be made of aluminum alloy and given a
EN60079-31:2009 protective polyurethane paint finish; however care should
be taken to protect it from impact or abrasion if located in
Markings: II 1/2 G, Ex db IIC T6...T4 Ga/Gb T6(–60 °C ≤ Zone 0.
Ta ≤ +70 °C), T4/T5(–60 °C ≤ Ta ≤ +80 °C);
II 1 D Ex ta IIIC T95 °C T500 105 °C Da 3. Some variants of the equipment have reduced markings
(–20 °C ≤ Ta ≤ +85 °C) on the nameplate. Refer to the Certificate for full
equipment marking.
Table 58. Process Temperature
IA ATEX FISCO
Temperature Class Process temperature Certificate: BAS97ATEX1089X
Standards: EN60079-0:2012, EN60079-11:2009
T6 –60 °C to +70 °C Markings: II 1 G Ex ia IIC Ga T4(–60 °C ≤ Ta ≤ +60 °C)
T5 –60 °C to +80 °C
T4 –60 °C to +120 °C

Special Conditions for Safe Use (X): Table 60. Input Parameters

1. This device contains a thin wall diaphragm. Installation, Voltage Ui FISCO
maintenance and use shall take into account the Current Ii 17.5 V
environmental conditions to which the diaphragm will be Power Pi 380 mA
subjected. The manufacturer’s instructions for installation Capacitance Ci 5.32 W
and maintenance shall be followed in detail to assure Inductance Li < 5 nF
safety during its expected lifetime. < 10 μH

2. Flameproof joints are not intended for repair

3. Non-standard paint options may cause risk from Special Conditions for Safe Use (X):
electrostatic discharge. Avoid installations that could
cause electrostatic build-up on painted surfaces, and only 1. The apparatus is not capable of withstanding the 500 V
clean the painted surfaces with a damp cloth. If paint is insulation test required by EN60079-11. This must be
ordered through a special option code, contact the taken into account when installing the apparatus.
manufacturer for more information.
2. The enclosure may be made of aluminum alloy and given a
4. Some variants of the equipment have reduced markings protective polyurethane paint finish; however care should
on the nameplate. Refer to the Certificate for full be taken to protect it from impact or abrasion if located in
equipment marking. Zone 0.

I1 ATEX Intrinsic Safety and Dust N1 ATEX Type n and Dust
Certificate: BAS00ATEX3105X; Baseefa11ATEX0275X
Certificate: BAS97ATEX1089X; Baseefa11ATEX0275X Standards: EN60079-0:2012, EN60079-15:2010,
EN60079-31:2009
Standards: EN60079-0:2012, EN60079-11:2012, Markings: II 3 G Ex nA IIC T5 Gc (–40 °C ≤ Ta ≤ +70 °C);
II 1 D Ex ta IIIC T95 °C T500105 °C Da
EN60079-31:2009 (–20 °C ≤ Ta ≤ +85 °C)

Markings: HART: II 1 G Ex ia IIC T5/T4 Ga Special Conditions for Safe Use (X):

T5(–60 °C ≤ Ta ≤ +40 °C), 1. This apparatus is not capable of withstanding the 500V
T4(–60 °C ≤ Ta ≤ +70 °C) insulation test that is required by clause 6.8.1 of
Fieldbus/PROFIBUS: II 1 G Ex ia Ga IIC EN60079-15. This must be taken into account when
installing the apparatus.
T4(–60°C ≤ Ta ≤ +60 °C)
DUST: II 1 D Ex ta IIIC T95 °C T500105 °C Da 2. Some variants of the equipment have reduced markings
(–20 °C ≤ Ta ≤ +85 °C) on the nameplate. Refer to the Certificate for full
equipment marking.
Table 59. Input Parameters

HART Fieldbus/PROFIBUS

Voltage Ui 30 V 30 V
Current Ii 200 mA 300 mA
Power Pi 1.3 W
Capacitance Ci 0.9 W
Inductance Li 0.012 μF 0 μF
0 mH
0 mH

Emerson.com/Rosemount 151

Rosemount DP Level March 2017

International Special Conditions for Safe Use (X):

E7 IECEx Flameproof and Dust 1. If the apparatus is fitted with an optional 90 V transient
Certificate: IECEx KEM 09.0034X; IECEx BAS 10.0034X suppressor, it is not capable of withstanding the 500 V
Standards: IEC60079-0:2011, IEC60079-1:2014-06, insulation test required by clause 6.3.12 of IEC60079-11.
This must be taken into account when installing the
IEC60079-26:2014-10, IEC60079-31:2008 apparatus.
Markings: Ex db IIC T6…T4 Ga/Gb T6(–60 °C ≤ Ta ≤ +70 °C),
2. The enclosure may be made of aluminum alloy and given a
T4/T5(–60 °C ≤ Ta ≤ +80 °C); protective polyurethane paint finish; however, care should
Ex ta IIIC T95 °C T500105 °C Da (–20 °C ≤ Ta ≤ be taken to protect it from impact or abrasion if located in
+85 °C) Zone 0.

Table 61. Process Temperature IECEx Mining (Special A0259)
Certificate: IECEx TSA 14.0001X
Temperature Class Process temperature Standards: IEC60079-0:2011, IEC60079-11:2011
Markings: Ex ia I Ma (–60 °C ≤ Ta ≤ +70 °C)
T6 –60 °C to +70 °C
T5 –60 °C to +80 °C Table 63. Input Parameters
T4 –60 °C to +80 °C HART Fieldbus/PROFIBUS FISCO

Special Conditions for Safe Use (X):

1. This device contains a thin wall diaphragm. Installation, Voltage Ui 30 V 30 V 17.5 V
maintenance and use shall take into account the Current Ii 200 mA 300 mA 380 mA
environmental conditions to which the diaphragm will be Power Pi 1.3 W 5.32 W
subjected. The manufacturer’s instructions for installation Capacitance Ci 0.9 W < 5 nF
and maintenance shall be followed in detail to assure Inductance Li 0.012 μF 0 μF < 10 μH
safety during its expected lifetime. 0 mH
0 mH
2. Flameproof joints are not intended for repair.
Special Conditions for Safe Use (X):
3. Non-standard paint options may cause risk from
electrostatic discharge. Avoid installations that could 1. If the apparatus is fitted with optional 90 V transient
cause electrostatic build-up on painted surfaces, and only suppressor, it is not capable of withstanding the 500 V
clean the painted surfaces with a damp cloth. If paint is insulation test required by IEC60079-11. This must be
ordered through a special option code, contact the taken into account when installing the apparatus.
manufacturer for more information.
2. It is a condition of safe use that the above input parameters
4. Some variants of the equipment have reduced markings shall be taken into account during installation.
on the nameplate. Refer to the Certificate for full
equipment marking. 3. It is a condition of manufacture that only the apparatus
fitted with housing, covers and sensor module housing
I7 IECEx Intrinsic Safety made out of stainless steel are used in Group I applications.
Certificate: IECEx BAS 09.0076X
Standards: IEC60079-0:2011, IEC60079-11:2011 N7 IECEx Type n
Markings: HART: Ex ia IIC T5/T4 Ga, Certificate: IECEx BAS 09.0077X
T5(–60 °C ≤ Ta ≤ +40 °C), Standards: IEC60079-0:2011, IEC60079-15:2010
T4(–60 °C ≤ Ta ≤ +70 °C) Markings: Ex nA IIC T5 Gc (–40 °C ≤ Ta ≤ +70 °C)
Fieldbus/PROFIBUS: Ex ia IIC Ga
T4(–60 °C ≤ Ta ≤ +60 °C) Special Condition for Safe Use (X):

Table 62. Input Parameters Fieldbus/PROFIBUS 1. The apparatus is not capable of withstanding the 500 V
HART insulation test required by IEC60079-15. This must be
taken into account when installing the apparatus.
Voltage Ui 30 V 30 V
Current Ii 200 mA 300 mA Brazil
Power Pi 1.3 W
Capacitance Ci 0.9 W E2 INMETRO Flameproof
Inductance Li 0.012 μF 0 μF Certificate: UL-BR 13.0643X
0 mH Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011,
0 mH ABNT NBR IEC60079-1:2009 + Errata 1:2011,
ABNT NBR IEC60079-26:2008 + Errata 1:2008
Markings: Ex db IIC T6…T4 Ga/Gb, T6(-60°C ≤ Ta ≤ +70°C),
T4/T5(-60°C ≤ Ta ≤ +80°C)

152 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Special Conditions for Safe Use (X): Table 65. Input Parameters

1. This device contains a thin wall diaphragm less than 1 mm Voltage Ui FISCO
thickness that forms a boundary between zone 0 (process Current Ii 17.5 V
connection) and zone 1 (all other parts of the equipment). Power Pi 380 mA
The model code and datasheet are to be consulted for Capacitance Ci 5.32 W
details of the diaphragm material. Installation, Inductance Li < 5 nF
maintenance and use shall take into account the < 10 μH
environmental conditions to which the diaphragm will be
subjected. The manufacturer's instructions for installation Special Conditions for Safe Use (X):
and maintenance shall be followed in detail to assure
safety during its expected lifetime. 1. If the equipment is fitted with an optional 90 V transient
suppressor, it is not capable of withstanding the 500 V
2. Flameproof joints are not intended for repair. insulation test required by ABNT NBR IEC 60079-11. This
must be taken into account when installing the
3. Non-standard paint options may cause risk from equipment.
electrostatic discharge. Avoid installations that could
cause electrostatic build-up on painted surfaces, and only 2. The enclosure may be made of aluminum alloy and given a
clean the painted surfaces with a damp cloth. If paint is protective polyurethane paint finish; however, care should
ordered through a special option code, contact the be taken to protect it from impact or abrasion if located in
manufacturer for more information. Zone 0.

I2 INMETRO Intrinsic Safety China
Certificate: UL-BR 13.0584X
Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011, E3 China Flameproof
ABNT NBR IEC60079-11:2009
Markings: HART: Ex ia IIC T5/T4 Ga, Certificate: GYJ14.1041X
T5(–60 °C ≤ Ta ≤ +40 °C),
T4(–60 °C ≤ Ta ≤ +70 °C) Standards: GB3836.1-2000, GB3836.2-2010,
Fieldbus/PROFIBUS: Ex ia IIC T4
Ga (–60 °C ≤ Ta ≤ +60 °C) GB12476-2000, GB3836.20-2010,

Table 64. Input Parameters Markings: Ex d IIC T6/T5 Ga/Gb, T6(–50 °C ≤ Ta ≤ +65 °C),
T5(–50 °C ≤ Ta ≤ +80 °C)
HART Fieldbus/PROFIBUS

Voltage Ui 30 V 30 V Special Conditions for Safe Use (X):
Current Ii 200 mA 300 mA
Power Pi 1.3 W 1. The relation between ambient temperature arrange and
Capacitance Ci 0.9 W temperature Class is as follows:
Inductance Li 0.012 μF 0 μF
0 mH Ta Temperature Class
0 mH
–50 °C~+80 °C T5
Special Conditions for Safe Use (X): –50 °C~+65 °C T6

1. If the equipment is fitted with an optional 90 V transient When used in a combustible dust environment, the
suppressor, it is not capable of withstanding the 500 V maximum ambient temperature is 80 °C.
insulation test required by ABNT NBR IRC 60079-11. This
must be taken into account when installing the 2. The earth connection facility in the enclosure should be
equipment. connected reliably.

2. The enclosure may be made of aluminum alloy and given a 3. Cable entry certified by notified body with type of
protective polyurethane paint finish; however, care should protection Ex d IIC in accordance with GB3836.1-2000 and
be taken to protect it from impact or abrasion if located in GB3836.2-2000, should be applied when installed in a
Zone 0. hazardous location. When used in combustible dust
environment, cable entry in accordance with IP66 or
IB INMETRO FISCO higher level should be applied.
Certificate: UL-BR 13.0584X
Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011, 4. Obey the warning “Keep tight when the circuit is alive.”
ABNT NBR IEC60079-11:2009
Markings: Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +60 °C) 5. End users are not permitted to change any internal
components.

6. During installation, use and maintenance of this product,
observe the following standards: GB3836.13-1997,
GB3836.15-2000, GB3836.16-2006, GB50257-1996,
GB12476.2-2006, GB15577-2007

Emerson.com/Rosemount 153

Rosemount DP Level March 2017

I3 China Intrinsic Safety 5. The product should be used with Ex-certified associated
Certificate: GYJ13.1362X apparatus to establish explosion protection system that
Standards: GB3836.1-2010, GB3836.4-2010, can be used in explosive gas atmospheres. Wiring and
GB3836.20-2010, GB12476.1-2000 terminals should comply with the instruction manual of
Markings: Ex ia IIC Ga T4/T5 the product and associated apparatus.

Special Conditions for Safe Use (X): 6. The cables between this product and associated apparatus
should be shielded cables (the cables must have insulated
1. Symbol “X” is used to denote specific conditions of use: shield). The shielded cable has to be grounded reliably in a
non-hazardous area.
a.If the apparatus is fitted with an optional 90 V
transient suppressor, it is not capable of 7. End users are not permitted to change any intern
withstanding the 500 V insulation test for 1 minute. components but to settle the problem in conjunction with
This must be taken into account when installing the the manufacturer to avoid damage to the product.
apparatus.
8. During installation, use and maintenance of this product,
b.The enclosure may be made of aluminum alloy and observe the following standards:
given a protective polyurethane paint finish; GB3836.13-1997, GB3836.15-2000, GB3836.16-2006,
however, care should be taken to protect it from GB50257-1996, GB12476.2-2006, GB15577-2007
impact or abrasion if located in Zone 0.
N3 China Type n
2. The relation between T code and ambient temperature Certificate: GYJ15.1105X
range is: Standards: GB3836.1-2010, GB3836.8-2003
Markings: Ex nA nL IIC T5 Gc (–40 °C ≤ Ta ≤ +70 °C)
Model T Temperature
Special Condition for Safe Use (X):
code range
1. Symbol “X” is used to denote specific conditions of use:
HART T5 –60 °C ≤ Ta ≤ +40 °C The apparatus is not capable of withstanding the 500V test
to earth for one minute. The must be taken into
HART T4 –60 °C ≤ Ta ≤ +70 °C consideration during installation.

Fieldbus/PROFIBUS/FISCO T4 –60 °C ≤ Ta ≤ +60 °C Japan

3. Intrinsically Safe parameters FISCO E4 Japan Flameproof
Table 66. Input Parameters Certificate: TC20577, TC20578, TC20583, TC20584
[HART]; TC20579, TC20580, TC20581,
HART Fieldbus/PROFIBUS TC20582 [Fieldbus]
Markings: Ex d IIC T5
Voltage Ui 30 V 30 V 17.5 V
Current Ii 200 mA 300 mA 380 mA Technical Regulations Customs Union (EAC)
Power Pi 1.3 W 5.32 W
Capacitance Ci 0.9 W < 5 nF EM EAC Flameproof
Inductance Li 0.012 μF 0 μF < 10 μH Certificate: RU C-US.GB05.B.01197
0 mH Markings: Ga/Gb Ex d IIC T5/T6 X, T5(–60 °C ≤ Ta ≤ +80 °C),
0 mH T6(–60 °C ≤ Ta ≤ +65 °C)

Note Special Condition for Safe Use (X):
FISCO parameters apply to both Group IIC and IIB.
[For Flowmeters] When 644 Temperature Transmitter is used, it 1. See certificate for special conditions.
should be used with Ex-certified associated apparatus to
establish explosion protection system that can be used in IM EAC Intrinsically Safe
explosive gas atmospheres. Wiring and terminals should comply Certificate: RU C-US.GB05.B.01197
with the instruction manual of both 644 Temperature Markings: HART: 0Ex ia IIC T4/T5 Ga X,
Transmitter and associated apparatus. The cables between 644 T4(–60 °C ≤ Ta ≤ +70 °C),
Temperatures Transmitter and associated apparatus should be T5(–60 °C ≤ Ta ≤ +40 °C)
shielded cables (the cables must have insulated shield). The Fieldbus/PROFIBUS: 0Ex ia IIC T4 Ga X
shielded cable has to be grounded reliably in a non-hazardous (–60 °C ≤ Ta ≤ +60 °C)
area.
Special Condition for Safe Use (X):
4. Transmitters comply with the requirements for FISCO field
devices specified in IEC60079-27:2008. For the connection 1. See certificate for special conditions.
of an intrinsically safe circuit in accordance with FISCO
Model, FISCO parameters are listed in the table above.

