The words you are searching are inside this book. To get more targeted content, please make full-text search by clicking here.
Discover the best professional documents and content resources in AnyFlip Document Base.
Search
Published by Hugo Naveros Mendoza, 2024-05-29 11:33:42

MT&O-8XL

MT&O-8XL

Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 47 Clear Silent Abort Clear Silent Abort is used to correct a Silent Abort error. This procedure is only used to clear specified errors and should not be performed unless directed by those error analysis procedures. DANGER Make sure static discharge wrist ground straps are worn when setting configuration switches or when working with the Steri-Vac 8XL Sterilizer covers removed. Static discharge will cause damage to equipment. Failure to wear static discharge straps can result in damage to equipment or injury to personnel. Note: Do not cycle power. 1. Set SW1 configuration switches 3 and 7 to the closed position (see Figure 28). 2. Set SW1 configuration switches 1, 2, 4-6, and 8 open (see Figure 28). During Clear Silent Abort the sterilizer automatically does the following: • Removes the error from the display and starts printing the report. • Punctures gas cartridge for one minute. • Runs a time based final vacuum for 18 minutes. • Opens the vent and purges for 30 minutes. • Closes the vent and locks up. • After Clear Silent Abort, the screen clears and the error is displayed. The report stops printing and the chamber parameters display. 3. If an error is still displayed after the gas is removed from the chamber, return to the error troubleshooting procedure to clear the error. 1 2 3 4 5 6 7 8 SW1 Dip Switch #s Clear Abort (System Reset Mode) Figure 28. SW1 Configuration Switch Positions to Clear Silent Abort.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 48  3M 2004 October Reset Site Setup Reset Site Setup is automatically performed by the sterilizer when an E1 or E2 checksum error occurs or when a new controller circuit board is installed. Reset Site Setup can be forced by the service representative for troubleshooting purposes or if one of the conditions previously mentioned did not automatically cause a reset to occur. Note: When a new controller board is installed and the sterilizer automatically performs a system reset, three long beeps will be heard when the sterilizer is powered up. This indicates the system parameters are set to their default settings. 1. Power Off the Steri-Vac 8XL Sterilizer. DANGER Make sure static discharge wrist ground straps are worn when setting configuration switches or when working with the Steri-Vac 8XL Sterilizer covers removed. Static discharge will cause damage to equipment. Failure to wear static discharge straps can result in damage to equipment or injury to personnel. 2. Set SW1 configuration switches 2 and 7 to the closed position (see Figure 29). 3. Set SW1 configuration switches 1, 3-6, and 8 open (see Figure 29). 4. Power On the sterilizer. 5. Six short beeps will be heard indicating the system parameters are set to their default settings. 6. A final vacuum, purge, and mandatory aeration are performed. 1 2 3 4 5 6 7 8 SW1 Dip Switch #s Reset Site Setup (System Reset Mode) Figure 29. SW1 Configuration Switch Positions to Reset Site Setup.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 49 Error Code Flow Description Each of the processes tested before or during the sterilization cycle has a specific error number associated with it. The error codes are selected to indicate to the operator or the service rep when the failure was detected (before the cycle, before gas was injected, or after gas was injected) and what process or component most likely failed (examples: RAM test, initial vacuum, final vacuum, or temperature during gas exposure).


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 50  3M 1997 April Error Code Listing Error Code Error Message Condition E1 Cycle Information Checksum Invalid Displayed when the cycle information checksum is invalid. This error automatically assumes the worst case scenario, gas in chamber. The final vacuum, purge, and mandatory aeration are performed. Under most circumstances, this error never occurs. Three medium beeps provide an audible indication of this error. E2 Setup Information Checksum Invalid Displayed when the site setup information checksum is invalid. This error automatically assumes the worst case scenario, gas in chamber. The final vacuum, purge, and mandatory aeration are performed. Under most circumstances, this error never occurs. Four medium beeps provide an audible indication of this error. E3 Electronics Failure This error does not allow the user access to the Steri-Vac 8XL Sterilizer. If this electronics failure occurs, video may not be initialized. Three long beeps provide an audible indication of this error. E4 Chamber Temperature Sensor Temperature sensor range is 5°C to 65°C. If the temperature exceeds this range at power up or at the start of a cycle, E4 is displayed. E5 Heatsink Temperature Sensor Temperature sensor range is 5°C to 99°C. If the temperature exceeds this range at power up or at the start of a cycle, E5 is displayed. E6 Vacuum Sensor Failure Indicates a vacuum sensor failure. This is only a range check and is designed to detect a disconnected sensor. E7 Vacuum Sensor With Door Open Failure Indicates a range failure while the door is open, out-of-range ambient pressure. E8 Wall Temp Sensor Failure, End Temperature sensor range is 5°C to 65°C. If the temperature exceeds this range at power up or at the start of a cycle, E8 is displayed. E9 Wall Temp Sensor Failure, Middle Temperature sensor range is 5°C to 65°C. If the temperature exceeds this range at power up or at the start of a cycle, E9 is displayed. E10 Low Water Error If the water is low, C2 caution appears during idle. If a cycle is started with C2 caution displayed, E10 occurs. E11 Compressed Air Failure If the compressed air pressure is low, C4 caution appears during idle. If a cycle is started with C4 caution displayed, E11 occurs. E12 No Corresponding Error Description E13 No Gas Cartridge Inserted No gas cartridge has been inserted or the last cartridge was not removed and replaced with a new cartridge. E14 Loading Door Open Load door open after the Start softkey is pressed. E15 Unloading Door Open Unload door open after the Start softkey is pressed. C6 caution is displayed until the unload door is closed. E16 Unloading Door Unlocked Unload door unlocked after the Start softkey is pressed. E17 Loading Door Unlock Load door unlocked after the Start softkey is pressed. E18 - E19 No Corresponding Error Description


