21 / 70 3 • Description of the Controls 80113-401 10.15 Chapter 3 Description of the Controls Controls on the front panel Fig. 3-1 Symbol (1) Read the user manual before switching on and using the unit. Power (2) Power switch Unit on / off. The unit is completely disconnected from the power supply only when the power connector is unplugged. Selection buttons (L1) (L2) (L3) (R1) (R2) (R3) The buttons have a different function depending on which window is shown on the display. Take note of the function the button refers to. Symbol (3) The equipment conforms to the requirements of Type CF (Cardiac Float) and is protected against the effects of a defibrillator discharge. Instrument receptacle (4) Instrument receptacle (applied part) For connecting the instruments to the unit. 1 2 3 L1 L2 L3 R1 R2 R3 ERBECRYO 2 4


3 • Description of the Controls 22 / 70 80113-401 10.15 Controls on the back Fig. 3-2 Gas connections (1) Waste gas nozzle For connecting the waste gas tube. Direct the waste gas into the anesthetic suction system or the open air. (2) Gas connection For connecting the gas hose. The gas hose directs the CO2 from the CO2 bottle into the unit. The input pressure may not exceed 65 bar (942 psi). Sockets (3) Footswitch socket For connecting the footswitch. (4) ECB socket (ECB means Erbe Communication Bus) Other compatible units can be connected to the unit. The ECB ensures communication between the units. Connect an ECB cable to these sockets and with one of the other units. Grounding terminal (5) Grounding terminal connection If required, attach a grounding terminal line and connect this to the grounding terminal system of the operating room. Power connection (6) Power connection Connect the unit to a properly installed grounded power outlet. Only use the Erbe power cord or an equivalent power cord for this purpose. The power cord must bear the national test symbol. Power fuses (7) Power fuses The unit is protected with power fuses. If one of these power fuses has blown, the unit may not be used on the patient again until it has been checked by a competent technician. The values of the power fuses are specified on the unit's rating plate. Only spare fuses with these values may be used. Symbol (8) Read the user manual before switching on and using the unit. 1 2 3 4 6 5 7 8


23 / 70 4 • Installation 80113-401 10.15 Chapter 4 Installation Ambient conditions WARNING Ignition of combustible anesthetics, skin cleansers, or disinfectants If you use the unit in the presence of combustible anesthetics, skin cleansers, or disinfectants, these substances could ignite. Risk of fire and explosion to the patient and medical personnel! Danger of material damage. Do not use the unit in the presence of combustible anesthetics, skin cleansers, or disinfectants. NOTICE Interference with the unit by portable and mobile HF communication devices (e.g. mobile phones, WLAN equipment) Electromagnetic waves emitted by portable and mobile HF communication devices can effect the unit. The unit may fail or not perform properly. Please see the table "Recommended separation distances between portable and mobile HF communications equipment and the equipment" at the end of this User Manual. NOTICE Unsuitable temperature or level of humidity during operation If you operate the equipment at an unsuitable temperature or level of humidity, it may sustain damage, fail, or not perform properly. Operate the equipment at a suitable temperature and level of humidity. You will find the tolerances for temperature and humidity in the Technical Data. If other ambient conditions must be observed for operation of the equipment, you will also find them in the Technical Data. NOTICE Unsuitable temperature or humidity in transit or storage If you transport or store the equipment at an unsuitable temperature or level of humidity, it may sustain damage and fail. Transport and store the equipment at a suitable temperature and level of humidity. You will find the tolerances for temperature and humidity in the Technical Data. If other ambient conditions must be observed for transport and storage of the equipment, you will also find them in the Technical Data.


4 • Installation 24 / 70 80113-401 10.15 NOTICE Insufficient acclimatization time, unsuitable temperature during acclimatization If the device was stored or transported below or above a certain temperature, it will take a certain time and temperature to acclimatize. If you do not observe the rules, the device can sustain damage and fail. Acclimatize the device according to the rules in the Technical Data. NOTICE Overheating of the device due to poor ventilation If ventilation is poor, the device can overheat, sustain damage, and fail. Install the device in such a way that there is an unobstructed circulation of air around the housing. Installation in confined wall recesses is prohibited. NOTICE Penetration of liquid into the device The housing is not absolutely watertight. If liquid penetrates, the device can sustain damage and fail. Make sure no liquid can penetrate the device. Ensure that liquid does not penetrate through the instrument socket. Do not place vessels containing liquids on top of the device. Unit The device may only be operated in rooms used for medical purposes. Furthermore, the ambient conditions for the transport, storage, acclimatization and operation of the unit must be observed. The ambient conditions are provided in the Technical Data chapter. Gas bottle Different ambient conditions to those for the unit apply for the operation, acclimatization, transport and storage of the gas bottle. Refer to the Technical Data, Safe handling of carbon dioxide and the gas bottle and Safety Instructions chapters in this User Manual and to the safety instructions provided by the gas supplier. Installing the unit on an Erbe equipment cart Please read the User Manual for the equipment cart concerned. There you will find instructions on how to secure the unit to the equipment cart. Connecting tubes and the footswitch CAUTION Using unsuitable accessories Danger of injury to patient and medical personnel. Only use Erbe instruments, footswitches, waste gas tubes, gas hoses and gas cylinder connections.


25 / 70 4 • Installation 80113-401 10.15 WARNING Dangerous CO2 concentration in room A high concentration of CO2 in the air can cause suffocation. CO2 is heavier than air. If discharge is not controlled, dangerous accumulations can develop close to the ground. Observe the national regulations regarding concentrations of CO2 permitted in the workplace. Connect the Erbe waste gas tube to the waste gas nozzle on the unit. Direct the waste gas into the anesthetic suction system or the open air. This procedure is recommended particularly where activation takes longer. If it is not possible to use the Erbe waste gas tube, ensure that the room is adequately ventilated: enable air to circulate by opening windows and doors or by switching on the air conditioner. Check the seal of all connections once the bottle valve has been opened. No gas may escape. If there are leaks, check that all connections are secure. If you cannot rectify the leak, close the bottle valve and inform Technical Service. Fig. 4-1 When the ERBECRYO 2 is installed on an ERBECRYO 2 Cart, the rear of the unit with the connections faces the ground. See Fig. 4-1. 6


