Technical Training Syringe Pump (Mindray BeneFusion eSP), JKN NS, 7 & 8 June 2022

02-BeneFusion nSP&eSP Syringe Pump Training_V3.0_EN (1) - JKN NS

Keywords: Technical training

Installation——Brand Management

Brand management

Common Brand

Select the frequently-used syringe as Common, which will help the nurse to select correctly syringe easily

Add Brand

Edit the Brand name, type and size, Confirm

Delete Brand

Select the unused brands, deleted

Modify Brand

Select the brands, modify the name and type.
Delete the calibration data or not

51 © 2016 Mindray Confidential

Installation——System Calibration

System Calibration

Menu——User Maintenance——System Calibration

Data Review

52 © 2016 Mindray Confidential

Installation——System Calibration

Accuracy Calibration

 Turn on
 Enter accuracy Calibration interface
 Select the brand, type and size
 Pull the Syringe piston to the full size
 Start the calibration
 Calibration successful

53 © 2016 Mindray Confidential

Installation——System Calibration

Pressure Calibration

1. Pressure Calibration.

Menu-User Maintenance-System Calibration

2. Select the correct brand, type and size

3. Connect the infusion tube to the pressure gauge

Purge until no air bubble in the tube

54 © 2016 Mindray Confidential

Installation——System Calibration

Pressure Calibration

4. The 1st pressure point 300 mmHg

Observe the pressure gauge, Stop the calibration when the
pressure value reach 150mmHg, input the value in system

5. The 2nd pressure point 900 mmHg

Observe the pressure gauge, Stop the calibration when the
pressure value reach 500 mmHg, input the value in system

55 © 2016 Mindray Confidential

Installation——System Calibration

Pressure Calibration

6. Pressure Calibration successful

56 © 2016 Mindray Confidential

Installation——Performance Test

Performance Test is required when the calibration is finished

Accuracy Test

Size(mL) VTBI(ml) Rate(mL/h) Time(min) Pass
1 1 20 3 ±1.8%
2 2 40 3
3 3 60 3
5 5 100 3
10 10 200 3
20 10 200 3
30 10 200 3
50 30 600 3

57 © 2016 Mindray Confidential

Installation——Performance Test

Occlusion Test

Size(mL) Occlusion threshold(mmHg) Rate(mL/h) Pass
1 975 475mmhg~1475mmhg
2 50
3 525 50 400mmhg~650mmhg
5 50
10 50
20 50
30 50
50 50
50
58 © 2016 Mindray Confidential

Installation——Performance Test

Empty position confirmation

 Use a syringe that passes the accuracy test, draw enough pure water inside the syringe, and push the piston to the
recommended position in the below table.

 Set the rate according to the table requirements, but there is no need to set the VTBI.
 Start the operation and make sure that the syringe pump emits the emptying alarm time, the syringe piston has run to the

bottom, and there is no visible gap between the piston and the base and the remaining pure water.
 When it is found that there are obvious gaps and pure water, it means that the syringe has not been calibrated correctly.

Please re-calibrate the syringe and test the emptying position again

Size(mL) Initial piston position(mL) Rate(mL/h) Estimate emptying time(min)

1 1 50 1.2min
2 2 150 0.8
3 3 150 1.2
5 5 300 1
10 10 800 0.75
20 10 1200 0.5
30 15 1500 0.6
50 25 2300 0.65

59 © 2016 Mindray Confidential

Installation——Export Brand File

After the calibration and common setting are complete, we can export the brand file to USB from
the pump
Export Brand Library

1. Insert the USB to USB port
2. Menu-User Maintenance-Import and Export
3. Edit the brand library file name
4. Export
5. Brand file(BrandFile.bin) will be saved in the USB

60 © 2016 Mindray Confidential

Installation——Export Configuration

Setting the General Option and System option base on the request from customer. Export the
configuration parameter after the setting is complete
Export Configuration