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 51 E20 Chamber Too Hot (more than 3°C above selected cycle temperature) The temperature of the chamber is 3°C hotter than the selected temperature cycle. E21 No Vacuum This error occurs within one minute after vacuum is started for early detection of vacuum failures. E22 Initial Vacuum Failure A time-out condition occurred while pulling a vacuum. E23 Chamber Preheat Failure A time-out during the preheating of the chamber to the temperature cycle selected. E24 Heatsink PreHeat Time-out Indicates a time-out in the preheating of the heatsink to the temperature of 95°C. E25 Chamber Over Temperature Error Indicates a temperature control problem. The chamber is at least 3°Cs hotter than the selected temperature cycle. E26 Chamber Under Temperature Error Indicates a temperature control problem. The chamber is at least 3°C cooler than the selected temperature cycle. E27 No Corresponding Error Description E28 No Moisture Injected Indicates problems with the water injection and RH chamber systems. If the heatsink doesn't decrease in temperature after the second moisture injection, this error is displayed. If the RH doesn't increase by at least 5% after three injections, this error is displayed. E29 Low Humidity Error Indicates too many moisture injections occur. E30 RH Sensor Failure RH sensor range is 1% to 99%. If the RH exceeds this range at power up or at the start of a cycle, E30 is displayed. E31 Chamber Leak Indicates a leak into the chamber. E32 Pressure Level Before Puncture Indicates an improper pressure before end of precondition and start of gas injection. E33 Interlock Relay Close Error Indicates a latching relay failure. This failure prevents the puncture solenoid from being energized. E34 - E37 No Corresponding Error Description E38 Nitrogen Pressure Error Indicates nitrogen tank pressure switch failed. E39 Nitrogen Injection Time-out Error Indicates a nitrogen injection problem by monitoring the pressure in the chamber. E40 User Interruption During Precondition Indicates Stop Button was pressed during the precondition stage. E41 - E49 No Corresponding Error Description E50 Gas Cartridge Empty Indicates no pressure increase in the chamber after the cartridge is punctured in the gas injection stage. E51 Chamber Vacuum Leakage Indicates chamber pressure too close to ambient pressure and vacuum can not be maintained. E52 Chamber Temperature Under Selected Cycle During Gas Expose Indicates a temperature control problem. The chamber is at least 3°C cooler than the selected temperature cycle.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 52  3M 1997 April E53 Chamber Temperature Over Selected Cycle During Gas Expose Indicates a temperature control problem. The chamber is at least 3°C hotter than the selected temperature cycle. E54 Extended Power Outage Indicates a chamber pressure problem after a power outage. E55 Undefined Error Condition Indicates an unknown condition error. This error occurs if (1) A self test error code occurred during a stage other than cycle start stage. (2) A precondition error code occurs during gas expose stage. (3) A unknown stage occurred. Verifies the error occurred in the proper stage of the cycle. E56 - E59 No Corresponding Error Description E60 User Interruption During Gas Expose Indicates the Stop button was pressed during the gas expose stage. E61 - E70 No Corresponding Error Description E71 Final Vacuum Pump down Error Indicates a time-out condition while pulling a vacuum in the final vacuum stage. E72 Obstructed Air Inlet Indicates a clogged air inlet filter. E73 Vacuum Sensor Gas Error Indicates a vacuum sensor failure while gas is in the chamber. E74 No Corresponding Error Description E75 Gas Low Pressure Error Indicates an insufficient pressure increase after the gas is injected. E76 Interlock Relay Open Error Interlock won't open to indicate no gas in chamber, or gas is in the chamber and the door unlock function was called, or the cycle was complete and gas was still in chamber. This is a safety precaution. Before access is allowed into the chamber, the interlock relay must be open. E77 Puncture Pin Stuck Error Indicates the puncture pin is stuck in the up position or switch is misaligned. E80 Abator Error Indicates a malfunctioning abator. Abator not ready in 60 minutes after final vacuum starts.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 53 Error Code Error Message Condition C1 Air Flow Caution Loss of air flow in exhaust hood. C2 Low Water Caution Low water in the container. C3 Power Interruption Caution Loss of system power during a cycle. C4 Compressed Air Caution Loss of compressed air. C5 Temperature Failed during Aeration Caution Improper temperature during aeration. C6 Unload Door Not Shut Caution The unload door is open. C7 Door Locked Caution The load door is locked when it should be unlocked. C8 Printer Out of Paper Caution No paper or printer ribbon cable is reversed on one end C9 Printer Failure Caution Printer problem. C10 Abator Caution Abator problem. C11 Replace Cartridge Caution Ethylene Oxide cartridge may not have been replaced. C12 Calibration Required Possible calibration problem. Check and adjust REF voltage.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 54  3M 1997 April E1 - Cycle Information Checksum Invalid The Cycle Information Checksum Invalid error is displayed when the cycle information checksum is invalid. This error automatically assumes the worst case scenario of gas in chamber. The final vacuum, purge, and mandatory aeration will be performed. Under most circumstances this error never occurs. Three medium beeps provide an audible indication of this error. Checked • At power up Possible Causes • A new device (U2, U23, U24, U27) was installed without resetting the setup information on the first power up • Unstable power supply • Power supply short Note: When replacing components, make sure you Reset Site Setup (page 48) to default all site setup information. Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button. E2 - Setup Information Checksum Invalid The Setup Information Checksum Invalid error is displayed when the site setup information checksum is invalid. This error will automatically assume the worst case scenario of gas in chamber. The final vacuum, purge, and mandatory aeration will be performed. Under most circumstances this error will never occur. Four medium beeps provide an audible indication of this error. Checked • At power up Possible Causes • New device (U2, U23, U24, U27) was installed without resetting the setup information on the first power up • Unstable power supply • Power supply short Note: When replacing components, make sure you Reset Site Setup (page 48) to default all site setup information. Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 55 E3 - Electronics Failure The Electronics Failure error will not allow the user access to the sterilizer. Note: If this error occurs during power up, the video display may not be initialized. Three long beeps provide an audible indication of this error. If this error occurs during precondition or purge, the sterilizer will lock up until the error is manually cleared by the service representative. See Reset Cycle to Idle (page 46) for clearing procedure. Checked • At power up • During precondition • During purge Possible Causes • EEPROM checksum failure U23 • RH set point saved failure • RAM failure U24 Clearing The sterilizer stops and ignores all softkeys until power is cycled. E4 - Chamber Temperature Sensor The valid range for this temperature sensor is 5°C to 65°C. If the temperature exceeds this range at either power up or at the start of a cycle, E4 will be displayed. Checked • At power up • At the start of each cycle Possible Causes • Disconnected sensor, faulty connection • Chamber too hot Clearing The sterilizer stops and ignores all softkeys until power is cycled.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 56  3M 1997 April E5 - Heatsink Temperature Sensor The valid range for this temperature sensor is 5°C to 99°C. If the temperature exceeds this range at either power up or at the start of a cycle, E5 will be displayed. Checked • At power up • At the start of each cycle Possible Causes: • Disconnected sensor, faulty connection • Chamber too hot Clearing The sterilizer stops and ignores all softkeys until power is cycled. E6 - Vacuum Sensor Failure The Vacuum Sensor Failure error indicates a vacuum sensor failure. This is only a range check and is designed to detect a disconnected sensor. Checked • At power up • At the start of each cycle Possible Causes: • Disconnected sensor, faulty connection Clearing The sterilizer stops and ignores all softkeys until power is cycled.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 57 E7 - Vacuum Sensor with Door Open Failure The Vacuum Sensor with Door Open Failure error indicates a range error while the door is open, out-of-range ambient pressure. Checked • At power up, with door open • At the start of each cycle, if door is open Possible Causes: • Disconnected sensor, faulty connection • Sensor failure Clearing The sterilizer stops and ignores all softkeys until power is cycled. E8 - Wall Temperature Sensor Failure, End The valid range for this temperature sensor is 5°C to 65°C. If the temperature exceeds this range at either power up or at the start of a cycle, E8 will be displayed. Checked • At power up • At the start of each cycle Possible Causes: • Disconnected sensor, faulty connection • Sensor failure Clearing The sterilizer stops and ignores all softkeys until power is cycled.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 58  3M 1997 April E9 - Wall Temperature Sensor Failure, Middle The valid range for this temperature sensor is 5°C to 65°C. If the temperature exceeds this range at either power up or at the start of a cycle, E9 will be displayed. Checked • At power up • At the start of each cycle Possible Causes: • Disconnected sensor, faulty connection • Sensor failure Clearing The sterilizer stops and ignores all softkeys until power is cycled. E10 - Low Water Error If the water is low, caution C2 will appear during idle. If a cycle is started with this caution displayed, E5 will occur. Checked • At the start of each cycle Possible Causes: • Water low • Disconnected sensor, faulty connection • Sensor failure Clearing 1. Raise the door handle. 2. After door opens, press Stop.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 59 E11 - Compressed Air Failure The Compressed Air Failure error indicates the compressor air pressure is low. Caution C4 will appear during idle. If a cycle is started with this caution displayed, E11 will occur. Checked • At the start of each cycle Possible Causes: • Compressed air disconnected • Compressed air pressure insufficient • Disconnected sensor, faulty connection Clearing 1. Raise the door handle. 2. After door opens, press Stop. E12 - No Corresponding Error Description


