MQ60 User manual 50 Graph 63 Experiment management (calibration experiment) Graph64 Main interface calibration setting make calibration test experiments Make experiments as normal process. Detailed method see Wrong! Can’t find quotes. enter calibration dialog and make calibration to the curve. After finishing experiment, clicking “reagent management” through menu or enter reagent management through access shortcut. Switch to” calibration management”page, choose relevant curve, click selection button.


MQ60 User manual 51 QC(only limited to ClassB and above authority) Open QC management Access menu” QC management” can open QC management interface, as graph65QCmanagement Graph 65QC management QC information display Reagent list: exhibit all reagent information of record. QC Batch No.: display all quality control batch No. under the reagent. QC experiment: Data list under reagent list shows all finished QC experiments. Q1~Q3 choice: provide single display function of Q1~Q3 under the condition of present screening. refresh quality control data: click to refresh the current selected reagent quality control chart. Quality Control graph: QC Graph: shows present quality control information under the scope of screening; after clicking some quality control dot, relevant information will choose automaticallyat the information list. Return button Return: return main interface


MQ60 User manual 52 Result query Test results query Graph 66 test results query Query condition: Make test result query through SID and programname assembly, or through screening date to query. Result displays: tested result will show at below data frame under the inputtingcondition Click “up” and “down” button can skip page; or long press “up” to page heading, longpress “down” to page bottom. “up” and “down” button will show when pagemorethan one, or will hide. Query today, quick check today test results information, in order to quick index. Print function: double click today test results, you can acquire “report print” dialog, print report. Date transfer: click “LIS” button, you can send the selected results to sever, LIS will hidewhen there no query information.


MQ60 User manual 53 print report graph 67 print query results graph 68 export query results


MQ60 User manual 54 Click “ today results query”, check the results stay for print. Click “LIS” button, you can send the selected query results to the server. Use short cut key “ctrl+R” can directly export the query results. As graph 68 export query results. Double click that information, will pop out “ print report”. As graph 69 report print. Graph 69 report print “case number” input; input patient case number, if the patient informationsavedbefore, the follow dialog can automatic entering patient information. “name”: input patients name “age”: input patients age “gender” drop-down box: select right gender “sending doctor” drop-down box; if you had selected on setting, it will showthedetail; if not, you need select on you need: detail information you can manager through


MQ60 User manual 55 report information edit. “test doctor” drop-down box: select on you need, detail information you can managerthrough report information edit. “review doctor” drop-down box: select on you need, detail information youcanmanager through report information edit. “diagnosis” input; you can input the diagnosis results. “inpatient area/ bed number” ; input patients inpatient area/bed number. “report date”; test day, can not be modified. “medical history”: input patients medical history. “save” button’ click this button can save the above information. If the patient information saved before will pop out the dialog as graph 70, if select “ok” will coverthe original information, if not will not cover. After save can use print function. Graph 70 patients information confirm


MQ60 User manual 56 Graph 71 report print “print” button; click this button, you can print the test report. ( preconditionthecomputer had connected the printer and can print A5 or A4 paper.)


MQ60 User manual 57 Calibration result query Graph 72 Calibration result query Query condition: make calibration result query through Project name andbatchassembly Results export: Use short cut key “ctrl+R” can directly export the query results. Result display: Calibrated result will show at below data frame under inputtingcondition.


MQ60 User manual 58 result query Graph 73 Calibration result query Query condition: item name and QC batch combination query the results. Results export: Use short cut key “ctrl+R” can directly export the query results. Click “LIS” button, you can send the selected query results to the server. Result display: QC result will show at below data frame under inputting condition. system setting Analyzer setting Graph 74 analyzer setting


MQ60 User manual 59 Can choose Chinese and English, switch the language, click back with immediate effect. Report information can be edit. The reagents list can be setting, to show the choose item. edit report information Graph 75 edit report information Sending doctor/test doctor/review doctor” : you can edit the doctor informationthrough drop-down box. “add”; can add the selected doctor name. “delete: delete the selected doctor. “save”: save the edited information and show those information in report, “direct report”, if choose “yes” when the test finish it will automatic print the test report, or you can print the report manually. “whether send LIS” if choose ‘YES” will automaticsending the test report to “LIS”. “Back”, click can back to analyzer setting.


MQ60 User manual 60 reagents setting Graph 76 reagents setting Available reagents list; display the common use reagents. Unavailable reagents list, display the uncommon use reagents. “Delete”; delete the available reagents, move to uncommon use reagents list. “add”; move the uncommon use reagents to available reagents list. “reset”; clear the available reagents list. operator behavior configuration Operator behavior configuration, are use for configuration the reagents management dependingonoperator using habit.


