CAPINTEC, INC. CAPTUS® 700t

Collimator with
Probe Detector

Pointer

Figure 4-2 Collimator Assembly with Pointer

If your system includes a Table Top Stand, installation consists of placing the components on
a sturdy table or bench as shown in Figure 4-3.

Collimator with Isolation
Probe Detector Transformer

Table Top Printer
Stand

Well Counter CAPTUS® 700t
(Optional) Readout

Figure 4-3 Complete Assembly with Table Top Stand

October 15 SYSTEM SETUP 4-3

CAPINTEC, INC. CAPTUS® 700t

If your system is a Well Counter system only, installation consists of placing the components
on a sturdy table or bench as shown in Figure 4-4.

Isolation
Transformer

Printer

Well Counter CAPTUS® 700t
Readout

Figure 4-4 Complete Well Counter System

Note: If the system includes a Well Counter, please verify that a Well Liner has been
inserted into the Well before use (refer to Figure 4-5 Well Counter Assembly). This
prevents contamination of the Well Detector if a spill or leak occurs.

Plug
Liner
Top Shield
Top Pipe Shield

Base Assembly

Figure 4-5 Well Counter Assembly

4-4 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

ENVIRONMENT REQUIREMENTS
Indoor use only. Pollution Degree 2, Altitude, and Installation Cat. II.

Operational
The instrument should be located where the level of the background radiation is as low and
as constant as possible.

The instrument should be located where the temperature is stable within a range of +50°F to
+85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-condensing to
warrant maximum reliability and accuracy.

The instrument should be located where the barometric pressure is within a range of 27–31
inches of mercury (91–105 kilopascals).

Storage
The instrument should be stored where the temperature is stable and the range is from
+39°F to +110°F (+4°C to +43°C) and the maximum relative humidity is 90% non-condensing
to warrant maximum reliability.

The instrument should be stored where the barometric pressure is within a range of 15–33
inches of mercury (51–112 kilopascals).

CAUTION: If these environmental requirements are not followed, the instrument may
display erroneous readings.

POWER REQUIREMENTS
CAUTION: If the input voltage to the following items is not within the stated
limits, the unit may not function correctly or may be damaged.

Input Power
Line Voltage:............. 115 or 230 Vac depending on model ordered
Line Frequency: ........ 60 or 50 Hz
Current:..................... 0.7 Ampere @ 115V
0.4 Ampere @ 230V
Fuse: For 115V use T 3.15AH, 250V, (Time Delay)
For 230V use T 2.5AH, 250V, (Time Delay)

TURN ON PROCEDURES

Verify all system components are mounted securely and all cables are connected. Connect
the Isolation Transformer main power cable to a medical grade wall socket.

October 15 SYSTEM SETUP 4-5

CAPINTEC, INC. CAPTUS® 700t

Note: To remove main power from the system, the Isolation Transformer main power cable
must be unplugged from the wall socket.

There are two power switches associated with the system:
• the Readout Unit has a power switch, and
• the printer has a power switch.

Turn on the printer first, then turn on the Readout Unit.

Your system will power up and display Figure 4-6 Startup Screen.

Note: After powering up, the screen may be white or have colored lines for up to 30
seconds before Figure 4-6 Startup Screen appears.

CAUTION: Changing the date or time while a timed test is in progress (for example,
a 24 hour Thyroid Uptake Test) might affect the results or introduce
unexpected error. Care should be taken to evaluate any potential impact
for ongoing tests before changing the date or time.

Figure 4-6 Startup Screen

Note: The screen will display the revision level of the installed software.

Note: If the Security Status is set to Traditional, the CONTINUE button will be replaced with
a LOGIN button. The Login Screen will include input boxes which will enable the user

4-6 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

to enter a User ID and a Password or to login as a guest. After logging in, the
CONTINUE button will appear. Reference the SECURITY STATUS section on page
4-8 for more information.

GENERAL OPERATIONAL SETUP
There are several things that must be done before using the CAPTUS® 700t Thyroid Uptake
System for the first time. The following briefly describes these steps:

• Verify or set the date/time: Although the date and time are set at the factory, you
should verify that the date and time are correct for your location. Reference
CHAPTER 5: SYSTEM INITIALIZATION, SECTION: SETTING DATE AND TIME.