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 60  3M 1997 April E13 - No Gas Cartridge Inserted The No Gas Cartridge Inserted error indicates no gas cartridge has been inserted, or the last cartridge was not removed and replaced with a new cartridge. Checked • At the start of each cycle Possible Causes: • No gas cartridge • Gas cartridge has not been removed • Disconnected sensor or faulty connection Clearing 1. Raise door handle. 2. After door opens, press Stop. E14 - Loading Door Open The Loading Door Open error indicates the load door is open after the Start softkey is pressed. Checked • At the start of each cycle Possible Causes: • Load door is open • Disconnected or misaligned switch, faulty connection Clearing 1. If the door is open: a. Press Stop b. Close door c. Restart cycle. 2. Otherwise: a. Raise door handle. b. After door opens, press Stop.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 61 E15 - Unloading Door Open Note: This error only occurs with the 2-door model of the Steri-Vac 8XL Sterilizer. The Unloading Door Open error indicates the unload door is open after the Start softkey is pressed. Caution C6 will be displayed until the unload door is closed. Checked • At the start of each cycle Possible Causes: • Unload door is open • Disconnected or misaligned switch, or faulty connection Clearing 1. Close the unload door. 2. If the door is open: a. Press Stop b. Close door c. Restart cycle. 3. Otherwise: a. Raise door handle. b. After door opens, press Stop. E16 - Unloading Door Locked The Unloading Door Locked error indicates the unload door is unlocked after the Start softkey is pressed. Checked • At the start of each cycle Possible Causes: • Unload door locking latch is stuck • Disconnected or misaligned switch, or faulty connection Clearing 1. If the door is open: a. Press Stop b. Close door c. Restart cycle. 2. Otherwise; a. Raise door handle. b. After door opens, press Stop.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 62  3M 1997 April E17 - Loading Door Unlock The Loading Door Unlock error indicates the load door is unlocked after the Start softkey is pressed. Checked • At the start of each cycle Possible Causes: • Load door locking latch is stuck • Disconnected or misaligned switch, or faulty connection • Door handle not in closed position Clearing 1. If the door is open: a. Press Stop b. Close door c. Restart cycle. 2. Otherwise: a. Raise door handle. b. After door opens, press Stop. E18 - E19 - No Corresponding Error Description


