associated with the axis that has the malfunctioning Overtravel limit switch. WARNING The overriding of the Overtravel limit switches should only be

Discover the best professional documents and content resources in AnyFlip Document Base.
Search
Published by , 2016-12-09 04:15:04

004-1055-00 Troubleshooting: 630 - midmark.com

associated with the axis that has the malfunctioning Overtravel limit switch. WARNING The overriding of the Overtravel limit switches should only be

ID Code 04 (Tilt / Motor Control Hub Board) - continued

Error Code Error Definition Probable Cause

0A Overtravel Error Loose wire connection.
Base Up Overtravel limit swit

0b Hall Sensors Loose wire connection.
0C Motor Current Motor Failure.

Excessive load on actuator

0d Motor Run On Error The motor has been detected moving for a
time than the run-on time allo

0E Motor Stall Error Excessive load on actuator
Loose wire connection.

0F Motor Direction Error The motor is being commanded to move i
but is actually moving in anoth

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

e Correction

tch. Check and secure wire connections at J8 on Motor Control Hub PC Board and
at Overtravel limit switch(es). Refer to: Fuses and Connections
r. Disconnect power cord for at least five seconds, then plug back in.
a longer period of Test / Replace switch. Refer to: Tilt Actuator / Limit Switches
ows. Disconnect power cord for at least five seconds, then plug back in.
r. Check and secure hall sensor wire connection at J6 on Motor Control Hub PC Board.
Refer to: Fuses and Connections
in one direction, Replace Tilt actuator motor.
her. Check / Reduce patient load. (650 lb. Maximum)
Check mechanical structure for binding.

Check actuator / motor and limit switches.

Check / Reduce patient load. (650 lb. Maximum)
Check mechanical structure for binding.
Check and secure wire connections at J7 on Motor Control Hub PC Board.
Refer to: Fuses and Connections

Check actuator / motor.

sh-26 004-1055-00 [Revised: 10/15/14]

ID Code 04 (Tilt / Motor Control Hub Board) - continued

Error Code Error Definition Probable Cause

10 Communication Error to Message / Communication fail
Machine Control

11 Corrupt Memory Last save to memory failed

12 Memory Save Error Saving to memory failed.

13 Memory Recall Error Recalling information from memor

14 Memory Erase Error Erasing information from memory

Loose wire connection.

15 I2C Slave Not Present Power Supply PC Board fuse(s)

Motor Control Hub PC Boar

Power Supply PC Board.

Loose wire connection.

16 I2C Slave Not Initialized Power Supply PC Board fuse(s)

Motor Control Hub PC Boar

Power Supply PC Board.

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

e Correction

lure. Error goes away by itself.
d. Disconnect power cord for at least five seconds, then plug back in.
Disconnect power cord for at least five seconds, then plug back in.
ry failed. Replace Motor Control Hub PC Board.
y failed. Disconnect power cord for at least five seconds, then plug back in.
Replace Motor Control Hub PC Board.
blown. Disconnect power cord for at least five seconds, then plug back in.
rd. Replace Motor Control Hub PC Board.
Disconnect power cord for at least five seconds, then plug back in.
blown. Replace Motor Control Hub PC Board.
rd. Check and secure wire connections at J3 on Power Supply PC Board
and J13 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Check and secure wire connections at J16 on Machine Control PC Board
and J14 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Test / Replace F3 & F4 fuses on Power Supply PC Board.
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power
Check and secure wire connections at J3 on Power Supply PC Board
and J13 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Check and secure wire connections at J16 on Machine Control PC Board
and J14 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Test / Replace F3 & F4 fuses on Power Supply PC Board.
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power

sh-27 004-1055-00 [Revised: mo/dd/yr]













ID Code 05 (Foot / Motor Control Hub Board) - continued

Error Code Error Definition Probable Cause

10 Communication Error to Message / Communication fail
Machine Control

11 Corrupt Memory Last save to memory failed

12 Memory Save Error Saving to memory failed.

13 Memory Recall Error Recalling information from memor

14 Memory Erase Error Erasing information from memory

Loose wire connection.

15 I2C Slave Not Present Power Supply PC Board fuse(s)

Motor Control Hub PC Boar

Power Supply PC Board.

