MQ60 User manual 100 Graph 137 Main interface (calibration experiment) Press “start” button or click software start button to begin test. Make sure not to do any edition to well-set specimen information. Test finished If calibration successful, the main interface bottom will show calibration success, or will show calibration fail. Graph 138 calibration successfully (calibration experiment) Solve the fault, must re-calibration. Calibration success, the current batch standard curve will replaced by the calibrationcurve. User can check through standard curve. Abandon the tips Take out tips from waste box, throw away used tips please followthelocal management rules. Return the waste box into position. Attention: experiments 20 times or one week should change the waste box. Graph 139 waste box dispose the waste box Take out reagent cassette rack from equipment, remove used reagent cassette. Paymore attention not to make reagent dropped or flashed fromreagent cassette.


MQ60 User manual 101 Throw away used cassettes please follow the local management rules. QC Products Test The follow chapter detailed described how to test the Quality Control reagents. Introduction Refer to experiment preparation and basic operation, after experiment preparation, proceed theQCexperiments as follow procedures. Quality Control purpose Quality control experiments are aim to guarantee the quality, check the calibration curve and storeddata, in order to ensure the sample result effective, quality control experiments are necessary. About QC experiment QC varieties Now the MQ60 system use Q1/Q2/Q3 three varieties. When need to proceed QC experiment QC designated time overdue After calibration experiments Suspect specimen results difference After maintaining optics/dropping system The requirements of Quality control QC DATA validity QC data validity are variety of test items, please refer to the reagents instructions. Select lane and edit QC information select one line, double click edit QC info.


MQ60 User manual 102 Graph 140 main interface (QC experiment) select specimen type


MQ60 User manual 103 Graph 141 specimen type select (QC experiment) After choosing, experiments management interface will become as follow


MQ60 User manual 104 Graph 142 experiment management (QC experiment) Input QC No.(also can input by manual) Graph 143 QC No.(QC experiments) Choose QC1/QC2/QC3


MQ60 User manual 105 Graph 144 choose QC11/QC2/QC3(QC experiment) After choosing, click confirmation button, finish setting. The line will show well set specimen and reagent information. Graph 145 QC information setting Instrument and QC reagents preparation install waste box Before preparing specimen and reagent, should check whether there are left disposal tips; if so, bring the box out, and take out the disposal tips and re-install waste box. Attention: experiments 20 times or one week, should change the waste box.


MQ60 User manual 106 Graph 146 install the waste box QC products preparation As the general precaution instruction prepare the QC products. During preparation the reagents, please follows below precaution description. Wear gloves and mask avoid other reagent or specimen pollution. After taking out reagent cassette and make sure doing the experiments immediately, do not put reagent cassette at room temperature for a long time. Follows precaution description at package instruction. Install reagents strips. Take out reagent from refrigerator, should make experiments at once; avoid exposingreagent in normal air long time. Should install reagent strip on the reagent cassette rack as following order, shouldnote below few points when installing: o When taking reagent cassette, take the edge of reagent cassette, avoidtouching aluminum seal probe slot, touch aluminum seal or probe slot will cause the test results difference. o Attention: avoid other kind of liquid touching probe slot, which will causethe test results difference. o Do not use the reagents fallen in the floor, that will cause the test resultsdifference. o Before testing, Inverting reagent lightly at horizon, eliminate bubble at thewall of reagent cassette or liquid inside the aluminum seal reagent box.


MQ60 User manual 107 Graph 147 Insert reagent tray into reagent slot Use pipette add specimen to reagent cassette specimen slot as requirements (100ul). eliminate bubble before testing. Graph 148 allocate demo to specimen slot Open instruments front cover, install reagent cassette rack in to right position Specimenslot


MQ60 User manual 108 Graph 149 set reagent cassette rack Install disposal tip Use MQ60 special design disposal tips. Graph 150 single used tips Take out new tips from tip box, install in the hole of fine head rack, and aimat thelinewhich reagent cassette installs. Caution tip bottom can not be polluted or damaged.


MQ60 User manual 109 Graph 151 place tips Place the tips into the rack, and close the front cover. Graph 152 close front cover Start test and test finished start test main interface setting


MQ60 User manual 110 Graph 153 Main interface (QC experiment) Press “start” button or click software start button to begin test; The reagents basicinformation will asthmatically displayed in the main interface. Attention: After finishing preparation, please start to make experiments at once. Precipitated red cells in whole blood and specimen vaporized will influence results. Make sure not to do any edition to well-set specimen information. After test finished, detected specimen concentration will indicate at relevant lineinformation. test finished Click menu “QC management” button Graph 154 menu QC management (QC experiment) Appear quality control management interface Through choosing QC batch under relevant reagent, QC variety, we can screentheQCexperiments. Abandon the tips Take out tips from waste box, throw away used tips please followthelocal management rules. Return the waste box into position. Attention: experiments 20 times or one week, should change the waste box.


MQ60 User manual 111 Graph 155 QC management (QC experiment) Graph 156 waste box dispose the waste box take out reagent cassette rack from equipment, remove used reagent cassette. Paymore attention not to make reagent dropped or flashed fromreagent cassette. Throw away used cassettes please follow the local management rules. Test data data QC test