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 63 E20 - Chamber Too Hot (more than 3°C above selected cycle temperature) The Chamber Too Hot error indicates the temperature of the chamber is 3°C hotter than the selected temperature cycle. Checked • At the start of precondition Possible Causes: • Chamber needs to cool down Clearing 1. Raise door handle. 2. After door opens, press Stop. E21 - No Vacuum The No Vacuum error occurs within one minute after vacuum is started for early detection of vacuum failures. Checked • At the start of precondition (pressure must decrease 52 mBars in 60 seconds) Possible Causes: • No compressed air • Chamber leak • Faulty vacuum sensor Clearing 1. Raise door handle. 2. After door opens, press Stop.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 64  3M 1997 April E22 - Initial Vacuum Failure The Initial Vacuum Failure error indicates a time-out condition occurred while pulling vacuum. Checked During precondition • Pressure must decrease to puncture pressure 40 minutes, after temperature is reached • Pressure must decrease to puncture pressure 10 minutes, before humidification • Pressure must decrease to puncture pressure 40 minutes, after nitrogen injection Possible Causes: • Chamber leak • Faulty vacuum sensor • Defective vacuum pump • Low compressor air pressure • Restricted vent line Clearing 1. Raise door handle. 2. After door opens, press Stop. E23 - Chamber Preheat Failure The Chamber Preheat Failure error indicates a time-out condition occurred while preheating the chamber to the temperature cycle selected. Checked • During precondition, time-out of 45 minutes • During Recovery of gas expose stage, time-out of 45 minutes Possible Causes: • Chamber High Limit thermostat • Solid state relay failure, SSR1 or SSR2, always open circuited • Calibration of end and middle temperature sensors • Disconnected Heater Blanket Clearing 1. Raise door handle. 2. After door opens, press Stop.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 65 E24 - Heatsink Preheat Time-out The Heatsink Preheat Time-out error indicates a time-out condition occurred while preheating the heatsink to the temperature of 95°C. Checked • During precondition, after initial pumpdown, time-out of 15 minutes Possible Causes: • Heatsink High Limit thermostat • Solid state relay failure, SSR3 Clearing 1. Raise door handle. 2. After door opens, press Stop. E25 - Chamber Over Temperature The Chamber Over Temperature error indicates a temperature control problem. The chamber temperature is at least 3°C s hotter than the selected temperature cycle. Checked • During precondition, humidification • At the end of precondition • During gas expose stage Possible Causes: • Chamber High Limit thermostat • Solid state relay failure, SSR1 or SSR2 shorted Clearing 1. Raise door handle. 2. After door opens, press Stop.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 66  3M 1997 April E26 - Chamber Under Temperature The Chamber Under Temperature error indicates a temperature control problem. The chamber temperature is at least 3°C cooler than the selected temperature cycle. Checked • During precondition, humidification • At the end of precondition Possible Causes: • Chamber High Limit thermostat • Solid state relay failure, SSR1 or SSR2 always open circuited Clearing 1. Raise door handle. 2. After door opens, press Stop. E27 - No Corresponding Error Description


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 67 E28 - No Moisture Injected The No Moisture Injected error indicates a problem with water injection and RH chamber systems. If the heatsink doesn’t decrease in temperature after the second moisture injection, this error will be displayed. If the RH doesn’t increase by at least 5% after three injections, this error will be displayed. Checked • During precondition, humidification • Second moisture injection - checks heatsink temperature decrease • Third moisture injection - RH value must increase by 5% over first injection RH value Possible Causes: • Disconnected heatsink temperature sensor, faulty connection • Heatsink temperature sensor failure • Clogged water line • RH sensor failure, RH must increase by 5% after first three injections • Leak in RH chamber Clearing 1. Raise door handle. 2. After door opens, press Stop. E29 - Low Humidity The Low Humidity error indicates too many moisture injections have occurred. Checked • During precondition, humidification Too many injections before reaching 60% RH. There are typically about 16 injections for a cool cycle and 15 injections for a warm cycle. Possible Causes: • Water low • RH chamber leak • RH sensor disconnected, faulty connection, or sensor failure • Water solenoid failure Clearing 1. Raise door handle. 2. After door opens, press Stop.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 68  3M 1997 April E30 - RH Sensor Failure The valid range for the RH sensor is 1% to 99%. The RH Sensor Failure error indicates the RH exceeds this range at end of initial vacuum in precondition. Checked • At end of initial vacuum in precondition Possible Causes: • Disconnected sensor, faulty connection • Saturated sensor • Leak in RH chamber Clearing 1. Raise door handle. 2. After door opens, press Stop. E31 - Chamber Leak The Chamber Leak error indicates a leak in the chamber. Checked • During precondition, before humidification • Chamber pressure increases more than 15 mBars in five minutes Possible Causes: • Door seal • Loose fittings • Check valve failure Clearing 1. Raise door handle. 2. After door opens, press Stop.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 69 E32 - Pressure Level Before Puncture The Pressure Level Before Puncture error indicates an improper pressure before end of precondition and start of gas injection. Checked • After humidification in the precondition stage (Time-out of 15 minutes) Possible Causes: • No compressed air • Compressed air sensor disconnected, faulty connection • Pressure transducer disconnected, faulty connection, or sensor failure Clearing 1. Raise the door handle. 2. After door opens, press Stop. E33 - Interlock Relay Close The Interlock Relay Close error indicates a latching relay failure. This failure prevents the puncture solenoid from being energized. Checked • Before gas injection Possible Causes: • U6 or U13 defective • 24-volt power supply failure Clearing 1. Raise the door handle. 2. After door opens, press Stop.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 70  3M 1997 April E34 - E37 - No Corresponding Error Description E38 - Nitrogen Pressure (Italian version only) The nitrogen pressure is low. Checked • During precondition Possible Causes: • Nitrogen tank is empty • Faulty pressure transducer or switch Clearing 1. Raise the door handle. 2. After door opens, press Stop.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 71 E39 - Nitrogen Injection Time-Out (Italian version only) The Nitrogen Injection Time-Out error indicates a nitrogen injection problem by monitoring the pressure in the chamber. Checked • During nitrogen injection (precondition) (Pressure did not reach 800 mBars in three minutes) Possible Causes: • Nitrogen tank is empty Clearing 1. Raise the door handle. 2. After door opens, press Stop. E40 - User Interruption During Precondition The User Interruption during precondition error indicates the Stop button was pressed during the precondition stage. Checked • During precondition Possible Causes: • User pressed Stop • Faulty Stop switch • Sudden ground spike Clearing 1. Raise the door handle. 2. After door opens, press the Stop button.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 72  3M 1997 April E41 - E49 - No Corresponding Error Description E50 - Gas Cartridge Empty The Gas Cartridge Empty error indicates there is no pressure increase in the chamber after the cartridge is punctured in the gas injection stage. Checked • During gas injection, pressure increased less than 50 mBars after puncture Possible Causes: • Cartridge not properly seated • Puncture pin failure • No cartridge inserted (see E13) • Empty cartridge Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 73 E51 - Chamber Vacuum Leakage The Chamber Vacuum Leakage error occurs when the chamber pressure approaches within 80 mBar of ambient pressure three times. Checked • During the gas injection and gas exposure stages Possible Causes: • Door seal • Loose fittings Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button. E52 - Chamber Temperature Under Selected Cycle During Gas Expose The Chamber Temperature Under Selected Cycle During Gas Expose error indicates a temperature control problem. The chamber is at least 3°C cooler than the selected temperature cycle. Checked • During gas injection and gas exposure stages Possible Causes: • Chamber High Limit thermostat • Solid state relay failure, SSR1 or SSR2 • Improper calibration of end and/or middle temperature sensors Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 74  3M 1997 April E53 - Chamber Temperature Over Selected Cycle During Gas Expose The Chamber Temperature Over Selected Cycle During Gas Expose error indicates a temperature control problem. The chamber is at least 3°C hotter than the selected temperature cycle. Checked • During gas injection and gas exposure stages Possible Causes: • Solid state relay failure, SSR1 or SSR2 • Improper calibration of end and/or middle temperature sensors Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button. E54 - Extended Power Outage The Extended Power Outage error indicates a chamber pressure problem after a power outage. Pressure inside chamber is too close to ambient. Checked • During recovery of gas injection or gas exposure stages, after a power outage Possible Causes: • Air leak into chamber Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 75 E55 - Undefined Error Condition The Undefined Error Condition error is an unknown error condition and occurs if: • A selftest error code occurred during a stage other than cycle start stage • A precondition error code occurs during gas expose stage • An unknown stage occurred This verifies the error occurred in the proper stage of the cycle. Checked • Error manager, after another error occurs Possible Causes: • See E2 and E3 Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button. E56 - 59 - No Corresponding Error Description


