CRC ® -35R Dose Calibrator Manual- Revision E

CRC-35R Manual - 9250-0054 Rev E

CAPINTEC, INC. CRC®-35R

ACCURACY TEST is the next part of Daily Test. It can also be reached directly from Test
Menu.

In order to perform Accuracy Test in Daily Test, a reference source designated as
ACCURACY or BOTH and DAILY must have been previously entered. See:

UTILITIES, SETUP, REFERENCE SOURCE DATA.

Daily test will step you through measuring every daily accuracy source.

Fig 8-12 will appear (for each daily accuracy source)

OK

[Fig 8-12]

Place the indicated source in the chamber and press OK to begin the measurement.

The accuracy source is measured and compared to the current activity of the accuracy
source. The current activity of the source is displayed and the deviation displayed.
If the deviation is too high, ERROR will appear.

SAVE

[Fig 8-13]
Press SAVE to end the measurement and save the data.
Fig 8-12 and 8-13 are repeated for each daily accuracy source.

August, 94 8-7 TESTS

CAPINTEC, INC. CRC®-35R

CONSTANCY TEST is the next part of Daily Test. It can also be reached directly from Test
Menu.

In order to perform Constancy Test in Daily Test, a reference source designated as
CONSTANCY or BOTH and DAILY must have been previously entered. See:

UTILITIES, SETUP, REFERENCE SOURCE DATA.

In order to compare results to a base value, the chosen nuclide must be a constancy nuclide.
See:

UTILITIES, SETUP, DEFINE TESTS, CONSTANCY TEST

Daily test will step you through measuring every daily constancy source.

Fig 8-14 will appear (for each daily constancy source)

OK

[Fig 8-14]
Place the indicated source in the chamber and press OK to begin the measurement.
Fig 8-15 appears.
Any nuclide may be chosen for constancy testing using NUCLIDE KEYS, USER KEYS or
NUCL KEY.

August, 94 8-8 TESTS

CAPINTEC, INC. CRC®-35R

DONE BASE SAVE

[Fig 8-15]

Press BASE to make the current activity the base activity
Press SAVE to save the data for this measurement.
Press DONE to end this test.

If the current nuclide is a constancy nuclide, the deviation from a base activity (decay
corrected to the present time) can be calculated and displayed. If there is no base activity or
if you want to change the base activity, press BASE to make the current activity the base
activity.

Data for constancy nuclides can be printed on the report. Press SAVE to save the current
result for the report.

Press DONE to end constancy test using this constancy source.

Figs 8-14 and 8-15 are repeated for each daily constancy source.

Daily Test results can now be printed:

EXIT PRINT

Press EXIT to return to test menu [Fig 8-16]
Press PRINT to print Daily Report.

August, 94 8-9 TESTS

CAPINTEC, INC. CRC®-35R

ACCURACY TEST

If Accuracy Test is chosen from the Test Menu, Fig 8-17 appears listing all the accuracy
sources:

[Fig 8-17]
Select the accuracy source to measure.

EXIT SAVE PRINT

[Fig 8-18]
Press EXIT to return to Test menu without saving test result.
Press SAVE to save test result and return to Test menu.
Press PRINT to print test result and remain on screen.

August, 94 8 - 10 TESTS

CAPINTEC, INC. CRC®-35R

CONSTANCY TEST

If Constancy Test is chosen from the Test Menu, Fig 8-19 appears listing all the constancy
sources:

[Fig 8-19]
Select the constancy source to measure.

DONE BASE SAVE

[Fig 8-20]
Press BASE to make the current activity the base activity
Press SAVE to save the data for this measurement.
Press DONE to end this test.

August, 94 8 - 11 TESTS

CAPINTEC, INC. CRC®-35R

After DONE is pressed, a constancy report may be printed:

EXIT PRINT

[Fig 8-21]
Press EXIT to return to Test menu
Press PRINT to print test result and remain on screen.

August, 94 8 - 12 TESTS

CAPINTEC, INC. CRC®-35R

LINEARITY TEST

There are three types of linearity test that can be performed with the CRC®-35. The type of
test must be selected and initialized in UTIL, DEFINE TESTS, LINEARITY

The Linearity test should be performed quarterly.

STANDARD TEST

The standard test measures the activity of Tc99m over a period of time.
The results are plotted on a semi-log graph and the best fit is found. The deviation of the
results from this fit are reported.

If 2 or more measurements have already been made, a summary screen will be displayed to
allow the previous results to be reviewed and the last measurement repeated if so desired.

Each measurement begins with Fig 8-22, showing the Test number and for all tests except
the first one the time at which the measurement should be made and the actual elapsed
time.