4 • Installation 26 / 70 80113-401 10.15 Connecting the waste gas tube 1. Attach the waste gas tube to the waste gas nozzle (1). This procedure is recommended particularly where activation takes longer. See warning: "Dangerous CO2 concentration in room". 2. Pull back the closure of the waste gas tube. 3. Hold the closure tightly. 4. Slide the waste gas tube fully on to the waste gas nozzle. 5. Release the closure. 6. Direct the waste gas into the anesthetic suction system or the open air. Connecting the gas hose to the gas connection 1. Attach the gas hose to the gas connection (2). 2. Unscrew the nut by hand. 3. Then tighten the nut using a 17 mm open-end wrench. Connecting the footswitch 1. Plug the footswitch connector into the footswitch socket (3). 2. Locate the correct angle of engagement by turning the connector (see illustration). 3. Tighten the retainer nut on the connector by hand. Connecting the gas bottle to the gas cylinder connection CAUTION Wrong gas e.g. N2O, Argon Risk of accidental tissue damage due to uncontrolled cooling with N2O. Risk of damage to the health of the patient and medical personnel if N2O is inhaled. Pressure in the system is too high. The unit can be damaged. Medical personnel and the patient could be injured by flying parts. Only use CO2. Observe the label on the gas bottle. CAUTION CO2 bottle with dip tube, horizontal CO2 bottle Liquid CO2 escapes. Pressure in the unit is too high. The unit can be damaged. Medical personnel and the patient could be injured by flying parts. Do not use a CO2 bottle with a dip tube. Operate the CO2 bottle upright. WARNING Damage to the gas bottle, to the valve. Gas escapes at high pressure. If the gas bottle has been weakened due to serious damage, this can lead to cracks in the wall of the bottle. Considerable stresses and strains occur at the cracks, causing the bottle to break open. Gas emerges at high pressure. The gas bottle is propelled like a rocket. Risk of injury or death for patients and medical staff! Risk of damage to property.


27 / 70 4 • Installation 80113-401 10.15 Do not apply force to gas bottles, bottle connectors and valves. When transporting, storing, and using the gas bottle, prevent it from falling over and dropping by securing it with chains, hoops, or retaining straps. Only transport gas bottles with the valve guard fitted: bottle cap or bottle collar. Operate and store the CO2 bottle upright. Do not place the CO2 bottle close to a heat source. Heaters or air conditioners are examples of heat sources. Do not actively heat the CO2 bottle, e.g. using a fan heater. Do not expose the CO2 bottle to direct sunlight. Observe the safety instructions provided by the gas supplier. Gas cylinder connection To connect the gas bottle and the gas hose you require a gas cylinder connection. The gas cylinder connection is specific to each country. Preparing and checking the gas bottle 1. Make sure that the gas bottle contains CO2. Check the label on the gas bottle. 2. The gas bottle may only be operated upright. Secure the gas bottle using retaining straps, chains, etc. to prevent it from falling over. 3. If you have stored the gas bottle at below 15 °C or over 25 °C, allow the gas bottle to acclimatize at room temperature for 8 hours. Do not actively heat the gas bottle, e.g. using a fan heater. 4. The room temperature should be between 15 °C and 25 °C. It is possible that these values may be exceeded or may not be reached for a short period, as the gas bottle only heats up or cools down slowly. 5. Check the thread of the bottle valve for damage. 6. Do not apply force to the gas bottle, gas cylinder connection and bottle valve. Connecting the gas cylinder connection and the gas hose Fig. 4-2 1. Place the sealing ring in the gas cylinder connection (2). Use of the sealing ring is mandatory. Without a sealing ring, the cryosystem is not leak-tight. 2. Use an open-end wrench to screw the nut on the gas cylinder connection (3) onto the bottle thread. 3. Use a 17 mm open-end wrench to screw the nut on the gas hose (4) onto the gas cylinder connection. 4. Attach the other end of the gas hose to the unit (see Connecting the gas hose to the gas connection p. 26). 5. Switch on the unit and press the Info button.


4 • Installation 28 / 70 80113-401 10.15 6. Open the bottle valve fully and note the input pressure that is displayed. 7. Close the bottle valve and monitor the pressure for approx. 120 seconds. The pressure may only fall by max. 2 bar. If the fall in pressure is greater, one of the connections is not leak-tight. Inform Technical Service. Do not work with a leaky gas connection. Connecting gas bottles with the switching valve CAUTION Wrong gas e.g. N2O, Argon Risk of accidental tissue damage due to uncontrolled cooling with N2O. Risk of damage to the health of the patient and medical personnel if N2O is inhaled. Pressure in the system is too high. The unit can be damaged. Medical personnel and the patient could be injured by flying parts. Only use CO2. Observe the label on the gas bottle. CAUTION CO2 bottle with dip tube, horizontal CO2 bottle Liquid CO2 escapes. Pressure in the unit is too high. The unit can be damaged. Medical personnel and the patient could be injured by flying parts. Do not use a CO2 bottle with a dip tube. Operate the CO2 bottle upright. WARNING Damage to the gas bottle, to the valve. Gas escapes at high pressure. If the gas bottle has been weakened due to serious damage, this can lead to cracks in the wall of the bottle. Considerable stresses and strains occur at the cracks, causing the bottle to break open. Gas emerges at high pressure. The gas bottle is propelled like a rocket. Risk of injury or death for patients and medical staff! Risk of damage to property. Do not apply force to gas bottles, bottle connectors and valves. When transporting, storing, and using the gas bottle, prevent it from falling over and dropping by securing it with chains, hoops, or retaining straps. Only transport gas bottles with the valve guard fitted: bottle cap or bottle collar. Operate and store the CO2 bottle upright. Do not place the CO2 bottle close to a heat source. Heaters or air conditioners are examples of heat sources. Do not actively heat the CO2 bottle, e.g. using a fan heater. Do not expose the CO2 bottle to direct sunlight. Observe the safety instructions provided by the gas supplier.


