7 & 8 June 2022
BeneFusion nSP/eSP Syringe Pump Training
BeneFusion nSP BeneFusion eSP
Applicable to nSP ex/nSP Neo Applicable to eSP ex/eSP Neo
2 © 2016 Mindray Confidential
Content
Introduction
Appearance
Technical Specification
Operation
Feature
Installation
Installation Process
Standard Pump
Mass Installation
Customer Training
3 © 2016 Mindray Confidential
Objective:
Be familiar with the nSP/eSP and the specification
Ability to perform the calibration and Installation
4 © 2016 Mindray Confidential
1 Introduction
5 © 2016 Mindray Confidential ✓ Appearance
Technical specification
Operation
Feature
Introduction
BeneFusion nSP/eSP Syringe Pump
Designed to accurately control and monitor the flow of solutions to the patient.
Rated for adults, children and newborns use
Apply to ICU, CCU, PICU, NICU, OR, ER, general ward
WARNING
This product must not be operated by anyone who has not received professional training.
6 © 2016 Mindray Confidential
Appearance——Front View Viewing Window
Stainless steel axle
nSP Electric push & pull box
Alarm Light On-Off
Red and Yellow. Observable STOP
7 inch capacitive
touchscreen
Weight:≤1.7Kg. Size:≤257x 150 x73(mm)
7 © 2016 Mindray Confidential
Appearance——Front View Infusion status
indicator
eSP
Stainless steel axle
Alarm Light
Red and Yellow. Observable Electric push & pull box
3.5 inch capacitive
touchscreen
STOP On-Off
Weight:≤1.6Kg. Size:≤257x 150 x73(mm)
8 © 2016 Mindray Confidential
Appearance——Rear View Speaker
1、DC supply USB Port
2、DOCK communication
3、Nurse Call FAT32 format USB
4、PCA Port 1、upgrade the system
software
Multifunction Interface 2、Export and import brand
library and configuration
9 © 2016 Mindray Confidential 3、Import drug library
4、Export history record
AC Port
Appearance——Bottom View
Placement for stacking pump Placement for pole
Max 3 pumps clamp
10 © 2016 Mindray Confidential
Appearance——Door Open Flange detection sensor
Magnet
Extension tube Syringe Holder
hook
Syringe installation guidance Syringe fixing clamp
label
11 © 2016 Mindray Confidential
Appearance——Accessary The step is designed for the
waterproof
Power
Cord
12 © 2016 Mindray Confidential
Appearance——Accessary
Standard stack Fast clamping clamp (high
structure configuration)
Standard pole Adjust the handle direction to match the
clamp Vertical and horizontal pole
13 © 2016 Mindray Confidential Manual fasten the screw to fix
the device and clamp
SN rule
N series XXX X X X XXXXXX(T)
TCN
Model code Product Serial SG1-0800001T
“-” number
nSP-SG1 Initiation
nSP ex-SG2 Component 000001
nSP Neo-SG3 “A”
Month
nVP-SG4 Year 1-9、
nVP ex-SG5 0-9 A(Oct.)、B、
nVP Neo-SG6 C
nDS-SG7
nDS ex-SG8
nCS-SG9
14 © 2016 Mindray Confidential
SN rule
E series XXX X X X XXXXXX(T)
TCN
Model code Product Month Serial
“-” 1-9、 number
eSP-SH1 A、B、C Initiation
eSP ex-SH2 Component 000001
eSP Neo-SH3 “A”
SG1-0800001T
eVP-SH4 Year
eVP ex-SH5 0-9
eVP Neo-SH6
eDS-SH7
eDS ex-SH8
eCS-SH9
15 © 2016 Mindray Confidential
1 Introduction
16 © 2016 Mindray Confidential Appearance
✓ Technical specification
Operation
Feature
Introduction——Technical Specification
No. Items Specification
1 Syringe ⚫ 1ml、2ml、3ml、5ml、10ml、20ml、30ml、50/60ml
2 Accuracy error ⚫ Accuracy error≤±1.8%,bolus accuracy error ≤±2%或0,05ml
3 Rate ⚫ 1ml:0.01-50ml/h
⚫ 2/3ml:0.01-150ml/h
⚫ 5/6ml:0.01-300ml/h
⚫ 10/12ml:0.1-800ml/h
⚫ 20ml:0.1-1200ml/h
⚫ 30/35ml:0.1-1500ml/h
⚫ 50/60ml:0.1-2300ml/h
4 Infusion Mode ⚫ 9 modes available
⚫ Rate/ Dose/ Time/ Dose Time/ Sequential/ Intermittent/ Loading Dose/ Ramp/
Micro-infusion
⚫ No Loading Dose in Ex Ver. No Ramp Mode in Neo
⚫ TCI/TIVA/PCA Optional。
5 Pressure detection ⚫ 15 level are adjustable