154 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Combinations Additional Certifications

K2 Combination of E2 and I2 SBS American Bureau of Shipping (ABS) Type Approval
K5 Combination of E5 and I5 Certificate: 09-HS446883A-5-PDA
K6 Combination of C6, E8, and I1 Intended: Marine and Offshore Applications -
K7 Combination of E7, I7, and N7 Measurement of either gauge or absolute
K8 Combination of E8, I1, and N1 pressure for liquid, gas and vapor.
KB Combination of E5, I5, and C6
KD Combination of E8, I1, E5, I5, and C6 SBV Bureau Veritas (BV) Type Approval
KM Combination of EM and IM Certificate: 23155
Requirements: Bureau Veritas Rules for the Classification of
Conduit plugs and adapters Steel Ships
Application: Class notations: AUT-UMS, AUT-CCS,
IECEx Flameproof and Increased Safety AUT-PORT and AUT-IMS; Pressure transmitter
Certificate: IECEx FMG 13.0032X type 3051 cannot be installed on diesel
Standards: IEC60079-0:2011, IEC60079-1:2007, engines

IEC60079-7:2006-2007 SDN Det Norske Veritas (DNV) Type Approval
Markings: Ex de IIC Gb Certificate: TAA000004F
Intended: Det Norske Veritas’ Rules for Classification of
ATEX Flameproof and Increased Safety Ships, High Speed and Light Craft and Det
Certificate: FM13ATEX0076X Norske Veritas’ Offshore Standards
Standards: EN60079-0:2012, EN60079-1:2007, Application:

IEC60079-7:2007 Location Classes
Markings: II 2 G Ex de IIC Gb

Table 67. Conduit Plug Thread Sizes Temperature D
Humidity B
Thread Identification mark Vibration A
EMC B
M20 x 1.5 M20 Enclosure D
1/2–14 NPT 1/2 NPT

Table 68. Thread Adapter Thread Sizes

Male thread Identification mark SLL Lloyds Register (LR) Type Approval
Certificate: 11/60002
M20 x1.5–6H M20 Application: Environmental categories ENV1, ENV2, ENV3
1/2 –14 NPT 1/2–14 NPT and ENV5
3/4 –14 NPT 3/4–14 NPT
Female thread Identification mark C5 Custody Transfer - Measurement Canada Accuracy
Approval
M20 x 1.5 –6H M20
1/2 –14 NPT 1/2–14 NPT Certificate: AG-0226; AG-0454; AG-0477
G 1/2
G 1/2

Special Conditions for Safe Use (X):

1. When the thread adapter or blanking plug is used with an
enclosure in type of protection increased safety “e” the
entry thread shall be suitably sealed in order to maintain
the ingress protection rating (IP) of the enclosure.

2. The blanking plug shall not be used with an adapter.

3. Blanking Plug and Threaded Adapter shall be either NPT or
Metric thread forms. G1/2 thread forms are only acceptable
for existing (legacy) equipment installations.

Emerson.com/Rosemount 155

Rosemount DP Level March 2017

Rosemount 2051

Rev 1.5 E6 Canada Explosion-Proof, Dust Ignition Proof
Certificate: 2041384
European Directive Information Standards: CAN/CSA C22.2 No. 0-10, CSA Std C22.2 No.
25-1966, CSA Std C22.2 No. 30-M1986,
A copy of the EU Declaration of Conformity can be found at the end CAN/CSA-C22.2 No. 94-M91, CSA Std C22.2
of the Quick Start Guide. The most recent revision of the EU No.142-M1987, CAN/CSA-C22.2 No.157-92,
Declaration of Conformity can be found at CSA Std C22.2 No. 213-M1987,
Emerson.com/Rosemount. CAN/CSA-E60079-0:07, CAN/CSA-E60079-1:07,
CAN/CSA-E60079-11-02, CAN/CSA-C22.2 No.
Ordinary Location Certification 60529:05, ANSI/ISA-12.27.01–2003
Markings: Explosion-Proof for Class I, Divisions 1, Groups
As standard, the transmitter has been examined and tested to B, C, and D. Dust-Ignition Proof for Class II and
determine that the design meets the basic electrical, mechanical, Class III, Division 1, Groups E, F, and G. Suitable
and fire protection requirements by a nationally recognized test for Class I, Division 2; Groups A, B, C, and D for
laboratory (NRTL) as accredited by the Federal Occupational Safety indoor and outdoor hazardous locations.
and Health Administration (OSHA). Class I Zone 1 Ex d IIC T5. Enclosure type 4X,
factory sealed. Single Seal
North America
I6 Canada Intrinsic Safety
E5 USA Explosionproof (XP) and Dust-Ignitionproof (DIP) Certificate: 2041384
Certificate: FM16US0232 Standards: CSA Std. C22.2 No. 142 - M1987, CSA Std.C22.2
Standards: FM Class 3600 – 2011, FM Class 3615 – 2006, No. 213 - M1987, CSA Std. C22.2 No.157 - 92,
FM Class 3616 – 2011, FM Class 3810 – 2005, CSA Std. C22.2 No. 213 - M1987, ANSI/ISA
ANSI/NEMA 250 –2008. 12.27.01 – 2003, CAN/CSA-E60079-0:07,
ANSI/IEC 60529 2004 CAN/CSA-E60079-11:02
Markings: XP CL I, DIV 1, GP B, C, D; DIP CL II, DIV 1, GP Markings: Intrinsically safe for Class I, Division 1, Groups
E, F, G; CL III; T5(–50 °C ≤ Ta ≤ +85 °C); A,B, C, and D when connected in accordance
Factory Sealed; Type 4X with Rosemount drawings 02051-1008.
Temperature code T3C. Ex ia IIC
I5 USA Intrinsic Safety (IS) and Nonincendive (NI) T3C. Single Seal. Enclosure Type 4X
Certificate: FM16US0231X
Standards: FM Class 3600 – 2011, FM Class 3610 – 2010, Europe
FM Class 3611 – 2004, FM Class 3810 – 2005,
ANSI/NEMA 250 –2008 E1 ATEX Flameproof
Markings: IS CL I, DIV 1, GP A, B, C, D; CL II, DIV 1, GP E,
F, G; Class III; DIV 1 when connected per Certificate: KEMA 08ATEX0090X
Rosemount drawing 02051-1009; Class I,
Zone 0; AEx ia IIC T4; NI CL 1, DIV 2, GP A, B, Standards: EN 60079-0:2012 + A11:2013,
C, D; T4(–50 °C ≤ Ta ≤+70 °C); Type 4x
EN 60079-1:2014, EN 60079-26:2015
Specific Condition of Use:
Markings: II 1/2 G Ex db IIC Ga/Gb T6(–60 °C ≤ Ta ≤ +70 °C),
1. The Model 2051 transmitter housing contains aluminum and T4/T5 (-60°C ≤ Ta ≤ +80°C)
is considered a potential risk of ignition by impact or friction.
Care must be taken into account during installation and use Temperature Process Ambient
to prevent impact and friction. Temperature
Class Temperature
IE USA FISCO –60°C to +70°C
Certificate: FM16US0231X T6 –60°C to +70°C –60°C to +80°C
Standards: FM Class 3600 – 2011, FM Class 3610 – 2010, T5 –60°C to +80°C –60°C to +80°C
FM Class 3611 – 2004, FM Class 3810 – 2005 T4 –60°C to +120°C
Markings: IS CL I, DIV 1, GP A, B, C, D when connected per
Rosemount drawing 02051-1009
(–50 °C ≤ Ta ≤ +60 °C); Type 4x

Specific Condition of Use:

1. The Model 2051 transmitter housing contains aluminum and
is considered a potential risk of ignition by impact or friction.
Care must be taken into account during installation and use
to prevent impact and friction.

156 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Special Conditions for Safe Use (X): Table 70. Input Parameters

1. Appropriate cable, glands and plugs need to be suitable for Parameters FISCO
a temperature of 5°C greater than maximum specified
temperature for location where installed. Voltage Ui 17.5 V
Current li 380 mA
2. Non- standard paint options may cause risk from Power Pi 5.32 W
electrostatic discharge. Avoid installations that could Capacitance Ci
cause electrostatic build-up on painted surfaces, and only Inductance Li 0 μF
clean the painted surfaces with a damp cloth. If paint is 0 mH
ordered through a special option code, contact the
manufacturer for more information. Special Conditions for Safe Use (X):

3. The device contains a thin wall diaphragm less than 1 mm 1. If the equipment is fitted with an optional 90V transient
thickness that forms a boundary between zone 0 (process suppressor, it is incapable of withstanding the 500V
connection) and zone 1 (all other parts of the equipment). isolation from earth test and this must be taken into
The model code and datasheet are to be consulted for account during installation.
details of the diaphragm material. Installation,
maintenance and use shall take into account the 2. The enclosure may be made of aluminum alloy and given a
environmental conditions to which the diaphragm shall be protective polyurethane paint finish; however care should
subjected. The manufacturer's instructions for installation be taken to protect it from impact and abrasion when
and maintenance shall be followed in detail to assure located in Zone 0.
safety during its expected lifetime.
N1 ATEX Type n
4. Flameproof joints are not intended for repair.
Certification: Baseefa08ATEX0130X
I1 ATEX Intrinsic Safety
Certificate: Baseefa08ATEX0129X Standards: EN60079-0:2012, EN60079-15:2010
Standards: EN60079-0:2012,
EN60079-11:2012 Markings: II 3G Ex nA IIC T4 Gc (–40 °C ≤ Ta ≤ +70 °C)
Markings: II 1 G Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +70 °C)
Special Condition for Safe Use (X):
Table 69. Input Parameters
1. If the equipment is fitted with an optional 90 V transient
Parameters HART Fieldbus/PROFIBUS suppressor, it is incapable of withstanding the 500 V
electrical strength test as defined in clause 6.5.1 of by EN
Voltage Ui 30 V 30 V 60079-15:2010. This must be taken into account during
Current li 200 mA 300 mA installation.
Power Pi 1.0 W 1.3 W
Capacitance Ci 0.012 μF ND ATEX Dust
Inductance Li 0 μF
0 mH 0 mH Certification: Baseefa08ATEX0182X

Standards: EN60079-0:2012, EN60079-31:2009

Markings: II 1 D Ex ta IIIC T95 °C T500 105 °C

Special Conditions for Safe Use (X): Da (–20 °C ≤ Ta ≤ +85 °C)

1. If the equipment is fitted with an optional 90 V transient Special Conditions for Safe Use (X):
suppressor, it is incapable of withstanding the 500 V
isolation from earth test and this must be taken into 1. If the equipment is fitted with an optional 90 V transient
account during installation. suppressor, it is incapable of withstanding the 500 V
isolation from earth test and this must be taken into
2. The enclosure may be made of aluminum alloy and given a account during installation.
protective polyurethane paint finish; however care should
be taken to protect it from impact and abrasion when International
located in Zone 0.
E7 IECEx Flameproof
IA ATEX FISCO Certificate: IECExKEM08.0024X
Certificate: Baseefa08ATEX0129X Standards: IEC 60079-0:2011, IEC 60079-1:2014-06,
Standards: EN60079-0:2012, EN60079-11:2012
Markings: II 1 G Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +60 °C) IEC 60079-26:2014-10
Markings: Ex d IIC T6/T5 IP66,

T6(–50 °C ≤ Ta ≤ +65 °C),
T5(–50 °C ≤ Ta ≤ +80 °C)

Emerson.com/Rosemount 157

Rosemount DP Level March 2017

Table 71. Process Temperature IG IECEx FISCO
Certificate: IECExBAS08.0045X
Temperature Process Ambient Standards: IEC60079-0:2011, IEC60079-11:2011
Temperature Markings: Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +60 °C)
Class temperature
–60 °C to +70 °C Table 73. Input Parameters
T6 –60 °C to +70 °C –60 °C to +80 °C
–60 °C to +80 °C
T5 –60 °C to +80 °C Parameters FISCO
T4 –60 °C to +120 °C
Voltage Ui 17.5 V
Special Conditions for Safe Use (X): Current li 380 mA
Power Pi 5.32 W
1. The device contains a thin wall diaphragm less than 1 mm Capacitance Ci
thickness that forms a boundary between zone 0 (process Inductance Li 0 μF
connection) and zone 1 (all other parts of the equipment). 0 mH
The model code and datasheet are to be consulted for
details of the diaphragm material. Installation, Special Conditions for Safe Use (X):
maintenance and use shall take into account the
environmental conditions to which the diaphragm shall be 1. If the equipment is fitted with an optional 90 V transient
subjected. The manufacturer's instructions for installation suppressor, it is incapable of withstanding the 500 V
and maintenance shall be followed in detail to assure isolation from earth test and this must be taken into
safety during its expected lifetime. account during installation.

2. Appropriate cable, glands and plugs need to be suitable for 2. The enclosure may be made of aluminum alloy and given a
a temperature of 5°C greater than maximum specified protective polyurethane paint finish; however care should
temperature for location where installed. be taken to protect it from impact and abrasion when
located in Zone 0.
3. Flameproof joints are not intended for repair.
N7 IECEx Type n
4. Non-standard paint options may cause risk from Certificate: IECExBAS08.0046X
electrostatic discharge. Avoid installations that could Standards: IEC60079-0:2011, IEC60079-15:2010
cause electrostatic build-up on painted surfaces, and only Markings: Ex nA IIC T4 Gc (–40 °C ≤ Ta ≤ +70 °C)
clean the painted surfaces with a damp cloth. If paint is
ordered through a special option code, contact the Special Condition for Safe Use (X):
manufacturer for more information.
1. If fitted with a 90 V transient suppressor, the equipment is
I7 IECEx Intrinsic Safety not capable of withstanding the 500 V electrical strength
Certificate: IECExBAS08.0045X test as defined in clause 6.5.1 of IEC60079-15:2010. This
Standards: IEC60079-0:2011, IEC60079-11:2011 must be taken into account during installation.
Markings: Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +70 °C)
Brazil
Table 72. Input Parameters
E2 INMETRO Flameproof
Parameters HART Fieldbus/PROFIBUS Certificate: UL-BR 14.0375X
Standards: ABNT NBR IEC60079-0:2008 + Errata 1:2011,
Voltage Ui 30 V 30 V ABNT NBR IEC 60079-1:2009 + Errata 1:2011,
Current li 200 mA 300 mA ABNT NBR IEC 60079-26:2008 + Errata 1:2009
Power Pi 1.0 W 1.3 W Markings: Ex d IIC T6/T5 Gb IP66,
Capacitance Ci 0.012 μF T6(–50 °C ≤ Ta ≤ +65 °C),
Inductance Li 0 μF T5(–50 °C ≤ Ta ≤ +80 °C)
0 mH 0 mH
Special Conditions for Safe Use (X):
Special Conditions for Safe Use (X):
1. The device contains a thin wall diaphragm. Installation,
1. If the equipment is fitted with an optional 90 V transient maintenance and use shall take into account the
suppressor, it is incapable of withstanding the 500 V environmental conditions to which the diaphragm will be
isolation from earth test and this must be taken into subjected. The manufacturer’s instructions for installation
account during installation. and maintenance shall be followed in detail to assure
safety during its expected lifetime.
2. The enclosure may be made of aluminum alloy and given a
protective polyurethane paint finish; however care should 2. The Ex d blanking elements, cable glands, and wiring shall
be taken to protect it from impact and abrasion when be suitable for a temperature of 90 °C.
located in Zone 0.
3. In case of repair, contact the manufacturer for information
on the dimensions of the flameproof joints.