OK

[Fig 8-22]
Place the source in the calibrator and press OK to begin the measurement

August, 94 8 - 13 TESTS

CAPINTEC, INC. CRC®-35R

OK

[Fig 8-23]
The measurement will begin and will be displayed.
Press OK to accept the measurement.
If this is the first , you will be returned to the Test Menu.
For subsequent tests, the predicted activity will be displayed as in Fig 8-24

MEAS OK

[Fig 8-24]
If this measurement seems wrong, press MEAS to repeat the measurement.

Press OK to accept the measurement.

August, 94 8 - 14 TESTS

CAPINTEC, INC. CRC®-35R

A summary screen appears with up to 4 tests per screen

NEXT

[Fig 8-25]

Press NEXT to view the next summary screen (if more tests have been made) or to view the
graph screen (Fig 8-26)

SAVE PRINT

[Fig 8-26]
This illustrates the screen after four measurements. This is a semi-log graph.
The X's are the measured activity.
The dotted line is the best fit calculated from a least squares fit.

Press SAVE to save the results and return to Test Menu screen.
Press PRINT to print and save the results. Remains on this screen.
Note: The least squares fit and predicted values are re-calculated after each measurement.

August, 94 8 - 15 TESTS

CAPINTEC, INC. CRC®-35R

LINEATOR TEST

The lineator test steps you through the measurement of each tube.
Figs 8-27 and 8-28 show the measurement for tube #5.
The tube number and name are shown and the measurement begins.

OK

[Fig 8-27]

Press OK to accept the measurement. The present factor and percent ratio will be filled in:

MEAS NEXT

[Fig 8-28]

Press MEAS if you think the measurement needs to be repeated
Press NEXT to accept the measurement and go to the next tube.

August, 94 8 - 16 TESTS

CAPINTEC, INC. CRC®-35R

After the last tube is measured summary screens will appear with 4 tubes per screen.
Fig 8-29 illustrates the summary of the first 4 tubes

NEXT

[Fig 8-29]
Press NEXT to display results for next 4 tubes.
The last summary screen is illustrated in Fig 8-30

SAVE PRINT

[Fig 8-30]
Press SAVE to save the results and return to Test Menu screen.
Press PRINT to print and save the results. Remains on this screen.

August, 94 8 - 17 TESTS

CAPINTEC, INC. CRC®-35R

CALICHECK TEST

The calicheck test steps you through the measurement of each tube. Figs 8-31 and 8-32
show the measurement for tube #5.
The tube number and name are shown and the measurement begins.

OK

[Fig 8-31]

Press OK to accept the measurement. The result (measurement x calibration factor) will be
filled in.

MEAS NEXT

[Fig 8-32]
Press MEAS if you think the measurement needs to be repeated
Press NEXT to accept the measurement and go to the next tube.

August, 94 8 - 18 TESTS

CAPINTEC, INC. CRC®-35R

After the last tube is measured summary screens will appear with 4 tubes per screen.
Fig 8-33 illustrates the summary of the first 4 tubes

NEXT

[Fig 8-33]
Press NEXT to display results for next 4 tubes.
The last summary screen is illustrated in Fig 8-34

SAVE PRINT

[Fig 8-34]
Press SAVE to save the results and return to Test Menu screen.
Press PRINT to print and save the results. Remains on this screen.

August, 94 8 - 19 TESTS

CAPINTEC, INC. CRC®-35R

GEOMETRY TEST

Performance of the Geometry Test is required on installation of the calibrator and determines
the effect of volume changes on the calibrator's accuracy.
The Geometry Test begins with Fig 8-35

S/V OK

[Fig 8-35]

Press S/V to toggle between Syringe and Vial

The test can be performed using a syringe or vial.
If a vial is used the original volume is entered and then the added volume is entered for each
measurement.
If a syringe is used, the total volume for each measurement is entered.

Select syringe or vial using the S/V key.

Press OK when selection is correct.

August, 94 8 - 20 TESTS

CAPINTEC, INC. CRC®-35R

When a Vial is used Fig 8-36 appears:

OK

[Fig 8-36]
Enter the volume of the vial in ml and press OK when finished.

Figs 8-37 and 8-38 appear for each sample to be measured.
The example is for the second sample.

OK

[Fig 8-37]

Press OK when correct added volume has been entered
The meaasurement will begin.

August, 94 8 - 21 TESTS

CAPINTEC, INC. CRC®-35R

DONE NEXT

[Fig 8-38]

Press DONE when all samples have been measured (will not appear on screen for
Sample A).

Press NEXT to measure next sample.

When DONE is pressed, a summary of the measurements appears:

[Fig 8-39]
Choose the source to be used as the reference.

