9.18 Y sensor | Technical data | 9 |
Y sensor • Y sensor module: W 154 x L 90 x H 21 mm
Size • Y sensor: W 18 x L 50 x H 27 mm
Weight • Y sensor module: 280 g
Connectors and cables • Y sensor: 11 g
• 15 mm male and female conical connector on flow sensor
Y sensor material
Power source according to ISO 5356-1
Power consumption • Pressure port on module
Y sensor measuring method
Parameters Pressure line, 2.0 m, phthalate free PVC
• Flow sensor cable, 2.0 m
Measuring range • Single use: PC, Polycarbonate
Y sensor resistance • Reusable: PEI, Polyetherimide or PSF, Polysulfone
Dead space Powered from the ventilator system
Pressure line connector
≤ 4.5 W, during normal operation
Hot Wire Anemometer (HWA)
• Airway flow
• Airway pressure
• Inspiratory volume
• Expiratory volume
• Trigger
• End inspiration
• Flow: 0.12 to 32 l/min
• Pressure: -40 to 120 cmH2O
10 cmH2O/l/s at 30 l/min
≤ 1 ml
Gable mounted bulk head connector to fit tubings with an
inner diameter of 3-4 mm
SERVO-U Ventilator System v2.0, User's Manual 201
| 9 | Technical data |
9.19 Communication/interface
Communication/interface Isolated
Serial ports RS-232C. For data communication via the Servo
Communication Interface (SCI)
Servo Communication Interface (SCI) Information regarding connector wiring is available from
Alarm output connection Maquet.
Data Transfer via USB port A protocol for data communication with external devices
VGA port
Isolated
Ethernet port 4-pin modular connector for communication of all active alarms
Switching capability: Max 40 V DC, Max 500 mA, Max 20 W
Information regarding connector wiring is available from
Maquet.
Non-isolated
For transfer of trends, logs, screen shots and recordings to a
USB memory stick
Non-isolated
External monitors should be isolated from the ventilator
system. This can be achieved if the connected monitor is
powered via a medical grade transformer.
The secondary monitor output is not to be relied on for alarms.
Isolated
The network connection (LAN) port is for service use, and
should only be used by personnel trained and authorized by
Maquet.
Connection of the ventilator system to other
equipment through the communication
interfaces, forming a medical electrical system,
could result in previously unidentified risks to
patient, users or third parties.
The responsible organization should identify,
analyze, evaluate and control these risks.
Subsequent changes to the medical electrical
system could introduce new risks and require
additional analysis.
Changes to the medical electrical system
include configuration changes, connection of
additional items, disconnection of items,
update or upgrade of connected equipment.
202 SERVO-U Ventilator System v2.0, User's Manual
| Technical data | 9 |
9.20 Accessories 13.0 kg
W 647 x L 547 x H 557 mm
Mobile cart (option) 5000 running hours
Weight
Dimensions
Preventive maintenance interval
Humidifier holder (option) 0.5 kg
Weight W 243 x L 38 x H 185
Dimensions 5 kg
Maximum load
Support arm 178 (option)
Weight 2.5 kg
Dimensions
Dimensions Length 900 mm
Maximum load • 1 kg at 180°
• 1.5 kg at 90°
• 3 kg at 45°
Refer to Support arm 178 Installation Instructions.
Note: When the knob on the support arm is loosened it also releases the lock to the column.
Water bag/IV pole (option) 0.4 kg
Weight W 148 x L 26 x H 1007
Dimensions 1.5 kg
Maximum load
Drawer (option) 0.6 kg
Weight W 247 x L 118 x H 302
Dimensions
Gas cylinder restrainer kit (option)
Make sure that the gas cylinder restrainer straps are placed on the middle of the gas cylinders.
Weight 0.5 kg
Dimensions Upper: W 104 x L 65 x H 48
Lower: W 106 x L 162 x H 76
Maximum load Two 4.5-liter bottles
SERVO-U Ventilator System v2.0, User's Manual 203
| 9 | Technical data |
Pendant/bed holder (option) 3.2 kg
Weight W 302 x L 302 x H 393
Dimensions
Shelf base (option)
Make sure that the shelf base is securely fixed on the table or shelf.
Weight 2.5 kg
Dimensions W 207 x L 302 x H 43
Y piece holder (option) W 26 x L 52 x H 46
Dimensions
User interface holder (option) 0.6 kg
Weight W 46 x L 90 x H 123
Dimensions
Cable holder for handle (option) 0.1 kg
Weight W 138 x L 92 x H 155
Dimensions 5 kg
Maximum load
Other accesssories (option) Refer to the Compressor Mini User's Manual.