Loose wire connection.

16 I2C Slave Not Initialized Power Supply PC Board fuse(s)

Motor Control Hub PC Boar

Power Supply PC Board.

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

e Correction

lure. Error goes away by itself.
d. Disconnect power cord for at least five seconds, then plug back in.
Disconnect power cord for at least five seconds, then plug back in.
ry failed. Replace Motor Control Hub PC Board.
y failed. Disconnect power cord for at least five seconds, then plug back in.
Replace Motor Control Hub PC Board.
blown. Disconnect power cord for at least five seconds, then plug back in.
rd. Replace Motor Control Hub PC Board.
Disconnect power cord for at least five seconds, then plug back in.
blown. Replace Motor Control Hub PC Board.
rd. Check and secure wire connections at J3 on Power Supply PC Board
and J13 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Check and secure wire connections at J16 on Machine Control PC Board
and J14 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Test / Replace F3 & F4 fuses on Power Supply PC Board.
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power
Check and secure wire connections at J3 on Power Supply PC Board
and J13 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Check and secure wire connections at J16 on Machine Control PC Board
and J14 on Motor Control Hub PC Board. Refer to: Fuses and Connections
Test / Replace F3 & F4 fuses on Power Supply PC Board.
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power
Check supply power to Motor Control Hub PC Board.
Refer to: Motor Control Hub PC Board Supply Power

sh-31 004-1055-00 [Revised: mo/dd/yr]

Testing

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

sh-32 004-1055-00 [Revised: mo/dd/yr]

Diagnostic LED’s

Isolating a Malfunction

This illustration shows only the LED’s used for diagnostic trouble shooting.

Activity Definition Foo
On Continuously Everything OK, No activity Actua

Slow Flashing Communicating
Fast Flashing Malfunction
OFF No power

TP210 Rev. A MA9093i

Note

The Machine Control PC Board is mou
on the Power Supply PC Board.

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Wireless Controls.......................................... 36
ot Tilt Back
ator Actuator Actuator

J14

Motor Control Hub PC Board

Machine Control PC Board

unted Foot Wireless
sh-33 Control Hand Controls
Control

004-1055-00 [Revised: mo/dd/yr]

Wired Controls

Isolating a Malfunction

To isolate a malfunction, try activating the inoperable function(s) from hand control and foot con

If function(s) are inoperable from the foot / hand control...
A) Plug the foot / hand control into another connection port.
B) Secure Inlet harness connection(s) on Machine Control PC Board.
Note: J2 thru J5 and J8 thru J11 are interchangeable.
If function(s) are operable: (LED illuminated)
C) Replace Inlet harness at malfunctioning connection port.
If function(s) are still inoperable: (LED not illuminated)
D) Perform the Foot / Hand Control Test.

J6
J1 J2 J3 J4 J5

J12
J8 J9 J10 J11

J16 Machine S2 Fo
Control Co
TP210 Rev. A Englis
PC Board

J18
J19

Foot Hand
Control Control

LED LED

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15

ntrol.

Connection Connection
Port Port
Connection
Port

Hand
Control
oot
ontrol

sh-34 004-1055-00 [Revised: mo/dd/yr]

Wired Controls - continued

Foot / Hand Control Test

Hand / Foot Control PC Board

(Located in controller)

Activity
On Continuous

OFF

Activity
On Continuous

Slow Flashing
Fast Flashing
OFF

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15

PWR D5 (LED) Solution
Definition Controller is good
sly Controller has power
Replace Controller
Malfunction, No power

COM D6 (LED) Solution
Definition Controller is good
sly No activity
Controller is good
Communicating Replace Controller
Malfunction Replace Controller
No power

sh-35 004-1055-00 [Revised: mo/dd/yr]

Wireless Controls

Isolating a Malfunction

Testing A

TP210 Rev. A • When any control button is pressed,
that controller sounds a single “beep”.
(All functions operate)
Cause: 1. Low battery
2. Faulty PC board in controller
Solution: 1. Replace batteries (size: AA)
2. Replace controller
• When any control button is pressed, nothing happens.
(No movement, controller “beeps”)
Check: 1. Perform Association Procedure
• When any control button is pressed, nothing happens.
(No movement, no “beeps”)
Check: 1. Perform Association Procedure
2. LEDs on base station PC board
If LED status is normal (see illustration):
• Check LEDs on controller PC board (see next page)
If no LEDs are illuminated, check the following:
• Modular harness (loose connections, damage, etc.)
• Power supply to chair
• Replace base station PC board if necessary

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15

Antenna Equipment Alert
Cable
Antenna connection is very delicate.
Do not kink wire or apply stress to connection.