August, 94 8 - 22 TESTS

CAPINTEC, INC. CRC®-35R

The variations with respect to the chosen source will be calculated and the summary will
appear:

SAVE PRINT

[Fig 8-40]
Press SAVE to save the test results and return to TEST MENU.
Press PRINT to print the test results and remain on this screen.
If Syringe had been chosen in Fig 8-35, Fig 8-36 would be omitted and Fig 8-41 would
appear instead of Fig 8-37:

OK

[Fig 8-41]
Press OK when correct total volume has been entered.
The rest of the test is similar to using a vial.

August, 94 8 - 23 TESTS

CAPINTEC, INC. CRC®-35R

QUALITY CONTROL
The CRC®-35R can perform radiochromatography strip tests using four methods.
Select the test method from the Quality Control Menu (Fig 8-42)

[Fig 8-42]

You will be stepped through the chosen test.

After viewing the result screen you may:

Press EXIT to return to Quality Control Menu
Press NEXT to perform another QC Test
Press PRINT to print the test results.

The result screens for each test are shown.

Single Strip / Single Solvent:

EXIT NEXT PRINT

[Fig 8-43]

August, 94 8 - 24 TESTS

CAPINTEC, INC. CRC®-35R

Two Strips / Two Solvents:

EXIT NEXT PRINT

[Fig 8-44]

Tc99m HMPAO (Exametazine):

EXIT NEXT PRINT

[Fig 8-45]
Tc99m Mertiatide (MAG3):

EXIT NEXT PRINT

[Fig 8-46]

August, 94 8 - 25 TESTS



CAPINTEC, INC. CRC®-35R

Chapter 9

INVENTORY

The CRC®-35R is capable of maintaining an inventory of 100 items. These can be multi-
dose vials or unit doses. When the desired dose and time are input, the CRC®-35R will
indicate the correct volume to be withdrawn from a multi-dose vial in inventory (Withdraw
Dose from Inventory). A dose can also be dispensed from inventory.

The Inventory functions are accessed from the Measurement screen by pressing
INV (F2) key.

The main Inventory Screen (Fig 9-1) will then appear displaying all items in inventory for the
current nuclide (Tc99m in the example).

The items appear on the screen in the order of reverse entry -- most recently entered is first
and oldest is last.

If an item has expired, an X will appear in front of it.

If there are no inventory items for the chosen nuclide, only the ADD and DEL keys will be
displayed.

Use the NUCLIDE KEYS, USER KEYS or NUCL KEY to display inventory for another
nuclide, if desired.

ADD DEL DRAW DISP

[Fig 9-1]

Press ADD to add a vial or unit dose to inventory
Press DEL to delete item(s) from inventory
Press DRAW to withdraw a dose from a vial in inventory (and optionally dispense it)
Press DISP to dispense an entire item in inventory

August, 94 9-1 INVENTORY

CAPINTEC, INC. CRC®-35R

ADDING AN ITEM TO INVENTORY

To add a vial or unit dose to inventory , press ADD key in Fig 9-1

A list of the possible compounds for the current nuclide is displayed (Fig 9-2). The first item
on the list is always the 'bare' nuclide.

Compounds for a different nuclide can be displayed by using the NUCLIDE KEYS, USER
KEYS or NUCL KEY

[Fig 9-2]

Use the arrows keys to highlight the compound to be added to inventory.
Press ENTER when the desired compound is highlighted.

Fig 9-3 now appears with the chosen compound filled in (the example shows the screen after
the lot # and quantity have been entered).

ml/cap ACTIV OK

[Fig 9-3]
Note: The Mo99/Tc99m field will only be displayed when the nuclide is Tc99m

August, 94 9-2 INVENTORY

CAPINTEC, INC. CRC®-35R

ml/cap Toggles the quantity field between ml and number of capsules
ACTIV Toggles between MEAS and INPUT

LOT #: Enter the item's lot number. This can be number and/or letters.
The maximum length is 16 characters. This field is optional.

QUANTITY: For a vial, enter the volume in ml. For capsules, enter the number of
capsules.
Press ml/cap key to change from ml to cap.
This is the only required field.

Mo99/Tc99m: Enter the ratio of Mo99 to Tc99m in µCi/mCi or MBq/GBq
This field is optional.
This field will only be displayed if the nuclide is Tc99m

EXP DATE: Enter the item's expiration date in the chosen date format. (optional)
EXP TIME: Enter the item's expiration time (optional).

The activity of the item can either be input by the user or measured in the calibrator.
Use the ACTIV key to select MEAS or INPUT

Press OK when all entries are correct.

If MEAS has been chosen, the measurement will begin when OK is pressed (Fig 9-4)

NEXT SAVE PRINT

[Fig 9-4]

The item being added to inventory is now being measured.
The measurement is accepted when the keys are pressed.

