CRC-55t R-PET Chamber Manual

CRC-55t R-PET Chamber Manual

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting Syringe Published (NCRP-58)
64Cu Copper Number Corr. % Data % ORNL76 Decays to 73As.
12.701 H 2 Estimation use only.
65Zn Zinc 015 or 2 ORNL76 Pure
66Ga Gallium 115 × 2 4 243.9 D 1 ORNL76 Pure
67Cu Copper 172 1.4 9.40 H 7 ORNL76 Reading gives act. of
67Ga Gallium 903 2.575 D 3 NBS78 81Rb or 81mKr in
68Ga Gallium 052 2 3.261 D 1 ORNL76 equilibrium sample.
69mZn Zinc 100 43 68.0 M 2 ORNL76 Eqb. after 2 minutes.
72As Arsenic 416 13.76 H 3 ORNL76 Decays to 85Kr
72Ga Gallium 143 5 26.0 H 1 ORNL76
73As Arsenic 795 2.5 14.10 H 1 ORNL76 Pure
73Se Selenium 470 ÷ 2 6 80.30 D 6 ORNL76 Pure
74As Arsenic 324 × 10 26 7.15 H 8 ORNL76
75Se Selenium 748 3 17.78 D 3 NBS75
76As Arsenic 304 3 119.8 D 1 ORNL76
77As Arsenic 258 26.32 H 7 ORNL76
77Br Bromine 110 38.8 H 3 ORNL76
79Kr Krypton 481 × 100 56 H 2 ORNL76
81Rb Rubidium 091 35.04 H 10 NM75
81mKr Krypton 050 4.58 H NM75
81Rb Rubidium 174 13 S 1
Eqb. 915 × 10
81mKr Krypton
270 NM75
82Br Bromine
82Rb Rubidium 536 ÷ 2 2 35.34 H 2 ORNL76
84Rb Rubidium 504
85mKr Krypton 347 1.273 M 2 NM75
85Kr Krypton 065
85Sr Strontium 031 × 100 32.77 D 4 NM75
86Rb Rubidium 193
86Y Yttrium 411 × 10 1 4.480 H 8 ORNL76
86Zr Zirconium 711 ÷ 2
87Kr Krypton 167 2 10.72 Y 1 ORNL76
87mSr Strontium 250
87Y Yttrium 095 1.0 64.854 D 3 NBS75
170
18.64 D 2 ORNL76

14.74 H 2 ORNL76

18 3 16.5 H 1 ORNL76

6 76.3 M 5 ORNL76

2.805 H 3 ORNL76

1 80.3 H 3 ORNL76

July 16 APPENDIX II – R CHAMBER CALIBRATION NUMBERS II-9

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting (NCRP-58) Reading gives 87Y Act.
Number Syringe Published in eqb. sample. Eqb.
Corr. % Data % after 18 hours.
357 Reading gives sum of
87Y Yttrium ORNL76 87Y & 87mSr activity in
(Eqb. 87mSr) equilibrium sample.

87Y Yttrium 341 2 ORNL76 Estimation use only.
Eqb. Almost pure β decay.
87mSr Strontium 189 14 17.8 M 1 ORNL76 Estimation use only.
88Rb 465 ÷ 2
88Y Rubidium 768 1.8 106.61 D 2 NBS73 Pure
89Rb Yttrium 48 × 10 Pure
90Y Rubidium 850 × 10 1 15.2 M 1 ORNL76 Reading gives sum of
91Y Yttrium 95mNb & 95Nb activity in
Yttrium 64.0 H 1 NIST92 equilibrium sample.
Eqb. after 2 years.
58.5 D 4 NDT70
Reading gives sum of
94Nb Niobium 673 2.03E4 Y 16 ORNL76 97Zr & 97mNb activity in
95Nb Niobium 285 34.97 D 1 NDT70 equilibrium sample.
95Zr Zirconium 271 64.02 D 5 NDT70 Eqb. after 10 minutes.
95Zr Zirconium 95mNb
Eqb. 145 3.61 D 1 NDT70 Decays to 97mTc
95m95Nb Niobium Estimation use only.
249
97Nb Niobium 72.1 M 7 ORNL76
97Zr 341
Zirconium 12 16.90 H 5 ORNL76
97mNb Eqb. 271
Niobium 116
256 × 10
97mNb Niobium 080 × 5 or 60 S 1 ORNL76
97Ru Ruthenium 246× 10 or
97mTc Technetium 030 × 3.5 15 2 2.9 D 1 ORNL76
99Mo Molybdenum 030 × 4 or 65
(in Std. Mo Kit) 204 × 10 91.0 D NM75
030 × 4 or 2
204 × 10 2 NTS78
165
99Mo Molybdenum NBS78
(in CAP-MAC) 080
99Mo 042 NBS78
Molybdenum
99Mo (in MAC-S) 1.9 65.92 H 2 NBS78
99mTc 2.1 6.007 H 1 NBS76
99mTc Molybdenum 2.1 6.007 H 1 NBS76
(Eqb. 99mTc)

Technetium

Technetium
(in CAP-MAC)

II-10 APPENDIX II – R CHAMBER CALIBRATION NUMBERS July 16

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting Syringe Published (NCRP-58)
Number Corr. % Data %
99mTc Technetium NBS78
(Eqb. 99Mo) 175 2
99Mo Molybdenum
Eqb. 145 2 NBS78 Reading gives sum of
99mTc Technetium 562 × 10 50 99Mo & 99mTc activity in
103Pd Palladium 634 × 10 50 equilibrium sample.
103Pd
Palladium 631 × 100 50 4 16.97 D 2 ORNL76 Pure
103mRh Eqb. 165 50 5 ORNL76
103mRh Rhodium 172 50 5 Reading gives sum of
103Ru 3 56.114 M 6 ORNL76 103Pd & 103mRh activity
103Ru Rhodium 027 or 39.26 D 2 ORNL76 in equilibrium sample.
140 × 2 15 Eqb. after 9 hours.
103mRh Ruthenium 4 ORNL76
106Ru Ruthenium Reading gives sum of
Eqb. 2 369 D 2 NDT70 103Ru & 103m,Rh activity
Rhodium 30 in equilibrium sample.
1.36 Eqb. after 9 hours.
Ruthenium 5 Reading gives 106Ru
(Eqb. 106Rh) 2 Act. in eqb. sample.
3.2 Eqb. after 5 minutes.
106Ru Ruthenium 480 × 10 2 369 D 2 NDT70 Reading gives sum of
Eqb. 106Ru & 106Rh activity
106Rh Rhodium 830 in equilibrium sample.
108mAg Silver 099 × 10
108Ag Silver 047 or 3 127 Y 21 ORNL76 Large β contribution.
109Cd Cadmium 180 × 2 6 2.37 M 1 ORNL76 Reading gives act of
Eqb. 40 462.6 D 4 109Cd, 109mAg, or Total
109mAg Silver ORNL76 Act. in eqb. sample.
109Pd 435 × 10 10 39.8 S ORNL76 Eqb. after 6 minutes.
Palladium 35 13.427 H 14 NBS77 Reading gives act of
109mAg Eqb. 554 ÷ 2 7 39.8 S MARTIN77 109Pd, 109mAg, or Total
110mAg Silver 054 × 10 Act. in eqb. sample.
111Ag 303 15 249.8 D 1 Eqb. after 6 minutes.
111In Silver 022 or 15 7.45 D 1
113Sn Silver 129 × 2 2.805 D 1 For pure 113Sn.
Indium 076 115.08 D 3
Tin
180
113mIn Indium 1.658 H 1 Separated for pure
113Sn 058 NBS73,77 113mIn.
113mIn Tin 4.486 H 4
115mIn Eqb. Reading gives act. of
Indium 113Sn, 113mIn, or Total
Act. in eqb. sample.
Indium Eqb. after 15 hours.