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 76  3M 1997 April E60 - User Interruption During Gas Expose The User Interruption During Gas Expose error indicates the Stop button was pressed during the gas injection or gas expose stage. Checked • During gas injection or gas exposure stages Possible Causes: • User pressed Stop • Stop switch failure • Sudden ground spike Clearing 1. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 2. After door opens, press the Stop button. E61 - E70 - No Corresponding Error Description


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 77 E71 - Final Vacuum Pump Down The Final Vacuum Pump Down error indicates a time-out condition occurred while pulling vacuum in the final vacuum stage. If an abator in installed (see Critical Options, page 20), a time-out occurs if the pressure does not reach 250 mBars within 60 minutes, and does not reach 160 mBars within an additional 30 minutes. If an abator is not installed, a time-out will occur if the pressure does not reach 160 mBars within 90 minutes. This error may also occur during gas expose, if the pressure can not be lowered to 180 mBars below ambient pressure within 2 minutes. Checked • During final vacuum • During gas expose Possible Causes: • Compressed air disconnected • Compressed air pressure insufficient • Disconnected pressure transducer or faulty connection • Vacuum pump failure • Restricted vent line • Air leak into chamber Clearing 1. After fixing the source of the error, press the Start softkey. 2. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 3. After door opens, press the Stop button. E72 - Obstructed Air Inlet The Obstructed Air Inlet error indicates a clogged air inlet filter. When the purge stage starts, the vent is opened and the vacuum remains On. If the pressure does not reach ambient pressure minus 400 mBars within eight minutes, then this error is displayed. Checked • At the start of the purge stage Possible Causes: • Clogged air inlet filter • Vent solenoid failure • 24-volt power supply failure Clearing 1. After fixing the source of the error, press the Start softkey. 2. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 3. After door opens, press the Stop button.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 78  3M 1997 April E73 - Vacuum Sensor Gas The Vacuum Sensor Gas error indicates a vacuum sensor failure while gas is in the chamber. Clearing This error locks the system with the vacuum running. 1. When the error occurs, the sterilizer automatically does the following: a. Stops the timer and stops printing the report. b. Closes the vent and pulls the puncture pin down. c. Turns Off the heaters and turns On the vacuum for 18 minutes. Note: Do not cycle power. 2. Place SW1 switches to Clear Silent Abort (see page 47). Note: During Clear Silent Abort the sterilizer automatically does the following: • Removes the error from the display and starts printing the report • Punctures gas cartridge for one minute. • Runs a time-based final vacuum for 18 minutes. • Opens the vent and purges for 30 minutes. • Closes the vent and locks up. a. After Clear Silent Abort starts, place SW1 configuration switches into Normal mode (see page 7). b. When Clear Silent Abort is complete the screen clears and displays the error, the report stops printing and the chamber parameters display. 3. To clear the error (after the gas is removed from the chamber): a. Turn the sterilizer power Off b. Perform Reset Cycle to Idle (see page 46). E74 - No Corresponding Error Description