Compressor Mini
Expiratory heater, Servo Duo Guard Refer to the Expiratory heater, Servo Duo Guard User's
Manual.
Servo Duo Guard
Servo Guard Refer to the Servo Duo Guard User's Manual.
Refer to the Servo Guard User's Manual.
204 SERVO-U Ventilator System v2.0, User's Manual
| Technical data | 9 |
9.21 Health and Environment
9.21.1 Pollution control
This product complies with environmental protection use period as defined in People's Republic
of China Electronic Industry Standard SJ/T11364-2006.
Toxic or hazardous substances will not leak
or mutate under normal operating conditions
for 50 years.
9.21.2 Hazardous substances
The following table shows the names and contents of toxic or hazardous substances in this
product as defined in People`s Republic of China Electronic Industry Standard SJ/T11364-2006.
Hazardous substances
Parts Pb Hg Cd Cr6+ PBB PBDE
Metal parts 000000
Plastic and polymeric parts 0 0 0 0 0 0
Electrical components 000000
LCD display 000000
0: Indicates that this toxic or hazardous substance contained in all of the homogeneous
materials for this part is below the limit and meets the requirement in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the
homogeneous materials used for this part is above the limit requirement in SJ/T11363-2006.
SERVO-U Ventilator System v2.0, User's Manual 205
| 9 | Technical data | Global standard for identifying Medical Equipment,
example:
9.22 UDI Label (01)07325710000007(11)140625(21)01311141
UDI Label Each UDI number can be divided into several parts,
Unique Device Identification number each referred to by their AI number '(#)'.
Application Identifier (AI) GTIN - Global Trade Item Number
(01) Part number
(241)
(10) Batch no.
(11)
(17) Manufacturing date (YYMMDD)
(20)
(21) Exp. date (YYMMDD)
(30)
The GTIN consists of four parts: Revision
a. Package level
b. GS-1 company prefix Serial number
c. Item reference
d. Check digit Count of items
0 732571 000021 2
ab cd
206 SERVO-U Ventilator System v2.0, User's Manual
| Technical data | 9 |
9.23 Information required by ISO 80601-2-12
9.23.1 Volume-controlled breath type
Volume-controlled breath type
Characteristics valid for patient categories and circuit configurations intended to provide the indicated
tidal volume, at conditions specified in ISO 80601-2-12, table 201.103.
Maximum inaccuracy of inspired tidal • Neonatal: ±(1 ml + 7 % of set volume) for VT ≤50 ml
volume (VT) • Pediatric: ±(4 ml + 7 % of set volume) for
50 ml ≤ VT ≤300 ml
• Adult: ±(4 ml + 7 % of set volume) for VT ≥300 ml
Maximum inaccuracy of PEEP • Neonatal:
±(1 cmH2O + 5 % of set value) for 20 < VT ≤50 ml
±(1.5 cmH2O + 5 % of set value) for VT ≤20 ml
• Pediatric: ±(1 cmH2O + 5 % of set value) for
50 ml ≤ VT ≤300 ml
• Adult: ±(1 cmH2O + 5 % of set value) for VT ≥300 ml
Maximum inaccuracy of inspired oxygen ±(3 % + 0 % of set value)
concentration (FiO2) at the patient
connection port
9.23.2 Pressure-controlled breath type
Pressure-controlled breath type
Characteristics valid for patient categories and circuit configurations intended to provide the indicated
tidal volume, at conditions specified in ISO 80601-2-12, table 201.104.
Maximum inaccuracy of airway pressure ±(1 cmH2O + 7 % of set value)
(PAW) at the end of the inspiratory phase
Maximum inaccuracy of PEEP ±(1 cmH2O + 5 % of set value)
Maximum inaccuracy of inspired oxygen ±(3 % + 0 % of set value)
concentration (FiO2) at the patient
connection port
9.23.3 O2 concentration response time • Maximum 35 s at VT = 150 ml, for patient
categories and circuit configurations
Characteristics valid at conditions specified intended to provide 50 ml ≤ VT ≤ 300 ml
in ISO 80601-2-12, table 201.105.
• Maximum 55 s at VT = 30 ml, for patient
Response time for oxygen concentration categories and circuit configurations
to change from 21 % to 90 % intended to provide VT ≤ 50 ml
• Maximum 30 s at VT = 500 ml, for patient
categories and circuit configurations
intended to provide VT ≥300 ml
SERVO-U Ventilator System v2.0, User's Manual 207
| 9 | Technical data |
9.24 Technical description
The technical description is intended for the
responsible organization and service
personnel.