• Select proper units: Ci or Bq. Although the system can be changed at any time, it is
recommended that the system be set for the proper units prior to use to prevent
confusion. Reference CHAPTER 5: SYSTEM INITIALIZATION, SECTION:
CHOOSING Ci or Bq.

• Enter Cs137 Test Source Data: The Test Source is used for the System Test.
Reference CHAPTER 6: DETECTOR INITIALIZATION, SECTION: TEST SOURCE
DATA.

• Perform Auto Calibration: The system must be calibrated before any measurements
can be performed. Reference CHAPTER 9: TESTS, SECTION: AUTO
CALIBRATION.

• Perform Efficiency measurements for nuclides: In order to select nuclides to look for
in Wipes, the efficiency of the desired nuclide must be input or measured. The system
includes a default efficiency for Cs137. Reference CHAPTER 6: DETECTOR
INITIALIZATION, SECTION: EFFICIENCY DATA.

ACCEPTANCE TESTING
The following tests should be performed prior to operational use of the unit.

Diagnostics Test
Reference CHAPTER 7: DIAGNOSTICS for instructions on how to perform this test.

Detector System Test
Reference CHAPTER 9: TESTS, SECTION: QUALITY ASSURANCE TESTS, System Test
for instructions on how to perform these tests.

October 15 SYSTEM SETUP 4-7

CAPINTEC, INC. CAPTUS® 700t

SECURITY STATUS
The purpose of the Security Status is to prevent unauthorized personnel from viewing patient
data and the BioAssay results of other Staff members.

In order to set the Security Status, you must be logged in as root. Reference the Login in as
root section on page 4-10.

The Security Status selections are:

• Disable.......... No password is required. Any user can access any part of the
system including Protected Modules. This is the default setting
when shipped from the factory.

• Traditional .... Login at power-up is required. Once logged in, the user can access
the modules as per their Security Role (see below).

• Enhanced ..... Login is at each Protected Module (see below). All users must
login for each of these modules and are logged-out when the
module is exited. Login is required each time the module is
accessed.

The following is a list of the Protected Modules that require a user to login if the Security
Status is set to Enhanced:

• Thyroid Uptake and Thyroid Uptake Reports

• BioAssay and BioAssay Reports

• Lab Test Reports

• Staff

The following is a list of the available Security Roles for users:

• Root ........ Root is a Super User and has complete access to all Protected
Modules and all privileges.

• Admin ..... These are users that have administrator privileges. Admins must be
added by the Root. These users can access all Protected Modules.
They can also access and view all Staff Members’ BioAssays. They
can add and edit users without Admin privileges.

• User ........ These are users that do not have administrator privileges. Users can
be added by the Root or Admins. These users can access all
Protected Modules. In BioAssay, they can only access and view their
own BioAssays. They cannot add or edit other users. They can edit
their own passwords and names but cannot change their Security
Role.

• Guest ...... A guest cannot access any Protected Module.

4-8 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Security Status and Staff Setup
From the Main Screen, touch the SETUP button. Figure 4-7 Setup Screen will appear.

Figure 4-7 Setup Screen

From Figure 4-7 Setup Screen, touch the STAFF button. Figure 4-8 Setup Staff Screen will
appear.

October 15 SYSTEM SETUP 4-9

CAPINTEC, INC. CAPTUS® 700t

Figure 4-8 Setup Staff Screen

Login in as root
In order to set the Security Status and add users with Admin privileges, you must be
logged in as root.

To log in as root, touch the LOGIN button. Figure 4-9 Login Screen will appear.

4-10 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 4-9 Login Screen

Touch the User Name: field box. Figure 4-10 Alphanumeric Keypad Screen will
appear.

October 15 SYSTEM SETUP 4-11

CAPINTEC, INC. CAPTUS® 700t

Figure 4-10 Alphanumeric Keypad Screen

Input root (all lower case letters) for the user name and touch the ACCEPT button.
Figure 4-9 Login Screen will re-appear with the User Name: field box populated with
“root”.

To cancel inputting the user name and return to Figure 4-9 Login Screen, touch the
CANCEL button.

Touch the Password: field box. Figure 4-10 Alphanumeric Keypad Screen will
appear.

The default password for root is AlphaDrive (case sensitive). Only root can change
this password.