29 / 70 4 • Installation 80113-401 10.15 Preparing and checking the gas bottle 1. Make sure that the gas bottle contains CO2. Check the label on the gas bottle. 2. The gas bottle may only be operated upright. Secure the gas bottle using retaining straps, chains, etc. to prevent it from falling over. 3. If you have stored the gas bottle at below 15 °C or over 25 °C, allow the gas bottle to acclimatize at room temperature for 8 hours. Do not actively heat the gas bottle, e.g. using a fan heater. 4. The room temperature should be between 15 °C and 25 °C. It is possible that these values may be exceeded or may not be reached for a short period, as the gas bottle only heats up or cools down slowly. 5. Check the thread of the bottle valve for damage. 6. Do not apply force to the gas bottle, switching valve and bottle valve. Connecting the switching valve and gas hose Fig. 4-3 1. Place the sealing rings in the nuts on the switching valve (1). Use of the sealing rings is mandatory. Without sealing rings, the cryosystem is not leak-tight. 2. Use an open-end wrench to screw the nuts on the switching valve (1) onto the bottle thread. 3. Use a 17 mm open-end wrench to screw the nut on the gas hose (2) onto the switching valve. 4. Turn the tap with the arrow towards the full gas bottle. In Fig. 4-3, bottle 2 is active. 5. Attach the other end of the gas hose to the unit (see Connecting the gas hose to the gas connection p. 26). 6. Switch on the unit and press the Info button. 7. Open the bottle valve fully and note the input pressure that is displayed. 8. Close the bottle valve and monitor the pressure for approx. 120 seconds. The pressure may only fall by max. 2 bar. If the fall in pressure is greater, one of the connections is not leak-tight. Inform Technical Service. Do not work with a leaky gas connection. 1 2


4 • Installation 30 / 70 80113-401 10.15 Changing the gas bottle (when using a gas cylinder connection) Loosening the gas cylinder connection and gas hose Fig. 4-4 When changing the bottle, you must observe the sequence of steps to be performed: First loosen the gas hose. Then loosen the gas cylinder connection. 1. Close the bottle valve (1). 2. Using a 17 mm open-end wrench, loosen the nut on the gas hose (4) slightly. The gas escapes from the gas hose. 3. Loosen the nut on the gas hose (4) completely. 4. Pull the gas hose off the gas cylinder connection. 5. Using an open-end wrench, loosen the nut on the gas cylinder connection (3). 6. Pull the gas cylinder connection (2) off the bottle thread. Do not lose the sealing ring. 7. Loosen the retaining straps, chains, etc. of the gas bottle. 8. Where bottles do not have a bottle collar, screw on a valve protection cap. 9. Change the gas bottle.


31 / 70 4 • Installation 80113-401 10.15 Changing gas bottles (when using a switching valve) Loosening the switching valve and gas hose Fig. 4-5 1. The bottle valves of both gas bottles must be closed. 2. Using an open-end wrench, loosen the nuts on the gas bottles (1) slightly. The gas escapes from the gas hose. 3. Loosen the nuts on the gas bottles (1) completely. 4. Pull the switching valve off the bottle threads. Do not lose the sealing rings. 5. Loosen the retaining straps, chains, etc. of the gas bottles. 6. Where bottles do not have a bottle collar, screw on a valve protection cap. 7. Change the gas bottles. Switching between two gas bottles using the switching valve If you have connected two gas bottles to the ERBECRYO 2 using a switching valve, you can use the tap to switch between the gas bottles. 1. Open the bottle valve of the full gas bottle. 2. Turn the tap with the arrow towards the full gas bottle. 3. Close the bottle valve of the empty gas bottle. 1 2


4 • Installation 32 / 70 80113-401 10.15 Fig. 4-6 In Fig. 4-6, bottle 1 is in operation. Fig. 4-7 In Fig. 4-7, bottle 2 is in operation. Electrical connection WARNING Defective grounded power outlet, power supply network without proper grounding, inferior-quality power cord, incorrect line voltage, multiple power outlets, extension cords Risk of electric shock and other injuries to the patient and medical personnel! Risk of damage to property. Connect the unit / the equipment cart to a properly installed grounded power outlet. 1 2 1 2


33 / 70 4 • Installation 80113-401 10.15 Only connect the unit to a power supply network with proper grounding. Only use the Erbe power cord or an equivalent power cord for this purpose. The power cord must bear the applicable national test symbol. Check the power cord for damage. You must not use a damaged power cord. The supply voltage must match the voltage specified on the unit's rating plate. Do not use multiple power outlets. Do not use extension cords. WARNING Incorrect line fuse, defective device Risk of electric shock to the patient and medical personnel! Risk of damage to property. Blown line fuses may only be replaced by a competent technician. Only replacement fuses that have the same rating as the one specified on the unit’s rating plate may be used. When a fuse has been changed, the function of the unit must be verified. If the unit does not function properly or if there are any concerns, please contact Erbe. WARNING Damaged device or damaged accessories, modified device or accessories Danger of injury for patient and medical personnel! Danger of material damage. Check the device and accessories for damage every time before using them (e.g. footswitch, instruments, instrument cables and equipment cart). Check the gas connection, waste gas nozzle and instrument connection in particular for damage. You must not use damaged equipment or damaged accessories. Replace defective accessories. If the unit or accessories are damaged, please contact Technical Service. For your safety and that of the patient: Never attempt to perform repairs or make modifications yourself. Any modification will invalidate liability on the part of Erbe Elektromedizin GmbH. Electrical connection Only use the Erbe power cord or an equivalent power cord. The power cord must bear the national test symbol. 1. Connect the power cord to the power connection at the back of the unit. 2. Plug the power cord into a properly installed grounded power outlet. The supply voltage must match the voltage specified on the unit's rating plate. Only connect the unit to a public low-voltage power supply network with grounding. Grounding terminal If necessary, connect the grounding terminal pin of the unit to the grounding terminal system of the operating room using a grounding terminal line.