ORNL76

July 16 APPENDIX II – R CHAMBER CALIBRATION NUMBERS II-11

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting Syringe Published (NCRP-58)
116mIn Indium Number Corr. % Data % ORNL76 Pure
117mSn Tin 54.15 M 6 ORNL76 Pure
117Sb Antimony 974 3 13.61 D 4 ORNL76 Reading gives act. of
119mSn Tin 2.80 H 1 ORNL76 121mTe or 121Te in eqb.
121mTe Tellurium 180 15 2 293.0 D 13 ORNL76 sample. Eqb. in 120
121Te Tellurium 154 D 7 ORNL76 days. See App. IV for
121mTe Tellurium 082 32 17 D 1 non-eqb. samples.
Eqb. Reading gives sum of
121Te Tellurium 657 × 10 35 4 121mTe or 121Te in eqb.
sample.
187 12 9
Ref. for 28 keV x-ray.
373 10 3
Pure
645 Reading gives 125Sb
Act. in eqb. sample.
121mTe Tellurium 572 Eqb. after 1 year.
Eqb. Reading gives sum of
121Te Tellurium 146 6 2.70 D 1 ORNL76 125Te and 125mTe in
122Sb Antimony eqb. sample. See App.
123I Iodine 277 15 1.9 13.221 H 3 NBS77 IV for non-eqb. activity.
123mTe Tellurium
124Sb Antimony 177 12 119.7 D 1 NDT70 Pure
124I Iodine Pure
125I Iodine 720 60.20 D 3 ORNL76
125Sb Antimony
125Sb Antimony 570 5 4.18 D 2 ORNL76
(Eqb. 125mTe)
125Sb 319 25 1.45 59.6 D 2 NBS76
Antimony
125mTe Eqb. 289 10 2.758 Y 1 NDT70
125mTe Tellurium
126I 371 12 2.758 Y 1 NDT70
127Xe Tellurium
129mTe Iodine 364 12 NDT70
129Te Xenon
Tellurium 259 25 57.40 D 5 NDT70
Tellurium 240 10
371 12 18 13.02 D 7 ORNL76
817 × 10 20
679 × 10 15 5 36.4 D 1 ORNL76

5 33.6 D 1 ORNL76

13 69.6 M 2 ORNL76

129mTe Tellurium 054 Reading gives act. of
Eqb. 129mTe or total act. in
129Te Tellurium 397 eqb. sample. Eqb. in
129Cs 166 10 hours.
129I Cesium 362
129mXe Iodine 15 20 32.06 H 6 ORNL76 NM75 gives 488
Xenon
20 1.57E7 Y 4 ORNL76

20 3 8.0 D 2 ORNL76

II-12 APPENDIX II – R CHAMBER CALIBRATION NUMBERS July 16

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting Syringe Published (NCRP-58)
130I Iodine Number Corr. % Data % ORNL76 Decays to 131mXe.
131I Iodine 12.36 H 1 NBS76 1.1% feeding.
984 1.65
8.021 D 1 Pure
151 Pure
Reading gives 132Te
131mXe Xenon 089 20 3 11.9 D 1 ORNL76 Act. in eqb. sample.
131Cs Cesium 148 20 Eqb. after 1 day.
131Ba Barium 505 10 3 9.69 D 1 ORNL76 Reading gives 132I Act.
132Te Tellurium 315 10 in eqb. sample.
132I Iodine 999 7 11.8 D 2 ORNL76 Reading gives sum of
132Te Tellurium 675 ÷ 2 132Te and 132I in eqb.
(Eqb. 132I) 5 76.3 H 2 ORNL76 sample.

2.30 H 3 ORNL76 Decays to 133mXe.
Decays to 133Xe. See
132I Iodine 653 ÷ 2 App. IV.
(Eqb. 132Te)
132Te 663 ORNL76 Decays to 133Ba.
Tellurium
132I Eqb. 485 10 2 6.475 D 10 ORNL76 Decays to 134Cs.
132Cs Iodine 225
133I 100 20 20.8 H 1 ORNL76 Decays to 135Xe. See
133mXe Cesium 12 App. IV.
188 12 3 2.19 D 1
Iodine 132 10 Ref: 661.6 & 32.9 keV.
591 3 Reading gives 137Cs or
Xenon 533 15 Total Activity of eqb.
037 or sample. Often referred
133Xe Xenon 160 × 2 3 1.95 5.243 D 1 NBS76 to as "Cs-137" Source.
133mBa Barium 726 2
133Ba Barium 181 15 3 38.9 H 1 ORNL76 Ref: 36.8 keV
134Te Tellurium
134mCs Cesium 085 3 10.5 Y 1 ORNL76
130
489 ÷ 2 6 41.8 M 8 ORNL76

220 2.91 H NM75

134Cs Cesium 2.3 2.065 Y 1 NBS73
135mXe Xenon
15.29 M 3 ORNL76

135Xe Xenon 9.09 H 1 ORNL76
135mBa Barium
136Cs Cesium 4 28.7 H 2 ORNL76
137Cs
Cesium 4 13.1 D 1 ORNL76
137mBa Eqb.
Barium 2.0 30.0 Y 2
NBS73

2.553 M 1

139Ba Barium 445 × 10 5 12 82.8 M 2 ORNL76
139Ce Cerium 352 5
141Ce Cerium 061 5 2.6 137.64 D 2 NBS73

5 32.50 D 1 ORNL76

July 16 APPENDIX II – R CHAMBER CALIBRATION NUMBERS II-13

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting Syringe Published (NCRP-58)
142Pr Praseodymium Number Corr. % Data % ORNL76 Estimation use only. β
144Pr Praseodymium 19.13 H 4 ORNL76 dominant.
226 × 10 14 Ref: 36.7 & 133.5 keV.
17.28 M 5 Reading gives sum of
137 × 10 5 144Ce & 144Pr Act. in
equilibrium sample.
144Ce Cerium 387 × 10 5 2.8 285.0 D 1 NBS73 Eqb. after 2 hours.
144Pr Eqb.
Praseodymium Decays to 188Re. Eqb.
after 7 days. See App.
145Pm Promethium 207 10 3 17.7 Y 4 ORNL76 IV
147Nd Neodymium 213 5 Reading gives 188W
157Dy Dysprosium 424 5 4 10.98 D 1 ORNL76 Act. in eqb. sample.
169Yb Ytterbium 948 3 Eqb. after 5 days.
171Tm Thulium 292 × 100 4 8.1 H 1 NM75 Reading gives 188Re
175Yb Ytterbium 308 × 10 2 Act. in eqb. sample.
177Lu Lutetium 450 × 10 2 2.5 32.03 D 1 NBS78 Reading gives sum of
181Hf Hafnium 387 188W & 188Re activity in
181W Tungsten 165 3 1.92 Y 1 NM75 equilibrium sample.
188W Tungsten 111 × 100
13 4.19 D 1 ORNL76 Ref. for 70 & 77 keV.

7 6.71 D 1 ORNL76

6 42.4 D 1 ORNL76

11 121.2 D 3 ORNL76

69.4 D 5 NM75

188Re Rhenium 496 × 10 16.98 H 2 NM75
188W 522 × 10
Tungsten
(Eqb. 188Re) 516 × 10

188Re Rhenium 217 × 10 9.90 M NM75
(Eqb. 188W)
188W Tungsten 858 2 13 15.4 D 2 ORNL76
Eqb. 250
188Re Rhenium 408 2 73.83 D 1 NDS73
190mOs Osmium 469 × 10 2
191Os Osmium 686 × 10 18 19.15 H 3 ORNL76
192Ir Iridium 197 2
194Ir Iridium 149 6 18.3 H 3 ORNL76
197Pt Platinum 053
197Hg Mercury 205 2.9 64.1 H 1 NBS76
198Au Gold 093
199Au Gold 344 1.65 2.696 D 2 NBS78
201Tl Thallium
203Hg Mercury 6 3.139 D 7 ORNL76
203Pb Lead
2.0 72.91 H 2 NBS76

1.1 46.60 D 1 NBS73

2 51.88 H 1 ORNL76

II-14 APPENDIX II – R CHAMBER CALIBRATION NUMBERS July 16

CAPINTEC, INC. CRC®-55t

Radioisotopes Calibration Uncertainty Due to Half-Life Ref. Comments
Setting Syringe Published (NCRP-58)
204Tl Thallium Number Corr. % Data % NDT70 Ref. for 1064, 569.7,
207Bi Bismuth 3.78 Y 2 NBS73 76.7, & 1772 keV.
420 × 100 22
32.2 Y 9 Decays to 212Bi. Eqb.
846 1.7 after 1 hr. See App. IV.