1. Insert the USB to USB port
2. Menu-User Maintenance-Import and Export
3. Edit the configuration file name
4. Export
5. Configuration file(Configuration.vp) will be saved in the USB

61 © 2016 Mindray Confidential

Installation——Create Drug Library

Two types of drug library
➢ Drug information library(SafeDose Info)
➢ Drug library(SafeDose Ders)

Need license in pump

➢ Password:druglib

62 © 2016 Mindray Confidential

Installation——Create Drug Library

Drug Color SafeDose Ders SafeDose Info
Drug capacity
Infusion view 30 30
Default pressure detection 5000 5000
Default infusion mode Yes Yes
Default infusion tube Yes Yes
DERS No Yes
No Yes
Yes No

63 © 2016 Mindray Confidential

Installation——Create Drug Library

Collect the drug information from the end-user, manual input the information to the drug management tool
⚫ Send the collection form to the end-user
⚫ End-user fill out the drug information and sing it
⚫ Engineer manual input the drug information to the drug management tool and create a drug library file
⚫ Upgrade the drug library on the pump

Note: The above procedure is not free of charge

64 © 2016 Mindray Confidential

2 Installation

65 © 2016 Mindray Confidential Installation Process
 Standard Pump
✓ Batch installation
Customer training

Installation——Mass Installation

OPEN BOX System Check Upgrade Brand Configurate Fixing
Information Parameter
Fix the device base on
1、Check the device Turn on customer requirement
and Accessary 1、System self-test pass
2、Check the 2、No Alarm
appearance

Accuracy
Confirmation

Select 1 unit for accuracy confirmation

66 © 2016 Mindray Confidential

Installation——Configuration

Configuration Brand File

Infusion
Parameter

Drug library

Note:
The configured file will be removed by the new configuration file
One Brand File/Infusion Parameter/Drug library can be used by nsp and esp

67 © 2016 Mindray Confidential

Installation——Configuration

 Save the configuration files to USB root
X:/BrandLib
X:/BatchConfig
X:/DrugLib

 Insert the USB to the USB port
 Menu-User Maintenance-Import and Export-Import
 Select the configuration file
 Confirm
 Restart the device and check the brand/ drug/ infusion parameter

Follow the above guidance to configure the other pumps which are required to be installed

68 © 2016 Mindray Confidential

Installation——Performance Test

Performance Test is required when the calibration is finished

Accuracy Test

Size(mL) VTBI(ml) Rate(mL/h) Time(min) Pass
1 1 20 3 ±1.8%
2 2 40 3
3 3 60 3
5 5 100 3
10 10 200 3
20 10 200 3
30 10 200 3
50 30 600 3

69 © 2016 Mindray Confidential

Installation——Performance Test

Performance Test is required when the calibration is finished

Occlusion Test Occlusion threshold(mmHg) Rate(mL/h) Pass
975 475mmhg~1475mmhg
Size(mL) 50
525 50 400mmhg~650mmhg
1 50
2 50
3 50
5 50
10 50
20 50
30
50

70 © 2016 Mindray Confidential

2 Installation

71 © 2016 Mindray Confidential  Installation Process
 Standard Pump
 Mass Installation
✓ Customer Training

Installation——Customer Training

Please follow the service training procedure to provide training service. The content should
include the below 3 parts:

 Device Operation
 Alarm Solution
 Cleaning and disinfection

72 © 2016 Mindray Confidential

Installation——Cleaning & Disinfection

Cleaning

To clean the equipment, follow this procedure:

 Dampen a soft lint-free cloth with water or ethanol (70%).
 Wring excess liquid from the cloth.
 Wipe the display screen of the equipment.
 Wipe the external surface of the equipment with the damp cloth, avoiding the connectors and metal parts.
 Dry the surface with a clean cloth. Allow the equipment air dry in a ventilated and cool place

Disinfection

The procedure is same with cleaning, but the disinfectants is different. For the disinfectants, please refer
to the Operator manual

73 © 2016 Mindray Confidential