4 • Installation 34 / 70 80113-401 10.15


35 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Chapter 5 Working with ERBECRYO 2 Putting the gas bottle into operation 1. Open the bottle valve. 2. Check the seal of all connections. No gas may escape. Switching on, self-test Switching on 1. Before using the cryosystem, make sure that it has been correctly installed and that the gas bottle is in operation. 2. Use the power switch to switch the unit on. Self-test Fig. 5-1 The unit performs a self-test. If the input pressure at the gas connection is too low or too high, you are requested to check the gas supply. Fig. 5-2 The unit is now ready for the instrument to be connected. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


5 • Working with ERBECRYO 2 36 / 70 80113-401 10.15 Instrument connection, main screen Connecting an instrument Plug the instrument connector into the instrument socket of the unit. Main screen Fig. 5-3 The instrument is equipped with a data carrier on which settings are saved. The instrument is detected. The settings are displayed on the main screen. You can work with the settings or change them. • In segment (1) of the display you will see an illustration of the instrument. • In segment (2) of the display you will see the effects. By selecting the effect you can influence the temperature of the probe tip. The higher the effect, the colder the probe tip will be. The bars beneath the effect show how many effects are supported by the instrument. The current instrument supports three effects. Effect 2 has been set. • In segment (3) of the display you will see status and error messages. They are described in the Status messages, error messages chapter. • In segment (4) of the display you will see the Timer. The timer provides various options for measuring the freeze interval. The current setting is Count Up & Beep. • In segment (5) of the display you will see the selected Program. With the current setting, no program is selected. You can save the effect and the timer setting in a program. 10 memory locations are provided. WARNING Damaged, leaky instrument CO2 can flow in an uncontrolled fashion and at high pressure into the patient's body. Risk of a gas embolism. Risk of death for the patient. Do not bend rigid instruments. Do not kink flexible instruments. Do not apply mechanical force to the instruments. Before each use, complete a visual check of the instrument and the instrument tube for mechanical damage. The instrument tube must not show any signs of abrasion. The rigid instrument must not be bent. Before each use, check the instrument function and seal. The check is described in the instructions for use of the instrument. The check is mandatory before each use! You must not use a damaged, leaky instrument! L1 L2 L3 R1 R2 R3 4 1 2 3 5


37 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Setting the effect, setting the timer By selecting the effect you can influence the temperature of the probe tip. The greater the effect, the colder the probe tip will be. The bars beneath the effect show how many effects are supported by the instrument. Setting an effect Fig. 5-4 Press the selection button next to the up/down arrows to set an effect. The effect is predefined for some instruments and cannot be changed. Instead of setting an effect and a timer manually, you can also use programs in which the effect and timer are saved. See: Selecting a program, Saving a program, Overwriting a program, Deleting a program. Setting a timer The timer provides various options for measuring the freeze interval: Timer off A freeze interval is not measured. 1. Press the selection button next to the Timer displayto call up the Timer menu. 2. Select No Timer with the selection buttons next to the up/down arrows. 3. Press the selection button next to OK on the display. Fig. 5-5 In the Timer segment (4) in the main screen you will see: Timer OFF. Count Up With the footswitch pressed, you activate Freeze mode. The timer counts the freeze interval in seconds. If you release the footswitch, the freeze interval remains visible on the display. The next time it is activated, the timer begins again at 0 seconds. 1. Press the selection button next to the Timer displayto call up the Timer menu. 2. Select Count Up with the selection buttons next to the up/down arrows. 3. Press the selection button next to OK on the display. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 4


5 • Working with ERBECRYO 2 38 / 70 80113-401 10.15 Fig. 5-6 In the Timer segment (4) in the main screen you will see the Count Up icon. Count Up & Beep If you press the footswitch, you activate Freeze mode. The timer counts the freeze interval in seconds. An acoustic signal sounds once every second. If you release the footswitch, the freeze interval remains visible on the display. The next time it is activated, the timer begins again at 0 seconds. 1. Press the selection button next to the Timer displayto call up the Timer menu. 2. Select Count Up & Beep with the selection buttons next to the up/down arrows. 3. Press the selection button next to OK on the display. Fig. 5-7 In the Timer segment (4) in the main screen you will see the Count Up & Beep icon. Count Down & Beep You select a Count Down interval. If you press the footswitch, you activate Freeze mode. The Count Down begins. Once the Count Down has been completed, an acoustic signal sounds. If you continue to press the footswitch after the Count Down has been completed, Freeze mode is active again. If you release the footswitch and interrupt the current Count Down, you will see the remaining Count Down interval on the display. The next time it is activated, the Count Down begins from the start. 1. Press the selection button next to the Timer displayto call up the Timer menu. 2. Select Count Down & Beep with the selection buttons next to the up/down arrows. 3. Press the selection button next to OK on the display. 4. Select a Count Down interval of 1 - 100 seconds with the selection buttons next to the up/down arrows. 5. Press the selection button next to OK on the display. L1 L2 L3 R1 R2 R3 4 L1 L2 L3 R1 R2 R3 4


39 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Fig. 5-8 In the Timer segment (4) in the main screen you will see the Count Down & Beep icon. Example Count Down interval: 10 seconds. Freezing, thawing Freezing 1. Position the instrument. Fig. 5-9 2. Press the footswitch. With the footswitch pressed, you activate Freeze mode. In segment (2) of the display you will see the freeze icon (Freeze mode). 3. Note the selected timer setting with the optical and acoustic signals in each case. Thawing Fig. 5-10 1. Release the footswitch. In segment (2) of the display you will see the thaw icon (Thaw mode) for instruments that can be thawed. 2. If required, you can interrupt the thawing phase and activate freezing. Bleed-off Fig. 5-11 Following the thawing phase, the unit performs bleed-off. In segment (2) of the display you will see the bleed-off icon (Bleed mode). The gas escapes from the system. If required, you can interrupt the bleed-off phase and activate freezing. L1 L2 L3 R1 R2 R3 4 L1 L2 L3 R1 R2 R3 2


5 • Working with ERBECRYO 2 40 / 70 80113-401 10.15 Selecting a program The effect and timer are saved in a program. Fig. 5-12 1. Press the selection button next to Program on the display. Fig. 5-13 2. Select a program with the selection buttons next to the up/down arrows. Example: program A. 3. Press the selection button next to OK on the display. Fig. 5-14 4. The name of the program is shown in segment (5) of the display. Example: program A. Saving a program You can save the Effect and the Timer setting in a program 1. Select an effect and a timer setting. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 5


41 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Fig. 5-15 2. Press the selection button next to Program on the display. Fig. 5-16 3. Select an empty memory location with the selection buttons next to the up/down arrows. Example: memory location 2. 4. Press the selection button next to the disk icon. The disk icon stands for "save". Fig. 5-17 5. Press the selection button next to SAVE. Fig. 5-18 6. Select a letter or a number with the L3/R3 selection buttons. Example: B. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


5 • Working with ERBECRYO 2 42 / 70 80113-401 10.15 Fig. 5-19 7. Move the cursor one place ahead with the R2 selection button. Fig. 5-20 8. Select the next letter or number with the L3/R3 selection buttons. Example: E. 9. When you have entered the name in full, press the selection button next to the disk icon. The disk icon stands for "save". Fig. 5-21 10. Press the selection button next to OK on the display. Fig. 5-22 11. The name of the program is shown in segment (5) of the display. Example: program BE. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 5