LEDs Normal Status

TX (OFF - except during Association Procedure)

RX (ON - except during Association Procedure)

COM (ON)

PWR (ON)

Base Station PC Board

Modular
Harness

sh-36 004-1055-00 [Revised: mo/dd/yr]

Wireless Controls - continued

Foot / Hand Control Test

Testing - continued

• When any control button is press
(No “beeps”, no movement, etc.)
(see previous page for Checks 1 &
Check: 3. LEDs on hand / foot
(press any button on
If LEDs function prop
• Check antenna con
• Replace antenna if
If LEDs do not functio
• Replace batteries i
• Replace controller

LEDs

TP210 Rev. A Hand / Foot C

(Locate

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15

sed, nothing happens. - continued

2)
control PC board
n hand / foot control)
perly (see illustration):
nnection. (see parts list for detail)
f necessary
on properly:
in controller (size: AA)
if necessary

s Normal Status

D2 (ON - only when a button on controller is pressed)
D3 (OFF - except during Association Procedure)

Control PC Board 004-1055-00 [Revised: mo/dd/yr]

ed in controller)

sh-37

Wireless Controls - continued

Association Procedure

The wireless controllers will only function with the table they have been “associated”
with. This prevents unwanted movement in locations with more than one table.
Each wireless controller’s association is preset at the factory. This procedure is only
required in the event of a malfunction, if a new controller is purchased, or if you are
attempting to use the controller with a table other then the one it was originally
associated with.

NOTE: The base station PC board only stores the last two associated controllers. To ensure proper
operation, Midmark recommends reassociating both controllers (hand & foot), even if only one new
controller is purchased.

Home

TP210 Rev. A Stop

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15

Association Procedure:
A) Disconnect power to the table for at least five seconds.
B) Reconnect power, wait for at least two seconds,
then press & hold the Stop & Home buttons on the
wireless controller until you hear three “beeps”.
C) Check for proper operation of wireless controller.
D) If associating a hand and foot control, repeat procedure
using the other wireless controller.

Note: Step B, do not wait for more than ten seconds to
press & hold the Stop & Home buttons.

Home Stop

sh-38 004-1055-00 [Revised: mo/dd/yr]

Smart Sensor Switches

Isolating a Malfunction

This illustration shows the Smart Sensor Switches and their location.

Treatment Pan
Switch (NO)

Seat Sensor
Switches (NC)

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Foot Smart Sensor Switch Access................ 77

Foot Extension
Switch (NO)

Foot Sensor
Switches (NC)

sh-39 004-1055-00 [Revised: mo/dd/yr]





Base Actuator / Limit Switches

This illustration shows the base limit switches and the serviceable components of the b
Use the table below to isolate the malfunction.

Problem Required Action
Motor runs, but makes grinding noise Clean / lube actuator threads
(Lube threads w/lithium grease)
Motor runs, but table does not move Replace actuator if necessary
Motor does not run
Inspect / replace motor coupler

Test Limit Switch(es)

Equipment Alert Up Lim
Switch (
Anytime the limit switch wires or the actuator wires
are disconnected, the PC Board must be calibrated.

Note

Anytime an Overtravel Limit Switch is detected open,
both Up and Down functions are deactivated.

Down Limit
Switch (NC)

TP210 Rev. A Table
Base
Englis
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |

base actuator. Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Base Function Override Switch..................... 55
Base Shrouds / Cladding Removal............... 73
Base Actuator / Motor Coupler .....003-2722-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

Motor Actuator
Coupler Motor
mit
(NC)

Up Overtravel
Limit Switch

(NC)

sh-41 004-1055-00 [Revised: mo/dd/yr]

Back Function / Fuses and Connections

Isolating a Malfunction

This illustration shows only the fuses and connections that affect the Back UP / DOWN

Equipment Alert

Anytime the limit switch wires or the actuator wires
are disconnected, the PC Board must be calibrated.