158 Emerson.com/Rosemount

March 2017 Rosemount DP Level

I2 INMETRO Intrinsic Safety Special Conditions for Safe Use (X):
Certificate: UL-BR 14.0759X
Standards: ABNT NBR IEC60079-0:2008, +Errata 1:2011 1. Symbol “X” is used to denote specific conditions of use:
ABNT NBR IEC60079-11:2009
Markings: Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +70 °C) a. The Ex d blanking elements, cable glands, and wiring
shall be suitable for a temperature of 90°C.
Table 74. Input Parameters
b. This device contains a thin wall diaphragm. Installation,
Parameters HART Fieldbus/PROFIBUS maintenance and use shall take into account the
environment conditions to which the diaphragm will be
Voltage Ui 30 V 30 V subjected.
Current li 200 mA 300 mA
Power Pi 1.3 W 2. The relation between T code and ambient temperature
Capacitance Ci 1W range is:
Inductance Li 0.012 μF 0 μF
0 mH Ta Temperature Class
0 mH
–50 °C ≤ Ta ≤ +80 °C T5
Special Conditions for Safe Use (X): –50 °C ≤ Ta ≤ +65 °C T6

1. If the equipment is fitted with an optional 90 V transient 3. The earth connection facility in the enclosure should be
suppressor, it is not capable of withstanding the 500 V connected reliably.
insulation test required by ABNT NBR IRC 60079-11:2008.
This must be taken into account when installing the 4. During installation, use and maintenance of the product,
equipment. observe the warning “Don’t open the cover when the circuit
is alive.”
2. The enclosure may be made of aluminium alloy and given a
protective polyurethane paint finish; however care should be 5. During installation, there should be no mixture harmful to
taken to protect it from impact and abrasion when located in flameproof housing.
atmospheres that require ELP Ga.
6. Cable entry and conduit, certified by NEPSI with type of
IB INMETRO FISCO protection Ex d IIC and appropriate thread form, should be
Certificate: UL-BR 14.0759X applied when installed in a hazardous location. Blanking
Standards: ABNT NBR IEC 60079-0:2008 + Errata 1:2011; elements should be used on the redundant cable entries.
ABNT NBR IEC 60079-11:2009
Markings: Ex ia IIC T4 Ga (–60 °C ≤ Ta ≤ +60 °C) 7. End users are not permitted to change any internal
Table 75. Input Parameters components, but to settle the problem in conjunction with
the manufacturer to avoid damage to the product.
FISCO
8. Maintenance should be done in a non-hazardous location.
Voltage Ui 17.5 V
Current li 380 mA 9. During installation, use and maintenance of this product,
Power Pi 5.32 W observe the following standards: GB3836.13-2013,
Capacitance Ci 0 μF GB3836.15-2000, GB3836.16-2006, GB50257-2014
Inductance Li 0 mH
I3 China Intrinsic Safety
Certificate: GYJ12.1295X
Standards: GB3836.1-2010, GB3836.4-2010,
GB3836.20-2010
Markings: Ex ia IIC Ga, T6(–60 °C ≤ Ta ≤ +70 °C)

Special Conditions for Safe Use (X): Special Conditions for Safe Use (X):

1. If the equipment is fitted with an optional 90 V transient 1. Symbol “X” is used to denote specific conditions of use:
suppressor, it is not capable of withstanding the 500 V
insulation test required by ABNT NBR IRC 60079-11:2008. a. If the apparatus is fitted with an optional 90 V transient
This must be taken into account when installing the suppressor, it is not capable of withstanding the 500 V
equipment. insulation test for 1 minute. This must be taken into
account when installing the apparatus.
2. The enclosure may be made of aluminium alloy and given a
protective polyurethane paint finish; however care should be b. The enclosure may be made of aluminum alloy and given a
taken to protect it from impact and abrasion when located in protective polyurethane paint finish; however, care
atmospheres that require ELP Ga. should be taken to protect it from impact or abrasion if
located in Zone 0.
China
2. The relation between T code and ambient temperature
E3 China Flameproof range is:
Certificate: GYJ13.1386X
Standards: GB3836.1-2010, GB3836.2-2010 Model T Temperature
Markings: Pressure Transmitter: Ex d IIC Gb,
T6(–50 °C ≤ Ta ≤ +65 °C), Code range
T5(–50 °C ≤ Ta ≤ +80 °C)
HART, Fieldbus, PROFIBUS, T4 –60 °C ≤ Ta ≤ +70 °C
and Low Power

Emerson.com/Rosemount 159

Rosemount DP Level March 2017

3. Intrinsically safe parameters: Fieldbus/ FISCO Special Condition for Safe Use (X):
HART PROFIBUS
17.5 V 1. See certificate for special conditions.
Voltage Ui 30 V 30 V 380 mA
Current li 200 mA 300 mA 5.32 W IM EAC Intrinsically Safe
Power Pi 1W 1.3 W 0 nF Certificate: RU C-US.GB05.B.01199
Capacitance Ci 0.012 μF 0 μF 0 μF Markings: 0Ex ia IIC T4 Ga X (-60 °C ≤ Ta ≤ +70 °C)
Inductance Li 0 mH 0 mH
Special Condition for Safe Use (X):
Note
FISCO parameters comply with the requirements for FISCO field 1. See certificate for special conditions.
devices in GB3836.19-2010.
Combinations
[For Flowmeters] When 644 temperature transmitter is used, the
644 temperature transmitter should be used with Ex-certified K1 Combination of E1, I1, N1, and ND
associated apparatus to establish explosion protection system that K2 Combination of E2 and I2
can be used in explosive gas atmospheres. Wiring and terminals K5 Combination of E5 and I5
should comply with the instruction manual of both 644 K6 Combination of E6 and I6
temperature transmitter and associated apparatus. The cables K7 Combination of E7, I7, N7 and IECEx Dust
between 644 temperatures transmitter and associated apparatus
should be shielded cables (the cables must have insulated shield). IECEx Dust
The shielded cable has to be grounded reliably in a non-hazardous Certificate: IECExBAS08.0058X
area. Standards: IEC60079-0:2011, IEC60079-31:2008
Markings: Ex ta IIIC T95 °C T500105 °C
4. The product should be used with Ex-certified associated
apparatus to establish explosion protection system that can Da (–20 °C ≤ Ta ≤ +85 °C)
be used in explosive gas atmospheres. Wiring and terminals
should comply with the instruction manual of the product Special Condition for Safe Use (X):
and associated apparatus.
1. If the equipment is fitted with an optional 90 V transient
5. The cables between this product and associated apparatus suppressor, it is incapable of withstanding a 500 V isolation
should be shielded cables (the cables must have insulated from earth test and this must be taken into account during
shield). The shielded cable has to be grounded reliably in a installation.
non-hazardous area.
KA Combination of E1, I1, and K6
6. End users are not permitted to change any internal KB Combination of K5 and K6
components, and needs to settle the problem in conjunction KC Combination of E1, I1, and K5
with the manufacturer to avoid damage to the product. KD Combination of K1, K5, and K6
KM Combination of EM and IM
7. During installation, use and maintenance of this product,
observe the following standards: GB3836.13-2013,
GB3836.15-2000, GB3836.16-2006, GB3836.18-2010,
GB50257-2014.

Japan

E4 Japan Flameproof

Certificate: TC20598, TC20599, TC20602, TC20603 [HART];
TC20600, TC20601, TC20604, TC20605
[Fieldbus]

Markings: Ex d IIC T5

Technical Regulations Customs Union (EAC)

EM EAC Flameproof

Certificate: RU C-US.GB05.B.01199

Markings: Ga/Gb Ex d IIC X, T5(–50 °C ≤ Ta ≤ +80 °C),
T6(–50 °C ≤ Ta ≤ +65 °C)

160 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Additional Certifications

SBS American Bureau of Shipping (ABS) Type Approval

Certificate: 09-HS446883B-3-PDA

Intended Use: Marine and Offshore Applications
Measurement of either Gauge or Absolute
Pressure for Liquid, Gas, and Vapor.

ABS Rules: 2013 Steel Vessels Rules 1-1-4/7.7,
1-1-Appendix 3, 4-8-3/1.7, 4-8-3/13.1

SBV Bureau Veritas (BV) Type Approval

Certificate: 23157/B0 BV

BV Rules: Bureau Veritas Rules for the Classification of Steel
Ships

Application: Class notations: AUT-UMS, AUT-CCS,
AUT-PORT and AUT-IMS; Pressure transmitter
type 2051 cannot be installed on diesel
engines.

SDN Det Norske Veritas (DNV) Type Approval

Certificate: TAA000004F

Intended Use: DNV GL Rules for Classification -Ships and
offshore units

Application:
Location Classes

Type 2051
Temperature D
Humidity B
Vibration A
EMC B
Enclosure D

SLL Lloyds Register (LR) Type Approval

Certificate: 11/60002

Application: Environmental categories ENV1, ENV2, ENV3
and ENV5

Emerson.com/Rosemount 161

Rosemount DP Level March 2017

Rosemount 3051 Wireless Markings: IS CL I, DIV 1, GP A, B, C, D T4;
CL 1, Zone 0 AEx ia IIC T4;
Rev 1.3 T4(–40 °C ≤ Ta ≤ +70 °C) when installed per
Rosemount drawing 03031-1062;
European Directive Information Type 4X/IP66/IP68

A copy of the EC Declaration of Conformity can be found at the Special Conditions for Safe Use (X):
end of the Quick Start Guide. The most recent revision of the EC
Declaration of Conformity can be found at 1. The Model 3051 Wireless Pressure Transmitter shall only
Emerson.com/Rosemount. be used with the 701PGNKF Rosemount SmartPower
Battery Pack.
Telecommunication compliance
2. The inline pressure sensor may contain more than 10%
All wireless devices require certification to ensure that they aluminum and is considered a potential risk of ignition by
adhere to regulations regarding the use of the RF spectrum. impact or friction. Care must be taken into account during
Nearly every country requires this type of product certification. installation and used to prevent impact and friction.
Emerson is working with governmental agencies around the
world to supply fully compliant products and remove the risk of 3. The surface resistivity of the transmitter housing is greater
violating country directives or laws governing wireless device than one gigaohm. To avoid electrostatic charge build-up,
usage. it must not be rubbed or cleaned with solvents or a dry
cloth.
FCC and IC
Canada
This device complies with Part 15 of the FCC Rules. Operation is
subject to the following conditions: This device may not cause I6 Canada Intrinsically Safe
harmful interference. This device must accept any interference Certificate: CSA2526009
received, including interference that may cause undesired Standards: CAN/CSA C22.2 No. 0-M91, CAN/CSA C22.2
operation. This device must be installed to ensure a minimum No. 94-M91, CSA Std C22.2 No. 142-M1987,
antenna separation distance of 20 cm from all persons. CSA Std C22.2 No. 157-92, CSA Std C22.2 No.
60529:05
Ordinary Location Certification from FM Markings: Intrinsically Safe for Class I, Division 1, Groups
Approvals A, B, C, D, T4 when installed per Rosemount
drawing 03031-1063; Type 4X/IP66/IP68
As standard, the transmitter has been examined and tested to
determine that the design meets the basic electrical, Europe
mechanical, and fire protection requirements by a nationally
recognized test laboratory (NRTL) as accredited by the Federal I1 ATEX Intrinsic Safety
Occupational Safety and Health Administration (OSHA). Certificate: Baseefa12ATEX0228X
Standards: EN 60079-0: 2012, EN 60079-11: 2012
Installing in North America Markings: II 1 G Ex ia IIC T4 Ga, T4(–40 °C ≤ Ta ≤ +70 °C)
IP66/IP68
The US National Electrical Code (NEC) and the Canadian
Electrical Code (CEC) permit the use of Division marked Special Conditions for Safe Use (X):
equipment in Zones and Zone marked equipment in Divisions.
The markings must be suitable for the area Classification, gas, 1. The plastic enclosure may constitute a potential
and temperature Class. This information is clearly defined in the electrostatic ignition risk and must not be rubbed or
respective codes. cleaned with a dry cloth.

USA 2. The Model 701PGNKF Power Module may be replaced in a
hazardous area. The power module has a surface resistivity
I5 U.S.A. Intrinsically Safe (IS) greater than 1GΩ and must be properly installed in the
Certificate: FM 3046325 wireless device enclosure. Care must be taken during
Standards: FM Class 3600 - 2011, FM Class 3610 - 2010, transportation to and from the point of installation to
FM Class 3810 - 2005, ANSI/ISA 60079-0 - prevent electrostatic charge build-up.
2009, ANSI/ISA 60079-11 - 2009, NEMA 250 -
2003, ANSI/IEC 60529 International

I7 IECEx Intrinsic Safety
Certificate: IECEx BAS 12.0124X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T4 Ga, T4(–40 °C ≤ Ta ≤ +70 °C)IP66/IP68

162 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Special Conditions for Safe Use (X) EAC - Belarus, Kazakhstan, Russia

1. The plastic enclosure may constitute a potential IM Technical Regulation Customs Union (EAC) Intrinsic Safety
electrostatic ignition risk and must not be rubbed or Certificate: RU C-US.ГБ05.B.00400
cleaned with a dry cloth. Markings: 0Ex ia IIC T4 Ga X;

2. The Model 701PGNKF Power Module may be replaced in a Special Condition for Safe Use (X):
hazardous area. The power module has a surface resistivity
greater than 1GΩ and must be properly installed in the 1. See certificate for special conditions.
wireless device enclosure. Care must be taken during
transportation to and from the point of installation to Korea
prevent electrostatic charge build-up.
IP Korea Intrinsic Safety
Brazil Certificate: 13-KB4BO-0295X
Markings: Ex ia IIC T4 (–40 °C ≤ Ta ≤ +70 °C)
I2 INMETRO Intrinsic Safety
Certificate: UL-BR 13.0534X Special Condition for Safe Use (X):
Standards: ABNT NBR IEC 60079-0:2008 + Errata 1:2011,
ABNT NBR IEC 60079-11:2009 1. See certificate for special conditions.
Markings: Ex ia IIC T4 IP66 Ga, T4(–40 °C ≤ Ta ≤ +70 °C)

Special Condition for Safe Use (X):

1. See certificate for special conditions.

China

I3 China Intrinsic Safety
Certificate: GYJ13.1362X
GYJ15.1367X [Flowmeters]
Standards: GB3836.1-2010, GB3836.4-20100,
GB3836.20-2010
Markings: Ex ia IIC T4 Ga, T4(–40 ~ +70 °C)

Special Condition for Safe Use (X):

1. See certificate for special conditions.

Japan

I4 TIIS Intrinsic Safety
Certificate: TC22022X (3051C/L)
TC22023X (3051T)
TC22024X (3051CFx)
Markings: Ex ia IIC T4 Ga, T4(–20 ~ +60 °C)

Special Condition for Safe Use (X):

1. See certificate for special conditions.

Emerson.com/Rosemount 163

Rosemount DP Level March 2017

Rosemount 2051 Wireless Special Conditions for Safe Use (X):

Rev 1.2 1. The Model 2051 Wireless Pressure Transmitter shall only
be used with the 701PGNKF Rosemount SmartPower
European Directive Information Battery Pack.

A copy of the EC Declaration of Conformity can be found at the 2. The inline pressure sensor may contain more than 10%
end of the Quick Start Guide. The most recent revision of the EC aluminum and is considered a potential risk of ignition by
Declaration of Conformity can be found at impact or friction. Care must be taken into account during
Emerson.com/Rosemount. installation and used to prevent impact and friction.

Telecommunication compliance 3. The surface resistivity of the transmitter housing is greater
than one gigaohm. To avoid electrostatic charge build-up,
All wireless devices require certification to ensure that they it must not be rubbed or cleaned with solvents or a dry
adhere to regulations regarding the use of the RF spectrum. cloth.
Nearly every country requires this type of product certification.
Emerson is working with governmental agencies around the Canada
world to supply fully compliant products and remove the risk of
violating country directives or laws governing wireless device I6 Canada Intrinsically Safe
usage. Certificate: CSA 2526009
Standards: CAN/CSA C22.2 No. 0-M91, CAN/CSA C22.2
FCC and IC No. 94-M91, CSA Std C22.2 No. 142-M1987,
CSA Std C22.2 No. 157-92, CSA Std C22.2 No.
This device complies with Part 15 of the FCC Rules. Operation is 60529:05
subject to the following conditions: This device may not cause Markings: Intrinsically Safe for Class I, Division 1,
harmful interference. This device must accept any interference Groups A, B, C, D, T4 when installed per
received, including interference that may cause undesired Rosemount drawing 03031-1063;
operation. This device must be installed to ensure a minimum Type 4X/IP66/IP68
antenna separation distance of 20 cm from all persons.
Europe
Ordinary Location Certification
I1 ATEX Intrinsic Safety
As standard, the transmitter has been examined and tested to Certificate: Baseefa12ATEX0228X
determine that the design meets the basic electrical, Standards: EN 60079-0: 2012, EN 60079-11: 2012
mechanical, and fire protection requirements by a nationally Markings: II 1 G Ex ia IIC T4 Ga, T4(–40 °C ≤ Ta ≤ +70 °C)
recognized test laboratory (NRTL) as accredited by the Federal IP66/IP68
Occupational Safety and Health Administration (OSHA).
Special Conditions for Safe Use (X):
Installing in North America
1. The plastic enclosure may constitute a potential
The US National Electrical Code (NEC) and the Canadian electrostatic ignition risk and must not be rubbed or
Electrical Code (CEC) permit the use of Division marked cleaned with a dry cloth.
equipment in Zones and Zone marked equipment in Divisions.
The markings must be suitable for the area Classification, gas, 2. The Model 701PGNKF Power Module may be replaced in a
and temperature Class. This information is clearly defined in the hazardous area. The power module has a surface resistivity
respective codes. greater than 1GΩ and must be properly installed in the
wireless device enclosure. Care must be taken during
USA transportation to and from the point of installation to
prevent electrostatic charge build-up.
I5 U.S.A. Intrinsically Safe (IS)
Certificate: FM 3046325 International
Standards: FM Class 3600 - 2011, FM Class 3610 - 2010,
FM Class 3810 - 2005, ANSI/ISA 60079-0 - I7 IECEx Intrinsic Safety
2009, Certificate: IECEx BAS 12.0124X
ANSI/ISA 60079-11 - 2009, NEMA 250 - 2003, Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
ANSI/IEC 60529 Markings: Ex ia IIC T4 Ga, T4(–40 °C ≤ Ta ≤ +70 °C)
Markings: IS CL I, DIV 1, GP A, B, C, D T4; CL 1, Zone 0 AEx IP66/IP68
ia IIC T4; T4(–40 °C ≤ Ta ≤ +70 °C) when
installed per Rosemount drawing 03031-1062;
Type 4X/IP66/IP68

164 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Special Conditions for Safe Use (X): Korea

1. The plastic enclosure may constitute a potential IP Korea Intrinsic Safety
electrostatic ignition risk and must not be rubbed or Certificate: 13-KB4BO-0220X
cleaned with a dry cloth. Markings: Ex ia IIC T4 (–40 °C ≤ Ta ≤ +70 °C)

2. The Model 701PGNKF Power Module may be replaced in a Special Condition for Safe Use (X):
hazardous area. The power module has a surface resistivity
greater than 1GΩ and must be properly installed in the 1. See certificate for special conditions.
wireless device enclosure. Care must be taken during
transportation to and from the point of installation to
prevent electrostatic charge build-up.