43 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Overwriting a program If you would like to keep the name of a program but change its settings, proceed as follows. Fig. 5-23 1. Select an effect and a timer setting. The asterisk next to the name of the program shows that you have temporarily changed the settings. 2. Press the selection button next to Program on the display. Fig. 5-24 3. Press the selection button next to the disk icon. The disk icon stands for "save". Fig. 5-25 4. Press the selection button next to SAVE. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


5 • Working with ERBECRYO 2 44 / 70 80113-401 10.15 Fig. 5-26 5. Press the selection button next to the disk icon. The disk icon stands for "save". Fig. 5-27 6. Press the selection button next to OK on the display. Fig. 5-28 7. The program BE with the new settings is shown on the display. Deleting a program Fig. 5-29 1. Press the selection button next to Program on the display. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


45 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Fig. 5-30 2. Select a program with the selection buttons next to the up/down arrows. Example: program BE. 3. Press the selection button next to the recycle bin icon. The recycle bin icon stands for "delete". Fig. 5-31 4. Press the selection button next to the recycle bin icon. The recycle bin icon stands for "delete". Calling up information on the instrument The Probe display provides information on the connected instrument. Fig. 5-32 1. Press the selection button next to the I icon. The I icon stands for "information". L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


5 • Working with ERBECRYO 2 46 / 70 80113-401 10.15 Fig. 5-33 2. Press the selection button next to return to go back to the main screen. The number of Cycles lets you know how often the instrument has been reprocessed. However, ERBECRYO 2 cannot count the actual number of times reprocessing has been performed. It checks how often you have connected the instrument. If the number of cycles is lost due to a malfunction, the following text is shown instead of the number of cycles: unknown. In such cases, please inform Technical Service. Checking input pressure The Input pressure displays the pressure applied to the unit. The input pressure must be between 45 - 65 bar. Activation is not possible outside of this range. If the input pressure is outside of this range, the unit displays an error message. Fig. 5-34 1. Press the selection button next to the I icon. The I icon stands for "information". Fig. 5-35 2. Press the selection button next to the bottle icon. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


47 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 Fig. 5-36 3. Press the selection button next to return to go back to the main screen. Calling up setup In SETUP you can adapt some of the unit's parameters to suit your requirements. You can, for example, adjust the unit to the light conditions in the room. 1. Press the selection button next to the tool icon. 2. Press the selection button next to SETUP. 3. Select a setup setting that you would like to change with the selection buttons next to the up/down arrows. 4. Press the selection button next to OK. 5. Select a setting with the selection buttons next to the up/down arrows. 6. Press the selection button next to return to go back to the main screen. Fig. 5-37 Fig. 5-38 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3


5 • Working with ERBECRYO 2 48 / 70 80113-401 10.15 Contrast Setting the display contrast (10 levels) Brightness Setting the display brightness (10 levels) Button volume An acoustic signal can be generated when pressing the selection buttons. Setting the button volume (10 levels) or Button volume off. System volume With the system volume you can control the acoustic signals of the timer and of error messages. Setting the system volume (10 levels). Activation volume Volume of the acoustic signal for activation. Adjust the signal so that it is clearly audible. Bleeding the cryosystem off after use With bleed-off, you relieve the pressure in the cryosystem, thus extending its service life. With gas cylinder connection Fig. 5-39 1. Close the bottle valve (1). 2. Using a 17 mm open-end wrench, loosen the nut on the gas hose (2) slightly. The gas escapes from the gas hose. 3. Tighten the nut on the gas hose (2) securely again.


49 / 70 5 • Working with ERBECRYO 2 80113-401 10.15 With switching valve Fig. 5-40 1. Close the bottle valves (1). 2. Using a 17 mm open-end wrench, loosen the nut on the gas hose (2) slightly. The gas escapes from the gas hose. 3. Tighten the nut on the gas hose (2) securely again. Disassembling the unit Disconnecting an instrument 1. Press the release button. Remove the instrument from the instrument connection. 2. Place the cover on the instrument. Switching off the unit 1. Use the power switch to switch the unit off. 2. Unplug the power cord. Disconnecting the waste gas tube 1. Pull back the closure of the waste gas tube. 2. Hold the closure tightly. 3. Pull the waste gas tube off the waste gas nozzle. 4. Release the closure. Storing the power cord and footswitch 1. The power cord and footswitch should ideally be stored on the ERBECRYO 2 Cart. 2. Avoid damaging the power cord and footswitch.


5 • Working with ERBECRYO 2 50 / 70 80113-401 10.15


51 / 70 6 • Cleaning and Disinfection 80113-401 10.15 Chapter 6 Cleaning and Disinfection Wipe disinfection For cleaning and disinfecting the surfaces of the unit or of the equipment cart, Erbe recommends a wipe disinfection. Use only disinfectant which complies with the relevant national standards. Instructions for cleaning and disinfection Mix the disinfectant in the concentration specified by the manufacturer. Clean surfaces contaminated with blood before using the disinfectant; otherwise it may be less effective. Wipe the surfaces. Make sure the surfaces are treated uniformly. Comply with the action time of the disinfectant specified by the manufacturer. Safety Instructions WARNING Connection of unit / equipment cart and power supply during cleaning and disinfection Risk of electric shock to the medical personnel! Switch off the device. Unplug the power cord of the device/equipment cart. NOTICE Penetration of liquid into the device The housing is not absolutely watertight. If liquid penetrates, the device can sustain damage and fail. Make sure no liquid can penetrate the device. Ensure that liquid does not penetrate through the instrument socket. Do not place vessels containing liquids on top of the device. NOTICE Alcohol-based spray disinfectant for fast disinfection With membrane keyboards and paint surfaces there is the risk of cracks. Propanol and ethanol will erode surfaces. Do not use these substances.


6 • Cleaning and Disinfection 52 / 70 80113-401 10.15 NOTICE Alternate use of disinfectant solutions based on different active ingredients A color reaction may occur with plastics. Do not use these substances alternately. Membrane keyboards Note: If alcohol-based disinfectants are used on units with membrane keyboards, this remove the anti-glare finish. However, the user surfaces remain fully functional. This does not present a hazard.