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 79 E75 - Gas Low Pressure The Gas Low Pressure error indicates an insufficient pressure increase after the gas is injected. If the pressure increase is less than 200 mBar after gas injection stage, this error will occur and a time-base final vacuum and purge are performed. Checked • During gas injection Possible Causes: • Partial puncture of a gas cartridge • Partially filled gas cartridge • Disconnected or defective pressure transducer • Faulty pressure transducer connection • 24-volt power supply failure • Gas cartridge popout (if E13 is displayed) Clearing This error locks the system with the vacuum running. 1. When the error occurs, the sterilizer automatically does the following: a. Stops the timer and stops printing the report. b. Closes the vent and pulls the puncture pin down. c. Turns Off the heaters and turns On the vacuum for 18 minutes. Note: Do not cycle power. 2. Place SW1 switches to Clear Silent Abort (see page 47). Note: During Clear Silent Abort the sterilizer automatically does the following: • Removes the error from the display and starts printing the report. • Punctures gas cartridge for one minute. • Runs a time-based final vacuum for 18 minutes. • Opens the vent and purges for 30 minutes. • Closes the vent and locks up. a. After Clear Silent Abort starts, place SW1 configuration switches into Normal mode (see page 7). b. When Clear Silent Abort is complete, the screen clears and displays the error, the report stops printing and the chamber parameters display. 3. To clear the error (after the gas is removed from the chamber): a. Turn the sterilizer power Off. b. Perform Reset Cycle to Idle (see page 46). E76 - Interlock Relay Open The Interlock Relay Open error indicates the interlock will not open to indicate no gas in chamber, or gas is in the chamber and the door unlock function was called, or the cycle was complete and gas was still in the chamber. This is a safety precaution, before access is allowed into the chamber, the interlock relay must be open. Checked • During aeration, if gas is in chamber • While unlocking doors Possible Causes: • U13 defective • Invalid stage - software failure • 24-volt power supply failure Clearing 1. After fixing the source of the error, press the Start softkey. 2. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 3. After door opens, press the Stop button.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 80  3M 1997 April E77 - Puncture Pin Stuck The Puncture Pin Stuck error indicates the puncture pin is stuck in the up position or puncture pin switch has failed. Checked • At end of gas injection and at end of purge Possible Causes: • Defective puncture pin mechanism • Disconnected or faulty puncture pin position switch Note: If E77 is printed on report after gas injection, this indicates puncture pin switch problem. If E77 is printed on report after purge, this indicates a stuck puncture pin. Clearing 1. After fixing the source of the error, press the Start softkey. If Start is not displayed, cycle power. 2. A puncture pin switch test is automatically performed. 3. After final vacuum, purge, and mandatory aeration, turn handle and wait for door to open. 4. After door opens, press the Stop button. E78 - E79 - No Corresponding Error Description


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair TROUBLESHOOTING AND REPAIR  3M 1997 April 81 E80 - Abator Time-Out The abator time-out error indicates the Abator Ready signal was not received within 90 minutes after the final vacuum stage starts. Caution code C10 will be displayed and printed. Checked • At the start of final vacuum stage Possible Causes: • Abator failure • Disconnected or faulty abator ready line Clearing This error locks the system, to clear the error: 1. Perform Clear Silent Abort (see page 47). Note: Do not cycle power. 2. After Clear Silent Abort the sterilizer returns to Idle. 3. Place SW1 configuration switches into Normal mode (see page 7).


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator TROUBLESHOOTING AND REPAIR 82  3M 1997 April This page intentionally left blank


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair SPECIAL CYCLES  3M 1997 April 83 Emergency Aeration Emergency aeration is used to aerate a load that needs further aeration. This option can be activated by configuring SW1 DIP switches (see page 45) or from the Main menu during Normal Mode (the following procedure). DANGER Make sure all precautions pertaining to the handling, use, and storage of Ethylene Oxide (EO) are observed. See the Site Planning and Installation Guide and the Operator’s Manual for detailed information in the handling, use, and storage of EO. Refer to the Ethylene Oxide Safety information at the front of this manual for the effects of EO exposure. 1. Place the items requiring further aeration in the sterilizer chamber. 2. With the sterilizer in Normal Mode, use the temperature select softkey (1) to set the cycle temperature. Use the Up arrow and Down arrow softkeys (2, 3) to set the cycle time. 3. Close the sterilizer door. 4. Press the Stop button (4) twice (within one second) to display the operator’s Site Setup menu. 5. Using the Up and Down softkeys, highlight the EMGY AER on the Site Setup menu. 6. Press the Select softkey to toggle EMGY AER to ON. 7. Press the Done softkey to start the emergency aeration cycle. 1 2 3 4