Topic Information
Filtering Refer to section Monitoring on page 179.
Detachable parts Refer to section Gas flow through the Patient unit on page 22.
Start and end the inspiratory phase Refer to chapter Ventilation modes and therapies on page 63.
Automatic check of alarm system. Refer to section Pre-use check on page 38.
Measurement uncertainty for disclosed Refer to sections Inspiratory channel on page 177, Expiratory
tolerances channel on page 178 and Monitoring on page 179.
Safe operation Refer to section Safety guidelines on page 8.
Transport and storage Refer to section Transport on page 35.
Measures or conditions for installing the Information regarding installation is available from Maquet.
ventilator system.
Operation overview Refer to chapter Operation overview on page 37.
Safety signs and symbols Refer to sections Symbols on patient unit on page 20,
• Marking on equipment Symbols on User Interface on page 31 and Symbols on
• Consult accompanying documents accessories and packaging on page 33.
• Mechanical stability
• Protective packaging
Identification of the ventilator system and Refer to sections SERVO-U Ventilator System v2.0 on page 1
software version and Version and Configurations on page 13.
Power sources Refer to section System on page 170.
IP classification Refer to section System on page 170.
Applied part — type of classification Refer to sections Symbols on accessories and packaging on
page 33 and System on page 170.
Mode of operation Refer to section System on page 170.
Fuses Refer to sections Operation overview on page 37 and System
on page 170.
External pressure source Refer to section Ventilator system on page 173.
Modification of the ventilator system Do not modify or remove any original parts.
208 SERVO-U Ventilator System v2.0, User's Manual
| Technical data | 9 |
Topic Information
Service and installation Refer to section Service on page 195.
• Qualifications for service personnel
• Replacement of parts Isolate the ventilator system from mains power by
• Installation requirements disconnecting the mains power cable from the outlet.
• Documentation
Isolate from mains power
Alarms preset Refer to section Alarms on page 166 and section Turning off
alarms on page 149.
Technical data Refer to chapter Technical data on page 169.
System overview Refer to chapter System Overview on page 15.
SERVO-U Ventilator System v2.0, User's Manual 209
| 9 | Technical data |
210 SERVO-U Ventilator System v2.0, User's Manual
| Index | 10 |
10 Index lifetime 197
medication cup 197
A medication temperature 197
Accessories particle size 197
residual volume 197
cable holder for handle 204 specification range 197, 197
drawer 203 weight 196, 197
expiratory heater, Servo Duo Guard 204 Aerosol output
gas cylinder restrainer kit 203 Aerogen Pro nebulizer 196, 196, 197
humidifier holder 203 Aerogen Solo nebulizer 197
mobile cart 203 Airway adapter dead space
pendant/bed holder 204 CO2 analyzer 199, 200
Servo Duo Guard 204 Airway pressure 180
Servo Guard 204 Alarm limits
shelf base 204 default values 189
support arm 203 parameters 189
water bag/IV pole 203 setting range 189
Y piece holder 204 Alarm output connection 14, 20, 160, 182, 202
Aerogen nebulizer unit 127, 132, 132 Alarm state test 182
Aerogen nebulizers Alarms
adding medication 130 default settings 144
assemble 128 handling 145
function test 130 high priority 152, 152, 159
Aerogen Pro nebulizer indication 145
aerosol output 196 low priority 158, 159
aerosol output rate 196, 197 medium priority 154, 159
control cable 196 sound level 44
dimensions 196 turn off 150, 150
flow rate 196 Ambient pressure 174
lifetime 196 Apnea
medication cup 196 audio delay 125
medication temperature 196 management 124
particle size 196 maximal time 113
residual volume 196 time 113, 124
specification range 196, 196 Atmospheric pressure 171, 171
Aerogen Solo nebulizer Audio off 189
aerosol output 197 Automode 89
average tested 196, 196, 197, 197 Average tested
continuous nebulization 132 Aerogen Pro nebulizer 196, 196
control cable 197 Aerogen Solo nebulizer 197, 197