208Tl Thallium 571 ÷ 2 3.053 M 4 NM75 Reading gives 212Pb
212Pb Lead 101 10.64 H 1 NM75 Act. in eqb. sample.
Eqb. after 8 hours.
212Bi Bismuth 489 × 10 60.55 M 6 NM75 Reading gives 212Bi
212Pb Lead 158 Act. in eqb. sample.
(Eqb. 212Bi) Reading gives sum of
212Bi 135 3.66 D 4 NM75 212Pb & 212Bi activity
Bismuth in equilibrium sample.
212Pb (Eqb. 212Pb) 030 or 0.5 1600 Y 7 NBS73
212Bi Lead 146 × 2 Reading in grams.
224Ra Eqb. Commonly referred to
226Ra Bismuth 646 × 100 as "Radium "Source.
Radium 778 1.025 g/Ci of Ra-226.
239Np Radium
241Am + chain of 147 6 2.355 D 4 ORNL76 Ref. for 59.5 & 14 keV.
daughters
055 4 1 432.2 Y 5 LMR69
Neptunium
Americium

July 16 APPENDIX II – R CHAMBER CALIBRATION NUMBERS II-15

CAPINTEC, INC. CRC®-55t

This page intentionally left blank.

CAPINTEC, INC. CRC®-55t

APPENDIX III

PET CHAMBER
CALIBRATION NUMBERS

CALIBRATION SETTING NUMBERS
The Isotope Calibration Numbers in Table I are applicable to the Capintec CRC®-55t PET
Chamber(s) only.

The CRC®-55t is a direct reading instrument. No manual multiplication or division should be
performed, even if the Calibration Setting Number is followed by a multiplication sign “× “ or a
division sign “÷ “ and a number.

If the sample contains radioactive impurities, the meter indication will always be higher than
the actual activity of the principal isotope. It will not, however, be the total activity of the
principal isotope and the impurities.

UNCERTAINTY DUE TO SYRINGE CORRECTION

The Calibration Setting Numbers are given for approximately 5 grams of radioactive solution
in a standard source ampoule made of about 0.6 mm. thick borosilicate glass. The standard
radioactive source in the ampoule is, however, a good approximation for a
radiopharmaceutical in a plastic syringe or a glass syringe (wall thickness about 1.2 mm) for
most radioisotopes.

RADIOISOTOPE TABLE

Radioisotopes Calibration Half-Life Ref.
Setting (NCRP-58)
11C Carbon 11 Number See Note
15O Oxygen 15 467 20.38 Minutes See Note
13N Nitrogen 13 122.24 Seconds See Note
18F Fluorine 18 474 9.965 Minutes
57Co Cobalt 57 109.71 Minutes NIST
60Co Cobalt 60 470 NIST
133Ba Barium 133 271.7 Days NIST
137Cs Cesium 137 480 5.271 Years NIST
10.5 Years NIST
112 30.0 Years

990

674

243

Note: These Calibration Numbers were determined by inter-comparison with a standard
Capintec CRC®-15R Dose Calibrator. No NIST standards are yet available. They

should be used for estimation purposes only.

October 15 APPENDIX III – PET CHAMBER CALIBRATION NUMBERS III-1

CAPINTEC, INC. CRC®-55t

This page intentionally left blank.

CAPINTEC, INC. CRC®-55t

APPENDIX IV

MULTIPLICATION FACTORS FOR NON-
EQUILIBRIUM RADIOISOTOPES

Ca47 Measurements .....................................................................................IV-2
Ni56 Measurements ......................................................................................IV-2
Zn62 Measurements .....................................................................................IV-3
Y87 Measurements .......................................................................................IV-3
Zr95 Measurements ......................................................................................IV-4
Te121m Measurements ................................................................................IV-4
Sb125 Measurements ...................................................................................IV-5
Te129m Measurements ................................................................................IV-5
Te132 Measurements....................................................................................IV-6
Xe133m Measurements ................................................................................IV-6
Xe135m Measurements ................................................................................IV-7
W188 Measurements ....................................................................................IV-7
Pb212 Measurements ...................................................................................IV-8

Note: These Multiplication Factors apply to the R Chamber only.

The activity in a non-equilibrium sample can be determined by using the following equation:

Activity of "A" = Meter Reading with Cal. Number for Pure "A"

1+ F TA  exp  - TA - TB × t ×1n2 RB
TA - TB 1 - TA × TB RA


where “A” and “B” are parent and daughter nuclide respectively, T's are half-lives, and RA
and RB are the chamber responses to the isotopes A and B, “t” is the elapsed time after the
pure parent isotope was produced, and “F” is the decay branching ratio.

The chamber response, ”R,” may be obtained from the calibration number “N,” using the
following equation:

R = N + 0.0797
1075.8

When the half-life of the parent nucleus (TA) is much shorter than that of the daughter
nucleus (TR), the parent decays down to the daughter. After about 10 × TA, we can assume
that only the daughter is left (except if the original activity of the parent was exceptionally
high). The equation above for the Activity of “A” can be used to obtain the activity of the
parent while it has measurable activity.

July 16 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES IV-1

CAPINTEC, INC. CRC®-55t

Multiplication factors obtained by using the equation are given in the following tables for
selected isotopes.

CA47 MEASUREMENTS
Use a Calibration Number of 373 and a multiplication factor as indicated in the following table
to obtain Ca47 activity at time “t” days after an initially pure source of Ca47 is obtained.

Ca47

t Multiplication t Multiplication
[days] Factor
Factor [days]
0 0.672
1 1.000 12 0.648
2 0.627
3 0.950 14 0.609
4 0.594
5 0.906 16 0.566
6 0.545
7 0.869 18 0.519
8 0.496
9 0.836 20 0.485
10 0.484
0.806 25

0.781 30

0.757 40

0.737 50

0.718 100

0.701 150

NI56 MEASUREMENTS
Use a Calibration Number of 844 and a multiplication factor as indicated in the following table
to obtain Ni56 activity at time “t” days after an initially pure source of Ni56 is obtained.

Ni56

t Multiplication t Multiplication
[days] Factor
Factor [days]
0 0.824
1 1.000 9 0.799
2 0.745
3 0.985 10 0.659
4 0.508
5 0.969 12 0.366
6 0.249
7 0.952 15 0.101
8 0.013
0.934 20

0.915 25

0.894 30

0.872 40

0.849 60

IV-2 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES July 16

CAPINTEC, INC. CRC®-55t

ZN62 MEASUREMENTS
Use a Calibration Number of 217 and a multiplication factor as indicated in the following table
to obtain Zn62 activity at time “t” minutes after an initially pure source of Zn62 is obtained.

Zn62

t Multiplication t Multiplication
[minutes] Factor
Factor [minutes]
0 0.368
10 1.000 45 0.361
20 0.359
30 0.526 60 0.358

0.425 75

0.389 90

Y87 MEASUREMENTS
Use a Calibration Number of 170 and a multiplication factor as indicated in the following table
to obtain Y87 activity at time “t” hours after an initially pure source of Y87 is obtained.