MQ60 User manual 61 Graph 77 operator behavior configuration Line select: configure SID input way: singe line or multi-line Setting mode: configure multi-line mode single item mode or single specimen mode. Reagents list: display the operator setting available reagents. Specimen type: setting reagents list corresponding default specimen type. “save”; save modified operator habit include SID input mode and reagents default specimen type.


MQ60 User manual 62 communication setting Graph 78 communication setting When communication can’t work, try to connect through “start” button. You can through setting parameter, to adjust present analyzer parameter.


MQ60 User manual 63 System management User management Graph 79 User management Information display: the dialogue shows” present user””password”, “passwordconfirmation” and “authority”. “password” and “password confirmation” will insteadof real character among them; three rank authority, “classA”, “ClassB”, “ClassC”( referto user ID and operation authority rank)。 Addition Button: use to add to new user in system; Revised button: can revise present name and password Cancel button: can cancel present operation, use can’t switch. Confirmation button: make present operation effective. Switch to user button: can switch present login user, click ”confirmation” returnmaininterface, relevant authority will be effective. Return button: return main interface, any operation will not effective.


MQ60 User manual 64 log management Graph 80 user log User can through date and user name combination to query correlation log.


MQ60 User manual 65 Graph 81 wrong log User can through date and user name combination to query correlation log.


MQ60 User manual 66 Graph 82 system log User can through date and user name combination to query correlation log. analyzer information Display analyzer serial number and software version Graph 83 analyzer information


MQ60 User manual 67 system maintenance resource debugging Graph 84 resource debugging This function only admin level can operate The list display all controllable resource operator interface, and current state( movement position, sensor state) This function can not be access during the experiment. All debugged parameters can saved to the data base.


MQ60 User manual 68 function debugging Graph 85 function debugging Temperature monitoring: can monitoring and test incubation sector and magneticseparation sector temperature. Graph 86 Temperature monitoring Barcode scanner: debugging barcode scanner


MQ60 User manual 69 Graph 87 barcode scanner PMT: debug six line reading function Graph 88 PMT debug


MQ60 User manual 70 Self-checking : self-checking will check all the parts whether working properly. Graph 89 self-checking Comprehensive test: though different process to check the analyzer.


MQ60 User manual 71 Graph 90 process debugging


MQ60 User manual 72 data export Graph 91 data export Current results export: export current test results, and save as excel. RLU VALUE: can export current experiment RLU results, save as excel, this functiononlyfor admin level. System explanation Experiment preparation and some basic operation. This chapter introduces some necessary preparation work before experiments and some basic knowledgein some system setting. system main interface Area explanation before connecting power Before connecting power, please confirm below information: Power has connected and inserted rightly. Confirm there is not any screen at front board cover


MQ60 User manual 73 logging in system and initialization Graph 92 appearance equipment open Close front board cover Open general power Turn on computer Open equipment power system open After opening software system, computer screen will show below starting-up pictures: Graph 93 starting up picture System login interface, input user and password, detailed operation see systemlogin. If there is something wrong with communication, system will have belowhint frame.


MQ60 User manual 74 Graph 94 Communication Failed After system login successfully, will initiate automatically and pop-updialogautomatically” system will be initialized” Graph 95 system will be initialized After finishing initializing, system will pop-up “system finish initializing” dialog box. Graph 96 system finish initializing Initializing information: Front board open &closure sensor Status display lamp open PMT mechanism return zero Magnetic separation mechanism return zero


MQ60 User manual 75 sampling mechanism return zero incubator mechanism return zero waste box test and waste TIP test PMT test If there is something wrong with spare parts, system will alarm with buzzer and will remind. If initializing is successful, the “start” and “stop” button will enable. Graph 97 button state ( add reagents) add user Login with the administrator identification and add to user information of relevant authority, seeuser management login with administrator identification. Graph 98 user log (add reagents) add reagents information choose reagent management through shortcut


MQ60 User manual 76 Graph 99 reagents management (add reagents) Manage reagent, click “addition: button Graph 100 reagents management (add reagents) Enter: reagent addition: dialogue; scanner scan barcode to add to reagent Graph 101 scan barcode (add reagents)


MQ60 User manual 77 After adding to all reagent information, need to check whether all reagent are withinperiod of validity. Graph 102 add reagents Click “add batch number” enter reagents batch number add interface youcanmanually add reagents batch number as graph 103, click back, return to “add reagents”(only for class A). Graph 103 add reagents batch number (add reagents)