⚫ 50/150/225/300/375/450/525/600/675/750/825/900/975/1050/1125 mmHg
⚫ Unit ,mmHg / Kpa / bar / psi
Note: 50mmHg and 15ul are not accessible in General Option, but in User maintenance
17 © 2016 Mindray Confidential
Introduction——technical Specification
No. Items Specification
7 KVO Rate ⚫ 0.1 ~5.0ml/h,Default:0.5ml/h
8 History Record
9 Drug Library ⚫ Max 3500
10 Audio Alarm ⚫ 5000 drugs
12 Brightness ⚫ 30 Colors
13 Battery run time ⚫ 50 Categories
14 Lock Time
⚫ 3 Sound
⚫ 8 level adjustable
⚫ 8 level adjustable
⚫ Brightness on Battery adjustable
⚫ Normal battery:Run time 5H,charge time5H. 7.2V,2600mAH
⚫ Intelligent battery:Run time 11H,charge time 6H. 10.8V,2900mAH
⚫ Lock Time For No Infusion:OFF/1/2/3/4/5min
⚫ Lock Time In Infusion :OFF/15s/30s/1/2/3/4/5min
18 © 2016 Mindray Confidential
1 Introduction
19 © 2016 Mindray Confidential Appearance
Technical Specification
✓ Operation
Feature
Operation——Turn On
1. Press [On-Off] button。
2. The loading guidance will be displayed on the screen
3. The push & pull box will move out automatically ,and the clamps will be open
20 © 2016 Mindray Confidential
Operation——Loading the syringe
3. Open the door Press this place
4. Pull down the syringe clamp
Press this place
21 © 2016 Mindray Confidential
Operation——Loading the syringe
5. Place the syringe into the syringe slot, ensure that the barrel flange is in the space between the pump
and the flange holder
6. Lift the syringe clamp until it locks the syringe
7. The push box automatically slides left until it reaches the plunger end, the push box automatically
squeeze the plunger flange and clamps hold the plunger.
Keep the scale mark Fool proof
outside If the syringe is not installed correctly,
it will not be detected
pull up the syringe fixing The sensor will detect the status(to touch or not to touch)
clamp
22 © 2016 Mindray Confidential
Operation——Drug & Brand Selection
nSP
If turn off the brand selection,
no this interface;
9:16
Common Other Drug A
10%KCl 10%NaCl 25%MgSO Bolus The pump can be configured
Belladonna with a drug library or a drug info
library
Search
50%Glucose Aminophylline Exit
1/9
23 © 2016 Mindray Confidential
Operation——Drug & Brand Selection Search
eSP
If turn off the brand selection, The pump can be configured
no this interface; with a drug library or a drug info
library
24 © 2016 Mindray Confidential
Operation——Parameter Setting
VTBI Bolus
Infusion Mode Rate Time Start
Syringe type
25 © 2016 Mindray Confidential
Operation——Parameter Setting
Rate VTBI
Time
Infusion Mode Syringe type
Bolus Start
26 © 2016 Mindray Confidential
Operation——Lock&Unlock
Lock Unlock
27 © 2016 Mindray Confidential
Operation——Running
Running progress bar You can modify rate, dose rate, drug name, pressure without stopping
the infusion.
VTBI: The remaining infusion
time is the expected time
No VTBI: the remaining
infusion time is empty time
28 © 2016 Mindray Confidential
Operation——Running
VTBI No VTBI
You can modify rate, dose rate, drug name, pressure without stopping
the infusion.
29 © 2016 Mindray Confidential
Operation——Bolus
Bolus: Manual or Automatic
1、Automatic Bolus Infusion
⚫ Select【Bolus】, the Bolus setting interface will be displayed in the screen.
⚫ Select bolus volume and select 【Bolus】to start a bolus infusion
2、Manual Bolus Infusion
⚫ Press and hold【Bolus】to deliver a bolus infusion
⚫ Bolus will be stopped when you release the【Bolus】or the bolus volume limit (0~20ml) is reached
30 © 2016 Mindray Confidential
Operation——Bolus
Remaining time
Automatic Bolus Infusion
⚫ Select【Bolus】, the Bolus setting interface will be displayed in the screen.