Note: It is strongly recommended that the default password be changed.

Input AlphaDrive (case sensitive) for the password and touch the ACCEPT button.
Figure 4-9 Login Screen will re-appear with the Password: field box populated with
asterisks.

To cancel inputting the password and return to Figure 4-9 Login Screen, touch the
CANCEL button.

Once both the User Name: and Password: field boxes are populated, touch the
ACCEPT button to complete the root login process. Figure 4-11 Setup Staff Screen –
root will appear.

4-12 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 4-11 Setup Staff Screen – root

Change root Password
Note: It is strongly recommended that the default password be changed.

To change the default password for root, touch the root user name. The entire line
root will become highlighted and an EDIT button will appear in the lower left portion of
the screen as shown in Figure 4-12 Setup Staff Screen – root selected.

October 15 SYSTEM SETUP 4-13

CAPINTEC, INC. CAPTUS® 700t

Figure 4-12 Setup Staff Screen – root selected

Touch the EDIT button to change the root password. Figure 4-13 Edit User Screen –
root will appear

4-14 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 4-13 Edit User Screen – root

Note: The UserName:, First Name: and Last Name: field boxes can not be edited
for root.

Touch the Password: field box. An alphanumeric keypad will appear. Input the
desired password (case sensitive) and touch the ACCEPT button. Figure 4-13 Edit
User Screen – root will re-appear with the Password: field box populated with
asterisks. Allowable input is any combination of 13 alphanumeric characters
maximum.

Touch the Password (again): field box. An alphanumeric keypad will appear. Input
the same password (case sensitive) and touch the ACCEPT button. Figure 4-13 Edit
User Screen – root will re-appear with the Password (again): field box populated
with asterisks. Allowable input is any combination of 13 alphanumeric characters
maximum.

To cancel changing the password and return to Figure 4-11 Setup Staff Screen –
root, touch the CANCEL button.

To accept the password change and return to Figure 4-11 Setup Staff Screen – root,
touch the CANCEL button. The new password will be saved.

October 15 SYSTEM SETUP 4-15

CAPINTEC, INC. CAPTUS® 700t

Set Security Status
From Figure 4-11 Setup Staff Screen – root, set the desired Security Status by
touching the radio button next to the preferred status.

Verify that the desired radio button is selected.

Touch the BACK button to save the changes and return to Figure 4-7 Setup Screen.

ADDING A USER

To add a User, touch the ADD button from Figure 4-8 Setup Staff Screen or Figure 4-11
Setup Staff Screen – root. If root is logged in, Figure 4-14 Add New User Screen – root will
appear.

Figure 4-14 Add New User Screen – root

If a user with Admin privileges is logged in, Figure 4-15 Add New User Screen – Admin will
appear.

4-16 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 4-15 Add New User Screen – Admin

If a user without Admin privileges is logged in, the ADD button will not be displayed.

Note: All fields (UserName:, Password:, Password (again):, First Name:, Last Name:,
Role:) for the user are required. UserName and Password are case sensitive.

To input the information for the user, touch the field’s box and input the appropriate data for
the selected field as described in the following sections.

UserName Field
To input the User Name which will be used to log in, touch the UserName: field box.
Figure 4-16 Alphanumeric Keypad Screen will appear.

October 15 SYSTEM SETUP 4-17

CAPINTEC, INC. CAPTUS® 700t

Figure 4-16 Alphanumeric Keypad Screen

Input the name of the user (case sensitive) and touch the ACCEPT button. The Add
New User screen will re-appear with UserName: field box populated with the entered
name. Allowable input is any combination of 13 alphanumeric characters maximum.

To cancel any changes and return to the Add New User screen, touch the CANCEL
button.

Password Field
To input the Password which will be used to log in, touch the Password: field box.
Figure 4-16 Alphanumeric Keypad Screen will appear.

Input the desired password (case sensitive) for the user and touch the ACCEPT
button. The Add New User screen will re-appear with Password: field box populated
with asterisks. Allowable input is any combination of 13 alphanumeric characters
maximum.

To cancel any changes and return to the Add New User screen, touch the CANCEL
button.

4-18 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Password (again) Field
To verify the Password that was input, touch the Password (again): field box. Figure
4-16 Alphanumeric Keypad Screen will appear.