Y87

t Multiplication t Multiplication
[hours] Factor
Factor [hours]
0 0.664
0.25 1.000 5 0.644
0.50 0.618
0.75 0.960 6 0.602
1.00 0.593
1.5 0.926 8 0.582
2.0 0.580
3.0 0.893 10 0.579
4.0 0.579
0.866 12

0.820 18

0.784 24

0.729 48

0.691 ∞

July 16 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES IV-3

CAPINTEC, INC. CRC®-55t

ZR95 MEASUREMENTS
Use a Calibration Number of 271 and a multiplication factor as indicated in the following table
to obtain Zr95 activity at time “t” days after an initially pure source of Zr95 is obtained.

Zr95

t Multiplication t Multiplication
[days] Factor
Factor [days]
0 0.592
2 1.000 40 0.548
4 0.513
6 0.961 50 0.462
8 0.426
10 0.925 60 0.374
15 0.347
20 0.893 80 0.322
25 0.309
30 0.863 100 0.308

0.836 150

0.776 200

0.727 300

0.685 600

0.650 ∞

TE121M MEASUREMENTS
Use a Calibration Number of 187 and a multiplication factor as indicated in the following table
to obtain Te121m activity at time “t” days after an initially pure source of Te121m is obtained.

Te121m

t Multiplication t Multiplication
[days] Factor
Factor [days]
0 0.437
1 1.000 40 0.415
2 0.401
5 0.943 50 0.392
10 0.386
15 0.895 60 0.382
20 0.379
25 0.782 70 0.376
30 0.374
0.661 80

0.586 90

0.535 100

0.499 120

0.472 140

IV-4 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES July 16

CAPINTEC, INC. CRC®-55t

SB125 MEASUREMENTS
Use a Calibration Number of 289 and a multiplication factor as indicated in the following table
to obtain Sb125 activity at time “t” days after an initially pure source of Sb125 is obtained.

Sb125

t Multiplication t Multiplication
[days] Factor
Factor [days]
0 0.878
10 1.000 90 0.861
20 0.849
30 0.977 120 0.841
40 0.824
50 0.958 150 0.822
60 0.821
0.941 180

0.927 360

0.914 540

0.903 ∞

TE129M MEASUREMENTS

Use a Calibration Number of 817 and a multiplication factor as indicated in the following table
to obtain Te129m activity at time “t” hours after an initially pure source of Te129m is
obtained.

Te129m

t Multiplication t Multiplication
[hours] Factor
Factor [hours]
0 6.55
1 10.0 6 6.50
2 6.49
4 8.04 8 6.48
Equilibrium
7.25 10

6.701 12

July 16 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES IV-5

CAPINTEC, INC. CRC®-55t

TE132 MEASUREMENTS
Use a Calibration Number of 315 and a multiplication factor as indicated in the following table
to obtain Te132 activity at time “t” hours after an initially pure source of Te132 is obtained.

Te132

t Multiplication t Multiplication
[hours] Factor
Factor [hours]
0 0.342
0.25 1.000 4 0.318
0.50 0.275
1.00 0.836 5 0.266
2.00 0.265
3.00 0.725 10 0.264
Equilibrium
0.586 15

0.447 20

0.380 24

XE133M MEASUREMENTS
Use a Calibration Number of 100 and a multiplication factor as indicated in the following table
to obtain Xe133m activity at time “t” days after an initially pure source of Xe133m is obtained.

Xe133m

t Multiplication t Multiplication
[days] Factor
Factor [days]
0 0.465
1 1.000 4 0.385
2 0.151
3 0.824 5 0.024

0.680 10

0.562 20

IV-6 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES July 16

CAPINTEC, INC. CRC®-55t

XE135M MEASUREMENTS

Use a Calibration Number of 181 and a multiplication factor as indicated in the following table
to obtain Xe135m activity at time “t” minutes after an initially pure source of Xe135m is
obtained.

Xe135m

t Multiplication t Multiplication
[minutes] Factor
Factor [minutes]
0 0.807
10 1.000 60 0.676
20 0.514
30 0.990 75 0.217
45 0.019
0.975 90

0.952 120

0.897 180

W188 MEASUREMENTS

Use a Calibration Number of 111 and a multiplication factor as indicated in the following table
to obtain W188 activity at time “t” hours (or days) after an initially pure source of W188 is
obtained.

W188

t Multiplication t Multiplication
[hours] Factor
Factor [days]
0 5.09
1 100.0 1 3.75
2 3.40
4 45.6 2 3.30
6 3.26
12 29.9 3 3.25
18 3.24
18.2 4

13.4 5

8.0 6

6.0 7

July 16 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES IV-7

CAPINTEC, INC. CRC®-55t

PB212 MEASUREMENTS
Use a Calibration Number of 101 and a multiplication factor as indicated in the following table
to obtain Pb212 activity at time “t” hours after an initially pure source of Pb212 is obtained.

Pb212

t Multiplication t Multiplication
[hours] Factor
Factor [hours]
0 0.772
0.5 1.000 5 0.768
1.0 0.766
2.0 0.924 6 0.765
3.0 0.765
4.0 0.875 7 0.764
Equilibrium
0.820 8

0.793 9

0.779 10

IV-8 APPENDIX IV – MULTIPLICATION FACTORS FOR NON-EQUILIBRIUM RADIOISOTOPES July 16

CAPINTEC, INC. CRC®-55t

INDEX

A Starting the Test, 11-16
ABORT button, 11-21
ACCEPT button, 6-4 AUTOLINEARITY button, 11-12
Acceptance Testing, 4-11, 7-1 AutoLinearity Test

Accredited Dosimetry Calibration Detailed Report, 11-29
Laboratory, 1-3
Inactivate, 11-30
ACCURACY button, 8-10, 8-12
Accuracy Test, 7-2, 7-3, 8-12 B
Background, 8-1, 8-7
in Daily Test, 8-10
Inactivate, 14-33
Individual Results Report, 14-27
Report, 7-2 BACKGROUND button, 8-1
Search Results Report, 14-25 Background Measurements

View an Individual Test Report, 14-26 Individual Results Report, 14-33
ACCURACY TEST button, 14-8
Accuracy Test Report, 14-24 Search Results Report, 14-31

from Daily Test, 14-8 View an Individual Test Report, 14-32
ACCURACY TEST Report button, 14-24
Accuracy/AutoConstancy Background Measurements Report, 14-30

Inactivate, 14-27 BACKGROUND Report button, 14-30
Battery, 16-13
ADCL, 1-3
ADD ITEM button, 12-3 Low, 16-6
Adding to Inventory, 12-3
adjustments, 1-4 Replacement, 16-6
ADVANCED CHAMBER button, 6-3, 6-10
Alphanumeric Data Brachytherapy, 1-3, 1-4

entering, 3-6 Brightness, 5-22

Assembly, 4-2 button
asterisk
ABORT, 11-21
Remote Display indicator, 3-4 ACCEPT, 6-4
Auto Zero, 8-6 ACCURACY, 8-10, 8-12
AUTOCONSTANCY button, 6-18, 8-16, ACCURACY TEST, 14-8
ACCURACY TEST Report, 14-24
14-9, 14-27 ADD ITEM, 12-3
AutoConstancy Test Report ADVANCED CHAMBER, 6-3, 6-10
AUTOCONSTANCY, 6-18, 8-16, 14-9,
from Accuracy Test, 14-27
from Daily Test, 14-9 14-27
AutoLinearity, 11-12
AUTOLINEARITY, 11-12
Aborting a Test, 11-21 BACKGROUND, 8-1
Pausing the Test, 11-17 BACKGROUND Report, 14-30
CAL#, 3-3, 10-2, 10-4
Power Failure, 11-26 CALICHECK, 6-42, 11-37
Reports, 11-27 CANCEL, 6-8
CDC, 5-17
Resuming the Test, 11-18 CH, 3-4, 10-2, 10-10, 12-4
Saving the Test, 11-23 CH (CHAMBER)/WELL, 3-4
Selecting Nuclide, 11-13 CHAMBER, 10-10
CHAMBER VOLTAGE Report, 14-18
Setting Interval, 11-15 CHAMBER VOLTS, 8-11
Setting Total, 11-16 CLEAR, 6-17, 6-27, 6-37, 11-45
CLEAR ALL MEASUREMENTS, 11-33