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator SPECIAL CYCLES 84  3M 1997 April Accelerate Cycle This SW1 DIP switch controllable option allows the service representative to accelerate the time of selected stages of a cycle by 60 during a sterilization cycle, in the normal operating mode (1 hour will equal 1 minute). See table at right for specific elements of the cycle that are accelerated. The acceleration DIP switch can be actuated anytime during the sterilization cycle, usually before the cycle starts. If the DIP switch is set to accelerate during a cycle, the time for the current stage will not be accelerated. During a cycle, times are first calculated then the stages are performed. After a stage has started, the only way to accelerate that stage is to cycle the power. When the sterilization cycle restarts, the acceleration option is detected and the new time is calculated. If the accelerated option is changed to non-accelerated during an accelerated cycle, the accelerated time continues until the next stage starts. Notes: Do not consider items loaded in the sterilizer chamber during accelerated times sterilized. Accelerate Cycle is a Normal Mode service operation. The sterilizer must be set to operate or be operating in the Normal Mode to activate this option. DANGER Make sure static discharge wrist ground straps are worn when setting configuration switches or when working with the Steri-Vac 8XL Sterilizer covers removed. Static discharge will cause damage to equipment. Failure to wear static discharge straps can result in damage to equipment or injury to personnel. 1. Make sure the sterilizer is in Normal Mode. 2. Close SW1 configuration switch 2 (see Figure 30). 3. If no additional cycles are to be run with accelerated time, set SW1 configuration switch 2 to open. Precondition Vacuum Test Not accelerated, approximately 1 minute Warm up Not accelerated, time depends on cycle temperature selected Pump down Not accelerated, time depends on puncture pressure selected and vacuum system installed Humidification Dwell times and RH read times are accelerated Leak Test Accelerated, approximately 5 seconds Gas Injection Not accelerated, always 50 seconds Gas Expose Accelerated ( 1 hour equals 1 minute) Final Vacuum Not accelerated, time depends on puncture pressure selected and vacuum system installed Purge Accelerated Locked Aeration Accelerated Unlocked Aeration Accelerated 1 2 3 4 5 6 7 8 SW1 Dip Switch #s Accelerate Cycle (Normal Mode) Figure 30. Accelerate Cycle Time SW1 Switch Positons.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair SPECIAL CYCLES  3M 1997 April 85 Simulating Low Pressure This SW1 DIP switch controllable option allows the service representative to simulate most pressure related checks during a normal sterilization cycle (see table at right for specific elements of the cycle that are simulated). Note: Do not consider items loaded in the sterilizer chamber during simulated pressure cycles sterilized. The simulated pressure DIP switch can be closed anytime during the cycle; usually the switch is closed before the cycle starts. If the DIP switch is set to simulate low pressure during a cycle, the pressure for the current stage may not be simulated. DANGER Make sure static discharge wrist ground straps are worn when setting configuration switches or when working with the Steri-Vac 8XL Sterilizer covers removed. Static discharge will cause damage to equipment. Failure to wear static discharge straps can result in damage to equipment or injury to personnel. Note: Simulating Low Pressure is a Normal Mode service operation. The sterilizer must be set to operate or be operating in the Normal Mode to activate this option. 1. Make sure the sterilizer is in Normal Mode. 2. Close SW1 configuration switch 1 (see Figure 31). All low-pressure stages are simulated during the sterilization cycle. 3. If no additional cycles are to be run with simulated low pressure, set SW1 configuration switch 1 to open. Precondition Vacuum Test No pressure-related checks - not simulated Warm up No pressure-related checks - not simulated Pump down Simulated Humidification One injection Leak Test Simulated Gas Injection Not simulated. If pressure level does not increase by 200 mBars a silent abort will occur Gas Expose Not simulated Final Vacuum Simulated Purge No pressure-related checks - Not simulated Locked Aeration No pressure-related checks - Not simulated Unlocked Aeration No pressure-related checks - Not simulated 1 2 3 4 5 6 7 8 SW1 Dip Switch #s Simulate low Press. (Normal Mode) Figure 31. Simulating Low Pressure SW1 Switch Positions.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator SPECIAL CYCLES 86  3M 1997 April Define Cycles Five different temperature cycles can be set in the Steri-Vac 8XL Sterilizer. CAUTION Only the 37°C, 180-minute and the 55°C, 60-minute cycles are approved for hospital use in the US. Additional cycles must be validated by the end user to meet FDA requirements. Use of unapproved cycles can result in inadequate sterilization. Note: Only the 37°C, 180-minute and 55°C, 60-minute cycles are active after Defaulting System. 1. The temperature and gas cycles and times are set through the SET CYCLES option of the Misc. Service menu. If the sterilizer is not at the Misc. Service menu perform the necessary steps from the following to display the Misc. Service menu: a. Place the sterilizer in the Service Mode - page 9. b. Place the sterilizer in Site Setup - page 11. c. Place the sterilizer in Misc. Service - page 23. 2. Using the Up and Down softkeys, highlight the SET CYCLES option on the Misc. Service menu. 3. Press the Select softkey. 4. Using the Up and Down softkeys, highlight the desired temperature cycle. Notes: A small circle (o ) to the left of the listed temperatures indicates that temperature cycle is selected. A one (1) indicates the temperature is not selected. Selecting temperature with a 1 next to it de-selects that temperature and it will not display on the Misc. Service menu. At least one of the Cycle Temps must be selected. 5. Press the Select softkey. Press YES or NO to define this cycle. See Figure 32. Figure 32. Define Cycle


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair SPECIAL CYCLES  3M 1997 April 87 Temperature 1. The Cycle Temperature selection list is displayed. Using the Up and Down softkeys, select the desired temperature for the currently selected temperature cycle. See Figure 33. Note: A custom temperature may be entered for any temperature cycle. 2. Press the Select softkey. The display advances to the Preheat selection, allowing Preheat Time to be set up. Preheat Option Two preheat options are available: 15-Minute Pre-Heat and No Pre-Heat. The 15-Minute Pre-Heat allows the load to be pre-heated an extra 15 minutes during the precondition stage. 1. Using the Up and Down softkeys, select the PREHEAT Time desired. 2. Press the Select softkey. The display advances to Vacuum. Figure 33. Temperature Figure 34. Preheat


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator SPECIAL CYCLES 88  3M 1997 April Vacuum Three puncture pressure options are available: 100, 128, and 160 mBar. The puncture pressure is the pressure at which the sterilizer pumps to during normal operation before puncturing the gas cartridge. Note: The standard puncture pressure for sterilizers in the US is 160 mBar. Some countries may require a lower puncture pressure. 1. Using the Up and Down softkeys, select the PUNCT PRES option. See Figure 35. 2. Press the Select softkey. The display advances to Humidity. Humidity The Relative Humidity level is preset to 70%. This level can be changed to suit special conditions. 1. Using the Up and Down softkeys, select the appropriate relative humidity level. See Figure 36. 2. Press the Select softkey. The display advances to Exposure time. Figure 35. Vacuum Figure 36. Humidity


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair SPECIAL CYCLES  3M 1997 April 89 Exposure The Gas Exposure Time selection is displayed. 1. Using the Up and Down softkeys, select the desired time. See Figure 37. 2. Press the Select softkey. The display will advance to Aeration. Aeration A locked aeration time can be customized for the needs of the installation. If the optional exhaust hood is not installed, three hours of aeration are recommended and preset. 1. Using the Up and Down softkeys, select the appropriate aeration time. See Figure 38 2. Press the Select softkey. The display advances to the Define Cycles and displays the selected options. Note: A small one (1) is now displayed next to the cycle temperature indicating that temperature cycle is selected and will be available during normal operation. 3. If additional Cycle Temps and Gas Exposure Times need to be selected, perform steps 3 – 6 to select additional temperature cycles. 4. After all of the desired temperature cycles are selected and the Cycle Temps selection list is displayed, press the Done softkey. Note: When the temperature selection softkey is pressed during normal operation, only the temperature cycles selected in this procedure will appear on the display. Figure 37. Exposure Figure 38. Aeration