53 / 70 7 • Status Messages, Error Messages 80113-401 10.15 Chapter 7 Status Messages, Error Messages Error messages on a blue screen The errors are not described individually in the table below. Follow the instructions on the screen. Note the error code (e.g. K-45). Switch off the device. Forward the error code to Technical Service. Fig. 7-1 Status messages Status messages are of an informative nature. Fig. 7-2 Example of a status message in segment (3) of the display. L1 L2 L3 R1 R2 R3 L1 L2 L3 R1 R2 R3 3 Status message Description Footswitch is ready for operation. The footswitch has been detected. Footswitch has been disconnected. The footswitch has been disconnected. Unit has cooled down. The unit has cooled down following overheating.


7 • Status Messages, Error Messages 54 / 70 80113-401 10.15 Error messages Error messages are also shown in segment (3) of the display (see Fig. 7-2). Complete the steps to rectify the fault in sequence for a possible cause. If there are several possible causes, use the table to identify the cause (e.g. footswitch not connected). Complete the corresponding steps to rectify the fault in sequence. Setting not changeable. You cannot change the setting. Maximal Setting achieved. Maximum setting value reached. Minimal Setting achieved. Minimum setting value reached. Connection to VIO-System recogn. The connection to the VIO system has been detected. Status message Description Error text Possible cause(s) What should you do? Please check gas outlet. Activation not possible. The waste gas tube is obstructed. Check whether the waste gas tube is kinked or blocked. Unit malfunction If the error continues to be displayed, inform Technical Service. Please check gas supply. Activation not possible. The input pressure is less than 5 bar: • No gas bottle is connected. • The gas bottle is empty. • The bottle valve is not open. The input pressure is greater than 75 bar: • The gas bottle is too warm. Unit malfunction Check the input pressure. If the input pressure is less than 5 bar, proceed as follows: • Connect a gas bottle. • Change the gas bottle. • Open the bottle valve. If the input pressure is greater than 75 bar, proceed as follows: • Change the gas bottle. • Allow the gas bottle to acclimatize at room temperature for approx. 8 hours. If the error continues to be displayed, inform Technical Service. Activation has been interrupted. Please use another probe. Probe malfunction Connect a different probe. Unit malfunction If the error continues to be displayed, inform Technical Service. Activation has been interrupted. Please use another probe. Probe malfunction Connect a different probe. If the error continues to be displayed, inform Technical Service. Check date/clock time adjustment. The date and time may not have been set correctly Have the date and time checked by Technical Service.


55 / 70 7 • Status Messages, Error Messages 80113-401 10.15 An Error has been detected. Activation not possible. The waste gas tube is obstructed. Check whether the waste gas tube is kinked or blocked. Unit malfunction If the error continues to be displayed, inform Technical Service. An Error has been detected. Please use another probe. Probe malfunction Connect a different probe. Unit malfunction If the error continues to be displayed, inform Technical Service. Footswitch is not compatible. Activation not possiblel. The software version of the footswitch is not suitable. Connect a footswitch with a suitable software version. You have connected a footswitch for a different type of unit. Connect the CRYO footswitch (white pedal). Unit malfunction If the error continues to be displayed, inform Technical Service. Unit is overheated. Activation not possible. The ambient temperature at which the unit was stored was too high. Allow the unit to acclimatize at room temperature for approx. 3 hours The ambient temperature during operation is too high. Operate the unit at a room temperature of between +10 °C - +40 °C. There is poor ventilation around the unit housing. Install the device in such a way that there is an unobstructed circulation of air around the housing. Installation in confined wall recesses is prohibited. Unit malfunction If the error continues to be displayed, inform Technical Service. Unit is too cold. Activation not possible. The ambient temperature at which the unit was stored was too low. Allow the unit to acclimatize at room temperature for approx. 3 hours The ambient temperature during operation is too low. Operate the unit at a room temperature of between +10 °C - +40 °C. Unit malfunction If the error continues to be displayed, inform Technical Service. Low input pressure. Please check gas supply. The gas bottle is almost empty. Connect a new gas bottle. The bottle valve is not fully open. Open the bottle valve completely. The gas bottle is too cold. Allow the gas bottle to acclimatize at room temperature for approx. 8 hours. Unit malfunction If the error continues to be displayed, inform Technical Service. High exhaust gas pressure. Please check gas outlet. The waste gas tube is obstructed. Check whether the waste gas tube is kinked or blocked. Unit malfunction If the error continues to be displayed, inform Technical Service. High input pressure. Please check gas supply. The gas bottle is too full or too warm. Allow the gas bottle to acclimatize at room temperature for approx. 8 hours. Unit malfunction If the error continues to be displayed, inform Technical Service. Error text Possible cause(s) What should you do?


7 • Status Messages, Error Messages 56 / 70 80113-401 10.15 No Footswitch connected. No footswitch is connected. Connect a footswitch. The connected footswitch is defective. Connect a different footswitch. Unit malfunction Switch the unit off and on again. If the error continues to be displayed, inform Technical Service. Lost connection to probe. Activation has been interrupted. The electrical contact to the probe was interrupted during activation. Do not loosen the probe from the unit during activation. Reconnect the probe. Connect a different probe. If the error continues to be displayed, inform Technical Service. Critical temperature reached. Let unit cool down. The ambient temperature at which the unit was stored was too high. Allow the unit to acclimatize at room temperature for approx. 3 hours or take an activation break. The ambient temperature during operation is too high. Operate the unit at a room temperature of between +10 °C - +40 °C. There is poor ventilation around the unit housing. Install the device in such a way that the circulation of air around the housing is not obstructed. Installation in confined wall recesses is prohibited. Unit malfunction If the error continues to be displayed, inform Technical Service. You have exceeded the max. ON time. In the Service Setup, an ON time limit (max. freeze interval) is set. You have exceeded the maximum ON time. Restart Freeze mode. If the ON time limit is too short for your needs, arrange to have the setting changed. Probe is not compatible. Please use another probe. You have connected a rigid instrument. In version 1.0.x, only flexible probes are supported. Connect a flexible probe. Probe is not ready for operation. Please use another probe. The probe data carrier contains invalid or corrupt data. Connect a different probe. If the error continues to be displayed, inform Technical Service. Keyboard error. You pressed a button during the self-test. Do not press any buttons when switching on the unit. A button is defective. If the error continues to be displayed, inform Technical Service. VIO-System has been disconnected. The connection to the VIO system is disconnected. Switch on the VIO system. Check the connection between the VIO system and ERBECRYO 2 (ECB cable). Switch the ERBECRYO 2 off and on again. If the error continues to be displayed, inform Technical Service. Error text Possible cause(s) What should you do?