Brazil

I2 INMETRO Intrinsic Safety
Certificate: UL-BR 13.0534X
Standards: ABNT NBR IEC 60079-0:2008 + Errata 1:2011,
ABNT NBR IEC 60079-11:2009
Markings: Ex ia IIC T4 IP66 Ga, T4(–40 °C ≤ Ta ≤ +70 °C)

Special Conditions for Safe Use (X):

1. See certificate for special conditions.

China

I3 China Intrinsic Safety
Certificate: GYJ12.1295X
GYJ15.1365X [Flowmeters]
Standards: GB3836.1-2010, GB3836.4-2010,
GB3836.20-2010
Markings: Ex ia IIC Ga T4, –40 ~ +70 °C

Special Conditions for Safe Use (X):

1. See certificate for special conditions.

Japan

I3 TIIS Intrinsic Safety
Certificate: TC22022X (2051C/L)
TC22023X (2051T)
TC22024X (2051CFx)
Markings: Ex ia IIC T4 Ga, T4(-20 ~ +60 °C)

Special Condition for Safe Use (X):

1. See certificate for special conditions.

EAC - Belarus, Kazakhstan, Russia

IM Technical Regulation Customs Union (EAC) Intrinsic Safety
Certificate: RU C-US.ГБ05.B.00390
Markings: 0Ex ia IIC T4 Ga X;

Special Condition for Safe Use (X):
1. See certificate for special conditions.

Emerson.com/Rosemount 165

Rosemount DP Level March 2017

Dimensional Drawings

Figure 8. Rosemount 3051S ERS Measurement Transmitter - Coplanar Style

5.17 4.20
(131) (107)

9.63
(245)

6.44
(164)

½– 14 NPT (2)
Dimensions are in inches (millimeters).

Figure 9. Rosemount 3051S ERS Measurement Transmitter - In-Line Style

5.17 (131) 4.20 3.45
4.51 (115) (107) (88)

8.04
(204)

Dimensions are in inches (millimeters).

166 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Figure 10. Rosemount 3051S Scalable Level Transmitter with FF(1)(2)- Coplanar Style

5.17
(131)

4.51 4.20
(114) (107)

8.57
(217,7)

3.62
(92)

+ Direct mount
extension length

Dimensions are in inches (millimeters).

Figure 11. Rosemount 3051S Scalable Level Transmitter with FF(1)(2)- In-Line Style

5.17 4.20
(131) (107)

8.04 167
(204)

1.50
(38)

Dimensions are in inches (millimeters).

1. FF (FFW) seal dimensions and pressure ratings can be found on page 174.
2. Lower housing (flushing ring) is available with FFW style flange.

Emerson.com/Rosemount

Rosemount DP Level March 2017

Figure 12. Rosemount 3051S Scalable Level Transmitter with Thermal Range Expander – Coplanar Style

5.17 4.20
(131) (107)

8.57
(217.7)

7.64
(194.2)

Dimensions are in inches (millimeters).

Figure 13. Rosemount 3051S Scalable Level Transmitter with Thermal Range Expander – In-Line Style

4.20 5.17
(107) (131)

8.04
(204)

13.51
(343.2)
1.5
(38.1)

Dimensions are in inches (millimeters). Emerson.com/Rosemount

168

March 2017 Rosemount DP Level

Figure 14. Rosemount 3051S Scalable Level Transmitter with RF(1)- Coplanar Style

5.17
(131)

4.55 4.20
(116) (107)

8.57
(217,7)

3.62
(92)
+ Direct mount
extension length

Dimensions are in inches (millimeters).

Figure 15. Rosemount 3051S Scalable Level Transmitter with RF(1)- In-Line Style

5.17 4.20
(131) (107)

8.04 169
(204)

1.5 (38)

Dimensions are in inches (millimeters).

1. RF (RFW) seal dimensions and pressure ratings can be found on page 183.

Emerson.com/Rosemount

Rosemount DP Level March 2017

Figure 16. Rosemount 3051S Scalable Level Transmitter with SS(1)- Coplanar Style

5.17
(131)

4.55 4.20
(116) (107)

8.57
(217,7)

3.62
(92)
+ Direct mount
extension length

Dimensions are in inches (millimeters).

Figure 17. Rosemount 3051S Scalable Level Transmitter with SS(1) - In-Line Style

5.17 4.20
(131) (107)

8.04 Emerson.com/Rosemount
(204)

1.50 (38)

Dimensions are in inches (millimeters).

1. SS (SSW) seal dimensions and pressure ratings can be found on page 200.

170

March 2017 Rosemount DP Level

Figure 18. Rosemount 3051S Scalable Level Transmitter with SC(1)- Coplanar Style

5.17
(131)

4.55 4.20
(116) (107)

8.57
(217,7)

3.62
(92)
+ Direct Mount
Extension Length

Dimensions are in inches (millimeters).

Figure 19. Rosemount 3051S Scalable Level Transmitter with SC(1) - In-Line Style

5.17 4.20
(131) (107)

8.04 171
(204)

1.50 (38)

Dimensions are in inches (millimeters).

1. SC (SCW) seal dimensions and pressure ratings can be found on page 199.

Emerson.com/Rosemount

Rosemount DP Level March 2017

Figure 20. Rosemount 3051L Level Transmitter with FF or EF Seal(1)

4.09 5.13
(104)
(130)
4.29

(109)

A 6.67 7.14
(181)
(169)
8.29
(211)

B

“A” B. Flange adapters (optional, differential configuration only)

A. 2-, 4-, or 6-in. extension (only available with 3- and 4-in. flange
configurations)

Dimensions are in inches (millimeters).

Table 76. Transmitter Direct Mount Extension

Flange rating Transmitter flange extension Extension dimension (“A”)
7.65-in. (194,3 mm)
ANSI/ASME B16.5 Class 600 2-in. 5.65-in. (143,5 mm)
All others 0-in.

Figure 21. Rosemount 2051L Level Transmitter with FF or EF Seal(1)

3.85 4.99
(97.8) 4.36 (127)
(110.7)

A 6.6 7.02
(167,7) (178,2)

B 8.12
(206,2)

A. 2-. 4-, or 6-in. extension (only available with 3- and 4-in. flange B. Flange adapters (optional, differential configuration only)
configurations) Dimensions are in inches (millimeters).

1. FF (FFW) and EF (EFW) seal and flange diameter dimensions can be viewed on page 174 and page 190. Emerson.com/Rosemount

172

March 2017 Rosemount DP Level

Figure 22. Tuned System(1)(2) Assembly shown with Rosemount 3051S Scalable Level Transmitter

5.31
(132)

4.55 4.20
(116) (107)

7.70
(196) 8.57

(217,7)

Dimensions are in inches (millimeters).

Figure 23. Rosemount 1199 Remote Seal System Assembly shown with Rosemount 3051S Scalable Transmitter

4.20 (107) 4.55 (116)

7.70 (196) 8.57 (217,7)

A A
2.5 (63,5) Dimensions are in inches (millimeters).

BB
A. Capillary connection only
B. Capillary connects to Rosemount 1199 Remote seals

1. Tuned System Assemblies require specification of capillary length and addition Rosemount 1199 Remote Seal. 173
2. Tuned System Assemblies are available on all Level Transmitters.

Emerson.com/Rosemount

Rosemount DP Level March 2017

Figure 24. Thermal Optimizer (D5) with FFW

4.55 4.20
(116) (107)

8.04
(204)

5.72
(145)

Dimensions are in inches (millimeters).

Figure 25. FFW Flush Flanged Seal - Two-Piece Design (Shown with Flushing Ring)

D

A
F

B

B E

C D. Connection to transmitter
E. Flushing ring
A. Process flange F. Lower housing alignment clamp (option code SA)
B. Diaphragm
C. Flushing connection
Dimensions are in inches (millimeters).

174 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 77. Dimensions for FFW Flush Flanged Raised Face Seals-Two Piece (Upper Housing and Flange) Design

Pipe Class Flange Flange Bolt Number Bolt hole Standard diaphragm Raised face
size diameter thickness circle of bolts diameter diameter outer diameter
“A” in. (mm) “B” in. (mm) “C” in. (mm) “D” in. (mm)
“F” in. (mm) “G” in. (mm)

ANSI/ ASME 2-in. 150 6.00 (152) 0.69 (18) 4.75 (121) 4 0.75 (19) 2.30 (58) 3.62 (92)
300 6.50 (165) 0.81 (21) 5.00 (127) 8 0.75 (19) 2.30 (58) 3.62 (92)
EN1092-1 3-in. 600 6.50 (165) 1.00 (25) 5.00 (127) 8 0.75 (19) 2.30 (58) 3.62 (92)
900 8.50 (216) 1.00 (25) 6.50 (165) 8 1.50 (38) 2.30 (58) 3.62 (92)
JIS 4-in. 1500 8.50 (216) 1.00 (25) 6.50 (165) 8 1.50 (38) 2.30 (58) 3.62 (92)
2500 9.25 (235) 1.13 (29) 6.75 (172) 8 2.00 (51) 2.30 (58) 3.62 (92)
DN 50 150 7.50 (191) 0.88 (22) 6.00 (152) 4 0.75 (19) 3.50 (89) 5.00 (127)
DN 80 300 8.25 (210) 1.06 (27) 6.62 (168) 8 0.88 (22) 3.50 (89) 5.00 (127)
600 8.25 (210) 1.25 (32) 6.62 (168) 8 0.88 (22) 3.50 (89) 5.00 (127)
DN 900 9.50 (241) 1.50 (38) 7.50 (191) 8 1.00 (25) 3.50 (89) 5.00 (127)
100 1500 10.50 (267) 1.88 (48) 8.00 (203) 8 1.25 (32) 3.50 (89) 5.00 (127)
50A 2500 12.00 (305) 2.62 (67) 9.00 (229) 8 1.38 (35) 3.50 (89) 5.00 (127)
80A 150 9.00 (229) 0.88 (22) 7.50 (191) 8 0.75 (19) 3.50 (89) 6.20 (157)
100A 300 10.0 (254) 1.19 (30) 7.88 (200) 8 0.88 (22) 3.50 (89) 6.20 (157)
600 10.75 (273) 1.50 (38) 8.50 (216) 8 1.00 (25) 3.50 (89) 6.20 (157)
900 11.50 (292) 1.75 (45) 9.25 (235) 8 1.25 (32) 3.50 (89) 6.20 (157)
1500 12.25 (311) 2.12 (54) 9.50 (241) 8 1.38 (35) 3.50 (89) 6.20 (157)
2500 14.00 (356) 3.00 (76) 10.75(274) 8 1.63 (41) 3.50 (89) 6.20 (157)
PN 40 6.50 (165) 0.67 (17) 4.92 (125) 4 0.71 (18) 2.30 (58) 4.00 (102)
PN 63 7.09 (180) 0.91 (23) 5.31 (135) 4 0.87 (22) 2.30 (58) 4.00 (102)
PN 100 7.68 (195) 0.99 (25) 5.71 (145) 4 1.02 (26) 2.30 (58) 4.00 (102)
PN 160 7.68 (195) 1.06 (27) 5.71 (145) 4 1.02 (26) 2.30 (58) 4.00 (102)
PN 40 7.87 (200) 0.83 (21) 6.30 (160) 8 0.71 (18) 3.50 (89) 5.43 (138)
PN 63 8.46 (215) 0.99 (25) 6.69 (170) 8 0.88 (22) 3.50 (89) 5.43 (138)
PN 100 9.06 (230) 1.15 (29) 7.09 (180) 8 1.02 (26) 3.50 (89) 5.43 (138)
PN 160 9.06 (230) 1.30 (33) 7.09 (180) 8 1.02 (26) 3.50 (89) 5.43 (138)
PN 10/16 8.66 (220) 0.67 (17) 7.09 (180) 8 0.71 (18) 3.50 (89) 6.20 (157)
PN 40 9.25 (235) 0.94 (24) 7.48 (190) 8 0.87 (22) 3.50 (89) 6.20 (157)
PN 63 9.84 (250) 0.83 (21) 7.87 (200) 8 1.02 (26) 3.50 (89) 6.20 (157)
PN 100 10.43 (265) 1.30 (27) 8.27 (210) 8 1.18 (30) 3.50 (89) 6.20 (157)
PN 160 10.43 (265) 1.46 (37) 8.27 (210) 8 1.18 (30) 3.50 (89) 6.20 (157)
10K 6.10 (155) 0.63 (16) 4.72 (120) 4 0.75 (19) 2.30 (58) 3.62 (92)
20K 6.10 (155) 0.71 (18) 4.72 (120) 8 0.75 (19) 2.30 (58) 3.62 (92)
40K 6.50 (165) 1.02 (26) 5.12 (130) 8 0.75 (19) 2.30 (58) 4.00 (102)
10K 7.28 (185) 0.71 (18) 5.91 (150) 8 0.75 (19) 3.50 (89) 5.00 (127)
20K 7.87 (200) 0.87 (22) 6.30 (160) 8 0.91 (23) 3.50 (89) 5.00 (127)
40K 8.27 (210) 1.26 (32) 6.69 (170) 8 0.91 (23) 3.50 (89) 5.43 (138)
10K 8.27 (210) 0.71 (18) 6.89 (175) 8 0.75 (19) 3.50 (89) 6.20 (157)
20K 8.86 (225) 0.95 (24) 7.28 (185) 8 0.91 (23) 3.50 (89) 6.20 (157)
40K 9.84 (250) 1.42 (36) 8.07 (205) 8 0.98 (25) 3.50 (89) 6.20 (157)

Emerson.com/Rosemount 175

Rosemount DP Level March 2017

Table 78. Dimensions for FFW Flush Flanged Raised Face Seals-Two Piece (Upper Housing and Flange) Design

Pipe Inner Beveled Thickness with Thickness with Minimum gasket
size diameter I.D. “N” in. (mm)
Class “K” in. (mm) edge 1/4-NPT F.C. “M” in. 1/2-NPT F.C. “M” in. Weight lb (kg)

“L” in. (mm) (mm) (mm)

ANSI/ASME 2-in. 150 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.50 (64) 7.40 (3,33)
300 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.50 (64) 8.99 (4,05)
EN1092-1 3-in. 600 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.50 (64) 10.44 (4,70)
900 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.50 (64) 24.62 (11,08)
JIS 4-in. 1500 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.50 (64) 24.62 (11,08)
2500 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.50 (64) 36.71 (16,52)
DN 50 150 3.60 (91) 0.97 (25) 1.30 (33) 3.70 (94) 13.79 (6,21)
DN 80 300 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 17.84 (8,03)
600 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 20.31 (9,14)
DN 900 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 33.21 (14,94)
100 1500 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 46.76 (21,04)
50A 2500 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 81.34 (36,60)
80A 150 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 19.56 (8,80)
100A 300 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 29.56 (13,30)
600 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 40.73 (18,33)
900 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 53.16 (23,92)
1500 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 71.72 (32,27)
2500 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 125.72 (56,57)
PN 40 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.50 (64) 9.02 (4,06)
PN 63 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.50 (64) 12.58 (5,66)
PN 100 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.50 (64) 15.23 (6,85)
PN 160 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.50 (64) 16.12 (7,25)
PN 40 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 15.03 (6,76)
PN 63 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 18.87 (8,49)
PN 100 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 23.34 (10,50)
PN 160 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 25.83 (11,62)
PN 10/16 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 16.08 (7,24)
PN 40 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 20.31 (9,14)
PN 63 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 26.74 (12,03)
PN 100 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 34.26 (15,42)
PN 160 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 37.44 (16,85)
10K 2.12 (54) N/A 0.97 (25) 1.30 (33) 2.5 (64) 6.93 (3,15)
20K 2.12 (54) N/A 0.97 (25) 1.30 (33) 2.5 (64) 7.11 (3,20)
40K 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.5 (64) 10.41 (4,68)
10K 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 10.52 (4,73)
20K 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 13.61 (6,12)
40K 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 20.08 (9,04)
10K 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 14.03 (6,31)
20K 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 19.16 (8,62)
40K 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.70 (94) 32.12 (14,45)
N/A

176 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Figure 26. FFW Flush Flanged Seal - One-Piece Design (Shown with Flushing Ring) D

A

B

B

C

E

A. Process flange D. Connection to transmitter
B. Diaphragm E. Flushing ring
C. Flushing connection

Dimensions are in inches (millimeters).