MQ60 User manual 78 Exit system: click exit when need exit the system Graph 104 exit menu( add reagents) Appear tooltip, click confirmation and finish exiting. Graph 105 exit system and confirm information(add reagent) Close equipment: when system exit, after turning off window, then turnoff equipments. General sample experiment The follow chapter detailed described how to test the patient samples. Login system and finish system initialization Login system and finish system initialization select test line and edit sample information This system version general sample test, line edition is for input sample ID SID, there are threetypes: single line, multi-line one sample, and muiti-line one test item, test item information, reagentsbatch, and specimen type will automatic scan when the experiment begin. Single line model: the operator need separately set each line information, normallyapply in calibration experiment and Quality control experiment. Eachlineinformation automatic input as the graph 107 experiment management( single-linegeneral sample test). how to operate: choose one line, double click enter intoexperiment management edit sample information as graph 106 main interface(sample test).


MQ60 User manual 79 Graph 106 main interface (sample test) Graph 107 experiment management (single line general sample test) Multi-line single specimen model, six line same sample ID SID one time setting, eachline information automatic scan input, normally apply in one patient samplefor Click


MQ60 User manual 80 different test items; how to operate: choose one line, double click enter intoexperiment management interface set sample information. Graph 108 experiment management( mutle-line one sample general sample test) Multi-line single test item model: multi-line different SID one time setting, eachlinereagents information will automatic scan input, normally used for multi-sampledoingsame or different test items. How to operate: chose one line, double click enter intoexperiment management setting sample information, line no. Can use the button“enter” switch to next line.


MQ60 User manual 81 Graph 109 experiment management (multi-line one test item general sample test) scan input SID ( Or manual input) Graph 110 SID Input ( general sample test)


MQ60 User manual 82 Graph 111 select doctor Single specimen mode choose finished, click ok, finish setting, other mode settingsame as single specimen mode setting. Single specimen mode, first line will show the specimen and reagents information. Graph 112 specimen information setting Delete line information: at main interface click right button select delete line or deleteall. Delete line means delete all selected line information, delete all means deleteall interface line information. Graph 113 line information delete


MQ60 User manual 83 instrument and specimen preparation safety protection The operator should wear safety protection, in order to avoid hazard of reagents andspecimen expose. install waste box Before preparing specimen and reagent, should check whether there are left disposal tips; if so, bring the box out, and take out the disposal tips and re-install waste box. Attention: experiments 20 times or one week should change the waste box. Graph 114 install the waste box specimen preparation Make sure at the adding specimen process no bubble occur, add 100ul specimen. Install reagents strips. Take out reagent from refrigerator, should make experiments at once; avoid exposingreagent in normal air long time. Should install reagent strip on the reagent cassette rack as following order, shouldnote below few points when installing: o When taking reagent cassette, take the edge of reagent cassette, avoidtouching aluminum seal probe slot, touch aluminum seal or probe slot will cause the test results difference. o Attention: avoid other kind of liquid touching probe slot, which will causethe test results difference. o Do not use the reagents fallen in the floor, that will cause the test resultsdifference. o Before testing, Inverting reagent lightly at horizon, eliminate bubble at thewall of reagent cassette or liquid inside the aluminum seal reagent box


MQ60 User manual 84 Graph 115 Insert reagent tray into reagent slot Use pipette add specimen to reagent cassette specimen slot as requirements (100ul). Eliminate bubble before testing. Graph 116 allocate demo to specimen slot Open instruments front cover, install reagent cassette rack in to right position. Specimenslot


MQ60 User manual 85 Graph 117 set reagent cassette rack Install disposal tips Use MQ60 special design disposal tips. Graph 118 single used tips Take out new tips from tip box, install in the hole of fine head rack, and aimat thelinewhich reagent cassette installs. Caution tip bottom can not be polluted or damaged.


MQ60 User manual 86 Graph 119 place tips Place the tips into the rack, and close the front cover. Graph 120 close front cover


MQ60 User manual 87 Start test and test finished start test main interface setting Graph 121 main interface (general sample experiment) Press “start” button or click software start button to begin test; The reagents basicinformation will asthmatically displayed in the main interface. Attention: After finishing preparation, please start to make experiments at once. Precipitated red cells in whole blood and specimen vaporized will influence results. Make sure not to do any edition to well-set specimen information. After test finished, detected specimen concentration will indicate at relevant lineinformation. test finished Abandon the tips Take out tips from waste box, throw away used tips please followthelocal management rules. Return the waste box into position. Attention: experiments 20 times or one week should change the waste box.