57 / 70 7 • Status Messages, Error Messages 80113-401 10.15 VIO-System has been disconnected. Activation not possible. The VIO system is switched off. Switch on the VIO system. The connection to the VIO system is disconnected. Check the connection between the VIO system and ERBECRYO 2 (ECB cable). VIO System is not compatible. Activation not possible. The software version of the VIO system is not suitable. Use a VIO system with a suitable software version. If the error continues to be displayed, inform Technical Service. Input pressure not sufficient. Please check gas supply. The gas bottle is almost empty. Connect a new gas bottle. The bottle valve is not fully open. Open the bottle valve completely. The gas bottle is too cold. Allow the gas bottle to acclimatize at room temperature for approx. 8 hours. Unit malfunction If the error continues to be displayed, inform Technical Service. Error text Possible cause(s) What should you do?


7 • Status Messages, Error Messages 58 / 70 80113-401 10.15


59 / 70 8 • Technical Data 80113-401 10.15 Chapter 8 Technical Data Power connection Rated supply voltage 100 V to 240 V (± 10%) Rated supply frequency 50 - 60 Hz Line current 0.4 – 0.8 A Terminal for grounding (potential equalization) yes Power fuses 2 x T 1.0 A H 250 V Data specific to the unit Effect setting Setting for indication-specific parameters Number of effect levels 1 - 5 (depending on the instrument) Number of program memory locations 10 Activation Footswitch Cooling gas CO2 Input pressure 45 - 65 bar Operating mode Continuous operation ON time 100% Dimensions and weight Width x height x depth 410 x 130 x 370 mm Weight 6.7 kg Ambient conditions for transport and storage of unit Temperature -20 °C to +55 °C Relative humidity 15 % – 80%


8 • Technical Data 60 / 70 80113-401 10.15 Ambient conditions for operation of unit Temperature +15 °C to +40 °C Relative humidity 30% - 75%, non-condensing Maximum operating altitude 4000 m above sea level Maximum operating altitude 4000 m above sea level Acclimatization of the unit If the unit has been stored or transported at temperatures below +15 °C or above +40 °C, the unit will require approx. 3 hours to acclimatize at room temperature. Gas supply Gas bottles Recommended: Gas bottle with 10 kg capacity Room temperature when operating the gas bottle Recommended: between +15 °C and +25 °C Storing the gas bottle Up to max. +50 °C Acclimatizing the gas bottle If gas bottles are stored at below +15 °C and above +25 °C, the acclimatization time at room temperature is approx. 8 hours. Bottle type Gas bottles without dip tube Standards Classification according to EC Directive 93/42/EEC II b Protection class as per EN 60 601-1 I Type as per EN 60 601-1 CF


61 / 70 9 • Information on electromagnetic compatibility (EMC) 80113-401 10.15 Chapter 9 Information on electromagnetic compatibility (EMC) Where EMC is concerned, medical electrical equipment is subject to special safety measures and must be installed and commissioned according to the EMC instructions stated herein. NOTICE Use of non-approved internal cables by Technical Service This can result in the increased emission of electromagnetic waves or reduce the immunity of the device. The unit may fail or not perform properly. Technical Service may only use the internal cables that are listed in the service manual for the device. NOTICE Stacked devices If you stack the device next to other equipment or with other equipment, the devices can affect each other. The unit may fail or not perform properly. If it is necessary to operate the device near other equipment or stacked together with other equipment, check whether the devices are affecting each other: Are the devices behaving unusually? Do errors occur? Guidance and manufacturer's declaration - electromagnetic emissions The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should ensure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance HF emissions CISPR 11 Group 1 The equipment or system uses HF energy only for its internal function. Therefore its HF emissions are very low and are not likely to cause any interference in nearby electronic equipment. HF emissions CISPR 11 Class B The equipment is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Class A Voltage fluctuations/flicker emissions IEC 61000-3-3 Complies


9 • Information on electromagnetic compatibility (EMC) 62 / 70 80113-401 10.15 Guidance and manufacturer's declaration - electromagnetic immunity The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should ensure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Discharge of static electricity (ESD) in accordance with IEC 61000-4-2 ±6 kV contact discharge ±8 kV air discharge ±6 kV contact dischargea) ±8 kV air dischargea) The floor should be of wood or concrete or be covered with ceramic tiles. If the floor is provided with non-conductive synthetic material, the relative humidity must be at least 30%. Electrical fast transient/burst IEC 61000- 4-4 ±2 kV for power supply lines ±1 kV for input/output lines ±2 kV for power supply lines ±1 kV for input/output lines Mains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5 ±1 kV differential mode ±2 kV common mode ±1 kV differential mode ±2 kV common mode Mains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 <5% U T (>95% dip in U T ) for 0.5 cycle 40% U T (60% dip in U T ) for 5 cycles 70% U T (30% dip in U T ) for 25 cycles <5% U T (>95% dip in U T ) for 5 s <5% U T (>95% dip in U T ) for 0.5 cycle 40% U T (60% dip in U T ) for 5 cycles 70% U T (30% dip in U T ) for 25 cycles <5% U T (>95% dip in U T ) for 5 s Mains power quality should be that of a typical commercial or hospital environment. If the user of the equipment requires continued operation during power mains interruptions, it is recommended that the equipment be powered from an uninterruptible power supply or a battery. Power frequency (50/ 60 Hz) magnetic field IEC 61000-4-8 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Note: UT is the a.c. mains voltage prior to application of the test level.