August, 94 9-3 INVENTORY

CAPINTEC, INC. CRC®-35R

Press NEXT to save the item to inventory and add another item to inventory.
Returns to Fig 9-2

Press SAVE to save the item to inventory. Returns to Measurement Screen.

Press PRINT to print a record of the item and save it to inventory.
Remains on this screen.

If INPUT had been selected in Fig 9-3, the calibration activity and calibration date/time must
be entered (Fig 9-5)

UNIT NEXT SAVE PRINT

[Fig 9-5]

Press UNIT to change Unit from µCi -> mCi -> Ci (MBq -> GBq if in Bq)

CALIB ACTIVITY: Enter the calibration activity of the item being entered into inventory.
DATE: TIME: Use UNIT key to change the activity unit if necessary.
Enter the date and time of the calibration.

Press NEXT to save the item to inventory and add another item to inventory.
Returns to Fig 9-2

Press SAVE to save the item to inventory. Returns to Measurement Screen.

Press PRINT to print a record of the item and save it to inventory.
Remains on this screen.

August, 94 9-4 INVENTORY

CAPINTEC, INC. CRC®-35R

INVENTORY DELETION

The inventory has to be 'cleaned up' from time to time by removing items which have expired
or which are empty.

To delete items from inventory, press DEL (F2) in Fig 9-1.
The entire inventory will be displayed (Fig 9-6).
The items will appear in the order of entry (oldest first) since these are the ones most likely to
be deleted.

If the inventory for only one nuclide is desired, use the NUCLIDE KEYS, USER KEYS or
NUCL KEY.

SELECT ALL DE SEL DONE

[Fig 9-6]

NOTE
An item selected for deletion is indicated by a # sign to the left of the item's name.

Press SELECT to add the highlighted item to the list of items to be deleted.

Press ALL to delete all the items displayed.

Press DE SEL to remove the highlighted item from the list of items to be deleted. (# will be
removed)

Press DONE when all items to be deleted have been selected.

When DONE is pressed, you will be given a chance to print the inventory before the selecting
items are deleted:

Do you want to print inventory
before deleting? (Yes or No)

August, 94 9-5 INVENTORY

CAPINTEC, INC. CRC®-35R

WITHDRAWING FROM INVENTORY

A portion of an item in inventory may be withdrawn, saved to inventory for future use or
dispensed to a patient. This facility will tell you how much to withdraw to achieve a specified
activity at the specified time.

To WITHDRAW FROM INVENTORY:
Use the NUCLIDE KEYS, USER KEYS or NUCL KEY to display the inventory for the
desired nuclide (Fig 9-1)

Highlight the item from which to withdraw.

Press DRAW (F3)

Fig 9-7 will be displayed with the current item's name, current activity and volume displayed.
The default date/time of 'now' will be filled in.

In the example in Fig 9-7, an activity of 10.0 mCi has been entered.

The current activity and volume are displayed. Input the desired dose (Rx) and the time.
The default time is the current time. Change the time/date of the dose if it will not be given
'at the current time (now).
When a valid Rx has been input, the WITHDRAW field will be filled in by the CRC®-35R with
the correct volume to achieve the dose.

MEAS UNIT

[Fig 9-7]
Press UNIT to toggle between mCi and µCi or MBq and GBq

August, 94 9-6 INVENTORY

CAPINTEC, INC. CRC®-35R

Rx: Enter the activity desired.
Use the UNIT key to change the activity unit if so desired.
TIME:
DATE: Enter time is activity entered is for a future time.
Enter if activity entered is not for 'today'.

When a valid Rx has been entered, the volume to be withdrawn is calculated and displayed.
In the example in Fig 9-8 , a dose of 10.0 mCi is desired from this item at 18:30 (6:30 pm) on
the same day.

MEAS UNIT

[Fig 9-8]

Remove the required volume from the vial and place it in the calibrator.
Press MEAS to begin the measurement.

August, 94 9-7 INVENTORY

CAPINTEC, INC. CRC®-35R

The measured activity is displayed after NOW:
If the time / date has been input, THEN: will be displayed with the activity at the input
TIME / DATE

OK

[Fig 9-9]

Press OK to accept the measurement and update the inventory of the item from which the
activity is being withdrawn.

The actual volume withdrawn is calculated from the measured volume and displayed
(Fig 9-10)

DISP SAVE PRINT

[Fig 9-10]

Press DISP to dispense this dose to a patient. The inventory will automatically be updated by
subtracting the withdrawn amount.

To add the withdrawn item to inventory press SAVE. The inventory will automatically be
updated by subtracting the withdrawn amount, and adding this withdrawn amount as a new
item to inventory.