Equipment Alert

Always remove power from table before disconnecting or connectin
J13 on Motor Control Hub PC Board or damage to PC Board may o

J1 J3 J5 J7
F3 F4 J6 J4
115V
J6
F1 F2 J1 J2 J3 J4 J5
J9 J2
J12
F5 J8 J9 J10 J11

J10 J16 S2
J131
J18
TP210 Rev. A J19

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

N function. Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Back Actuator / Limit Switches...................... 43
Power Supply PC Board................003-2715-00
Motor Control Hub PC Board........003-2717-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

ng TPA
occur.

TPB3 TPB1 TPC3 TPC1 TPD3 TPD1

J14 J9 J5 J1
J15 TPB2
F3 TPC2 TPD2
J13
J12 J11 J10 J8 J7 J6 J4 J3 J2

sh-42 004-1055-00 [Revised: 6/5/14]

Back Actuator / Limit Switches

Isolating a Malfunction

This illustration shows the back limit switches and the serviceable components of the b
Use the table below to isolate the malfunction.

Problem Required Action
Motor runs, but makes grinding noise
Clean / lube actuator threads
Motor runs, but table does not move (Lube threads w/lithium grease)
Motor does not run Replace actuator if necessary

Inspect / replace motor coupler

Test Limit Switch(es)
Refer to: Back Actuator Test

Up Limit
Switch (NC)

Down Limit
Switch (NC)

Up Overtravel
Limit Switch

(NC)

Down Overtravel
Limit Switch (NC)

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Englis

back actuator. Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Back Function Override Switch..................... 56
Back Actuator / Motor Coupler .....003-2721-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

Equipment Alert

Anytime the limit switch wires or the actuator wires
are disconnected, the PC Board must be calibrated.

Note

Anytime an Overtravel Limit Switch is detected open,
both Up and Down functions are deactivated.

Motor
Coupler

Actuator
Motor

Limit Switch 004-1055-00 [Revised: mo/dd/yr]
Stops

Equipment Alert

Limit Switch Stops are pre set at factory.
Do not adjust or damage to table may occur.

sh-43

Back Actuator Test

Isolating a Malfunction
Equipment Alert

The wire connections and wire routing must be put
back in original locations before installing covers.

Equipment Alert

Failure to move all three connections
may result in damage to table.

Step 1:
A) Disconnect power to the table.
B) Disconnect Tilt actuator connections.
C) Move Back wire connection J10 to Tilt PC Board connection J6.
D) Move Back wire connection J11 to Tilt PC Board connection J7.
E) Move Back wire connection J12 to Tilt PC Board connection J8.
F) Connect power to table.

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Back Actuator / Motor Coupler .....003-2721-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

Step 2: Using the hand / foot control:
A) Press & hold Tilt UP button briefly.
B) Press & hold Tilt DOWN button briefly.

Did Back move Up and Down in Step 2? Required Action
YES Actuator is OK

NO Replace actuator

J14

Motor Control Hub PC Board

Equipment Alert

The wire connections and wire routing must be put
back in original locations before installing covers.

sh-44 Step 3:
A) Reconnect Back actuator connections.
B) Reconnect Tilt actuator connections.
C) Calibrate Table. Refer to: Calibration Procedure

004-1055-00 [Revised: mm/dd/yr]

Tilt Function / Fuses and Connections

Isolating a Malfunction

This illustration shows only the fuses and connections that affect the Tilt UP / DOWN f
Equipment Alert

Anytime the limit switch wires or the actuator wires
are disconnected, the PC Board must be calibrated.

Equipment Alert

Always remove power from table before disconnecting or connecting
J13 on Motor Control Hub PC Board or damage to PC Board may occur.

