Installation——Brand Management
Brand management
Common Brand
Select the frequently-used syringe as Common, which will help the nurse to select correctly syringe easily
Add Brand
Edit the Brand name, type and size, Confirm
Delete Brand
Select the unused brands, deleted
Modify Brand
Select the brands, modify the name and type.
Delete the calibration data or not
51 © 2016 Mindray Confidential
Installation——System Calibration
System Calibration
Menu——User Maintenance——System Calibration
Data Review
52 © 2016 Mindray Confidential
Installation——System Calibration
Accuracy Calibration
Turn on
Enter accuracy Calibration interface
Select the brand, type and size
Pull the Syringe piston to the full size
Start the calibration
Calibration successful
53 © 2016 Mindray Confidential
Installation——System Calibration
Pressure Calibration
1. Pressure Calibration.
Menu-User Maintenance-System Calibration
2. Select the correct brand, type and size
3. Connect the infusion tube to the pressure gauge
Purge until no air bubble in the tube
54 © 2016 Mindray Confidential
Installation——System Calibration
Pressure Calibration
4. The 1st pressure point 300 mmHg
Observe the pressure gauge, Stop the calibration when the
pressure value reach 150mmHg, input the value in system
5. The 2nd pressure point 900 mmHg
Observe the pressure gauge, Stop the calibration when the
pressure value reach 500 mmHg, input the value in system
55 © 2016 Mindray Confidential
Installation——System Calibration
Pressure Calibration
6. Pressure Calibration successful
56 © 2016 Mindray Confidential
Installation——Performance Test
Performance Test is required when the calibration is finished
Accuracy Test
Size(mL) VTBI(ml) Rate(mL/h) Time(min) Pass
1 1 20 3 ±1.8%
2 2 40 3
3 3 60 3
5 5 100 3
10 10 200 3
20 10 200 3
30 10 200 3
50 30 600 3
57 © 2016 Mindray Confidential
Installation——Performance Test
Occlusion Test
Size(mL) Occlusion threshold(mmHg) Rate(mL/h) Pass
1 975 475mmhg~1475mmhg
2 50
3 525 50 400mmhg~650mmhg
5 50
10 50
20 50
30 50
50 50
50
58 © 2016 Mindray Confidential
Installation——Performance Test
Empty position confirmation
Use a syringe that passes the accuracy test, draw enough pure water inside the syringe, and push the piston to the
recommended position in the below table.
Set the rate according to the table requirements, but there is no need to set the VTBI.
Start the operation and make sure that the syringe pump emits the emptying alarm time, the syringe piston has run to the
bottom, and there is no visible gap between the piston and the base and the remaining pure water.
When it is found that there are obvious gaps and pure water, it means that the syringe has not been calibrated correctly.
Please re-calibrate the syringe and test the emptying position again
Size(mL) Initial piston position(mL) Rate(mL/h) Estimate emptying time(min)
1 1 50 1.2min
2 2 150 0.8
3 3 150 1.2
5 5 300 1
10 10 800 0.75
20 10 1200 0.5
30 15 1500 0.6
50 25 2300 0.65
59 © 2016 Mindray Confidential
Installation——Export Brand File
After the calibration and common setting are complete, we can export the brand file to USB from
the pump
Export Brand Library
1. Insert the USB to USB port
2. Menu-User Maintenance-Import and Export
3. Edit the brand library file name
4. Export
5. Brand file(BrandFile.bin) will be saved in the USB
60 © 2016 Mindray Confidential
Installation——Export Configuration
Setting the General Option and System option base on the request from customer. Export the
configuration parameter after the setting is complete
Export Configuration
1. Insert the USB to USB port
2. Menu-User Maintenance-Import and Export
3. Edit the configuration file name
4. Export
5. Configuration file(Configuration.vp) will be saved in the USB
61 © 2016 Mindray Confidential
Installation——Create Drug Library
Two types of drug library
➢ Drug information library(SafeDose Info)
➢ Drug library(SafeDose Ders)
Need license in pump
➢ Password:druglib
62 © 2016 Mindray Confidential
Installation——Create Drug Library
Drug Color SafeDose Ders SafeDose Info
Drug capacity
Infusion view 30 30
Default pressure detection 5000 5000
Default infusion mode Yes Yes
Default infusion tube Yes Yes
DERS No Yes
No Yes
Yes No
63 © 2016 Mindray Confidential
Installation——Create Drug Library
Collect the drug information from the end-user, manual input the information to the drug management tool
⚫ Send the collection form to the end-user
⚫ End-user fill out the drug information and sing it
⚫ Engineer manual input the drug information to the drug management tool and create a drug library file
⚫ Upgrade the drug library on the pump
Note: The above procedure is not free of charge
64 © 2016 Mindray Confidential
2 Installation
65 © 2016 Mindray Confidential Installation Process
Standard Pump
✓ Batch installation
Customer training
Installation——Mass Installation
OPEN BOX System Check Upgrade Brand Configurate Fixing
Information Parameter
Fix the device base on
1、Check the device Turn on customer requirement
and Accessary 1、System self-test pass
2、Check the 2、No Alarm
appearance
Accuracy
Confirmation
Select 1 unit for accuracy confirmation
66 © 2016 Mindray Confidential
Installation——Configuration
Configuration Brand File
Infusion
Parameter
Drug library
Note:
The configured file will be removed by the new configuration file
One Brand File/Infusion Parameter/Drug library can be used by nsp and esp
67 © 2016 Mindray Confidential
Installation——Configuration
Save the configuration files to USB root
X:/BrandLib
X:/BatchConfig
X:/DrugLib
Insert the USB to the USB port
Menu-User Maintenance-Import and Export-Import
Select the configuration file
Confirm
Restart the device and check the brand/ drug/ infusion parameter
Follow the above guidance to configure the other pumps which are required to be installed
68 © 2016 Mindray Confidential
Installation——Performance Test
Performance Test is required when the calibration is finished
Accuracy Test
Size(mL) VTBI(ml) Rate(mL/h) Time(min) Pass
1 1 20 3 ±1.8%
2 2 40 3
3 3 60 3
5 5 100 3
10 10 200 3
20 10 200 3
30 10 200 3
50 30 600 3
69 © 2016 Mindray Confidential
Installation——Performance Test
Performance Test is required when the calibration is finished
Occlusion Test Occlusion threshold(mmHg) Rate(mL/h) Pass
975 475mmhg~1475mmhg
Size(mL) 50
525 50 400mmhg~650mmhg
1 50
2 50
3 50
5 50
10 50
20 50
30
50
70 © 2016 Mindray Confidential
2 Installation
71 © 2016 Mindray Confidential Installation Process
Standard Pump
Mass Installation
✓ Customer Training
Installation——Customer Training
Please follow the service training procedure to provide training service. The content should
include the below 3 parts:
Device Operation
Alarm Solution
Cleaning and disinfection
72 © 2016 Mindray Confidential
Installation——Cleaning & Disinfection
Cleaning
To clean the equipment, follow this procedure:
Dampen a soft lint-free cloth with water or ethanol (70%).
Wring excess liquid from the cloth.
Wipe the display screen of the equipment.
Wipe the external surface of the equipment with the damp cloth, avoiding the connectors and metal parts.
Dry the surface with a clean cloth. Allow the equipment air dry in a ventilated and cool place
Disinfection
The procedure is same with cleaning, but the disinfectants is different. For the disinfectants, please refer
to the Operator manual
73 © 2016 Mindray Confidential
Thank You!