MXR-451HP/11
Product Description
The MXR-451HP/11 is a bipolar oil-cooled
X-Ray tube assembly with integrated radiati-
on protection. It comprises dual focal spot, a
high power tungsten anode and a directional
beam.The tube is specifically designed for
non-destructive imaging applications.
Product Specifications
Nominal tube voltage..............................................................................................................450 kV
Continuous rating..................................................................................................... 700 W | 1500 W
Focal spot acc. EN 12543.........................................................................d = 0.4 mm¹) | d = 1.0 mm
Focal spot acc. ASTM E1165-12................................................................................................... N/A
Filament current, max.................................................................................................... 4.1 A | 4.2 A
Filament voltage, typical.................................................................................................2.3 V | 6.2 V
Inherent filtration...................................................................................................3 mm + 2 mm Be
Target material................................................................................................................................ W
Target angle....................................................................................................................................11°
Radiation coverage.............................................................................................................. 40° x 30°
Leakage radiation, max. at loading factors in 1m distance..................5 mSv/h (450 kV; 3.3 mA)
Weight.........................................................................................................................................95 kg
Terminal type............................................................................................................................... R28
Gapping spring-loaded HV-cable............................................................... 2 rings visible (~7 mm)
Gapping non-spring-loaded HV-cable.............................................................................5.5 - 6 mm
Grease quantity for HV-cable terminal....................................................................................1.4 ml
1) Deviation from EN 12543-2: evaluation of the focal spot based on 25% threshold
COMET AG, Herrengasse 10, CH-3175 Flamatt Part No. 915369.51 Page 1
T +41 31 744 90 00, www.comet-xray.com Version 6, 12/2014
Power rating charts Connection of filament (C = Common)
Power rating at given filament current
20 mA
18 1500 W
16
14
12
10
8
6
4 700 W
2
0
0 100 200 300 400 500
kV
Cooling
Cooling medium............................................................................................................................ Oil
Cooling medium flow, min.................................................................................................. 14 l/min
Cooling medium temperature at inlet, max............................................................................. 50°C
Pressure at cooling medium inlet, max................................................................................. 10 bar
Post-cooling time after switchoff, min.................................................................................... 2 min
COMET AG, Herrengasse 10, CH-3175 Flamatt Part No. 915369.51 Page 2
T +41 31 744 90 00, www.comet-xray.com Version 6, 12/2014
Outline drawing R1/2"
4 x M8 847
130 ±0.2 ~28
756.5 Angriffspunkt 110 ±0.15
355 Mounting point
R28
Ausstrahlwinkel 10° 30
Emission angle 4x M6/14 deep 110 ±0.15
20°
210 ±0.3 Angriffspunkt
Mounting point
R28
Angriffspunkt
Mounting point
Ausstrahlwinkel Öl aus
Emission angle Oil out
40° Erdung
80 ±0.15 Ground
160
292
4x M5/11 deep Masse ohne Toleranzangabe sind nur zur Information.
Dimensions without tolerances are for information only.
Öl ein Die radiographischen Kenndaten sind dem Prüfschein
Oil in des Strahlers zu entnehmen.
For radiographic characteristics of the tube refer to
299 Final Test Report.
Gegenstand Menge Material Modell Bemerkung
08.0040
Änderungen 1 07.0094 02 Artikel Nr. 915369.51
Auftrag Nr. (H07.0094)
Artikelname Massstab Gezeichnet 08.04.2008 dano
Strahler MXR-451HP/11COMET AG, Herrengasse 10, CH-3175 Flamatt1:6Part No. 915369.51 Geprüft 13.10.2008 plal
Tube-assemblyT +41 31 744 90 00, www.comet-xray.com Freigegeben 14.11.20P0a8gem3ima
Version 6, 12/2014
Zeichnungs Nr. Blatt Nr.: 1 Herrengasse 10
MXR-451HP/11 bipolarer Röntgenstrahler Revisions No. 4
Bestellnummer 915369.51 Date of release
07/2012
COMET AG Herrengasse 10, CH-3175 Flamatt Page 4 of 5
Tel. +41 31 744 9000, www.comet-xray.com Page 4
COMET AG, Herrengasse 10, CH-3175 Flamatt Part No. 915369.51
T +41 31 744 90 00, www.comet-xray.com Version 6, 12/2014
General information
General Cooling
The control of the high voltage and the The customer is in charge that the cooling
filament current as well as the design circuit is properly connected to the cooling
of the cooling system and the radiation system. Prior to operating the tube, the
protection are within the responsibility coolant must be turned on. Make sure that
of the original equipment manufacturer the coolant flow meets the required cooling
(OEM). For detailed instructions refer to conditions. When the tube is switched off,
our „X-RayTube Manual“. Provisions in the the coolant flow must continue for at least
generator must assure that the X-Ray tube 2 minutes in order to protect the anode and
will be protected against overcurrent, over- the lead protection from overheating.
voltage and lack of cooling. Otherwise the
tube and/or the radiation protection may
be damaged and become a hazard.
Connection of the X-ray tube Radiation protection
High-voltage cables must be installed by When installing an X-ray tube assembly into
trained personnel only. Please proceed a X-ray equipment and/or operating an X-ray
with the installation as follows: Cleaning, equipment, the responsibility for radiati-
gapping & greasing (see values page 1), on protection is with the user. Compliance
assembly and maintaining. Detailed inst- with local regulatory requirements and limit
ructions can be found in our „X-RayTube values must be assured. After each tube
Manual“ exchange, repair, modification or upgrade
of the unit a radiation protection integrity
should be performed. Do not modify the
tube assembly itself!
Return of Goods to COMET
Prior to shipment to COMET a Field Failure Report (FFR) will be required.The completed
FFR has to be be added to the shipping documents.The FFR can be downloaded from:
www.comet-xray.com/Service
COMET AG, Herrengasse 10, CH-3175 Flamatt Part No. 915369.51 Page 5