63 / 70 9 • Information on electromagnetic compatibility (EMC) 80113-401 10.15 Guidance and manufacturer's declaration - electromagnetic immunity The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should ensure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile HF communications equipment should be used no closer to any part of the equipment, including cables, than the recommended separation distance. The separation distance is calculated from various equations depending on the frequency of the portable and mobile HF communications equipment: Recommended separation distance Conducted HF IEC 61000- 4-6 3 Vrms 150 kHz to 80 MHz 3 Vrms Equation 1) d=1.2 P1/2 Radiated HF IEC 61000-4- 3 3 V/m 80 MHz to 800 MHz 3 V/m Equation 2) d=1.2 P1/2 3 V/m 800 MHz to 2.5 GHz 3 V/m Equation 3) d=2.3 P1/2 P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. d is the recommended separation distance in meters (m). Field strengths from fixed transmitters, as determined by an electromagnetic site surveya) should be less than the compliance level in each frequency rangeb). Interference may occur in the vicinity of equipment marked with the following symbol:


9 • Information on electromagnetic compatibility (EMC) 64 / 70 80113-401 10.15 Note 1: At 80 MHz equation 2) applies. At 800 MHz equation 3) applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed HF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment is used exceeds the applicable compliance level above, the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment. b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Recommended separation distances between portable and mobile HF communications equipment and the equipment The equipment is intended for use in an electromagnetic environment in which radiated HF disturbances are controlled. The customer or the user of the equipment can help prevent electromagnetic interference. This can be achieved by maintaining the minimum distance recommended below between the communications equipment (transmitters) and the equipment. The minimum distance depends on the maximum output power and the frequency of the communications equipment. Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m) 150 kHz to 80 MHz d=1.2 P1/2 80 kHz to 800 MHz d=1.2 P1/2 800 MHz to 2.5 GHz d=2.3 P1/2 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For transmitters rated at a maximum output power not listed above, the recommended separation distance can be determined using the equation applicable to the frequency of the transmitter. P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the frequency bands between 80 MHz and 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Guidance and manufacturer's declaration - electromagnetic immunity


65 / 70 10 • Maintenance, Customer Service, Warranty, Disposal 80113-401 10.15 Chapter 10 Maintenance, Customer Service, Warranty, Disposal Maintenance Modifications and repairs Modifications and repairs must not impair the safety of the equipment or equipment cart and accessories for the patient, user and the environment. This condition is met when changes to the structural and functional characteristics are not detrimental to safety. Authorized persons Modifications and repairs may only be undertaken by Erbe or by persons expressly authorized by Erbe. Erbe accepts no liability if modifications and repairs to the unit or accessories are made by unauthorized persons. This will also invalidate the warranty. Technical safety checks The technical safety checks determine whether the safety and operational readiness of the unit or the equipment cart and accessories conform to a defined technical required status. Technical safety checks must be performed at least once a year. What technical safety checks must be performed? For this device the following technical safety checks have been stipulated: • Checking of labels and User Manual • Visual inspection of unit and accessories for damage • Testing the grounded conductor in accordance with IEC 60601-1 Section 18 • Testing the leakage current in accordance with IEC 60601-1 Section 19 • Functional testing of all the unit's operating and control elements • Testing footswitch activation • Seal test, pneumatic test program • Obstruction test, pneumatic test program • Flow measurement, pneumatic test program The results of the safety checks must be documented. If during the safety checks any defects are found which might endanger patients, staff or third parties, the device may not be operated until the defects have been remedied by competent service technicians. Customer service If you are interested in a maintenance contract, please contact Erbe Elektromedizin in Germany, or your local contact in other countries. This may be an Erbe subsidiary, an Erbe representative or a distributor. Warranty The General Terms and Conditions or the conditions of the purchase contract apply.


10 • Maintenance, Customer Service, Warranty, Disposal 66 / 70 80113-401 10.15 Disposal Your product bears a crossed-out garbage can icon (see image). Meaning: In all EU countries this product must be disposed of separately in accordance with the national laws implementing EU Directive 2002/96/EC of January 27, 2003, WEEE. In non-EU countries the local regulations must be observed. If you have any questions about disposal of the product, please contact Erbe Elektromedizin or your local distributor.


67 / 70 11 • Accessories 80113-401 10.15 Chapter 11 Accessories Introduction The following offers an overview of example accessories for each accessory category. A complete overview is available in the Erbe accessories catalog and on the Erbe website. We recommend the use of Erbe accessories.


11 • Accessories 68 / 70 80113-401 10.15 ERBECRYO 2 example accessories Cryoprobes for ERBECRYO 2 Accessories for ERBECRYO 2


69 / 70 Index 80113-401 10.15 Index A Activating the volume setting 48 B Bleed-off after use 48 as a function of the unit 39 Brightness setting 48 C Carbon dioxide dangerous concentration 15, 25 escapes through leaky instrument 13, 36 properties, chemical 17 properties, physical 17 properties, physiological 18 Contrast setting 48 E Effect 37 F Footswitch connecting 26 damage not permitted 12, 33 only use Erbe accessories 13, 24 socket for the footswitch 22 Freezing 39 G Gas wrong type 15, 26, 28 Gas bottle changing (when using a gas cylinder connection) 30 changing (when using a gas switching tap) 31 checking input pressure 46 connecting to the gas cylinder connection 26 connecting with the gas switching tap 28 damage 13, 26, 28 do not operate horizontally 14, 26, 28 gas pressure too high 14 preparing and checking 27, 29 putting into operation 35 storage 18 store and operate upright 17 transport 18 using the gas switching tap to switch between two gas bottles 31 vacuum 14 with dip tube 14, 26, 28 working with 18 Gas connections gas connection 22 waste gas nozzle 22 Gas cylinder connection connecting 27 loosening to change gas bottles 30 national variations 27 removal for transport 18 Gas hose connecting on the gas switching tap 29 connecting to the gas connection 26 connecting to the gas cylinder connection 27 loosening from the gas cylinder connection 30 loosening from the gas switching tap 31 only use Erbe accessories 13, 24 Gas switching tap connecting 29 loosening to change gas bottles 31 Grounding terminal connecting 22 I Instrument calling up information 45 connected incorrectly 13 connecting 36 damage due to excessive gas pressure 14 damaged, leaky 13, 36 disconnecting 49 instrument socket 21 only use Erbe accessories 13, 24 P Power connection 22 electrical connection 33 technical data 59 Power cord, footswitch storage 49 Power fuses 22 Program deleting 44 overwriting 43 saving 40 selection 40 T Thawing 39 Timer Count Down & Beep 38 Count Up 37 Count Up & Beep 38 off 37 setting 37 U Unit disassembly 49 switching off 49 switching on 35 V Volume of alarm sound setting 48 Volume of button sound setting 48 W Waste gas tube connecting 26 disconnecting 49 kinking, trapping 11 only use Erbe accessories 13, 24