To save and print a ticket, press PRINT.

August, 94 9-8 INVENTORY

CAPINTEC, INC. CRC®-35R

DISPENSING WITHDRAWN DOSE TO A PATIENT

Press DISP in Fig 9-10 to directly dispense this dose to a patient (at the time in Fig 9-10).
Fig 9-11 appears with the RX is automatically filled in from the data input above
.

NOTE
The RX cannot be changed since the dose has already been determined.

OK

[Fig 9-11]

PAT ID and TECH ID may be input. (not required)
Press OK when all entries are correct.
The current activity and activity at the input time will be displayed as in Fig 9-12

SAVE PRINT

[Fig 9-12]

Press SAVE to save patient data and return to Fig 9-1
Press PRINT to print ticket and saves patient data. Fig 9-12 remains on this screen.

August, 94 9-9 INVENTORY

CAPINTEC, INC. CRC®-35R

NOTE
The activity at the Rx time will be checked against the patient's Rx to see if it is within the
allowed range input by the user in utilities.

DISPENSING A DOSE FROM INVENTORY

An entire inventory item may be dispensed from inventory to a patient.
After the item is dispensed, it is removed from inventory.
This procedure is similar to dispensing dose from Measurement Screen.
To DISPENSE A DOSE FROM INVENTORY

Use the NUCLIDE KEYS, USER KEYS or NUCL KEY to display the inventory for the
desired nuclide (Fig 9-1)
Highlight the item from which to withdraw.
Press DISP (F4)

The screen for inputting time/date of dose will appear (Fig 9-13)

DAY OK

[Fig 9-13]
Input time of dose and press ENTER, and the current day will appear as shown in Fig 9-14.

August, 94 9 - 10 INVENTORY

CAPINTEC, INC. CRC®-35R

APR 22 92 14:21
CHAMBER #3

THEN: 14:00 April 22, 1992

DAY OK .

[Fig 9-14]
The (F1) DAY key will toggle between today’s and tomorrow’s date. When the date and time
are correct. press OK.
The Patient Screen (Fig 9-15) will appear:

UNIT OK .

[Fig 9-15]

Tech's ID, Patient's ID and/or Patient's Rx may be entered (all are optional).

Press UNIT to toggle between mCi and µCi or MBq and GBq if necessary.
Press OK when all entries are correct.

August, 94 9 - 11 INVENTORY

CAPINTEC, INC. CRC®-35R

SAVE PRINT

[Fig 9-16]

Press SAVE to save patient data . Returns to Fig 9-1
Press PRINT to print ticket and save data. Fig 9-16 remains on this screen.

August, 94 9 - 12 INVENTORY

CAPINTEC, INC. CRC®-35R

Chapter 10

UTILITIES

The Utilities are a group of functions including: Reports, Calculator, Setup and Diagnostics.
The Utilities are reached by pressing the UTIL key from the measurement screen.
After pressing UTIL the menu in Fig 10-1 appears.

[Fig 10-1]
Use Arrows to highlight selection. Press ENTER to make selection.

REPORTS

When Report Menu is chosen the Menu in Fig 10-2 appears:

August, 94 [Fig 10-2] UTILITIES
10 - 1

CAPINTEC, INC. CRC®-35R

TEST REPORTS

When Test Reports is chosen the Menu in Fig 10-3 appears:

[Fig 10-3]

Summaries of the Accuracy, Linearity and Geometry Tests may be printed. The last test
performed, or all saved tests may be printed. After printing, you will be given a chance to
delete the printed tests from memory.

The next screen will be similar to Fig 10-4 for all tests.
The example is for Accuracy Test.

[Fig 10-4]

If PRINT LAST TEST is selected, only the last Accuracy test done will be printed.
If PRINT ALL TESTS is selected, all the accuracy tests in memory will be printed.

The user will then have the opportunity to delete all the accuracy tests from memory, and the
screen will return to the report selection screen.

August, 94 10 - 2 UTILITIES

CAPINTEC, INC. CRC®-35R

INVENTORY REPORT

Select Inventory Report to print a summary of all the items in inventory.
Items are listed in the order of oldest to newest. All activities are the values at the time of the
printing. If an item has expired, an X appears in front of its name.

PATIENT REPORT

Select Patient Report to print a summary of doses given to patients.
The patient report prints out the Rx, measured activity, and the dispensed dose for all saved
patients.
After the report is printed, you will be given an opportunity to delete all patient data.

NUCLIDE LIST

Select Nuclide List to print a list of all nuclides which can be used with the CRC®-35R.
The nuclide list contains all of the nuclides provided with the system, and the nuclides added
by the user. The Nuclide name, it's half-life and calibration number are printed. If a
calibration number has been changed, an asterisk (*) will appear to the left of the nuclide
name.

