DAILY BIOLOGIC CHECK OF AUDIOMETERS
An audiometer must be checked before each day of use or if a problem is suspected with the
equipment during use. After the audiometer is turned on, the operator should listen to the signal
at different dial settings through each transducer that may be used in testing. In order to do a
daily biologic check, the operator should:
Check the cords for signs of wearing or cracking.
Check all connection points for secure insertion.
Check the audiometer for loose dials or dials that are out of alignment.
Check that the earphones are securely attached to the headset.
After putting on the headsets, present a tone and make sure it comes through the
appropriate earphone.
Turn the attenuator dial from minimum to maximum levels and listen for increases and
decreases in loudness. Be sure to increase the attenuator to high intensity levels. In
addition to presenting the tone at these high intensity levels, allow the tone to be
absent to listen for hum.
For screening purposes, check the screening levels to be used at each frequency. Make
sure that the sounds, which the client is expected to hear, can be heard in the
environment.
If the equipment is to be used for threshold testing, then the threshold of the clinician
(or a person with known thresholds) should be checked.
Turn the frequency dial with the attenuator set at a comfortable loudness and listen to
each frequency.
Jiggle the earphone cords and listen for continuous tone presentation. A defective cord
will produce static or cause the tone to be intermittent.
All transducers that will be used in testing must be checked.
In all of the above situations listen for audible clicks, discontinuous presentation of the tone,
crosstalk between the earphones, static, hum, background noise, a dead earphone, or any
undesired interference with the presentation of the tone. If problems exist that cannot be
solved, then testing cannot be done. Document the quality control data on the Audiometer -
Daily Listening Check form.
Students and clinical educators are responsible for the biologic check of the screening
audiometers. Students will be assigned to do the daily biologic check of the equipment in the
Audiologic Suite. Completed forms are filed in the Calibration Book in the Audiologic Suite.
WAIT, DON’T TOUCH THAT DIAL UNTIL YOU HAVE…
1. Checked the earphones and bone oscillator cords for signs of wear.
2. Jiggled the cords while listening to a tone to see if the signal is intermittent. Check each
earphone.
3. Listened for audible clicks through each earphone when the dial switches are manipulated.
4. Listened for static hum when the tone is present or absent (at 20 and 50 dB HL).
Run your fingers along the cord while listening especially checking where it enters
the earphone.
5. Listened for "cross-talk" between earphones. These are sounds that are heard in the
opposite earphone from the test ear.
6. Listened to the signal as the attenuator dial is turned from maximum to minimum to
insure linearity and loudness perception appropriate to dial changes through each
earphone.
7. Checked for loose or out-of-alignment dials.
8. Checked all connection points for secure insertion.
9. Cleaned the earphone cushions and headband. Check for any signs of wear or
cracking.
10. Checked an adult with normal hearing to ensure the screening levels are audible.
Listen for room noise while wearing the headsets.
11. Documented the results on the Daily Listening Check form.
12. Reported any equipment problems to the audiologist to arrange for repairs.
CARE AND HANDLING OF THE AUDIOMETER
• Avoid exposing the audiometer to extreme temperature. Store in a cool, dry place.
• Do not drop the audiometer.
• Handle the audiometer with care. Secure the audiometer within the case. Fasten the case
securely. Place the audiometer in the car with a seat belt or place it securely on the floor.
• Keep the audiometer covered when not in use to avoid dust.
• Wipe the audiometer clean with a soft cloth.
• If the audiometer has a bone oscillator, secure it in foam rubber for traveling.
• Never force the dials or the tone interrupter switch. They should turn freely and quietly. If
there is a problem, repair is needed.
• The earphones and headband must be cleaned carefully. Keep cleaning solutions out of the
diaphragm or center of the earphone.
• Check the earphone cushions for cracks.
• Do not twist the electrical cords or the cords to the headphones. Keep the cords untangled.
Never bend or twist the headphone cords where they enter the earphone. Make certain
cords and wires are stored in the appropriate compartment and are not hanging outside the
case.
• Refer to "The Handling of Earphones."
• Audiometers must be electroacoustically calibrated to meet the current ANSI standards at
least annually and when the examiner suspects a problem. The last date of electroacoustic
calibration should be noted on the audiometer.
TROUBLESHOOTING THE AUDIOMETER
PROBLEM REPAIR IS NEEDED IF...
...sounds other than pure tones
Audible Click/Static/Hum? are heard
Press interrupter bar and
elicit tone ...the tone can be heard in the
opposite earphone
Cross Talk?
Present a 70 dB HL continuous ...the tone is not heard at
tone through one earphone the intensity levels expected
Low Intensity Function of the ...static, crackling, hum, or
Attenuator? intermittent sound is heard
Present 1000 Hz tone at 30, 25
and 20 dB HL ...the frequencies do not sound
as expected
Cords, Connections to Phones?
Present a 50 dB HL continuous ...the tones do not get steadily
tone through each earphone softer
Frequencies? ...the signals are not equal in
Present a 60 dB HL continuous intensity
tone changing the frequency dial
through the octaves
Linearity of the attenuator?
Present a 2000 Hz tone at 90
dB HL, decrease intensity in 5
dB steps
Matched earphones?
Switch signal from one phone
to the other
At least annual electroacoustic calibration to meet current ANSI standards of every
audiometer is required. If a problem is noted, repair and additional electroacoustic
calibration is needed.
THE HANDLING OF EARPHONES
Care must be taken in handling audiometer earphones. They are delicate devices designed to
perform a task requiring great sensitivity.
1. The internal construction of earphones makes use of very fine wires. If the
earphones are mistreated or handled roughly, these wires may break.
2. Damage can occur if the earphones are placed face down on a table and pressed
downward until the cushion is flattened. In this case a quantity of air will be trapped
and pressed against the earphone diaphragm resulting in damage.
3. Damage can be caused by slapping an earphone against an open hand or against the
other earphone.
4. Earphones can be damage by sharp objects, which may pick holes in the diaphragm.
5. Dropping an earphone severely jars the internal mechanism. Dropping the earphones
can cause permanent damage or change the response of the earphone.
6. The earphones are an integral part of the audiometer. They must produce a
particular sound at the appropriate level. In order to accomplish this, the audiometer
must be calibrated with a particular set of earphones. Another set of earphones
cannot be substituted without compromising the accuracy of testing. Whenever the
earphones are replaced, the audiometer must be recalibrated.