⚫ Select bolus volume and select 【Bolus】to start a bolus infusion
31 © 2016 Mindray Confidential
Operation——Bolus
Manual Bolus Infusion
⚫ Press and hold【Bolus】to deliver a bolus infusion
⚫ Bolus will be stopped when you release the【Bolus】or the bolus volume limit (0~20ml) is reached
32 © 2016 Mindray Confidential
Operation——Turn Off
12. Remove the syringe and turn off 。
➢ The pump cannot be turned off if the syringe is not removed
➢ The plunger holder automatically slides left
Pull down the syringe clamp ,the push box
automatically slider right and release the syringe
33 © 2016 Mindray Confidential
Operation——Alarm
Alarm Priority Alarm Lamp Alarm Lamp Alarm Sound Alarm Message Alarm priority Duty cycle
Color Flashing Frequency Interval indicator
High priority Red 2.0±0.6Hz 5s(±2s) White text or symbol *** 20% to 80%
alarm inside a red box
Not flashing
Low priority Yellow 20s(±2s) Black text or Symbol * 100%
alarm inside a yellow box
High Priority Alarm
Air In Line, Accumulated Air, Empty, Drop Error, Upstream Occlusion, Downstream Occlusion, Infusion Set Disengaged,
No Infusion Tube, Infusion Set Error, No Drop Sensor, Battery Depleted, VTBI Complete, KVO finish, Relay Invalid,
System Error
Low Priority Alarm
KVO Running, Battery in Use, Battery Error, CMS/eGW Disconnected, Standby Time Expired, Dock Connection
interrupted, System Time Error, Relay Invalid soon, Time Near End, Reminder, Low battery
34 © 2016 Mindray Confidential
Operation——Alarm
√ ***Downstrem Occlusion 9:16
10% NaCl Check the infusion set or adjust the limit
Rate Mode Rate 55.0 Peak Pressure:383>=Pressure Threshold:375 Alarm Reset
AudioPause
Regular-Bbraun Pause Pressure Setup
ml/h
Audio Pause Alarm Reset
Touch to silence the alarm sound Touch to acknowledge and reset the alarm
Alarms will be silenced for 2 min,except for Battery depleted √ appears in the front of the alarm message
Alarm sound is silenced
will be displayed in the behind of alarm message
is displayed in the behind of alarm message
If a new alarm is triggered during the audio pause, No alarm sound
35 © 2016 Mindray Confidential
Operation——Alarm
36 © 2016 Mindray Confidential
Operation——Relay & Volume
Relay mode is available in n/e series pump, nDS or eDS is required
9:16
10% NaCl VTBI Vloume
Audio Pause Relay Lock 30Standby .6P4urge 19Volume.36 Menu
5.0 ml ml Bolus
Rate Mode Rate Time Pressure 450
mmHg
Regular-Bbraun Pause 06:07:57 0 Start
ml/h h:m:s
37 © 2016 Mindray Confidential
Operation——Menu
Menu
38 © 2016 Mindray Confidential
Operation——General Option
General Option
39 © 2016 Mindray Confidential
Operation——System Option
System Option
40 © 2016 Mindray Confidential
Operation——User Maintenance
Log in
Password is required, 8868@m
41 © 2016 Mindray Confidential
1 Introduction
42 © 2016 Mindray Confidential Appearance
Technical Specification
Operation
✓ Feature
Feature——Easy to use
UI Friendly
7/3.5 inch Capacitive Touchscreen
Syringe loading Easy
Electric plunder holder system
Alarm Clear
Navigating Alarm Solution
Infusion Smooth
Circular & Customized Relay
43 © 2016 Mindray Confidential
Feature——Safety & Efficiently
Drug Name with different Color, easy to distinguish
Infusion View, easy to observe the treatment
DERS drug library, safety to use the drug
Common drug setting, easy to start the infusion
44 © 2016 Mindray Confidential
Feature——New Technology
⚫ Response Fast
2 seconds to turn on and 3 seconds to start infusion
⚫ Infusion accurate
Accuracy error ±1.8%
⚫ Intelligent occlusion management system
• Occlusion alarm triggered, the pump can automatically restart when it detects the occlusion has released.
• If infusion restart fails 5 times, anti-bolus will help to release the pressure in the tube
⚫ extension Tube detection
45 © 2016 Mindray Confidential
2 Installation
46 © 2016 Mindray Confidential ✓ Installation Process
Standard Pump
Mass Installation
Customer Training
Installation——Installation Process Tool Required
1、USB
2、Maintenance Kit
1. Calibrate 1 pump and 3、knife
configurate the infusion Prepare: Prepare Standard Pump
parameter Collect the IV eSP or nSP
2. Export the configuration set information
information Mass Installation
OPEN BOX System Check Upgrade Brand Configurate Fixing
Information Parameter
Select 1 unit for accuracy confirmation Accuracy
Confirmation
47 © 2016 Mindray Confidential
2 Installation
48 © 2016 Mindray Confidential Installation Process
✓ Standard Pump
Mass Installation
Customer Training
Installation——Prepare Standard Pump
Brand information
Standard Infusion Parameter
pump
Drug Library
eSP and nSP use same standard pump
49 © 2016 Mindray Confidential
Installation——Brand Management
Brand management Take nSP as an example
Menu——User Maintenance——Brand Management
Add or modify the infusion IV set Brand information in system according to the customer requirement
50 © 2016 Mindray Confidential