Emerson.com/Rosemount 177

Rosemount DP Level March 2017

Table 79. Dimensions for FFW Flush Flanged Seals- One Piece (Upper Housing and Flange) Design
(Option code E)

Pipe Class Flange Flange Bolt Number Bolt hole Standard Raised Raised Minimum
size diameter thickness circle of bolts diameter diaphragm face face gasket I.D. Weight
“C” in. diameter
“A” in. “B” in. (mm) “D” in. diameter height “N” in. lb (kg)
(mm) (mm) (mm) “F” in. “G” in. “J” in. (mm)
(mm) (mm) (mm)
2.5 7.40
ANSI/ASME 150 6.00 0.69 4.75 4 0.75 2.30 3.62 0.06 (64) (3,33)
(152) (18) (121) 8 (19) (58) (92) (1,50)
EN 1092-1 300 8 3.62 2.5 8.99
6.50 0.81 5.00 8 0.75 2.30 (92) 0.06 (64) (4,05)
2-in. 600 (165) (21) (127) 8 (19) (58) 3.62 (1,50)
4 (92) 2.5 10.44
900/1500 6.50 1.00 5.00 8 0.75 2.30 3.62 0.25 (64) (4,70)
(165) (25) (127) 8 (19) (58) (92) (6,40)
2500 8 3.62 2.5 24.62
8.50 1.50 6.50 8 1.00 2.30 (92) 0.25 (64) (11,08)
150 (216) (38) (165) 8 (25) (58) 5.00 (127) (6,40)
8 2.5 36.71
300 9.25 2.00 6.75 8 1.13 2.30 5.00 (127) 0.25 (64) (16,52)
(235) (51) (172) 8 (29) (58) (6,40)
600 8 5.00 (127) 3.70 13.79
3-in. 7.50 0.88 6.00 8 1.13 3.50 0.06 (94) (6,21)
(191) (22) (152) 8 (25) (89) 5.00 (127) (1,50)
900 4 3.70 17.84
8.25 1.06 6.62 4 0.88 3.50 5.00 (127) 0.06 (94) (8,03)
1500 (210) (27) (168) 4 (22) (89) (1,50)
4 5.00 (127) 3.70 20.31
2500 8.25 1.25 6.62 0.88 3.50 0.25 (94) (9,14)
(210) (32) (168) (22) (89) 6.20 (157) (6,40)
150 3.70 33.21
9.50 1.50 7.50 1.00 3.50 6.20 (157) 0.25 (94) (14,94)
300 (241) (38) (229) (25) (89) (6,40)
6.20 (157) 3.70 46.76
600 10.50 1.88 8.00 1.25 3.50 0.25 (94) (21,04)
4-in. (267) (48) (203) (32) (89) 6.20 (157) (6,40)
3.70 81.34
900 12.00 2.62 9.00 1.38 3.50 6.20 (157) 0.25 (94) (36,60)
(305) (67) (229) (35) (89) (6,40)
1500 6.20 (157) 3.70 19.56
9.00 0.88 7.50 0.75 3.50 0.06 (94) (8,80)
2500 (229) (22) (191) (19) (89) 4.00 (102) (1,50)
3.70 29.56
PN 40 10.00 1.19 7.88 0.88 3.50 4.00 (102) 0.06 (94) (8,80)
(254) (30) (200) (22) (89) (1,50)
PN 63 4.00 (102) 3.70 40.73
DN5 10.75 1.50 8.50 1.00 3.50 0.25 (94) (18,33)
(273) (38) (216) (25) (89) 4.00 (102) (6.40)
0 3.70 53.16
PN 100 11.50 1.75 9.25 1.25 3.50 0.25 (94) (23,92)
(292) (45) (235) (32) (89) (6.40)
PN160 3.70 71.72
12.25 2.12 9.50 1.38 3.50 0.25 (94) (32,27)
(311) (54) (241) (35) (89) (6.40)
3.70 125.72
14.00 3.00 10.75 1.63 3.50 0.25 (94) (56,57)
(356) (76) (274) (41) (89) (6.40)
2.50 9.02
6.50 0.67 4.92 0,71 2.30 0.12 (64) (4,06)
(165) (17) (125) (18) (58) (3,00)
2.50 12,58
7.08 0.91 5.31 0.87 2.30 0.12 (64) (5,66)
(180) (23) (135) (22) (58) (3,00)
2.50 15.23
7.68 0.99 5.71 1.02 2.30 0.12 (64) (6,85)
(195) (25) (145) (26) (58) (3,00)
2.50 16.12
7.68 1.06 5.71 1.02 2.30 0.12 (64) (7,25)
(195) (27) (145) (26) (58) (3,00)

178 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 79. Dimensions for FFW Flush Flanged Seals- One Piece (Upper Housing and Flange) Design
(Option code E)

Pipe Class Flange Flange Bolt Number Bolt hole Standard Raised Raised Minimum
size diameter thickness circle of bolts diameter diaphragm face face gasket I.D. Weight
“C” in. diameter
“A” in. “B” in. (mm) “D” in. diameter height “N” in. lb (kg)
(mm) (mm) (mm) “F” in. “G” in. “J” in. (mm)
(mm) (mm) (mm)
3.70 15.03
PN 40 7.87 0.83 6.30 8 0.71 3.50 5.43 (138) 0.12 (94) (6,76)
(200) (21) (160) 8 (18) (89) (3,0)
8 3.70 18.87
PN 63 8.46 0.99 6.69 8 0.88 3.50 5.43 (138) 0.12 (94) (8,49)
DN8 (215) (25) (170) 8 (22) (89) (3,0)
0 8 3.70 23.34
9.06 1.15 7.09 8 1.02 3.50 5.43 (138) 0.12 (94) (10,50)
PN 100 (230) (29) (180) 8 (26) (89) (3,0)
8 3.70 25.83
EN 1092-1 PN160 9.06 1.30 7.09 4 1.02 3.50 5.43 (138) 0.12 (94) (11,62)
PN (230) (33) (180) 8 (26) (89) (3,0)
8 3.70 16.08
10/16 8.66 0.67 7.09 8 0.71 3.50 6.20 (157) 0.12 (94) (7,24)
PN 40 (220) (17) (180) 8 (18) (89) (3,0)
8 3.70 20.31
PN 63 9.25 0.83 7.48 8 0.87 3.50 6.20 (157) 0.12 (94) (9,14)
(235) (21) (190) 8 (22) (89) (3,0)
PN 100 8 3.70 26.74
DN1 9.84 1.07 7.87 1.02 3.50 6.20 (157) 0.12 (94) (1203)
00 (250) (27) (200) (26) (89) (3,0)
3.70 34.26
10.43 1.30 8.27 1.18 3.50 6.20 (157) 0.12 (94) (15,42)
(265) (33) (210) (30) (89) (3,0)
3.70 37.44
PN 160 10.43 1.46 8.27 1.18 3.50 6.20 (157) 0.12 (94) (16,85)
(265) (37) (210) (30) (89) (3,0)
2.50 6.93
10K 6.1 0.63 4.72 0.75 2.30 3.62 0.08 (64) (3,15)
(155) (16) (120) (19) (58) (92) (2,0)
2.50 7.11
50A 20K 6.1 0.71 4.72 0.75 2.30 3.62 0.08 (64) (3,20)
(155) (18) (120) (19) (58) (92) (2,0)
2.50 10.41
40K 6.5 1.02 5.12 0.75 2.30 4.00 0.08 (64) (4,68)
(165) (26) (130) (19) (58) (102) (2,0)
3.70 10.52
10K 7.28 0.71 5.91 0.75 3.50 5.00 0.08 (94) (4,73)
(185) (18) (150) (19) (89) (127) (2,0)
3.70 13.61
JIS 80A 20K 7.87 0.87 6.3 0.91 3.50 5.00 0.08 (94) (6,12)
(200) (22) (160) (23) (89) (127) (2,0)
3.70 20.08
40K 8.27 1.26 6.69 0.91 3.50 5.43 0.08 (94) (9,04)
(210) (32) (170) (23) (89) (138) (2,0)
3.70 14.03
10K 8.27 0.71 6.89 0.75 3.50 6.20 0.08 (94) (6,31)
(210) (18) (175) (19) (89) (157) (2,0)
3.70 19.16
100A 20K 8.86 0.95 7.28 0.91 3.50 6.20 0.08 (94) (8,62)
(225) (24) (185) (23) (89) (157) (2,0)
3.70 32.12
40K 9.84 1.42 8.07 0.98 3.50 6.20 0.08 (94) (14,45)
(250) (36) (205) (25) (89) (157) (2,0)

Emerson.com/Rosemount 179

Rosemount DP Level March 2017

Figure 27. FFW Flush Flanged Seal - Flushing Connection Ring (Lower Housing)

M

K
L
G

Table 80. Dimensions for FFW Flushing Connection Ring (Lower Housing)

Raised face Inner diameter Beveled edge Thickness with Thickness with Weight lb
diameter “G” in. “K” in. (mm) “L” in. (mm) (kg)
Pipe size Class 1/4 NPT F.C. “M” 1/2 NPT F.C. “M”
(mm)
in. (mm) in. (mm)

ANSI/ASME 2-in. 150 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 7.41 (3,33)
3-in. 300 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 8.99 (4,05)
EN1092-1 600 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 10.44 (4,70)
4-in. 900/1500 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 24.62 (11,08)
DN 50 2500 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 36.71 (16,52)
DN 80 150 5.00 (127) 3.60 (91) 0.97 (25) 1.30 (33) 13.79 (6,21)
DN 100 300 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 17.84 (8,03)
600 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 20.31 (9,14)
900 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 33.21 (14,94)
1500 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 46.76 (21,04)
2500 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 81.34 (36,60)
150 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 19.56 (8,80)
300 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 29.56 (13,30)
600 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 40.73 (18,33)
900 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 53.16 (23,92)
1500 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 71.72 (32,27)
N/A
2500 6.20 (157) 3.60 (91) 0.97 (25) 1.30 (33) 125.72
N/A (56,57)
PN 40 4.00 (102) 2.40 (61) 0.97 (25) 1.30 (33) 9.02 (4,06)
PN 63 4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 12.58 (5,66)
PN 100 4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 15.23 (6.85)
PN 160 4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 16.12 (7,25)
PN 40 5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 15.03 (6,76)
PN 63 5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 18.87 (8,49)
PN 100 5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 23.34 (10.50)
PN 160 5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 25.83 (11,62)
PN 10/16 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 16.08 (7,24)
PN 40 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 20.31 (9,14)
PN 63 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 26.74 (12,03)
PN 100 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 34.26 (15,42)
PN 160 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 37.44 (16,85)
N/A

180 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 80. Dimensions for FFW Flushing Connection Ring (Lower Housing)

Raised face Inner diameter Beveled edge Thickness with Thickness with Weight lb
diameter “G” in. “K” in. (mm) “L” in. (mm) (kg)
Pipe size Class 1/4 NPT F.C. “M” 1/2 NPT F.C. “M”
(mm)
in. (mm) in. (mm)

10K 3.62 (92) 2.12 (54) N/A 0.97 (25) 1.30 (33) 6.93 (3,15)

50A 20K 3.62 (92) 2.12 (54) N/A 0.97 (25) 1.30 (33) 7.11 (3,20)

40K 4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 10.41 (4,68)

10K 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 10.52 (4,73)

JIS 80A 20K 5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 13.61 (6,12)

40K 5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 20.08 (9,04)

10K 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 14.03 (6,31)

100A 20K 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 19.16 (8,62)

40K 6.20 (157) 3.60 (91) N/A 0.97 (25) 1.30 (33) 32.12 (14,45)

Emerson.com/Rosemount 181

Rosemount DP Level March 2017

Figure 28. RFW Flanged Seal Standard Design D

A

B
C

B

A. Process flange C. Lower housing or flushing connection
B. Diaphragm D. Connection to transmitter

Dimensions are in inches (millimeters).

Table 81. RFW Flanged Seal Standard Design Dimensions(1)

Flange Flange Overall height “C” in. (mm) Bolt circle Bolt hole Lower
diameter diameter
Pipe Class thickness No or 1/4-in. NPT 1/2-in.NPT flush diameter “D” housing Weight lb (kg)
size “A” in. “E” in. diameter
(mm) “B” in. (mm) flush connection connection in. (mm) (mm)
“F” in. (mm)

ANSI/ASME 1/2-in. 2500 5.25 (133) 1.19 (30) 2.45 (62) 2.79 (71) 3.25 (83) 0.75 (19) 2.62 (67) 8.49 (3,82)
3/4-in. 300/600 4.62 (117) 0.62 (16) 2.45 (62) 2.79 (71) 3.50 (89) 0.88 (22) 2.62 (67) 4.99 (2,25)
EN 1092-1 900/1500 5.12 (130) 1.00 (25) 2.45 (62) 2.79 (71) 3.50 (89) 0.88 (22) 2.62 (67) 7.25 (3,26)
1-in. 5.50 (140) 1.25 (32) 2.45 (62) 2.79 (71) 3.75 (95) 0.88 (22) 2.62 (67) 9.52 (4,28)
2500 4.25 (108) 0.50 (13) 2.45 (62) 2.79 (71) 3.25 (83) 0.75 (19) 2.62 (67) 4.19 (1,89)
11/2-i 150 4.88 (124) 0.62 (16) 2.45 (62) 2.79 (71) 3.12 (79) 0.63 (16) 2.62 (67) 5.30 (2,39)
n. 300 4.88 (124) 0.69 (18) 2.45 (62) 2.79 (71) 3.50 (89) 0.75 (19) 2.62 (67) 5.58 (2,51)
600 5.88 (150) 1.12 (29) 2.45 (62) 2.79 (71) 4.00 (102) 1.00 (25) 2.62 (67) 9.68 (4,36)
DN 900/1500 6.25 (159) 1.38 (35) 2.45 (62) 2.79 (71) 4.25 (108) 1.00 (25) 2.87 (73) 13.68 (6,16)
25 2500 5.00 (127) 0.62 (16) 2.45 (62) 2.79 (71) 3.50 (89) 0.63 (22) 2.62 (67) 5.63 (2,53)
DN 150 6.12 (155) 0.75 (19) 2.45 (62) 2.79 (71) 3.88 (99) 0.75 (19) 2.88 (73) 8.20 (3.69)
40 300 6.12 (155) 0.88 (22) 2.45 (62) 2.79 (71) 4.50 (114) 0.88 (22) 2.88 (73) 9.09 (4,09)
20A 600 7.00 (178) 1.25 (32) 2.45 (62) 2.79 (71) 4.50 (114) 0.88 (22) 2.88 (73) 14.48 (6,52)
25A 900 7.00 (178) 1.25 (32) 2.45 (62) 2.79 (71) 4.88 (124) 1.13 (29) 2.88 (73) 14.48 (6,62)
1500 8.00 (203) 1.75 (45) 2.45 (62) 2.79 (71) 5.75 (146) 1.25 (32) 2.88 (73) 25.34 (11,40)
40A 2500
4.53 (115) 0.71 (18) 2.45 (62) 2.79 (71) 3.35 (85) 0.55 (14) 2.68 (68) 5.09 (2,29)
PN 40
JIS 5.91 (150) 0.71 (18) 2.45 (62) 2.79 (71) 4.33 (110) 0.71 (18) 3.47 (88) 8.04 (3,62)
PN 40
4.72 (120) 0.79 (20) 2.45 (62) 2.79 (71) 3.35 (85) 0.75 (19) 2.62 (67) 5.59 (2,52)
40K 4.92 (125) 0.55 (14) 2.45 (62) 2.79 (71) 3.54 (90) 0.75 (19) 2.62 (67) 5.00 (2,25)
10K 4.92 (125) 0.63 (16) 2.45 (62) 2.79 (71) 3.54 (90) 0.75 (19) 2.62 (67) 5.31 (2,39)
20K 5.12 (130) 0.87 (22) 2.45 (62) 2.79 (71) 3.74 (95) 0.75 (19) 2.76 (70) 6.86 (3,09)
40K 5.51 (140) 0.63 (16) 2.45 (62) 2.79 (71) 4.13 (105) 0.75 (19) 3.19 (81) 6.20 (2,79)
10K 5.51 (140) 0.71 (18) 2.45 (62) 2.79 (71) 4.13 (105) 0.75 (19) 3.19 (81) 7.36 (3,31)
20K 6.30 (160) 0.94 (24) 2.45 (62) 2.79 (71) 4.72 (120) 0.91 (23) 3.54 (90) 11.06 (4,98)
40K

1. Lower housing is loose on standard design, consult factory for retained lower housing options.

182 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Figure 29. RFW Flanged Seal Stud Bolt Design E

A A
B
C
C E
D B
D
B
D. Bolts
A. Upper housing E. Connection to transmitter
B. Diaphragm
C. Lower housing or flushing connection

Dimensions are in inches (millimeters).