dimensions 197
flow rate 197 211
SERVO-U Ventilator System v2.0, User's Manual
| 10 | Index |
B classification 198
Backup ventilation components 134
digitizing rate 199, 200
deactivating 126 full calibration 137
trends 125 inaccuracy 199, 200
Barometer test 181 measured parameters 198
Barometric pressure compensation measurement conditions 198
CO2 analyzer 198 oxygen concentration compensation 198
monitoring 180 standard compliance 198, 199, 200
Battery system response time 199, 200
backup 172 use guidelines 134
disposal 172 warm-up time 199, 200
lifetime 172 weight 199, 200
warnings 11 zero offset adjustment 137
Battery switch test 182 CO2 monitoring 133, 133
Bias flow during expiration 178 Communication/Interface
Biomed 162 alarm output connection 202
Bi-Vent/APRV 72 ethernet port 202
Breathing parameters SCI 202
default values 183, 186 serial ports 202
parameters 186 USB port 202
setting range 186 VGA port 202
BTPS 174 Compensation
circuit 118, 174
C leakage 117
Cable holder for handle Compliance 176
Compressor Mini 204
dimensions 204 Configuration
maximum load 204 Biomed code 167
weight 204 set date 167
Calibration set time 167
CO2 analyzer 40, 135, 137, 137 view 60
Y sensor 40, 140 Conical fittings 173
Cautions Connection standards 173
general 9 Connectors and cables
power supply 11 Y sensor 201
Classification Continuous nebulization 132
CO2 analyzer 198 Control cable
CO2 Aerogen Pro nebulizer 196
monitoring 133 Aerogen Solo nebulizer 197
CO2 analyzer user interface 17
airway adapter dead space 199, 200
barometric pressure compensation 198 SERVO-U Ventilator System v2.0, User's Manual
calibration 40, 135, 137, 137
212
D | Index | 10 |
Dead space
tidal volume 179
Y sensor 201 Expiratory channel
Decelerating flow 70
Default alarm settings 144 airway pressure 180
Digitizing rate barometric pressure compensation 180
bias flow during expiration 178
CO2 analyzer 199, 200 expiratory flow measurements 178
Dimensions filtering 181
internal compressible factor 178
cable holder for handle 204 O2 concentration 180
drawer 203 PEEP regulation 178
gas cylinder restrainer kit 203 PEEP setting range 178
humidfier holder 203 pressure drop 178
mobile cart 203 respiratory rate 180
pendant/bed holder 204 supply pressure 180
shelf base 204 system response time O2 180
support arm 203 Export & delete 59, 176
user interface adapter 204
ventilator 173, 203 F
water bag/IV pole 203 Filtering 181
Y piece holder 204 Flow adaptation 69
Disconnection/Suction 120 Flow rate
Drawer
dimensions 203 Aerogen Pro nebulizer 196
weight 203 Aerogen Solo nebulizer 197
Driving pressure 114 Flow transducer test 182
Functions
E circuit compensation 118
Edi disconnection/suction 120
Edi monitoring 119
cable 192 leakage compensation 117
catheters 83, 84, 84, 192 manual breath 115
module 192, 192 max. inspiration time 191
monitoring 119 nebulization 116
Edi module NIV disconnection flow 191
function check 40 O2 boost level 116
EMC 170 static measurements 115
End inspiration 108
Essential performance 175 G
Expiratory Gas
flow measurements 178
heater, Servo Duo Guard 204 delivery device 177
hold 116 delivery system 177
minute volume 180 exhaust port 173
Gas cylinder restrainer kit
SERVO-U Ventilator System v2.0, User's Manual
213
| 10 | Index | Ingress protection 171
Inlet
dimensions 203
maximum load 203 gas 173
weight 203 pressure 174
Gas flow 22 Inspiration time 108
Gas quality 173 Inspiratory
Gas supply hold 116
connection standards 173 minute volume 177
gas quality 173 resistance 176
inlet gas 173 rise time 109
max. levels 173 tidal volume 177, 179
test 181 Inspiratory channel
General gas delivery device 177
caution 9 gas delivery system 177
important 10 inspiratory minute volume 177
note 10 inspiratory tidal volume 177
warning 8 internal compressible factor 177
max. airway pressure 177
H NIV max leakage compensation level 177
Height O2 concentration 177
pressure drop 177
units 164 Intended use 6
High Flow therapy 49, 50, 103 Internal
High priority alarms 152, 159 compressible factor 177, 178
Holders leakage test 181