COMPOUND LIST

Select Compound List to print a list of the compounds provided with the system and the
compounds added by the user. The compounds which will be displayed are indicated by
YES.

REFERENCE DECAY CALENDAR

Select Reference Decay Calendar to list the daily activity of any reference source which has
been entered. The activity is printed in the form of a calendar which covers a period of 1
year beginning with the current month.

CALCULATOR / Ci ↔ Bq 10 - 3 UTILITIES

August, 94

CAPINTEC, INC. CRC®-35R

The calculator feature allows the user to convert an activity from Ci to Bq or from Bq to Ci
and to find the activity of a source given the initial activity.

The calculator menu is shown in Fig 10-5:

[Fig 10-5]

CONVERT Ci TO Bq
If Conversion of Ci to Bq is selected, Fig 10-6 appears:

UNIT DONE

[Fig 10-6]

Enter the activity in Curies.
The activity in Bequerels will appear after the = sign.

Press UNIT to change unit from mCi to Ci to µCi.
Press DONE when all desired conversions from Ci to Bq have been made, and returns to Fig
10-5.

CONVERT Bq TO Ci
If Conversion of Bq to Ci is selected, Fig 10-7 appears:

August, 94 10 - 4 UTILITIES

CAPINTEC, INC. CRC®-35R

UNIT DONE

[Fig 10-7]

Enter the activity in Bequerels.
The activity in Curies will appear after the = sign.

Press UNIT to change unit from MBq to GBq.
Press DONE when all desired conversions from Bq to Ci have been made and return to Fig
10-5.

August, 94 10 - 5 UTILITIES

CAPINTEC, INC. CRC®-35R

DECAY CALCULATION
If Decay Calculation is chosen, Fig 10-8 appears with the current nuclide displayed:

DONE OK

[Fig 10-8]

Change the nuclide if so desired by pressing a nuclide selection key, user selection key or
NUCL key.

Press OK when the correct nuclide is displayed.
Press DONE when all desired decay calculations have been performed.

When OK is pressed, Fig 10-9 appears:

DONE UNIT NEXT OK

[Fig 10-9]

Enter the calibrated activity and the date and time of calibration.
Enter the date and time for which you want the activity calculated.
Press UNIT to change the unit of activity from mCi to Ci to µCi or from MBq to GBq.
Press OK to display the new activity.

When UNIT is pressed or a new calibration activity or date/time is entered, the currently
displayed new activity will be erased.

Press NEXT to return to Fig 10-8 and select another nuclide.
Press DONE when all desired decay calculations have been performed.

August, 94 10 - 6 UTILITIES

CAPINTEC, INC. CRC®-35R

SETUP

Setup consists of a group of utilities which should only be changed by an authorized person
and a password must therefore be entered after SETUP is selected on the Utility Menu
Screen (Fig 10-1).

When Setup Menu is selected, PASSWORD: appears on the screen. Enter the last three
digits of the Main Unit's serial number. The Setup Menu (Fig 10-10) will appear:

[Fig 10-10]

Ci / Bq Allows locking in Ci or locking in Bq or allows Ci \ Bq
to be user selectable

REFERENCE SOURCES Brings up Reference Source Menu

NUCLIDE DATA Brings up Nuclide Data Menu

USER KEY ASSIGNMENTS The 5 user keys may be assigned to a nuclide stored in
memory or a user input nuclide

COMPOUND MENU Brings up Compound Menu

DEFINE TESTS Brings up menu to define linearity test and constancy
test

PARAMETERS / USER PREFERENCES
Used to setup Mo Assay; input allowed variations;
choose type of printing

CONFIGURATION MENU Menu for system configuration of chambers and
remotes

TIME / DATE SETUP Setup of Time and Date

HOSPITAL NAME Up to 30 alpha numeric characters can be entered.
This name will appear on all report headings.

August, 94 10 - 7 UTILITIES

CAPINTEC, INC. CRC®-35R

The choices listed are described in Chapter 12, System Setup.

PERFORM DIAGNOSTICS

Diagnostics is used to print some system parameters and to test the system's integrity.
When the tests are completed, Fig 10-11 will appear. In this example, the system has
passed all the tests. Note: the checksum shown after the PROM test is just an example and
is not the value which will be shown.

Press any key to continue

[Fig 10-11]

If any of the tests fail, repeat diagnostics. If there is a failure, call Capintec Service.

A diagnostics report will automatically be printed. This report shows the system
configuration, system parameters, the user key assignments, a summary of the reference
source data and the user parameters. An example is shown in the two pages of Fig 10-12.