J1 J3 J5 J7
F3 F4 J6 J4
115V
J6
F1 F2 J1 J2 J3 J4 J5
J9 J2
J12
F5 J8 J9 J10 J11

J10 J16 S2
J131
J18
J19

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

function. Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Smart Sensor Switches................................. 39
Tilt Actuator / Limit Switches......................... 46
Power Supply PC Board................003-2715-00
Motor Control Hub PC Board........003-2717-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

TPA TPB3 TPB1 TPC3 TPC1 TPD3 TPD1

J14 J9 J5 J1
J15 F2
TPB2 TPC2 TPD2
J13
J12 J11 J10 J8 J7 J6 J4 J3 J2

sh-45 004-1055-00 [Revised: 6/5/14]

Tilt Actuator / Limit Switches

Isolating a Malfunction

This illustration shows the tilt limit switches and the serviceable components of the tilt
Use the table below to isolate the malfunction.

Problem Required Action
Motor runs, but makes grinding noise
Clean / lube actuator threads
Motor runs, but table does not move (Lube threads w/lithium grease)
Motor does not run Replace actuator if necessary

Inspect / replace motor coupler

Test Limit Switch(es)
Refer to: Tilt Actuator Test

Motor
Coupler

Limi
S

TP210 Rev. A D
Li
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

actuator. Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Tilt Function Override Switch......................... 56
Tilt Actuator / Motor Coupler .........003-2724-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

Equipment Alert

Anytime the limit switch wires or the actuator wires
are disconnected, the PC Board must be calibrated.

Note

Anytime an Overtravel Limit Switch is detected open,
both Up and Down functions are deactivated.

Equipment Alert

Limit Switch Stops are pre set at factory.
Do not adjust or damage to table may occur.

Actuator Limit Switch
Motor Stops

it Switch Up Limit
Stops Switch (NC)

Down Limit Up Overtravel
Switch (NC) Limit Switch
Down Overtravel
imit Switch (NC) (NC)

sh-46 004-1055-00 [Revised: mo/dd/yr]

Tilt Actuator Test

Isolating a Malfunction
Equipment Alert

The wire connections and wire routing must be put
back in original locations before installing covers.

Equipment Alert

Failure to move all three connections
may result in damage to table.

Step 1:
A) Disconnect power to the table.
B) Disconnect Back actuator connections.
C) Move Tilt wire connection J6 to Back PC Board connection J10.
D) Move Tilt wire connection J7 to Back PC Board connection J11.
E) Move Tilt wire connection J8 to Back PC Board connection J12.
F) Connect power to table.

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

Step 2: Using the hand / foot control:
A) Press & hold Back UP button briefly.
B) Press & hold Back DOWN button briefly.

Did Tilt move Up and Down in Step 2? Required Action
YES Actuator is OK
Refer to: Tilt Actuator Motor Resistance Test
NO

J14

Motor Control Hub PC Board

Equipment Alert

The wire connections and wire routing must be put
back in original locations before installing covers.

sh-47 Step 3:
A) Reconnect Tilt actuator connections.
B) Reconnect Back actuator connections.
C) Calibrate Table. Refer to: Calibration Procedure

004-1055-00 [Revised: 8/14/14]

Tilt Actuator Motor Resistance Test

Equipment Alert

Anytime actuator wires are disconnected,
the table must be calibrated.

Step 1:
A) Disconnect power to the table.
B) Disconnect tilt actuator from J6 on PC Board.
C) Set multimeter to Ohms to check resistance.

J2 J3 J4 J6

TPD2 F1

J1

TPD3

TPD1

M

TP210 Rev. A

© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | Englis

Troubleshooting Chart..................................... 2
Error Code Chart........................................... 15
Tilt Actuator / Motor Coupler .........003-2724-00
Wiring Diagrams............................004-1045-00
Theory of Operation......................004-1060-00

Step 2:
A) Place the Red meter probe on the Black wire,
and the Black meter probe on the Blue wire.
B) Place the Red meter probe on the Black wire,
and the Black meter probe on the Green wire.
C) Place the Red meter probe on the Black wire,
and the Black meter probe on the White wire.
If any of the readings are less than two Mega Ohms
after three seconds, the motor is defective.

6 J7 J8 J10 J11 J12 J13

TPC2 F2 TPB2 F3 J15
J14
J5 J9

TPC3 TPB3 TPA

TPC1 TPB1

Motor Control Hub PC Board

sh-48 004-1055-00 [Revised: 10/15/14]