MQ60 User manual 88 Graph 122 waste box dispose the waste cassette take out reagent cassette rack from equipment, remove used reagent cassette. Paymore attention not to make reagent dropped or flashed fromreagent cassette. Throw away used cassettes please follow the local management rules. Calibration test The follow chapter detailed described how to test the Calibration reagents. Calibration purpose In order to get a credible result, we need proceed calibration experiments to confirmcalibrationcurve, before the patients specimen experiments. Standard curve or calibration curve exceed thelimit time need re-calibration, before the calibration experiment please make sure all reagents and calibratornot expire. About calibration calibrator(CAL) Use recommended calibrator, frozen calibration must firstly handle with diluent andbuffer in ready. Repeat calibration make repeat calibration experiment(general twice) CAL starting time Bring new experiment project Reagent batch change


MQ60 User manual 89 CAL expire QC over control scope Even the same batch, need to use new purchased reagent Optics/after maintaining drop and add system CAL validity According to different projects, CAL has different date of validity; please refer to packageexplanation in details. Attention CAL experiments should make at the same batch reagent and experiment project. During double holes testing, calibration can choose access to make experiments; but calibration variety and quantity must meet experiments requirements. During making new experiment project or new batch reagent experiment, inputtingbarcode information including reagent cassette. calibration valid The same calibration dot, different test line, difference is equal or greater thanplus-minus 30% Low concentration test value higher than higher concentration. There is 20% difference between new test value and original value. When above situation occurred, confirm this calibration experiment failed. Check calibration status Before making calibration, input calibration information into system through scanning reagent barcode。Click “Reagent management” button at main interface of system (as graph main interface (calibrationexperiment)) Select the reagents kind and relevant batch no. Which you need check, at ‘reagents management’, afterchoosing, switch label page to “calibration management” ”calibration management” ; Throughchecking whether show calibration under the batch No., distinguish the reagent whether registeredcalibration in system. If no relevant calibration, before making calibration, should register calibration, details refer to add to reagent information.


MQ60 User manual 90 Graph 123 main interface (calibration experiment) Graph 124 calibration management (calibration experiment)


MQ60 User manual 91 select line and edit calibration information select one line, double click enter experiment management edit calibrationinformation. Graph 125 main interface (calibration experiment) Select specimen type


MQ60 User manual 92 Graph 126 specimen type select (calibration experiment)


MQ60 User manual 93 Graph 127 Assay management (calibration experiment) choose cal1/cal2


MQ60 User manual 94 Graph 128 choose cal1/cal2 (calibration experiment) Finish chose calibrator, click ok, finish setting. In sequence set each line information, as graph 129 calibration information setting


MQ60 User manual 95 Graph 129 calibration information setting Instruments and calibrator preparation install waste box Before preparing specimen and reagent, should check whether there are left disposal tips; if so, bring the box out, and take out the disposal tips and re-install waste box. Attention: experiments 20 times or one week should change the waste box. Graph 130 install the waste box


MQ60 User manual 96 calibrator preparation The calibrator will provide with the reagents. Use the calibrator please refer to the user manual. During preparation the reagents, please follows below precaution description. Wear gloves and mask avoid other reagent or specimen pollution. After taking out reagent cassette and make sure doing the experiments immediately, do not put reagent cassette at room temperature for a long time. Follows precaution description at package instruction. Install reagents strips. Take out reagent from refrigerator, should make experiments at once; avoid exposingreagent in normal air long time. Should install reagent strip on the reagent cassette rack as following order, shouldnote below few points when installing: o When taking reagent cassette, take the edge of reagent cassette, avoidtouching aluminum seal probe slot, touch aluminum seal or probe slot will cause the test results difference. o Attention: avoid other kind of liquid touching probe slot, which will causethe test results difference. o Do not use the reagents fallen in the floor, that will cause the test resultsdifference. o Before testing, Inverting reagent lightly at horizon, eliminate bubble at thewall of reagent cassette or liquid inside the aluminum seal reagent box Graph 131 Insert reagent tray into reagent slot Use pipette add specimen to reagent cassette specimen slot as requirements (100ul). Eliminate bubble before testing.


MQ60 User manual 97 Graph 132 allocate demo to specimen slot Open instruments front cover, install reagent cassette rack in to right position Graph 133 set reagent cassette rack Install disposal tips Use MQ60 special design disposal tips. Specimenslot


MQ60 User manual 98 Graph 134 single used tips Take out new tips from tip box, install in the hole of fine head rack, and aimat thelinewhich reagent cassette installs. Caution tip bottom can not be polluted or damaged. Graph 135 place tips Place the tips into the rack, and close the front cover.


MQ60 User manual 99 Graph 136 close front cover Start test and test finished start test main interface setting