Operating Instructions | HERAcell® 150i / 240i
13 Annex: Data communication
Date & Time DATE & TIME
The submenu DATE & TIME is used for adjusting date and time to the
desired time zone.
1. Data in the two text boxes must be entered in the format DD.MM.YY
(day, month, year).
2. Accept the input data:
> Press the SET button.
Error Logger ERROR LOGGER
The submenu ERROR LOGGER is used for reading the error messages into
the text box of the user surface.
50115191 B / 11.2010 The data sets can be stored in the metaformat *.CSV.
151
Operating Instructions | HERAcell® 150i / 240i
13 Annex: Data communication
• Saving data sets as files:
> Press the SAVE TO FILE button.
Data Logger DATA LOGGER
The submenu DATA LOGGER is used for reading the event entries into the
text box of the user surface.
The data sets can be stored in the metaformat *.CSV.
1. Read the data sets:
> Press the READ button.
2. The progress of the data transmission is indicated in the three text
boxes:
> AMOUNT DATA: Total number of transmitted data sets.
> AMOUNT TELEGRAMS: Therein the number of transmitted event
entries.
> AMOUNT ERRORS: Therein the number of transmitted error
messages.
Duration of the data transmission:
As the data logger can contain up to 10,000 data sets, the data
transmission to a PC may take some time.
50115191 B / 11.2010 152
Operating Instructions | HERAcell® 150i / 240i
13 Annex: Data communication
Servicefile SERVICEFILE
The submenu SERVICEFILE is used for reading error informations and for
creating a service file from it, saved with the proprietary extension *srf. The
service file is transmitted to the Technical Service of Thermo Fisher Scien-
tific for fault analysis.
1. Create the service file :
> Press the CREATE SERVICEFILE button.
> Define file name and saving directory in the Windows dialog box.
2. Start saving process:
> Press OK button.
Duration of the procedure:
The compilation of device informations and the creation of the
service file may take some time.
50115191 B / 11.2010 153
Operating Instructions | HERAcell® 150i / 240i
13 Annex: Data communication
Password PASSWORD
The submenu PASSWORD can only be accessed by the service personnel
of Thermo Fisher Scientific.
50115191 B / 11.2010 154
Operating Instructions | HERAcell® 150i / 240i Part number: Date Signature
Service number:
14 Device log Operator's note:
Please list carried out works here: Notes
Device type:
Serial number:
Location:
Work carried out
50115191 B / 11.2010 155