July 16 INDEX i-1

CAPINTEC, INC. CRC®-55t

CLEAR SELECTED NUCLIDE, 6-8, 6- REPORTS, 14-2
9, 6-20, 6-53 RESUME TEST, 11-19
SAVE, 11-23, 11-25
CONTINUE, 2-4, 8-3, 8-6 SCREEN CALIB, 5-23
DAILY, 8-6 SEARCH, 11-28, 14-6, 14-13, 14-18,
DAILY TEST Report, 14-4
DATE FORMAT, 5-5 14-24, 14-30
DATE/TIME, 3-3, 10-2
DELETE ALL, 12-29 SETUP, 3-3, 6-3, 6-9, 6-21
DELETE CURRENT SETTINGS, 6-41, SETUP CALNUM, 6-27
SETUP LINEARITY, 6-33
6-46 SETUP MOLY, 6-21
SETUP PET CHAMBER HOTKEYS, 6-
DELETE INVENTORY, 12-28
DELETE ITEM, 12-28 5
DIAGNOSTICS, 9-2 SETUP R CHAMBER HOTKEYS, 6-5
DOSE DECAY, 3-4, 10-2, 10-3, 10-6, SETUP REMOTE, 6-50
SETUP SOURCES, 6-12
10-7 SINGLE STRIP TEST, 11-46
SPANISH, 5-18
DOSE TABLE, 13-2 STANDARD, 11-33
DOWN ARROW (▼), 3-7, 6-7, 6-8, 6- STANDARD PET CHAMBER, 6-35
STANDARD R CHAMBER, 6-35
19, 6-28, 6-35, 6-52, 11-14, 11-28, START, 11-16, 11-42
START TEST, 11-12
12-5, 13-4, 14-6, 14-14, 14-19, 14-25, STOP, 11-44
SUBMIT, 12-15
14-31, 15-4 TEST, 3-3, 5-24, 10-2
TWO STRIP TEST, 11-49
ENGLISH, 5-18 UNITS, 3-3, 10-2, 10-5
ENHANCED TESTS, 11-1 UP ARROW (▲), 3-7, 6-19, 6-28, 6-35,
ESPANOL, 5-20
FINISHED, 11-9 6-52, 11-14, 11-28, 12-5, 13-4, 14-6,
FULL, 10-7 14-14, 14-19, 14-25, 14-31, 15-4
GEOMETRY, 11-4
HALF-LIFE CALCULATOR, 11-40 UTIL, 9-1, 13-1, 14-2, 15-1
HMPAO TEST, 11-52 VIEW, 11-29, 14-7, 14-14, 14-20, 14-26,
INACTIVATE, 11-30, 14-9, 14-16, 14-
14-32
22, 14-28, 14-33
WELL, 2-5, 10-2
INGLES, 5-20 WELL/CH (CHAMBER), 3-4
INVENTORY, 12-2 WITHDRAW, 12-23, 12-24
LEGACY, 5-17 YES, 6-42, 11-32, 14-11, 14-17, 14-23,
LINEARITY, 11-11
LINEATOR, 6-37, 11-35 14-29, 14-35
LOGIN, 2-4, 4-10
MAG3 TEST, 11-54 ZERO Report, 14-12
MAKE KIT, 12-16
MEASURE ACTIVITY, 12-22, 12-26 C
MEASUREMENT, 3-3, 10-2, 10-5 CAL# button, 3-3, 10-2, 10-4
MOLY ASSAY, 6-21, 10-12 Cal#(P), 6-25
MORE MEASUREMENTS, 11-9
NO, 6-42, 11-32, 14-11, 14-17, 14-23, Cal#(R), 6-25

14-29, 14-35 Calculation Utilities, 15-1

NUCLIDE, 3-2, 6-7, 10-2, 10-4 Ci /Bq Conversion, 15-2
PRINT, 3-3, 5-7, 10-2, 11-30, 11-45, 14-
Decay Calculator, 15-4
7, 14-9, 14-14, 14-15, 14-19, 14-21,
Calibration Numbers, 6-25, 6-27, 6-29, 10-
14-25, 14-27, 14-31, 14-33
5
QC, 11-46
QUICK, 10-7 adding, 6-27

i-2 INDEX July 16

CAPINTEC, INC. CRC®-55t

changing, 4-10, 6-27 Source, 4-10, 6-18
Contamination Test, 7-4
determining, 6-26 CONTINUE button, 2-4, 8-3, 8-6

entering, 6-25, 10-4 D
DAILY button, 8-6
Limits, 6-26 Daily Chamber Test, 7-2
Daily Test, 7-3, 8-6
new, 6-27
Accuracy Test, 8-10
restoring, 6-29 Auto Zero, 8-6
CALICHECK button, 6-42, 11-37 Background, 8-7
Calicheck Linearity Test, 11-37 Chamber Voltage, 8-9
Constancy Test, 8-10
defining, 6-42 Data Check, 8-9
Inactivate, 14-9
Report, 11-39 Individual Results Report, 14-9
CANCEL button, 6-8 Report, 7-2, 14-4
Canister, 4-11, 10-12 Search Results Report, 14-7
View an Individual Test Report, 14-7
CapMac, 4-11, 10-12 DAILY TEST Report button, 14-4
Data Check, 8-9
Caution Labels, 1-2 Date
choosing format, 5-5
CDC, 5-17 entering, 5-1
CDC button, 5-17 DATE FORMAT button, 5-5
CH (CHAMBER)/WELL button, 3-4 DATE/TIME
CH button, 3-4, 10-2, 10-10, 12-4 button, 3-3, 10-2
CHAMBER button, 10-10 Screen, 3-7
Chamber Hotkeys, 2-2 Decay Calculator, 15-4
Activity, 15-5
Assignment, 6-2 Date, 15-5
Results, 15-9
Chamber Initialization, 6-1 Select Nuclide, 15-4
DELETE ALL button, 12-29
Chamber Measurement Delete All Items, 12-29
DELETE CURRENT SETTINGS button, 6-
Procedures, 10-3 41, 6-46
Delete Individual Items, 12-28
Screen, 3-2 DELETE INVENTORY button, 12-28
Chamber Voltage, 8-9, 8-11 DELETE ITEM button, 12-28
Deleting from Inventory, 12-28
Inactivate, 14-21 Description
Functional, 2-2
Individual Results Report, 14-21 Technical, 2-6
Detailed Report
Report, 14-18 AutoLinearity Test, 11-29
Diagnostics, 9-1
Search Results Report, 14-19 Test, 7-2, 7-5
DIAGNOSTICS button, 9-2
View an Individual Test Report, 14-20 Dimensions, 2-8
CHAMBER VOLTAGE Report button, 14-