test 181
cable holder for handle 204
gas cylinder restrainer kit 203 L
humidifier holder 203 Leakage 174
pendant/bed holder 204 Lifetime
shelf base 204
support arm 203 Aerogen Pro nebulizer 196
water bag/IV pole 203 Aerogen Solo nebulizer 197
Y piece 204 Logs
Humidifier event log 194, 194
holder Low priority alarms 158, 159
Lowest pressure in patient circuit 171
dimensions 203
M
I Maneuvers
I:E ratio 107, 174
Important manual breath 115
static measurements 115
general 10 Material
Important settings 106
Inaccuracy SERVO-U Ventilator System v2.0, User's Manual
CO2 analyzer 199, 200
214
| Index | 10 |
Y sensor 201 Volume Control 69
Maximum Volume Support 77
Module
airway pressure 177 handling 19
gas levels 173 Monitoring
inspiration time 191 expiratory minute volume 180
Maximum load expiratory tidal volume 179
cable holder for handle 204 inspiratory tidal volume 179
gas cylinder restrainer kit 203
humidifier holder 203 N
support arm 203 Nasal CPAP 101
water bag/IV pole 203 NAVA
Measured parameters
CO2 analyzer 198 level 113
Measurement conditions workflow 82
CO2 analyzer 198 Nebulization 127, 131, 132
Measuring method Nebulization
Y sensor 201 continuous 132
Measuring range on/off 131
Y sensor 201 Nebulizer
Media Pro 127, 132
recording waveforms 58 Solo 127, 132
saving screenshots 58 use guidelines 128, 128
Medication cup NIV
Aerogen Pro nebulizer 196 Pressure Control 98
Aerogen Solo nebulizer 197 Pressure Support 99
Medication temperature NIV disconnection flow 191
Aerogen Pro nebulizer 196 NIV disconnection function
Aerogen Solo nebulizer 197 edit 165
Medium priority alarms 154, 159 NIV max leakage compensation level 177
Minute volume 106 NIV NAVA 100
Mobile cart Noise 171
dimensions 203 Non invasive ventilation 95
preventive maintenance interval 203 Non operating conditions
weight 203 atmospheric pressure 171
Modes relative humidity 171
Automode 89 temperature 171
Bi-Vent/APRV 72 Notes
NAVA 80, 88 general 10
Pressure Control 65 power supply 11
Pressure Regulated Volume Control 67 Numerical values 53
Pressure Support/CPAP 74
SIMV 92 215
SERVO-U Ventilator System v2.0, User's Manual
| 10 | Index | exporting 59
Patient system connectors
O
O2 conical fittings 173
gas exhaust port 173
boost function 117 pressure line connector 173
cell adjustment 141 Patient unit 18
concentration 177 Patient unit
O2 boost level 116 on/off switch 19
O2 cell adjustment 141 serial number 19
O2 cell/sensor test 182 PBW 114
O2 concentration 106, 180 Pdrive 114
Operating conditions PEEP regulation 178
atmospheric pressure 171 PEEP setting range 178
lowest pressure in patient cirucuit 171 Pendant/bed holder
relative humidity 171 dimensions 204
temperature range 171 weight 204
Oxygen concentration compensation Power consumption
CO2 analyzer 198 Y sensor 201
Power supply
P cautions 11
Parameters external 12 V DC 172
notes 11
Y sensor 201 range 172
Particle size warnings 11
Y sensor 201
Aerogen Pro nebulizer 196 Predicted body weight 114
Aerogen Solo nebulizer 197 Pressure Control 65
Patient category 41, 41, 170 Pressure drop 177, 178
Patient category Pressure level 106
change 41 Pressure line connector 173
select 41 Pressure line connector
set 41, 41 Y sensor 201
Patient circuit 23 Pressure Regulated Volume Control 67
Patient circuit Pressure Support/CPAP 74
10 mm dual heat 26 Pressure transducer test 181
10 mm single heat 25 Pre-use check 38, 39, 174, 181
15-22 mm single heat 23 Pre-use check
22 mm dual heat 24 start 39
Patient circuit test 40, 182 Pre-use check test
Patient circuit test alarm output connection 14, 20, 182
compliance 176 alarm state test 182
expiratory resistance 176 barometer test 181
inspiratory resistance 176
Patient data SERVO-U Ventilator System v2.0, User's Manual
deleting 59
enter 30
216
| Index | 10 |
battery switch test 182 inspiratory rise time 109
flow transducer test 182 minute volume 106
gas supply test 181 NAVA level 113
internal leakage test 181 O2 concentration 106
internal test 181 pressure level 106
O2 cell test 182 settings-related information 114
patient circuit test 182 SIMV breath cycle time 114