Table 82. RF/RFW Flanged Seal Standard Design Dimensions(1)

Overall height “A” in.

(mm) Stud circle Stud (size, length) Lower housing Raised face Weight lb
diameter “C” in. (mm) (kg)
Pipe size Class No or 1/4-in. 1/2-in. NPT “B” in. (mm) diameter “D” diameter “E”

NPT flush flush in. (mm) in. (mm)

connection connection

EN 1092-1 ANSI/ASME 1/2-in. 150 2.52 (64) 2.82 (72) 2.38 (61) 1/2–13NC, 2.5-in. 3.74 (95) 1.38 (35) 6.28 (2,83)
300/600 2.77 (70) 2.87 (73) 2.62 (67) 1/2–13NC, 2.5-in. 3.75 (95) 1.38 (35) 6.53 (2,94)

3/4-in. 150 2.52 (64) 2.82 (72) 2.75 (70) 1/2–13NC, 2.5-in. 3.88 (99) 1.69 (43) 6.46 (2,91)

JIS DN 15 PN 40 2.52 (64) 2.82 (72) 2.56 (65) M12 x 1.75, 60 mm 3.74 (95) 1.77 (45) 6.27 (2,82)
DN 20 PN 100/160 2.52 (64) 2.82 (72) 2.95 (75) M12 x 1.75, 60 mm 4.13 (105) 1.77 (45) 6.92 (3,11)
10A
PN 40 2.52 (64) 2.82 (72) 2.95 (75) M12 x 1.75, 60 mm 4.13 (105) 2.28 (58) 6.90 (3,11)
15A
20A 10/20K 2.52 (64) 2.82 (72) 2.56 (65) M12 x 1.75, 60 mm 3.74 (95) 1.81 (46) 6.30 (2,84)
40K 2.52 (64) 2.82 (72) 2.95 (75) M16 x 2.00, 70 mm 4.33 (110) 2.05 (52) 7.70 (3,47)
10K 2.52 (64) 2.82 (72) 2.76 (70) M12 x 1.75, 60 mm 3.74 (95) 2.01 (51) 6.39 (2,88)
20K 2.52 (64) 2.82 (72) 2.76 (70) M12 x 2.00, 60 mm 3.74 (95) 2.01 (51) 6.39 (2,88)
40K 2.52 (64) 2.82 (72) 3.15 (80) M16 x 2.00, 70 mm 4.53 (115) 2.17 (55) 8.26 (3,72)
2.52 (64) 2.82 (72) 2.95 (75) M12 x 1.75, 60 mm 3.94 (100) 2.21 (56) 6.68 (3,01)
10/20K

1. Lower housing is loose on standard design, consult factory for retained lower housing options.

Emerson.com/Rosemount 183

Rosemount DP Level March 2017

Figure 30. EFW Extended Flanged Seal - Extended Flanged Assembly D

A

B E
C C

A. Process flange D. Connection to transmitter
B. Extension E. Extension length
C. Diaphragm

Dimensions are in inches (millimeters).

Table 83. EFW Extended Flanged Seal Dimensions

Pipe Class Flange diameter Flange thickness Bolt circle “C” in. Number Bolt hole Raised face
size “A” in. (mm) diameter “D” diameter “F” in.
“B” in. (mm) (mm) of bolts
in. (mm) (mm)

150 5.00 (127) 0.62 (16) 0.63 (16) 4 3.88 (99) 2.88 (73)
300 6.12 (156) 0.75 (19) 0.88 (22)
600 6.12 (156) 0.88 (22) 0.88 (22) 4 4.50 (114) 2.88 (73)
900/1500 7.00 (178) 1.25 (32) 1.13 (28)
11/2-in. 2500 8.00 (203) 1.75 (45) 1.25 (32) 4 4.50 (114) 2.88 (73)
2-in. 150 6.00 (152) 0.69 (18) 0.75 (19)
3-in. 300 6.50 (165) 0.82 (21) 0.75(19) 4 4.88 (124) 2.88 (73)
4-in. 600 6.50 (165) 1.00 (25) 0.75 (19)
900/1500 8.50(216) 1.50 (38) 1.00 (25) 4 5.75 (146) 2.88 (73)
2500 9.25(235) 2.00 (51) 1.13 (29)
150 7.50(191) 0.88 (22) 0.75 (19) 4 4.75 (121) 3.62 (92)
300 8.25(210) 1.06 (27) 0.88 (22)
600 8.25(210) 1.25 (32) 0.88 (22) 8 5.00 (127) 3.62 (92)
900 9.50(241) 1.50 (38) 1.00 (25)
1500 10.50(267) 1.88 (48) 1.25(32) 8 5.00 (127) 3.62 (92)
2500 12.00(305) 2.62 (67) 1.38 (35)
ANSI/ASME 150 9.00(229) 0.88 (22) 0.75 (19) 8 6.50 (165) 3.62 (92)
300 10.00(254) 1.19 (30) 0.88 (22)
600 10.75(273) 1.50 (38) 1.00 (25) 8 6.75 (172) 3.62(92)
900 11.50(292) 1.75 (45) 1.25 (32)
1500 12.25(311) 2.12 (54) 1.38 (35) 4 6.00 (152) 5.00 (127)
2500 14.00 (356) 3.00 (76) 1.63 (41)
8 6.62 (168) 5.00 (127)

8 6.62 (168) 5.00 (127)

8 7.50 (191) 5.00 (127)

8 8.00 (203) 5.00 (127)

8 9.00 (229) 5.00 (127)

8 7.50 (191) 6.20 (158)

8 7.88 (200) 6.20 (158)

8 8.50 (216) 6.20 (158)

8 9.25 (235) 6.20 (158)

8 9.50 (241) 6.20 (158)

8 10.75 (274) 6.20 (158)

184 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 83. EFW Extended Flanged Seal Dimensions

Pipe Class Flange diameter Flange thickness Bolt circle “C” in. Number Bolt hole Raised face
size “A” in. (mm) diameter “D” diameter “F” in.
“B” in. (mm) (mm) of bolts
in. (mm) (mm)

PN 40 6.50 (165) 0.67 (17) 0.71 (18) 4 4.92 (125) 4.02 (102)
PN 63 7.08 (180) 0.91 (23) 0.87 (22)
DN 50 PN 100 7.68 (195) 0.98 (25) 1.02 (26) 4 5.31 (135) 4.02 (102)
DN 80 PN 160 7.68 (195) 1.06 (27) 1.02 (26)
PN 40 7.87 (200) 0.83 (21) 0.71 (18) 4 5.71 (145) 4.02 (102)
DN 100 PN 63 8.46 (215) 0.98 (25) 0.88 (22)
50A PN 100 9.06 (230) 1.14 (29) 1.02 (26) 4 5.71 (145) 4.02 (102)
80A PN 160 9.06 (230) 1.30 (33) 1.02 (26)
EN 1092-1 100A PN 10/16 8.66 (220) 0.67 (17) 0.71 (18) 8 6.30 (160) 5.43 (138)
PN 40 9.25 (235) 0.83 (21) 0.87 (22)
PN 63 9.84 (250) 1.06 (27) 1.02 (26) 8 6.69 (170) 5.43 (138)
PN 100 10.43 (265) 1.30 (33) 1.18 (30)
PN 160 10.43 (265) 1.46 (37) 1.18 (30) 8 7.09 (180) 5.43 (138)
10K 6.10 (155) 0.63 (16) 0.75 (19)
20K 6.10 (155) 0.71 (18) 0.75 (19) 8 7.09 (180) 5.43 (138)
40K 6.50 (165) 1.02 (26) 0.75 (19)
10K 7.28 (185) 0.71 (18) 0.75 (19) 8 7.09 (180) 6.20 (158)
20K 7.87 (200) 0.87 (22) 0.91 (23)
40K 8.27 (210) 1.26 (32) 0.91 (23) 8 7.48 (190) 6.20 (158)
10K 8.27 (210) 0.71 (18) 0.75 (19)
20K 8.86 (225) 0.94 (24) 0.91 (23) 8 7.87 (200) 6.20 (158)
40K 9.84 (250) 1.42 (36) 0.98 (25)
8 8.27 (210) 6.20 (158)

8 8.27 (210) 6.20 (158)

4 4.72 (120) 3.62 (92)

8 4.72 (120) 3.62 (92)

8 5.12 (130) 4.00 (102)

8 5.91 (150) 5.00 (127)

JIS 8 6.30 (160) 5.00 (127)

8 6.69 (170) 5.43 (138)

8 6.89 (175) 6.20 (158)

8 7.28 (185) 6.20 (158)

8 8.07 (205) 6.20 (158)

Table 84. EFW Extended Flanged Seal Dimensions

Process connection size Diameter “E”
in. (mm)
ANSI B16.5 EN 1092-1 JIS B2238
2.58 (66)
3-in. DN 80 80A 3.50 (89)
4-in. DN 100 100A 1.45 (37)
11/2-in. DN 40 40A 1.90 (48)
2-in. DN 50 50A 2.88 (73)
3-in. Headbox DN 80 Headbox N/A 3.78 (96)
4-in. Headbox DN100 Headbox N/A

Table 85. EFW Extended Flanged Seal Weights Pounds (Kilograms)

Extension length

Pipe size Class 1-in. 2-in. 3-in. 4-in. 5-in. 6-in. 7-in. 8-in. 9-in.

(25 mm) (51 mm) (76 mm) (102 mm) (127 mm) (152 mm) (178 mm) (203 mm) (229 mm)

ANSI/ASME 11/2-in. 150 5.53 5.99 6.46 6.92 7.38 7.85 8.31 8.78 7.47
300 (2,49) (2,70) (2,91) (3,11) (3,32) (3,53) (3,74) (3,95) (3,36)
600
900/1500 8.11 8.57 9.04 9.50 9.96 10.43 10.89 11.36 10.05
2500 (3,65) (3,86) (4,07) (4,28) (4,48) (4,69) (4,90) (5,11) (4,52)

9.00 9.46 9.93 10.39 10.86 11.32 11.78 12.25 10.94
(4,05) (4,56) (4,47) (4,68) (4,89) (5,09) (5,30) (5,51) (4,92)

15.19 15.66 16.12 16.59 17.05 17.51 17.98 18.44 18.70
(6,86) (7,05) (7,25) (7,47) (7,67) (7,88) (8,09) (8,30) (8,42)

25.38 25.84 26.31 26.77 27.23 27.70 28.16 28.63 28.89
(11,42) (11,63) (11,84) (12,05) (12,25) (12,47) (12,67) (12,88) (13,00)

Emerson.com/Rosemount 185

Rosemount DP Level March 2017

Table 85. EFW Extended Flanged Seal Weights Pounds (Kilograms)

Extension length

Pipe size Class 1-in. 2-in. 3-in. 4-in. 5-in. 6-in. 7-in. 8-in. 9-in.

(25 mm) (51 mm) (76 mm) (102 mm) (127 mm) (152 mm) (178 mm) (203 mm) (229 mm)

150 8.22 8.80 9.41 10.00 10.60 11.19 11.79 12.38 11.16
(3,70) (3,96) (4,23) (4,50) (4,77) (5,04) (5,31) (5,57) (5,02)
300
9.81 10.39 11.00 11.60 12.19 12.79 13.38 13.98 12.75
2-in. 600 (4,41) (4,68) (4,95) (5,22) (5,49) (5,76) (6,02) (6,29) (5,74)

900/1500 11.26 11.84 12.44 13.05 13.64 14.23 14.83 15.42 14,20
(5,07) (5,33) (5,60) (5,87) (6,14) (6,40) (6,67) (6,94) (6.39)
2500
25.50 26.31 27.12 27.92 28.73 29.54 30.34 31.15 31.32
150 (11,48) (11,84) (12,20) (12,56) (12,93) (13,29) (13,65) (14,02) (14,09)

300 36.58 37.38 38.19 39.00 39.80 40.61 41.42 42.22 42.40
(16,46) (16,82) (17,19) (17,55) (17,91) (18,27) (18,64) (19,00) (19,08)
600
3-in. 15.89 17.64 19.48 21.27 23.08 24.88 26.69 28.50 22.47
(7,15) (7,94) (8,77) (9,57) (10,39) (11,20) (12,01) (12,83) (10,11)
900
19.94 21.69 23.53 25.32 27.13 28.93 30.74 32.54 26.52
1500 (8,97) (9,76) (10,59) (11,39) (12,21) (13,02) (13,83) (14,64) (11,93)

ANSI/ASME 2500 22.43 24.18 26.02 27.81 29.62 31.42 33.23 35.03 29.01
(10,09) (10,88) (11,71) (12,51) (13,33) (14,14) (14,95) (15,76) (13,05)
150
33.26 35.10 36.90 38.71 40.51 42.32 44.12 45.93 48.80
300 (14,97) (15,80) (16,61) (17,42) (18,23) (19,04) (19,85) (20,67) (21,96)

3-in. Headbox 600 47.88 49.71 51.52 53.33 55.13 56.94 58.74 60.55 63.42
900 (21,55) (22,37) (23,18) (24,00) (24,81) (25,62) (26,43) (27,25) (28,54)

1500 83.46 85.30 87.10 88.91 90.71 92.52 94.33 96.13 99.00
(37,56) (38,39) (39,20) (40,01) (40,82) (41,63) (42,45) (43,26) (44,55)
2500
15.76 17.40 19.07 20.90 22.40 24.07 25.74 27.41 23.24
150 (7,09) (7,83) (8,58) (9,41) (10,08) (10,83) (11,58) (12,33) (10,46)

300 19.81 21.45 23.12 24.95 26.45 28.12 29.79 31.45 27.29
(8,91) (9,65) (10,40) (11,23) (11,90) (12,65) (13,41) (14,15) (12,28)
600
4-in. 22.30 23.94 25.61 27.44 28.94 30.61 32.28 33.94 29.78
(10,04) (10,77) (11,52) (12,35) (13,02) (13,77) (14,53) (15,27) (13,40)
900
33.13 34.83 36.50 38.17 39.84 41.51 43.15 44.85 47.58
1500 (14,91) (15,67) (16,53) (17,18) (17,93) (18,68) (19,42) (20,18) (21,41)

2500 47.75 49.45 51.12 52.79 54.46 56.13 57.76 59.46 62.20
(21,49) (22,25) (23,00) (23,76) (24,51) (25,26) (25,99) (26,76) (27,99)

83.33 85.03 86.70 88.37 90.04 91.71 93.35 95.05 97.78
(37,50) (38,26) (39,02) (39,77) (40,52) (41,27) (42,01) (42,77) (44,00)

28.61 39.17 49.62 60.07 70.52 80.94 91.42 101.88 31.74
(12,87) (17,63) (22,33) (27,03) (31,73) (36,42) (41,14) (45,85) (14,28)

38.62 49.18 59.63 70.08 80.54 90.96 101.44 111.89 41.75
(17,38) (22,13) (26,83) (31,54) (36,24) (40,93) (45,65) (50,35) (18,79)

48.37 58.93 69.38 79.83 90.28 100.70 111.19 121.64 51.50
(21,77) (26,52) (31,22) (35,92) (40,63) (45,32) (50,04) (54,74) (23,18)

55.27 58.50 61.73 64.96 67.31 70.34 73.36 76.38 80.30
(24,87) (26,33) (27,78) (29,23) (30,29) (31,65) (33,01) (34,37) (36,14)

72.28 75.51 78.74 81.97 84.33 87.35 90.37 93.39 97.31
(32,53) (33,98) (35,43) (36,89) (37,95) (39,31) (40,67) (42,03) (43,79)

126.52 129.75 132.98 136.20 138.57 141.59 144.61 147.63 151.55
(56,93) (58,39) (59,84) (61,29) (62,36) (63,72) (65,07) (66,43) (68,20)

186 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 85. EFW Extended Flanged Seal Weights Pounds (Kilograms)

Extension length

Pipe size Class 1-in. 2-in. 3-in. 4-in. 5-in. 6-in. 7-in. 8-in. 9-in.