18
CHAMBER VOLTS button, 8-11
Changing Chamber, 10-10

Ci / Bq

Choosing, 5-3

Conversion, 15-2

Cleaning, 16-1, 16-2

Instructions, 16-2
CLEAR ALL MEASUREMENTS button,

11-33
CLEAR button, 6-17, 6-27, 6-37, 11-45
CLEAR SELECTED NUCLIDE button, 6-

8, 6-9, 6-20, 6-53

Communications Device Class, 5-17

Constancy Channels, 6-18

adding nuclides, 6-19

removing nuclides, 6-20

Constancy Test, 8-10, 8-17

July 16 INDEX i-3

CAPINTEC, INC. CRC®-55t

Cables, 2-8 Functional Description, 2-2
Chamber, 2-8 functions, 2-4
Console, 2-8 Fuse Servicing, 16-5
Dipper, 4-7
cleaning, 16-2 G
disinfecting, 16-3 general operating instructions, 3-1
Disinfecting, 16-2 Geometry, 7-1
Instructions, 16-3 GEOMETRY button, 11-4
Disposal, 16-4 Geometry Test, 11-4
division factors, 1-4
Dose Decay, 10-6 Measurement Results, 11-10
Entering Date/Time, 10-7 Report, 11-10
Entry Modes, 10-6 Ground Current, 2-7
DOSE DECAY button, 3-4, 10-2, 10-3, 10-
6, 10-7 H
Dose Table, 13-1 Half-life, 6-24
Activity, 13-5 Half-life Calculator, 11-40
Dose, 13-6
Interval, 13-7 Calculated Half-life, 11-44
Nuclide, 13-3 Clear Current Half-Life Result, 11-45
printing, 13-8 Setting Interval, 11-40
report, 13-8 Setting Total, 11-41
Results, 13-7 Starting the Measurements, 11-42
Volume, 13-5 Stopping the Measurements Early, 11-
DOSE TABLE button, 13-2
DOWN ARROW (▼) button, 3-7, 6-7, 6-8, 44
6-19, 6-28, 6-35, 6-52, 11-14, 11-28, 12- HALF-LIFE CALCULATOR button, 11-40
5, 13-4, 14-6, 14-14, 14-19, 14-25, 14- HID, 5-17
31, 15-4 HMPAO Test, 11-52

E report, 11-53
Electromagnetic Interference, 1-5 HMPAO TEST button, 11-52
Element, 6-24 Holders, optional, 4-7
EMC, 2-9 Home Screen Hotkeys, 3-2, 6-2, 10-4
EMI, 1-5, 5-7 Hotkeys
English, 5-18
ENGLISH button, 5-18 Home Screen, 3-2, 6-2, 10-4
Enhanced Tests, 11-1 Select Nuclide Screen, 6-8
Environment Requirements, 2-6, 4-8 Human Interface Device, 5-17
Error Message
I
Moly Assay Inactivate
Activity Too Low, 10-19
Caution, 10-20 Accuracy/AutoConstancy Test, 14-27
MO Too High, 10-20 AutoLinearity Test, 11-30
Background Measurement, 14-33
ESPANOL button, 5-20 Chamber Voltage Measurement, 14-21
ETL Listed, 2-9 Daily Test, 14-9
Zero Measurement, 14-15
F INACTIVATE button, 11-30, 14-9, 14-16,
FINISHED button, 11-9 14-22, 14-28, 14-33
FULL button, 10-7 INGLES button, 5-20
Installation, 4-1
Intended Use, 2-1
Inventory, 12-1
Activity, 12-10

i-4 INDEX July 16

CAPINTEC, INC. CRC®-55t

Input, 12-10 Linearity Test, 7-1, 7-2, 7-5
Measure, 12-12, 12-26 Calicheck, 6-42
Measure Kit, 12-22 Calicheck-defining, 6-42
adding to, 12-3 Calicheck-deleting definition, 6-46
Date, 12-12 Calicheck-Number of Measurements, 6-
Delete All Items, 12-29 43
Delete Individual Items, 12-28 Calicheck-Serial Number, 6-43
deleting from, 12-28 definition, 6-30
ID, 12-7 Lineator, 6-37
Kit Activity, 12-20 Lineator-Calibration Procedure, 6-38
Kit ID, 12-19 Lineator-defining, 6-37
Kit Lot Number, 12-19 Lineator-deleting definition, 6-40
Kit Study, 12-18 Lineator-Serial Number, 6-38
Kit Volume, 12-21 performing, 11-11
Lot Number, 12-8 Standard, 6-34
Make Kit, 12-16 Standard- Time Interval, 6-36
Mo/Tc, 12-14 Standard-defining, 6-34
printing, 12-27 Standard-deleting definition, 6-37
Save Item, 12-15 Standard-Number of Measurements, 6-
Select Nuclide, 12-4 36
Study, 12-6 Standard-Select Nuclide, 6-35
Time of Use, 12-25
Volume, 12-9 LINEATOR button, 6-37, 11-35
Withdraw Activity, 12-24 Lineator Linearity Test
withdrawing activity from, 12-23
INVENTORY button, 12-2 report, 11-36
Inventory List Report, 12-27 Lineator Test, 11-35
IVBT, 1-3, 1-4
Calibration Procedure, 6-44
K defining, 6-37
Kit Liner, 1-4, 4-7
cleaning, 16-2
Activity, 12-20 disinfecting, 16-3
ID, 12-19 liners, 16-2
Lot Number, 12-19 List Screen, 3-7
making, 12-16 LOGIN button, 2-4, 4-10
measure activity, 12-22 Low Activity Measurements, 10-3
Study, 12-18 Low Battery, 16-6
Volume, 12-21
M
L MAG3 Test, 11-54
Labels
report, 11-56
Caution, 1-2 MAG3 TEST button, 11-54
High Voltage, 1-3 maintenance, 16-1
Shipping, 16-15 Make Kit, 12-16
Warning, 1-2 MAKE KIT button, 12-16
Language, 5-18 MEASURE ACTIVITY button, 12-22, 12-
Legacy, 5-17
LEGACY button, 5-17 26
Line Filter, 2-7 Measurement
LINEARITY button, 11-11
Calibration Number, 10-4
July 16 Nuclide Specification, 10-4
Procedures:, 10-3
MEASUREMENT button, 3-3, 10-2, 10-5

INDEX i-5

CAPINTEC, INC. CRC®-55t

Mo/Tc Printing, 5-7
Inventory, 12-14 Dose Table, 13-8
Limit, 4-11 Inventory, 12-27
Record of measurment, 10-11
Mo99
Assay, 10-16 Profile
Background, 10-13 Operator, 2-1

Moly Assay, 4-11, 10-12 Program Flow, 2-5
CapMac or Canister, 6-20
Limit, 6-22 Q
Results, 10-19 QC button, 11-46
Selecting Method, 6-22 QC Tests, 11-46
Setup, 6-20
Volume Entry, 10-19 HMPAO Test, 11-52
MAG3 Test, 11-54
MOLY ASSAY button, 6-21, 10-12 Single Strip Test, 11-46
MORE MEASUREMENTS button, 11-9 Two Strip Test, 11-49
multiplication factors, 1-4 Quality Assurance Tests, 7-3, 16-3
Quality Control, 11-46
N HMPAO Test, 11-52
NO button, 6-42, 11-32, 14-11, 14-17, 14- MAG3 Test, 11-54
Single Strip Test, 11-46
23, 14-29, 14-35 Two Strip Test, 11-49
Nuclide, 6-24 Quarterly Tests, 7-5
QUICK button, 10-7
specification, 10-4
NUCLIDE button, 3-2, 6-7, 10-2, 10-4 R
Nuclide Screen Hotkeys, 2-2 Radiochromatography, 11-46
Nuclides, 6-23 Range

adding, 4-10, 6-23 limits, 2-8, 2-9
deleting, 6-27 recalibration, 1-5
Numeric Data Receiving Condition Examination, 4-1
entering, 3-5 Regulatory Listings, 2-9