August, 94 10 - 8 UTILITIES

CAPINTEC, INC. CRC®-35R

CRC-35R RADIOISOTOPE CALIBRATOR REV 2.08 Aug 29 1994 14:22
S/N MAIN UNIT: 350020 Capintec, Inc. USA

DIAGNOSTICS

PROM VALUE: AEB91D94

MEMORY TESTS
RAM: PASS

PROM: PASS
EEPROM: PASS

SYSTEM CONFIGURATION

Number of Chambers :1

Number of Remote Displays :0

Chamber ID# S/N Rev Assigned to Remote ID# S/N Rev

1 1 350020 1.18

SYSTEM PARAMETERS

Gain Response Linearity Nominal
Corr. (%) Corr. (%)
Chamber -1.44 Corr. (%) Volts
1 3.02
0.50 155.2

-------------------------------------------------------------------

CRC-35R RADIOISOTOPE CALIBRATOR REV 2.08 Aug 29 1994 14:22
S/N MAIN UNIT: 350020 Capintec, Inc. USA

USER KEY ASSIGNMENTS

U1: None
U2: None
U3: None
U4: None
U5: None

REFERENCE SOURCES CALIBRATION CURRENT
NUCLIDE S/N ACTIVITY
DATE ACTIVITY USE

A: Accuracy Source C: Constancy Source

USER PARAMETERS

Calibrator Test Results: 1.0%
Patient Dose: 10.0%
0.150µCi/mCi
Mo99 / Tc99m Ratio:

Printing Method: TICKET

Date Input Format: MMDDYY

[Fig 10-12]

August, 94 10 - 9 UTILITIES



CAPINTEC, INC. CRC®-35R

Chapter 11

SETUP

The selections in Setup Menu are described in this chapter. See the preceding chapter for
directions for entering password.

Ci / Bq

During measurements, the user may switch between Ci and Bq or the key may be locked out
so that only Ci or only Bq may be used. Select USER SELECTABLE to be able to switch
between Ci and Bq. Select Ci ONLY to lock out Bq. Select Bq ONLY to lock out Ci.
If USER SELECTABLE is not selected, a beep will sound when Ci / Bq Key is pressed.

If Ci / Bq is chosen from the SETUP MENU, the Fig 11-1 appears:

Ci/Bq SAVE

[Fig 11-1]

The current setting is displayed.

Press Ci/Bq to change the setting from USER SELECTABLE to Ci ONLY to Bq ONLY.
Press SAVE to accept the displayed setting and return to SETUP menu.

August, 94 11 - 1 SETUP

CAPINTEC, INC. CRC®-35R

REFERENCE SOURCES

Before an accuracy or constancy test can be performed, a description of the Reference
Source(s) to be used with the test must be entered.

Reference Source Menu
The first selection on this menu is for entering Standard Reference Source Data for Co60,
Co57, Ba133, or Cs137.
The second selection which is for additional reference nuclides must be used before entering
other isotopes as reference sources.

Refer to Fig 11-2:

[Fig 11-2]
To enter a Standard Reference Source into memory, select REFERENCE SOURCE DATA,
Fig 11-3 appears:

SOURCES

ADD DONE

[Fig 11-3]

August, 94 11 - 2 SETUP

CAPINTEC, INC. CRC®-35R

ADDING A REFERENCE SOURCE
To add a reference source, press ADD key.
Fig 11-4 will appear with all the built-in reference nuclides and user added nuclides listed:

[Fig 11-4]

Use arrow keys to highlight desired reference nuclide and then press ENTER.
Select the desired isotope and press ENTER. Fig 11-5 will appear:

UNIT USE DAILY OK

[Fig 11-5]

The nuclide chosen will appear after SOURCE.

Press UNIT to toggle between µCi, mCi, Ci or Bq units.
Press USE to toggle between use for Accuracy, for Constancy , for both or for neither.
Press DAILY to toggle between YES (use daily) and NO (not for daily use).

SERIAL NUMBER: Enter the serial number (numbers and letters). SETUP
11 - 3
August, 94

CAPINTEC, INC. CRC®-35R

Maximum length is 20 characters.

CALIBRATION DATE: Enter the calibration date in the format chosen.

ACTIVITY: Enter the calibration activity. Use the UNIT key if necessary.

USE: Default is BOTH (used for Constancy and Accuracy tests).
DAILY: If the source is to be used for only one of these or neither,
press the USE key to change this setting.
The default is to use the source in the daily test. If it is not to
be used daily, use the DAILY key to change this setting.

Press OK when all entries are correct.

NOTE
If an error is made, the data may be edited at a later time.

DELETING A REFERENCE SOURCE

If a reference source is no longer in use, it may be deleted from the list of reference sources.