(25 mm) (51 mm) (76 mm) (102 mm) (127 mm) (152 mm) (178 mm) (203 mm) (229 mm)

150 22.84 25.85 28.90 31.99 35.00 38.06 41.11 44.13 32.00
(10,28) (11,63) (13,01) (14,40) (15,75) (17,13) (18,50) (19,86) (14,40)
300
32.85 35.87 38.92 42.00 45.02 48.07 51.12 54.14 42.02
ANSI/ASME 4-in. 600 (14,78) (16,14) (17,51) (18,90) (20,26) (21,63) (23,00) (24,36) (18,91)
Headbox 900
42.60 45.62 48.67 51.75 54.77 57.82 60.8 63.89 51.7 7
1500 (19,17) (20,53) (21,90) (23,29) (24,65) (26,02) 7(27,39) (28,75) (23,30)

2500 55.24 58.32 61.37 64.41 67.47 70.52 73.5 76.62 80.74
(24,86) (26,24) (27,62) (28,98) (30,36) (31,73) 7(33,11) (34,48) (36,33)
PN 40
72.25 75.33 78.38 81.43 84.48 87.53 90.58 93.63 97.75
DN 40 PN 63/100 (32,51) (33,90) (35,27) (36,64) (38,02) (39,39) (40,76) (42,13) (43,99)

PN 160 126.49 129.57 132.62 135.67 138.72 141.78 144.83 147.88 152.00
(56,92) (58,31) (59,68) (61,05) (62,42) (63,80) (65,17) (66,55) (68,4)
PN 40
7.46 7.92 8.38 8.85 9.31 9.77 10.24 10.70 9.39
DN 50 PN 63 (3,36) (3,56) (3,77) (3,98) (4,19) (4,40) (4,61) (4,82) (4,23)
PN 100
11.52 11.98 12.44 12.91 13.37 13.84 14.30 14.76 13.45
PN 160 (5,18) (5,39) (5,60) (5,81) (6,23) (6,34) (6,44) (6,64) (6,05)

PN 40 13.17 13.63 14.10 14.56 15.03 15.49 15.95 16.42 16.83
(5,93) (6,13) (6,35) (6,55) (6,76) (6,97) (7,18) (7,39) (7,57)
EN 1092 - 1 Schedule PN 63
40 PN 100 9.87 10.45 11.06 11.66 12.25 12.84 13.44 14.03 12.81
(4,44) (4,70) (5,00) (5,25) (5,51) (5,78) (6,05) (6,31) (5,76)
PN 160
13.37 13.96 14.56 15.16 15.75 16.35 16.94 17.54 16.31
PN 40 (6,02) (6,28) (6,55) (6,82) (7,09) (7,36) (7,62) (7,89) (7,34)

DN Schedule PN 63 16.05 16.63 17.23 17.83 18.43 19.02 19.61 20.21 18.99
80 80 PN 100 (7,22) (7,48) (7,75) (8,02) (8,29) (8,56) (8,82) (9,09) (8,55)

PN160 18.14 18.95 19.76 20.56 21.37 22.18 22.98 23.79 23.96
(8,16) (8,53) (8,89) (9,25) (9,62) (9,98) (10,34) (10,71) (10,78)
PN 40
16.85 18.47 20.08 21.70 23.32 24.94 26.56 28.18 23.97
Headbox PN 63 (7,58) (8,31) (9,04) (9,77) (10,49) (11,22) (11,95) (12,68) (10,79)
PN 100
20.70 22.32 23.93 25.55 27.17 28.79 30.41 32.03 27.82
PN 160 (9,32) (10,04) (10,77) (11,50) (12,23) (12,96) (13,68) (14,41) (12,52)

25.29 26.90 28.51 30.13 31.75 33.37 34.99 36.61 32.40
(11,38) (12,11) (12,83) (13,56) (14,29) (15,02) (15,75) (16,47) (14,58)

29.45 31.10 32.72 34.33 35.95 37.57 39.17 40.81 43.50
(13,25) (14,00) (14,72) (15,45) (16,18) (16,91) (17,64) (18,36) (19,58)

16.53 17.76 19.07 20.36 21.65 22.93 24.22 25.51 21.12
(7,44) (7,99) (8,58) (9,16) (9,74) (10,32) (10,90) (11,48) (9,50)

20.38 21.61 22.92 24.21 25.50 26.78 28.07 29.36 24.97
(9,17) (9,72) (10,31) (10,89) (11,48) (12,05) (12,63) (13,21) (11,24)

24.97 26.20 27.51 28.79 30.08 31.37 32.65 33.94 29.56
(11,24) (11,79) (12,38) (12,96) (13,54) (14,12) (14,69) (15,27) (13,30)

29.17 30.67 32.17 33.67 35.17 36.66 38.16 39.66 40.51
(13,13) (13,80) (17,48) (15,15) (15,83) (16,50) (17,17) (17,85) (18,23)

16.92 18.56 20.23 22,06 23.56 25.23 26.90 28.56 24.40
(7,61) (8,35) (9,10) (9,93) (10,60) (11,35) (12,11) (12,85) (10,98)

20.77 22.41 24.08 25.91 27.41 29.08 30.75 32.41 28.25
(9,35) (10,08) (10,84) (11,66) (12,33) (13,09) (13,84) (14,58) (12,71)

25.35 26.99 28.66 30.49 31.99 33.66 35.33 37.00 32.84
(11,41) (12,15) (12,90) (13,72) (14,40) (15,15) (15,90) (16,65) (14,78)

29.49 31.19 32.86 34.53 36.20 37.87 39.50 41.20 43.94
(13,27) (14,04) (14,79) (15,54) (16,29) (17,04) (17,78) (18,54) (19,77)

Emerson.com/Rosemount 187

Rosemount DP Level March 2017

Table 85. EFW Extended Flanged Seal Weights Pounds (Kilograms)

Extension length

Pipe size Class 1-in. 2-in. 3-in. 4-in. 5-in. 6-in. 7-in. 8-in. 9-in.

(25 mm) (51 mm) (76 mm) (102 mm) (127 mm) (152 mm) (178 mm) (203 mm) (229 mm)

PN 10/16 19.23 22.07 24.95 27.85 30.73 33.62 36.50 39.39 29.81
(8,65) (9,93) (11,23) (12,53) (13,83) (15,13) (16,43) (17,73) (13,41)
Schedule PN 40
40 PN 63 23.32 26.16 29.05 31.94 34.83 37.71 40.60 43.48 33.90
PN 100 (10,50) (11,77) (13,07) (14,37) (15,67) (16,97) (18,27) (19,57) (15,26)

PN 160 29.83 32.67 35.56 38.45 41.34 44.22 47.11 50.00 40.41
DN (13,42) (14,70) (16,00) (17,30) (18,60) (19,90) (21,20) (22,50) (18,18)
100
37.37 40.21 43.10 45.99 48.88 51.76 54.65 57.53 47.95
PN 16 (16,82) (18,09) (19,40) (20,70) (22,00) (23,29) (24,59) (25,89) (21,58)

EN 1092 - 1 Schedule PN 40 42,48 45.4 48.29 51.17 54.05 56.94 59.82 52.71 66.63
80 PN 63 (19,12) (20,43) (21,73) (23,03) (24,32) (25,62) (26,92) (28,22) (29,98)

PN 100 18.85 21.43 23.98 26.53 29.08 31.66 34.17 36.72 26.81
(8,48) (9,64) (10,79) (11,94) (13,09) (14,25) (15,38) (16,52) (12,06)
PN 160
22.95 25.53 28.07 30.62 33.17 35.75 38.27 40.82 30.90
PN 16 (10,33) (11,49) (12,63) (13,78) (14,93) (16,09) (17,22) (18,37) (13,91)

DN Headbox PN 40 29.46 32.04 34.58 37.13 39.68 42.26 44.78 47.33 37.41
100 PN 63 (13,26) (14,42) (15,56) (16,71) (17,86) (19,02) (20,15) (21,30) (16,83)
PN 100
36.99 39.57 42.12 44.67 47.22 49.80 52.32 84.87 44.95
PN 160 (16,65) (17,81) (18,95) (20,10) (21,25) (22,41) (23,54) (24,69) (20,23)

42.18 44.73 47.30 49.85 52.40 54.94 57.49 60.03 63.62
(18,98) (20,13) (21,29) (22,43) (23,58) (24,72) (25,87) (27,01) (28,63)

19.38 22.40 25.45 28.53 31.55 34.60 37.65 40.67 28.55
(8,72) (10,08) (11,45) (12,84) (14,20) (15,57) (16,94) (18,30) (12,85)

23.48 26.49 29.54 32.63 35.65 38.70 41.75 44.77 32.64
(10,57) (11,92) (13,29) (14,68) (16,04) (17,42) (18,79) (20,15) (14,69)

29.99 33.00 36.05 39.14 42.16 45.21 48.26 51.28 39.15
(13,50) (14,85) (16,22) (17,61) (18,97) (20,34) (21,72) (23,08) (17,62)

37.52 40.54 43.59 46.68 49.69 52.74 55.80 58.81 46.69
(16,88) (18,24) (19,62) (21,01) (22,36) (23,73) (25,11) (26,46) (21,01)

42.68 45.76 48.81 51.86 54.91 57.96 61.01 64.06 68.15
(19,21) (20,59) (21,96) (23,34) (24,71) (26,08) (27,45) (28,83) (30,67)

188 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 85. EFW Extended Flanged Seal Weights Pounds (Kilograms)

Extension length

Pipe size Class 1-in. 2-in. 3-in. 4-in. 5-in. 6-in. 7-in. 8-in. 9-in.

(25 mm) (51 mm) (76 mm) (102 mm) (127 mm) (152 mm) (178 mm) (203 mm) (229 mm)

JIS 40A 10K 6.09 6.55 7.01 7.48 7.94 8.41 8.87 9.33 8.02
20K (2,74) (2,95) (3,15) (3,37) (3,57) (3,78) (3,99) (4,20) (3,61)
50A 40k
10K 6.52 6.98 7.45 7.91 8.38 8.84 9,30 9.33 8.02
Schedule 20K (2,93) (3,14) (3,35) (3,56) (3,77) (3,98) (4,19) (4,20) (3,81)
40 40K
10K 9.64 10.10 10.57 11.03 11.50 11.96 12.43 12.89 11.85
80A 20K (4,34) (4,55) (4,76) (4,96) (5,18) (5,38) (5,59) (5,80) (5,21)
Schedule 40K
80 10K 7.73 8,31 8,91 9,51 10,11 10.70 11.30 11.89 10.67
20K (3.48) (3.74) (4.01) (4.28) (4,55) (4,82) (5,08) (5,35) (4,80)
Schedule 40K
40 10K 7.91 8.49 9.10 9.70 10.29 10.89 11,48 12.07 10,85
20K (3,56) (3,82) (4,10) (4,37) (4,63) (4,90) (5,17) (5,43) (4,88)
100 40K
A 10K 11.18 11.76 12.37 13.00 13.56 14.16 14.75 15.35 14.12
Schedule 20K (5,03) (5,29) (5,57) (5,85) (6,10) (6,37) (6,64) (6,91) (6,35)
80 40K
12.41 14.02 15.63 17.25 18.87 20.49 22.11 23.73 19.52
(5,58) (6,31) (7,03) (7,76) (8,49) (9,22) (9,95) (10,68) (8,78)

15.51 17.12 18.73 20.35 21.97 23.59 25.21 26.83 22.62
(6,98) (7,70) (8,43) (9,16) (9,89) (10,62) (11,34) (12,07) (10,18)

21.92 23.53 25.15 26.77 28.39 30.00 31.62 33.24 29.04
(9,86) (10,59) (11,32) (12,05) (12,78) (13,50) (14,23) (14,96) (13,07)

12.09 13.32 14.63 15.91 17.20 18.49 19.78 21.06 16.68
(5,44) (5,99) (6,58) (7,16) (7,74) (8,32) (8,90) (9,48) (7,51)

15.19 16.42 17.73 19.01 20.30 21.59 22.88 24.16 19.78
(6,84) (7,39) (7,98) (8,55) (9,14) (9,72) (10,30) (10,87) (8,90)

21.60 22.83 24.14 25.43 26.72 28.00 29.29 30.58 26.19
(9,72) (10,27) (10,86) (11,44) (12,02) (12,60) (13,18) (13,76) (11,79)

17.15 19.99 22.87 25.77 28.65 31.54 34.42 37.31 27.73
(7,72) (9,00) (10,29) (11,60) (12,89) (14,19) (15,49) (16,79) (12,48)

22.16 24.99 27.88 30.78 33.66 36.55 39.43 42.31 32.73
(9,97) (11,25) (12,55) (13,85) (15,15) (16,45) (17,74) (19,04) (14,73)

35.21 38.05 40.94 43.83 46.72 49.60 52.49 55.37 45.79
(15,84) (17,12) (18,42) (19,72) (21,02) (22,32) (23,62) (24,92) (20,61)

16.77 19.35 21.90 24.45 27.00 29.58 32.09 34.64 24.73
(7,55) (8,71) (9,86) (11,00) (12,15) (13,31) (14,44) (15,59) (11,13)

21.78 24.36 26.91 29.46 32.00 34.59 37.10 39.65 29.73
(9,80) (10,96) (12,11) (13,26) (14,40) (15,57) (16,70) (17,84) (13,38)

34.83 37.41 39.96 42.51 45.06 47.64 50.16 52.71 42.79
(15,67) (16,83) (17,98) (19,13) (20,28) (21,44) (22,57) (23,72) (19,26)

Emerson.com/Rosemount 189

Rosemount DP Level March 2017

Figure 31. PFW Pancake Seal

A

B D
C
C
G

F

A. Process flange D. Connection to transmitter
B. Flushing connection F. Flushing connection
C. Diaphragm G. Lower housing alignment clamp (option code SA)

Dimensions are in inches (millimeters).

Table 86. PFW Pancake Seal Dimensions

Pipe Class Flange diameter Flange Number Bolt circle Bolt hole Standard diaphragm
size “A” in. (mm) thickness of bolts “C” in. diameter diameter “F” in. (mm)
“B” in. (mm) (mm) “D” in. (mm)

150 6.00 (152) 0.69 (18) 4 4.75 (121) 0.75 (19) 2.30 (58)
6.50 (165) 0.81 (21) 8 5.00 (127) 0.75 (19) 2.30 (58)
300 6.50 (165) 1.00 (25) 8 5.00 (127) 0.75 (19) 2.30 (58)
8.50 (216) 1.50 (38) 8 6.50 (165) 1.00 (25) 2.30 (58)
2-in. 600 9.25 (235) 2.00 (51) 8 6.75 (172) 1.13 (29) 2.30 (58)
7.50 (191) 0.88 (22) 4 6.00 (152) 0.75 (19) 3.50 (89)
ANSI/ASME 900/1500 8.25 (210) 1.06 (27) 8 6.62 (168) 0.88 (22) 3.50 (89)
8.25 (210) 1.25 (32) 8 6.62 (168) 0.88 (22) 3.50 (89)
2500 10.50 (267) 1.50 (38) 8 8.00 (203) 1.25 (32) 3.50 (89)
10.50 (267) 1.88 (48) 8 8.00 (203) 1.25 (32) 3.50 (89)
150 12.00 (305) 2.62 (67) 8 9.00 (229) 1.38 (35) 3.50 (89)
6.50 (165) 0.67 (17) 4 4.92 (125) 0.71 (18) 2.30 (58)
300 7.09 (180) 0.91 (23) 4 5.31 (135) 0.87 (22) 2.30 (58)
7.68 (195) 0.98 (25) 4 5.71 (145) 1.10 (28) 2.30 (58)
3-in. 600 7.87 (200) 0.83 (21) 8 6.30 (160) 0.71 (18) 3.50 (89)
900 8.46 (215) 0.98 (25) 8 6.69 (170) 0.87 (22) 3.50 (89)
9.06 (230) 0.98 (25) 8 7.09 (180) 1.10 (28) 3.50 (89)
1500

2500

DN PN 40
50
EN1092-1 PN 63

PN 100

DN PN 40
80
PN 63

PN 100

190 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 87. PFW Pancake Seal Dimensions

Pipe Outer Inner Beveled Thickness with Thickness with Minimum Weight
size diameter “G” diameter lb (kg)
diameter “L” 1/4 NPT F.C. “M” 1/2 NPT F.C. “M” gasket I.D.
in. (mm) “K” in.
(mm) in. (mm) in. (mm) in. (mm) “N” in. (mm)

ANSI/ASME 2-in. 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.5 (64) 8.61 (3,87)
3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.5 (64) 10.20 (4,59)
EN1092-1 3-in. 3.62 (92) 2.12 (54) 2.48 (63) 0.97 (25) 1.30 (33) 2.5 (64) 11.65 (5,24)
DN 50 3.62 (92) 2.12 (54) 0.97 (25) 1.30 (33) 2.5 (64) 24.84 (11,18)
DN 80 3.62 (92) 2.12 (54) N/A 0.97 (25) 1.30 (33) 2.5 (64) 36.92 (16,61)
5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 16.83 (7,57)
5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 20.88 (9,40)
5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 23.35 (10,51)
5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 33.83 (15,22)
5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 47.39 (19,98)
5.00 (127) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 81.97 (36,89)
4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.5 (64) 10.67 (4,80)
4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.5 (64) 14.24 (6,41)
4.00 (102) 2.40 (61) N/A 0.97 (25) 1.30 (33) 2.5 (64) 16.89 (7,60)
5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 18.76 (8,44)
5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 22.60 (10,17)
5.43 (138) 3.60 (91) N/A 0.97 (25) 1.30 (33) 3.7 (94) 27.07 (12,18)
N/A

Emerson.com/Rosemount 191

Rosemount DP Level March 2017

Figure 32. FCW Flush Flanged Seal – Ring Type Joint (RTJ) Gasket Surface Two-Piece Design
(Shown with Flushing Ring)
D

A

B B
C

A. Process flange C. Flushing connection
B. Diaphragm D. Connection to transmitter

Dimensions are in inches (millimeters).