pressure transducer test 181 tidal volume 106
safety valve test 182 trigger 110
Y sensor calibration 182 Shelf base
Preventive maintenance interval dimensions 204
mobile cart 203 weight 204
Previous mode 123 Shortcuts 30
SIMV breath cycle time 114
R Size
Relative humidity 171, 171 CO2 analyzer 199, 200
Remote Services 168 Y sensor 201
Residual volume Specification range
Aerogen Pro nebulizer 196, 196
Aerogen Pro nebulizer 196 Aerogen Solo nebulizer 197, 197
Aerogen Solo nebulizer 197 Standard conditions
Resistance ambient pressure 174
Y sensor 201 BTPS 174
Respiratory rate 180 circuit compensation 174
Room temperature 174 I:E 174
inlet pressure 174
S pre-use check 174
Safety scales 29, 42 room temperature 174
Safety valve test 182 settings 174
Select STPD 174
ventilatory frequency 174
ventilation 42 Standards
Service 167, 195 CO2 analyzer 198
SERVO COMPASS 54 Ventilator 170
Servo Duo Guard 204 Start
Servo Guard 204 pre-use check 39
Set ventilation 45
Startup configuration 164
alarm limits 43 Stop
patient category 41, 41 ventilation 45
Settings 174 Storage temperature 171
Settings STPD 174
apnea time 113
end inspiration 108 217
I:E ratio 107
SERVO-U Ventilator System v2.0, User's Manual
| 10 | Index | User interface 27
User interface
Suction
closed suctioning 122 interactive areas 28
open suctioning 121 light frame 27
navigating 28
Supply pressure 180 symbols 31
Support arm USB port 27
User interface adapter
dimensions 203 dimensions 204
maximum load 203 weight 204
weight 203
Symbols V
accessories and packaging 33 Ventilation
patient unit 20
user interface 31 select 41, 42
System start 45
EMC 170 stop 45
ingress protection 171 Ventilator 16
noise 171 Ventilator
System response time O2 180, 199, 200 configuration 60
System status 60 dimensions 173, 203
weight 173
T Ventilatory frequency 174
Technical alarms Ventilatory settings 183
Views
high priority alarms 152 Advanced 49, 55
low priority alarms 158, 159 Basic 48
medium priority alarms 154, 159 Distance 51, 56
Technical error messages 159 Family 51
Temperature range 171 Loops 50, 55
Temporary O2 increase SERVO COMPASS 54
edit 165 Volume Control 69
Tidal volume 106 Volume Support 77
Transport 35 VT/PBW 114
Transport
before transport 35 W
during transport 35 Warm-up time
Trends 56, 193
Triggering CO2 analyzer 199, 200
Edi trigger 112 Warnings
pneumatic trigger 110
battery 11
U fire hazard 12
Units general 8
power supply 11
height 164
weight 164 SERVO-U Ventilator System v2.0, User's Manual
218
Water bag/IV pole resistance 201 | Index | 10 |
dimensions 203 size 201 219
maximum load 203 test 182
weight 203 use guidelines 139
weight 201
Waveform
estimated pressure 53
Waveforms
displaying 52
scaling 52
Weight
Aerogen Pro nebulizer 196, 196
Aerogen Solo nebulizer 197, 197
cable holder for handle 204
CO2 analyzer 199, 200
drawer 203
gas cylinder restrainer kit 203
humidifier holder 203
mobile cart 203
pendant/bed holder 204
shelf base 204
support arm 203
units 164
user interface adapter 204
ventilator 173
water bag/IV pole 203
Y sensor 201
Y
Y piece holder
dimensions 204
Y sensor
calibration 40, 140
components 139
connectors and cables 201
dead space 201
material 201
measuring method 201
measuring range 201
parameters 201
power consumption 201
power supply 201
pressure line connector 201
SERVO-U Ventilator System v2.0, User's Manual
| 10 | Index |
220 SERVO-U Ventilator System v2.0, User's Manual
x
Manufacturer: Maquet Critical Care AB © Maquet Critical Care AB 2016. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice.
Röntgenvägen 2 • Order No. 68 85 331 • User's Manual • Printed in Sweden • 160621 • Rev: 02 English •
171 54 Solna, Sweden
Phone: +46 (0) 10 335 73 00
www.maquet.com
For local contact:
Please visit our website
www.maquet.com
Getinge Group is a leading global provider of products and systems that
contribute to quality enhancement and cost efficiency within healthcare
and life sciences. We operate under the three brands of ArjoHuntleigh,
Getinge and Maquet. ArjoHuntleigh focuses on patient mobility and wound