When Fig 11-6 appears, use the arrow keys to highlight the source to be deleted and press
DEL key.

The data shown in Fig 11-8 will appear and you will be asked if you want to delete that
source.
Press YES to delete, NO to keep.

EDITING A REFERENCE SOURCE

Data entered for a reference source may be changed.

When Fig 11-9 appears, use the arrow keys to highlight the source to be edited and press
EDIT key.

Fig 11-5 will appear with the current data filled in. Make the desired changes and press OK.

If the wrong nuclide was chosen, select PREV SCREEN by pressing the 2nd key and the 1
key, to return to Fig 11-3.

If the maximum number of additional reference nuclides (4) has already been added, an

error message will appear: NO ROOM TO ADD REF NUCLIDE

If a reference source is not one of the standard nuclides (Co57, Co60, Ba133, Cs137)
select ADDITIONAL REFERENCE NUCLIDES to add that nuclide to the list of
reference nuclides.

August, 94 11 - 4 SETUP

CAPINTEC, INC. CRC®-35R

DONE

[Fig 11-6]
The standard reference nuclides and current additional nuclides are shown.
Use NUCLIDE KEYS, NUCL KEYS or USER KEYS to choose the nuclide to add or delete.
Press DONE when all desired changes to Reference Nuclides have been made.

ADDING A REFERENCE NUCLIDE
If the Nuclide is not a Reference nuclide (Ra226 in the example) Fig 11-7 appears:

DONE ADD

[Fig 11-7]

Press ADD to add Ra226 to the list of additional reference nuclides and return to Fig 11-6
(Ra226 will now be listed under ADDITIONAL).

DELETING A REFERENCE NUCLIDE
If the nuclide chosen is already an additional reference nuclide, Fig 11-8 appears:

August, 94 11 - 5 SETUP

CAPINTEC, INC. CRC®-35R

DONE DEL

[Fig 11-8]

Press DEL to delete Ra226 from the list of additional reference nuclides and return to
Fig 11-7 (Ra226 will no longer be listed under ADDITIONAL).

If the wrong nuclide was chosen, press PREV SCREEN to return to Fig 11-7.
If a Co57 or Co60 or Ba133 or Cs137 is chosen, an error message will appear because
these nuclides can neither be added nor deleted.

Select REFERENCE SOURCE DATA to input, edit, or delete calibration data for a reference
source.

Fig 11-9 appears with a list of the current reference sources, showing nuclide number and
serial number.

If no reference sources have been entered, only ADD and DONE keys appear:

ADD EDIT DEL DONE

[Fig 11-9]

Press ADD to enter data for a new reference source.
Press EDIT to make changes in the highlighted reference source data.
Press DEL to delete the highlighted reference source.
Press DONE when all changes to reference sources have been made.

NUCLIDE DATA MENU

Calibration data for over 80 nuclides are built into the CRC®-35R.
Calibration numbers for up to 10 of these nuclides may be changed.

August, 94 11 - 6 SETUP

CAPINTEC, INC. CRC®-35R

Data for up to 10 nuclides can be added by the user.
To obtain a list of all nuclides, select NUCLIDE LIST from the REPORT MENU.
To edit Nuclide Data, select NUCLIDE DATA from the Setup Menu. Fig 11-10 will appear:

[Fig 11-10]

Select ADDITIONAL NUCLIDES to add a nuclide to memory or delete an existing additional
nuclide.
Fig 11-10 will appear with the current additional nuclides (if any) displayed.

DONE

[Fig 11-11]

Press DONE when all changes to additional nuclides have been made.

Enter the name of the additional nuclide (in alphanumeric characters) to add or delete using
the number / letter keys and the arrow keys.
The entry will be forced to correspond to a legitimate nuclide format.

If the nuclide is one of the built in nuclides an error message will appear since those nuclides
can neither be added nor deleted.

August, 94 11 - 7 SETUP

CAPINTEC, INC. CRC®-35R

ADDING A NUCLIDE
If the entered nuclide name is not an additional nuclide (AL28 in the example) Fig 11-12 will
appear:

DONE ADD

[Fig 11-12]

Press ADD to add the nuclide.

If the maximum number of nuclides (10) have been added, an error message will appear.
If the nuclide can be added, the Fig 11-13 will appear:

UNIT OK

[Fig 11-13]

Press UNIT to select the half-life unit : M (minutes), H (hours), D (days), Y (years)

HALF-LIFE: Enter the half-life of the nuclide. Use the UNIT key to select the correct unit.
CALIBRATION NUMBER: Enter the calibration number for the nuclide.
Press OK when all entries are correct.

DELETING A NUCLIDE 11 - 8 SETUP