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator SPECIAL CYCLES 90  3M 1997 April This page intentionally left blank


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair ILLUSTRATED PARTS BREAKDOWN  3M 1997 April 91 Illustrated Parts Breakdown (IPB) This Illustrated Parts breakdown (IPB) section contains information necessary to locate and identify assemblies, subassemblies, and specific parts of the Model 8XL Steri-Vac. Part Numbers All components available on a replacement basis are listed with their 3M identification number. The parts identification list opposite each illustration provides the description adequate to identify components. If the component is available only as a part of the next higher assembly, the description column provides the nest higher assembly description or identification number. Revision Bars Revision bars placed in the margins indicate where additions or changes occurred. Refer to the Revision History to identify specific additions and changes. Locating and Identifying a Part If physical appearance and general location of an item are known: 1. Refer to an assembly or sub-assembly to locate the item on the explodes view. If machine serial numbers determine a component, the part ID number in this manual are from the most recent data available. Make sure part ID numbers for your component have not been amended by changes and updates made to your machine. Verify the serial number on the serial plate located on the Frame Chamber Assembly. Ordering Parts When ordering parts, include the following information: 1. Machine model number and serial number. 2. Part identification number. 3. Part Description 4. Quantity required. For orders call 800 292-6298 between 8:00 a.m. and 4:30 p.m., CST, except Saturday, Sunday, and holidays. Abbreviations PWA printed wiring assembly Mach Machine Dia diameter


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator ILLUSTRATED PARTS BREAKDOWN 92  3M 1997 April This page intentionally left blank.


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair ILLUSTRATED PARTS BREAKDOWN  3M 1997 April 93 Illustrated Parts Breakdown Contents Title_________________________________________________________________________________________Page Final Assembly ………………………………………………………………………………………………………….94 Frame-Chamber Assembly ……………………………………………………………………………………………...96 Door Lock Assembly (Load Side) ………………………………………………………………………………………98 Water Filler, Printer Assembly …………………………………………………………………………………………100 Power Supply Assembly ………………………………………………………………………………………………..102 Services-In Assembly …………………………………………………………………………………………………..104 Manifold Assembly …………………………………………………………………………………………………….106 Display Panel Assembly ……………………………………………………………………………………………….108 Water Tank Assembly …………………………………………………………………………………………………110 Chamber Assembly ……………………………………………………………………………………………………112 Chamber-Heater Assembly ……………………………………………………………………………………………116 Heat Sink Assembly ………………………………………………………………………………………………….. 118 Door Assembly (Load Side ) …………………………………………………………………………………………..120 Local Exhaust Assembly ………………………………………………………………………………………………122 Controller Bracket Assembly ………………………………………………………………………………………….124 Controller Assembly …………………………………………………………………………………………………..126 IMPORTANT NOTICE 3M reserves the right to sell or furnish NEW OR RECONDITIONED PARTS. New and reconditioned parts are Covered by the same 3M parts limited warranty.


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator ILLUSTRATED PARTS BREAKDOWN 94  3M 1997 April IPB Figure 1 FINAL ASSEMBLY


Steri-Vac 8XL Gas Sterilizer/Aerator Service and Repair ILLUSTRATED PARTS BREAKDOWN  3M 1997 April 95 FINAL ASSEMBLY Item No. Part Number Description Qty. 0 78-8078-3971-3 FINAL ASSEMBLY REF 1 78-8078-3866-5 PANEL, Upper Rear 1 2 12-1003-9521-4 Screw, Mach, Hex,1/4-20 X .375 41 3 78-8078-3878-0 PANEL, Upper Side 2 4 Not Used 5 26-1011-6138-3 HOLE PLUG (For 11/16 Hole 28 6 78-8078-5714-5 CENTER SIDE PANEL, CONTROLLER ASSEMBLY 1 7 78-8078-3986-1 PANEL, Lower Services 1 8 78-8078-3970-5 FRAME CHAMBER ASSEMBLY (See IPB Figure 2) REF 9 78-8078-3864-0 PANEL, Bottom 1 10 26-1004-3140-7 BUSHING, Strain Relief 2 11 78-8078-3876-4 PANEL, Lower Side 1 12 78-8078-5715-2 CENTER SIDE PANEL ASSEMBLY, Hinge Side 1 13 78-8078-3865-7 PANEL, front, Upper 1 14 78-8078-3871-5 PANEL, Top 1 15 78-8078-3998-6 INSULATION, Rear panel 2 16 78-8005-1710-0 TAPE, Cloth, 1.00 W AR 17 78-8078-5999-2 COVER, Hose, Exhaust Hood 2 18 26-1011-7366-9 FASTENER 8 19 78-8078-5678-2 REAR PANEL, Bottom 1 20 78-8078-5235-1 DECAL, 8XL 1 21 26-1004-2619-1 SCREW, Mach, 10-32 X .35 Phillips, Black 3 22 26-1002-5278-7 SCREW, Mach, 8-32 X .500, Pan Head, Phillips 9 23 26-1000-6654-2 NUT, Mach, 8-32, ST, Zinc 8 24 26-1011-7368-5 STUD, ¼ Turn 4 25 26-1011-7369-3 SPRING 4 26 26-1000-1451-8 Washer, Cupped Wear 4 27 26-1011-7367-7 RETAINER 8 28 26-1011-7371-9 STUD, ¼ Turn 4


Service and Repair Steri-Vac 8XL Gas Sterilizer/Aerator ILLUSTRATED PARTS BREAKDOWN 96  3M 1997 April IPB Figure 2 FRAME-CHAMBER ASSEMBLY


Click to View FlipBook Version