Table 88. Dimensions for FCW 2-Piece Flange Type Flush Diaphragm Seal

Pipe Flange Flange Bolt circle Bolt hole Overall height Raised face
size diameter “A” thickness “B” diameter “C” diameter “D”
Class “H” height “J”
in. (mm) in. (mm) in. (mm) in. (mm)
in. (mm) in. (mm)
0.75 (19)
ANSI/ASME 150 6.00 (152) 0.69 (18) 4.75 (121) 0.75 (19) 2.43 (62) 0.68 (17)
300 6.50 (165) 0.82 (21) 5.00 (127) 0.75 (19) 2.43 (62) 0.68 (17)
2-in. 600 6.50 (165) 1.00 (25) 5.00 (127) 1.00 (25) 2.43 (62) 0.68 (17)
1500 8.50 (216) 1.50 (38) 6.50 (165) 1.14 (29) 2.57 (65) 0.82 (21)
2500 9.25 (235) 2.00 (51) 6.75 (171) 0.75 (19) 3.07 (78) 0.82 (21)
150 7.50 (191) 0.88 (22) 6.00 (152) 0.88 (22) 2.43 (62) 0.68 (17)
300 8.25 (210) 1.06 (27) 6.62 (168) 0.88 (22) 2.43 (62) 0.68 (17)
600 8.25 (210) 1.25 (32) 6.62 (168) 1.00 (25) 2.43 (62) 0.68 (17)
3-in. 9.50 (241) 1.50 (38) 7.50 (191) 1.25 (32) 2.57 (65) 0.82 (21)
900 10.50 (267) 1.88 (48) 8.00 (203) 1.38 (35) 3.07 (78) 0.82 (21)
1500 12.00 (305) 2.62 (67) 9.00 (229) 4.07 (103) 0.82 (21)
2500

192 Emerson.com/Rosemount

March 2017 Rosemount DP Level

Table 89. Dimensional Table for FCW 2-Piece Flange Type Flush Diaphragm Seal

Pipe RTJ diameter “E” Diaphragm Raised face Inner Thickness with Thickness Weight
size in. (mm) diameter “F” diameter “G” diameter “K” 1/4 NPT F.C. “L” with 1/2 NPT lb (kg)

in. (mm) in. (mm) in. (mm) in. (mm) F.C. “L”
in. (mm)

ANSI/ASME 3.25 (83) 2.30 (58) 4.00 (102) 2.12 (54) 1.40 (36) 1.70 (43) 8.78 (3,95)
3.25 (83) 2.30 (58) 4.25 (108) 2.12 (54) 1.40 (36) 1.70 (43) 10.56 (4,75)
2-in. 3.25 (83) 2.30 (58) 4.25 (108) 2.12 (54) 1.40 (36) 1.70 (43) 12.01 (5,40)
3.75 (95) 2.30 (58) 4.88 (124) 2.12 (54) 1.40 (36) 1.70 (43) 26.81 (12,06)
4.00 (102) 3.50 (89) 5.25 (133) 2.12 (54) 1.40 (36) 1.70 (43) 39.98 (17,99)
4.50 (114) 3.50 (89) 5.25 (133) 3.60 (91) 1.50 (38) 1.80 (46) 16.04 (7,22)
4.88 (124) 3.50 (89) 5.75 (146) 3.60 (91) 1.50 (38) 1.80 (46) 20.72 (9,32)
4.88 (124) 3.50 (89) 5.75 (146) 3.60 (91) 1.50 (38) 1.80 (46) 23.19 (10,44)
3-in. 3.50 (89) 6.12 (155) 3.60 (91) 1.50 (38) 1.80 (46) 35.56 (16,00)
4.88 (124) 3.50 (89) 6.62 (168) 3.60 (91) 1.50 (38) 1.80 (46) 50.72 (22,82)
5.38 (137) 3.50 (89) 6.62 (168) 3.60 (91) 1.50 (38) 1.80 (46) 86.12 (38,75)
5.00 (127)

Figure 33. RCW Flanged Remote Seal Ring Type Joint (RTJ) and Flushing Connection Ring

D

A BC B

A. Process flange C. Flushing connection
B. Diaphragm D. Connection to transmitter
Dimensions are in inches (millimeters).

Emerson.com/Rosemount 193

Rosemount DP Level March 2017

Table 90. RCW Flanged Remote Seal Dimensions

Pipe Class Flange Flange Bolt circle Bolt hole Lower RTJ Lower Overall height Weight
size diameter thickness diameter diameter housing groove housing “H” in. (mm) lb (kg)
diameter
“A” in. “B” in. “C” in. “D” in. inner “F” in. outer No or 1/4-in. 1/2-in. NPT
(mm) (mm) (mm) (mm) diameter (mm) diameter NPT flush flush
connection connection
“E” in. “G” in.
(mm) (mm)

1/2-in. 2500 5.25 1.19 (30) 3.50 (89) 0.88 (22) 0.62 (16) 1.69 (43) 2.64 (67) 2.88 (73) 3.18 (81) 1.49
300/600 (133) 3.18 (81) (0,67)
0.62 (16) 3.25 (83) 0.75 (19) 0.82 (21) 1.69 (43) 2.64 (67) 2.88 (73) 3.18 (81)
3/4-in. 900/1500 4.62 3.18 (81) 5.22
2500 (117) 1.00 (25) 3.50 (89) 0.88 (22) 0.82 (21) 1.75 (45) 2.64 (67) 2.88 (73) 3.18 (81) (2,35)
150 3.18 (81)
300 5.12 1.25 (32) 3.75 (95) 0.88 (22) 0.82 (21) 2.00 (51) 2.90 (74) 2.88 (73) 3.18 (81) 7.45
(130) 3.18 (81) (3,35)
1-in. 600 0.50 (13) 3.12 (79) 0.63 (16) 1.05 (27) 1.88 (48) 2.64 (67) 2.88 (73) 3.18 (81)
900/1500 5.50 3.18 (81) 10.11
2500 (140) 0.62 (16) 3.50 (89) 0.75 (19) 1.05 (27) 2.00 (51) 2.77 (70) 2.88 (73) 3.18 (81) (4,55)
150 3.18 (81)
ANSI/ ASME 300 4.25 0.69 3.50 (89) 0.75 (19) 1.05 (27) 2.00 (51) 2.77 (70) 2.88 (73) 3.18 (81) 4.38
(108) (183) 1.00 (25) 1.05 (27) 2.00 (51) 2.83 (72) 2.88 (73) 3.18 (81) (1,97)
11/2-in. 600 1.12 (29) 4.00 1.00 (25) 1.05 (27) 2.38 (60) 3.27 (83) 2.88 (73)
900/1500 4.88 (102) 5.67
2500 (124) 1.38 (35) 4.25 (2,55)
(108)
4.88 5.95
(124) 0.62 (16) 3.88 (98) 0.63 (16) 1.61 (41) 2.56 (65) 3.27 (83) 2.88 (73) (2,68)

5.88 0.75 (19) 4.50 0.88 (22) 1.61 (41) 2.69 (68) 3.58 (91) 2.88 (73) 10.15
(149) 0.88 (22) (114) (4,57)
1.25 (32) 0.88 (22) 1.61 (41) 2.69 (68) 3.58 (91) 2.88 (73)
6.25 1.75 (45) 4.50 14.55
(159) (114) 1.12 (28) 1.61 (41) 2.69 (68) 3.64 (93) 2.88 (73) (6,55)

5.00 4.88 1.25 (32) 1.61 (41) 3.25 (83) 4.52 2.88 (73) 6.78
(127) (124) (115) (3,05)

6.12 5.75 10.01
(156) (146) (4,50)

6.12 10.90
(156) (4,91)

7.00 16.43
(178) (7,39)

8.00 29.39
(203) (13,23)

194 Emerson.com/Rosemount

March 2017 Rosemount DP Level
Figure 34. FUW Flush Flanged Type Seal - EN1092-1 Type D
C

B

AB C. Connection to transmitter

A. Process flange
B. Diaphragm
Dimensions are in inches (millimeters).

Table 91. FUW Flush Flanged Type Seal Dimensions

Flange Flange Bolt Bolt hole Standard Raised Groove Groove Groove
diameter thickness face O.D. I.D. depth
Pipe Class circle diameter Number diaphragm “J” “K” Weight
size “A” “B” diameter “L” lb (kg)
in. (mm) in. (mm) “C” “D” of bolts diameter “F” “G”

in. (mm) in. (mm) in. (mm) in. (mm)

EN 1092-1 DN PN 6.50 (165) 0.79 (20) 4.92 0.71 (18) 4 2.30 (58) 4.00 (102) 3.46 2.83 0.16 6.29
50 40 (125) 0.71 (18) 8 3.50 (89) 5.43 (138) (88) (72) (4,00) (2,83)

DN PN 7.87 (200) 0.94 (24) 6.30 4.76 4.13 0.16 11.29
80 40 (160) (121) (105) (4,00) (5,08)

Emerson.com/Rosemount 195

Rosemount DP Level March 2017

Figure 35. FVW Flush Flanged Type Seal - EN1092-1 Type C

A

C

B
B

A. Process flange C. Connection to transmitter
B. Diaphragm

Dimensions are in inches (millimeters).

Table 92. FVW Flush Flanged Type Seal Dimensions

Pipe Class Flange Flange Bolt circle Bolt hole Number Standard Groove Tongue Tongue Weight
size diameter thickness “C” in. diameter of bolts diaphragm O.D. I.D. depth lb (kg)
(mm) diameter “J” in. “L” in.
“A” in. “B” in. “D” in. (mm) “K” in. (mm)
(mm) (mm) (mm) “F” in. (mm)
(mm)

EN 1092-1 DN 50 PN 6.50 0.79 (20) 4.92 0.71 (18) 4 2.30 (58) 3.43 2.87 0.18 5.52
DN 80 40 (165) 0.94 (24) (125) 0.71 (18) 8 3.50 (89) (87) (73) (4,50) (2.48)

PN 7.87 6.30 4.72 4.17 0.18 10.01
40 (200) (160) (120) (106) (4,50) (4,50)

196 Emerson.com/Rosemount

March 2017 Rosemount DP Level
Figure 36. RTW Threaded Seal
D
A
B

C

B

A. Upper housing C. Lower housing or flushing connection
B. Diaphragm D. Connection to transmitter

Dimensions are in inches (millimeters).

Table 93. RTW Threaded Seal Dimensions

Rating Overall diameter 'A' in. Overall height “B” in. (mm)

(mm) No or 1/4-in. NPT flush connection 1/2-in. NPT flush connection

2500 psi (173 bar) 3.74 (95) 2.47 (63) 2.82 (72)
5000 psi (345 bar) 3.74 (95) 1.95 (50) 2.31 (59)
10000 psi (690 bar) 4.00 (102) 1.95 (50)
N/A

Table 94. RTW Threaded Seal Weights Pounds (Kilograms)

ANSI/ASME Pipe size 1500 psi 2500 psi 5000 psi Class 103 bar 172 bar 344 bar

EN 1092 - 1 1/4–18 NPT 10.73 6.15 5.72 10000 psi N/A N/A N/A
(4,83) (2,77) (2,57)
3/8–18 NPT 10.72 6.13 5.70 6.95 N/A N/A N/A
(4,82) (2,76) (2,57) (3,13)
1/2–14 NPT 10.67 6.09 5.66 6.93 N/A N/A N/A
(4,80) (2,74) (2,55) (3,12)
3/4–14 NPT 10.62 6.03 5.60 6.89 N/A N/A N/A
(4,78) (2,71) (2,52) (3,10)
1–11.5 NPT 10.52 5.93 5.50 6.83 N/A N/A N/A
(4,73) (2,67) (2,48) (3,07)
1 1/4–11.5 NPT 10.38 5.76 5.33 6.73 N/A N/A N/A
(4,67) (2,59) (2,40) (3,03)
1 1/2–11.5 NPT 10.23 5.61 5.18 6.56 N/A N/A N/A
Parallel thread: G 1/2 A (4,60) (2,52) (2,33) (2,95) 12.93 7.07 6.64
6.41 (5,82) (3,18) (3,00)
DIN 16288 N/A N/A N/A (2,88) 10.67 6.10 5.67
Tapered thread: R1/2 per (4,80) (2,75) (2,55)
N/A N/A N/A
ISO 7/1
N/A N/A

Emerson.com/Rosemount 197

Rosemount DP Level March 2017

Figure 37. HTS Male Threaded Seal

2.03 (52)

A

B

A. Connection to transmitter
B. Diaphragm

Dimensions are in inches (millimeters).

Table 95. HTS Male Threaded Seal Dimensions

Process type Connection Outer diameter Diaphragm diameter Length “C” Overall height Weight lb
size in. (mm) “D” in. (mm) (kg)
“A” in. (mm) “B” in. (mm)
2.50 (63,5) 1.60 (0,72)
ANSI NPT 1-in. NPT 2.03 (51,6) 1.09 (27,9) 1.24 (31,5) 2.50 (63,5) 2.32 (1,04)
ISO 228-1 BSP 11/2-in. NPT 2.36 (59,9) 1.70 (43,2) 1.24 (31,5) 2.50 (63,5) 3.09 (1,39)
2.74 (69,6) 1.90 (48,3) 1.24 (31,5) 2.15 (54,6) 1.48 (0,67)
2-in. NPT 2.03 (51,6) 1.09 (27,9) 0.87 (22,0) 2.24 (56,9) 2.10 (0,95)
G1 BSP 2.36 (59,9) 1.70 (43,2) 0.98 (24,9) 2.50 (63,5) 3.06 (1,38)
G11/2 BSP 2.74 (69,6) 1.90 (48,3) 1.24 (31,5)
G2 BSP

198 Emerson.com/Rosemount

March 2017 Rosemount DP Level
Figure 38. SCW Tri Clamp Seal
A A
1.30 (33)
1.30 (33) 1.30 (33)

B B
A

B

A. Connection to transmitter
B. Diaphragm

Dimensions are in inches (millimeters).

Table 96. SCW Tri Clamp Seal Dimensions

Pipe size Outer diameter “A” O-ring groove diameter “B” Diaphragm diameter “C” Weight
in. (mm) lb (kg)
in. (mm) in. (mm)
0.97 (0,44)
11/2-in. 2.00 (51) 1.72 (44) 1.21 (31) 1.23 (0,55)
2-in. 2.50 (64) 2.22 (56) 1.68 (43) 1.56 (0,70)
3.05 (77) 2.78 (71) 2.07 (53) 1.98 (0,89)
21/2-in. 3.58 (91) 3.28 (83) 2.58 (66) 3.02 (1,36)
3-in. 4.68 (119) 4.35 (110) 3.66 (93)
4-in.

Emerson.com/Rosemount 199

Rosemount DP Level March 2017

Figure 39. SSW Tank Spud Seal

5.22 (133) A

1.43 (36)

B

3.63 (92) C
3.97 (101)

D 0.50 (13)

C. Diaphragm 3.70 (94)
D. Tank spud 4.156 (106)

5.98 (152)

A.Connection to transmitter
B. Extension length

Note
Wetted surfaces of Spud are 32 Ra max.

Dimensions are in inches (millimeters).

Table 97. SSW Tank Spud Seal Dimensions

Pipe size Extension length “A” in. (mm) Weight lb (kg)

4-in. SCH 5 2-in. 2.10 (53) 9.20 (4,14)
6-in. 6.10 (155) 12.66 (5,70)

200 Emerson.com/Rosemount


Click to View FlipBook Version