O Electrical, 2-9
On / Off Switch, 2-6 EMC, 2-9
Operational Setup, General, 4-10 ETL Listed, 2-9
Operator Related Products, 16-14
Remote Display
Profile, 2-1 Indicator, 3-4
Training, 2-2 Remote Display (Legacy Systems Only)
Adding or Changing Nuclides, 6-51
P Deleting Nuclides, 6-53
Performance, 2-8 Nuclide Assignment, 6-47
Power Requirements, 4-8 report
Precision Accuracy Test, 14-24
Accuracy Test from Daily Tests, 14-8
Selecting Measurement, 10-5 Accuracy Test Individual Results, 14-27
Preventative Maintenance, 16-3 Accuracy Test Search Results, 14-25
AutoConstancy Test from Accuracy
Quality Assurance Tests, 16-3
PRINT button, 3-3, 5-7, 10-2, 11-30, 11- Test, 14-27
AutoConstancy Test from Daily Tests,
45, 14-7, 14-9, 14-14, 14-15, 14-19, 14-
21, 14-25, 14-27, 14-31, 14-33 14-9
printers, 4-10
power requirements, 5-7
selection, 5-8

i-6 INDEX July 16

CAPINTEC, INC. CRC®-55t

AutoLinearity Test Detailed, 11-29 Accuracy Test No Source Data, 8-12
Background Measurements, 14-30 Accuracy Test Results, 8-16, 14-26
Background Measurements Individual Add Inventory Item, 12-4
Advanced Chamber Setup, 6-5, 6-12, 6-
Results, 14-33
Background Measurements Search 33, 6-50, 10-6
Alphanumeric Keypad, 3-6, 11-32, 14-
Results, 14-31
Calicheck Linearity Test, 11-39 11, 14-17, 14-23, 14-29, 14-35
Chamber Voltage, 14-18 Auto Zero Measurement Results, 8-7
Chamber Voltage Search Results, 14-19 AutoLinearity After Measurements
Chamber Voltage Test Individual
Begin, 11-16
Results, 14-21 AutoLinearity Menu Screen after Pause,
Daily Test Individual Results, 14-9
Daily Test Search Results, 14-7 11-19
Daily Tests, 14-4 AutoLinearity Report, 11-30
Diagnostics, 9-2 AutoLinearity Setup, 11-13
Dose Decay, 10-9 AutoLinearity Test Finished, 11-25
Dose Table, 13-8 AutoLinearity with Save Button, 11-23
Geometry, 11-10 Background Measurement, 8-3
HMPAO Test, 11-53 Background Measurement Results, 8-8,
Inventory List, 12-27
Lineator Linearity Test, 11-36 14-32
MAG3 Test, 11-56 Background Remove Sources, 8-2
Moly Assay, 10-22 Background Results, 8-5
Single Strip Test, 11-49 Background Too High, 8-4
Standard Linearity Test, 11-34 Calicheck Linearity Next Measurement,
Two Strip Test, 11-52
Zero Measurement Search Results, 14- 11-39
Calicheck Linearity Test First
14
Zero Measurements, 14-12 Measurement, 11-38
Zero Measurements Individual Results, Calicheck Test Measurement, 6-44
Chamber Measurement, 3-2, 6-2, 8-2, 9-
14-15
REPORTS button, 14-2 1, 10-2, 11-2, 12-2, 13-1, 15-1
Reports Module, 14-1 Chamber Measurement Screen with
Requirements
Dose Decay, 10-9
Environment, 2-6, 4-8 Chamber Voltage Measurement
Power, 2-7, 4-8
RESUME TEST button, 11-19 Results, 8-11
Chamber Voltage Test Results, 14-20
S Constancy Test Results, 8-17
Safety, 1-1 Daily Test, 8-6
Safety Tips, 1-6 Daily Test Report, 14-8
SAVE button, 11-23, 11-25 Daily Test Results, 8-9
Screen Date Format Selection, 5-5
Date/ Time, 3-8
Abort AutoLinearity Test After Finished, Date/Time, 6-16, 12-13, 12-26, 15-5, 15-
11-22
8
Abort AutoLinearity Test In-progress, Date/Time for Full Entry Mode, 10-8
11-21 Date/Time Quick Entry Mode, 10-8
Decay Calculator Results, 15-9
Accuracy Test, 8-13 Default Linearity Tests, 11-11
Accuracy Test Measurement, 8-14 Delete Calicheck Settings, 6-47
Accuracy Test Measurement Error, 8-15 Delete Lineator Settings, 6-42
Dose Decay Entry Mode Selection, 10-7
Dose Table, 13-3

July 16 INDEX i-7

CAPINTEC, INC. CRC®-55t

Dose Table Results, 13-7 MAG3 Test – Cartridge Measurement,
Enhanced Tests for PET Chambers, 11- 11-55

3 MAG3 Test – Fraction #1 Measurement,
Enhanced Tests for R Chambers, 11-3 11-54
Enter Activity, 6-15, 12-11, 12-20, 12-
MAG3 Test Results, 11-56
25, 13-6, 15-3, 15-7 Main, 2-5, 14-1
Enter Cal#, 6-25 Main showing Low Battery, 16-7
Enter Current Time, 5-3 Make Inventory Kit, 12-18
Enter Password, 5-2, 6-4, 6-11, 6-32, 6- Make Kit – Please Select Study, 12-19
Make Kit High Activity Message, 12-21
49 Make Kit Too Little Volume Message,
Enter Serial Number, 6-14
Geometry Measurement, 11-6 12-22
Geometry Measurement with Added Measure Activity, 13-5
Moly Assay Background Choice, 10-14
Volume, 11-8 Moly Assay Background Measurement
Geometry Report, 11-10
Geometry Test, 11-4 Accept, 10-15
Geometry Test Added Volume Entry, Moly Assay Background Measurement

11-7 Too High, 10-16
Geometry Test Initial Volume Entry, 11- Moly Assay Eluate Accept, 10-17
Moly Assay Method Selection, 6-21, 10-
5
Geometry Test More Measurements / 13
Moly Assay Results – Less than 12
Finished, 11-9
Half-life Calculator, 11-40 hours, 10-21
Half-life Calculator After Measurements Moly Assay Results – More than 12

Begin, 11-43 hours, 10-22
Half-life Calculator After Measurements Moly Assay Tc99m Accept, 10-19, 10-

Finished, 11-44 20
HMPAO Test Results, 11-53 Numeric Keypad, 3-5, 11-15, 11-41, 11-
Inactivate Record, 11-31, 14-10, 14-16,
42
14-22, 14-28, 14-34 Pause AutoLinearity Test – Back Button,
Inventory – Add Item with Submit
11-17
Button, 12-15 Pause AutoLinearity Test – Home
Inventory – Added Tc99m with Study
Button, 11-18
Field Box, 12-6 Please Measure Withdrawal, 12-27
Inventory – ID QC Test Selection, 11-46
Reports, 14-3
Field Box, 12-8 Resume AutoLinearity Test, 11-20
Inventory – Lot Save AutoLinearity Test After Finished,

Field Box, 12-9 11-26
Inventory – Measure Activity, 12-12 Save AutoLinearity Test In-progress, 11-
Inventory – Tc99m Select Study, 12-7
Inventory – Volume 24
Search Accuracy Tests after Searching,
Field Box, 12-10
Inventory Main, 12-3 14-25
Inventory Main with Added Item, 12-16 Search AutoLinearity Tests, 11-27
Inventory with All Buttons, 12-24, 12-28 Search AutoLinearity Tests with Test
Kit Please Measure Withdrawal, 12-23
Linearity Test First Measurement, 11-36 Highlighted, 11-29
Linearity Test Message, 6-30 Search Background Measurements after
Lineator Test Measurement, 6-39
List, 3-7 Searching, 14-31
Search Chamber Voltage Tests after

Searching, 14-19

i-8 INDEX July 16

CAPINTEC, INC. CRC®-55t

Search Daily Tests, 14-4 Two Strip Test – Top of Strip B
Search Daily Tests after Searching, 14-
Measurement, 11-51
6
Search Daily Tests with Highlighted Test Two Strip Test Results, 11-52

Result, 14-7 USB PC Driver Select, 5-17
Search Zero Measurements, 14-12
Search Zero Measurements after Utility, 9-2, 13-2, 14-2, 15-2

Search, 14-13 Utility Screen with Diagnostics System
Select Chamber, 10-10
Select Language, 5-18 Test PASS, 9-3
Select Language – Spanish, 5-20
Select Nuclide, 6-7, 6-19, 6-52, 10-4, Zero Measurement Results, 14-15

11-14, 12-5, 13-4, 15-4 Screen Adjustments, 5-21
Select Tc99m for Kit, 12-17
Setup, 5-4, 5-21, 6-3, 6-10, 6-31, 6-48 Brightness, 5-22
Setup – after Day Month Year Selected,
Calibrarion, 5-23
5-6
Setup – after Year Month Day Selected, Sleep Brightness, 5-22

5-6 Sleep Timeout, 5-21
Setup – Spanish, 5-19 SCREEN CALIB button, 5-23
Setup CalNum, 6-28 Screen Calibration, 5-23
Setup Linearity, 6-34 SEARCH button, 11-28, 14-6, 14-13, 14-
Setup Linearity Calicheck, 6-43
Setup Linearity Calicheck Results, 6-45 18, 14-24, 14-30
Setup Linearity Calicheck View Results,
Select Nuclide Screen Hotkeys, 6-8
6-46
Setup Linearity Lineator, 6-38 Selecting Measurement Precision, 10-5
Setup Linearity Lineator Results, 6-40
Setup Linearity Lineator View Results, Servicing, 16-5

6-41 Fuse, 16-5
Setup Linearity Standard, 6-35
Setup Moly, 6-22 Setup
Setup Nuclide, 6-23
Setup PET Chamber Hotkeys, 6-6 General Operational, 4-10
Setup R Chamber Hotkeys, 6-6 SETUP button, 3-3, 6-3, 6-9, 6-21
Setup Remote Nuclides, 6-51 SETUP CALNUM button, 6-27
Setup Sources, 6-13, 6-18 SETUP LINEARITY button, 6-33
Single Strip Test – Bottom of Strip SETUP MOLY button, 6-21
SETUP PET CHAMBER HOTKEYS
Measurement, 11-48
Single Strip Test – Top of Strip button, 6-5
SETUP R CHAMBER HOTKEYS button,
Measurement, 11-47
Single Strip Test Results, 11-49 6-5
Standard Linearity Test Remaining SETUP REMOTE button, 6-50
SETUP SOURCES button, 6-12
Measurements, 11-33 Shipping, 16-15
Standard Linearity Test Results, 11-34
Start AutoLinearity, 11-12 Single Strip Test, 11-46
Startup, 2-4, 4-9
Touchscreen Calibration, 5-23 report, 11-49
Two Strip Test – Bottom of Strip A SINGLE STRIP TEST button, 11-46
Sleep Brightness, 5-22
Measurement, 11-50
Sleep Timeout, 5-21

Spanish, 5-18
SPANISH button, 5-18
Speaker Volume, 5-24
STANDARD button, 11-33
Standard Linearity Test, 11-32

Defining, 6-34

report, 11-34
STANDARD PET CHAMBER button, 6-35
STANDARD R CHAMBER button, 6-35
START button, 11-16, 11-42
START TEST button, 11-12

July 16 INDEX i-9

CAPINTEC, INC. CRC®-55t

STOP button, 11-44 Report, 11-52
SUBMIT button, 12-15 TWO STRIP TEST button, 11-49
Symbol Definitions, 1-1
System Initialization, 5-1 U
System Setup, 4-1 UNITS button, 3-3, 10-2, 10-5
Unpacking, 4-1
T UP ARROW (▲) button, 3-7, 6-19, 6-28,
Table
6-35, 6-52, 11-14, 11-28, 12-5, 13-4, 14-
Calibration Number Limits, 6-26, 10-5 6, 14-14, 14-19, 14-25, 14-31, 15-4
Disposal Information, 16-4 USB PC driver, 5-17
Test Source calibration activity limits, 6- UTIL button, 9-1, 13-1, 14-2, 15-1

15 V
Tc99m VIEW button, 11-29, 14-7, 14-14, 14-20,

Assay, 10-18 14-26, 14-32
Technical Description, 2-6 View Individual Test Report
Test
Accuracy, 14-26
Report, 7-3 Background Measurements, 14-32
TEST button, 3-3, 5-24, 10-2 Chamber Voltage, 14-20
Test Source, 4-10 Daily Test, 14-7
Zero Measurements, 14-14
Activity Field, 6-14 Voltage
adding, 6-13 High, 1-3, 1-4
Constancy, 6-18
Constancy Field, 6-18 W
Daily Field, 6-17 Warm Up Period, 2-6
Date Field, 6-16 Warning Labels, 1-2
deleting, 6-17 WELL button, 2-5, 10-2
editing, 6-17 Well Counter System Test
S/N Field, 6-13
Setup, 6-9 Report, 7-3, 7-4
Testing Well Test, 7-3, 7-4
Acceptance, 4-11, 7-1 WELL/CH (CHAMBER) button, 3-4
Accuracy, 7-2, 7-3, 8-12 Withdraw Activity, 12-24
Chamber Daily, 7-2 WITHDRAW button, 12-23, 12-24
Contamination, 7-4 Withdrawing activity from Inventory, 12-23
Daily, 7-3, 8-6
Diagnostics, 7-2, 7-5 Y
Linearity, 7-2, 7-5 Yearly Tests, 7-5
Quality Assurance, 7-3 YES button, 6-42, 11-32, 14-11, 14-17, 14-
Quality Control (QC), 11-46
Quarterly, 7-5 23, 14-29, 14-35
Well, 7-3, 7-4
Yearly, 7-5 Z
Time Zero Measurements
entering, 5-1
Time of Use, 12-25 Inactivate, 14-15
Training Individual Results Report, 14-15
Operator, 2-2 Report, 14-12
Troubleshooting, 16-13 Search Results Report, 14-14
Two Strip Test, 11-49 View an Individual Test Report, 14-14
ZERO Report button, 14-12

i - 10 INDEX July 16

CAPINTEC, INC. CRC®-55t

NEW PRODUCT WARRANTY

Conditions and Limitations

CAPINTEC warrants our products to be free from defects in material and
workmanship for a period of 24 months after delivery to the original buyer
(unless indicated otherwise). Our obligation under this warranty is
limited to servicing or adjusting products returned to our factory for that
purpose promptly upon any claim under this warranty, provided our
examination discloses to our satisfaction that the affected parts were
originally defective and have not been subjected to misuse, abuse,
mishandling or improper operation. This warranty extends to every part
of the product except batteries, fuses, ink, magnetic storage media,
toner, transistors, tubes or any other normally consumable item. In no
event shall we liable for transportation, installation, adjustment or any
other expenses which may arise in connection with such products or
their servicing or adjustment.

This warranty is expressly in lieu of any other express or implied
warranties, and we make no warranty that the products sold herein are
merchantable or are suitable for any particular purpose. The benefits of
this warranty shall extend only to the buyer and to none other. We shall
have no liability of any nature arising out of the use or application of the
product in conformity with this warranty. If the product shall fail to
perform in conformity with the warranty made herein, we shall be liable
solely for the repair and replacement of the product as provided
hereinabove, and shall have no other liability in respect of such failure or
performance, or any consequences therefrom, including, without
limitation, business impairment or loss, personal injury or property
damage.

No agent, employee, or representative of ours has any authority to bind
us to any representation or warranty concerning the product sold herein,
and unless a representation or warranty made by an agent, employee or
representative is specifically included herein, it shall not be enforceable
by the buyer. No waiver, alteration or modification of the foregoing
conditions shall be valid, unless made in writing and signed by an
executive officer of our corporation.

CAPINTEC, INC. CRC®-55t

This page intentionally left blank.