Input the same password (case sensitive) for the user and touch the ACCEPT button.
The Add New User screen will re-appear with Password: field box populated with
asterisks. Allowable input is any combination of 13 alphanumeric characters
maximum.

To cancel any changes and return to the Add New User screen, touch the CANCEL
button.

First Name Field
Note: The first and last name will be used for the BioAssay.

To input the First Name for the new user, touch the First Name: field box. Figure
4-16 Alphanumeric Keypad Screen will appear.

Input the desired first name for the user and touch the ACCEPT button. The Add New
User screen will re-appear with First Name: field box populated with the entered
name. Allowable input is any combination of 13 alphanumeric characters maximum.

To cancel any changes and return to the Add New User screen, touch the CANCEL
button.

Last Name Field
Note: The first and last name will be used for the BioAssay.

To input the Last Name for the new user, touch the Last Name: field box. Figure 4-16
Alphanumeric Keypad Screen will appear.

Input the desired last name for the user and touch the ACCEPT button. The Add New
User screen will re-appear with Last Name: field box populated with the entered
name. Allowable input is any combination of 13 alphanumeric characters maximum.

To cancel any changes and return to the Add New User screen, touch the CANCEL
button.

Role Field
To assign non-Admin privileges to the user, touch the User radio-button.

To assign Admin privileges to the user, touch the Admin radio-button.

Touch the CANCEL button to abort any changes. The Setup Staff screen will re-appear
without saving the new user information.

October 15 SYSTEM SETUP 4-19

CAPINTEC, INC. CAPTUS® 700t

Touch the ACCEPT button to save the new user information. The Setup Staff screen will re-
appear with the added user shown in the list.

EDITING USER DATA

To edit user data, touch the desired user’s name. The entire line will become highlighted and
an EDIT button will appear in the lower left portion of the screen as shown in Figure 4-17
Setup Staff with Name Highlighted – root.

Figure 4-17 Setup Staff with Name Highlighted – root

Note: If an Admin user is logged in, the first line (root-Super-User) will not be displayed.
Only root can edit these data. If a non-Admin user is logged in, only that user’s name
will be displayed.

Touch the EDIT button. Figure 4-18 Edit User Screen – root or Figure 4-19 Edit User Screen
– Admin will appear.

4-20 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 4-18 Edit User Screen – root

October 15 Figure 4-19 Edit User Screen – Admin 4-21

SYSTEM SETUP

CAPINTEC, INC. CAPTUS® 700t

The password, first name and last name can be changed. If root is logged in, the user’s role
can be changed. Reference the appropriate sections under ADDING A USER above for
instructions for each field box.

Touch the CANCEL button to abort any changes. The Setup Staff screen will re-appear
without saving the modified user information.

Touch the ACCEPT button to save the modified user information. The Setup Staff screen will
re-appear.

Inactivating a User
Users can be inactivated. An inactivated user cannot log in and Bio Assay cannot be run for
that user. However, in the Bio Assay report, that user can be searched for and past tests
displayed and printed. The user can always be re-activated.

When a user is inactivated, a line will appear through their name on the Staff Member list.

An Admin user can inactivate a non-admin user but not an admin user (including
themselves).

To inactivate a user, from Figure 4-11 Setup Staff Screen – root, touch the desired user’s
name. The entire line will become highlighted and the EDIT button will appear in the lower
left portion of the screen as shown in Figure 4-17 Setup Staff with Name Highlighted – root.

Touch the EDIT button. Figure 4-18 Edit User Screen – root or Figure 4-19 Edit User Screen
– Admin will appear.

Touch the Inactive checkbox. A check mark will appear in the box.

Touch the ACCEPT button to save the modified user information. The Setup Staff screen will
re-appear with a line will through the selected Staff Member’s name.

Touch the CANCEL button to abort any changes. The Setup Staff screen will re-appear
without saving the modified user information.

Reactivating a User
To reactivate a user, touch the checked Inactive checkbox. The check mark will disappear.

Touch the ACCEPT button to save the modified user information. The Setup Staff screen will
re-appear.

Touch the CANCEL button to abort any changes. The Setup Staff screen will re-appear
without saving the modified user information.

4-22 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

CHANGING USERS
When the Security Status is set to Traditional, the currently logged-in user can be changed
from the either the Utility screen or the Setup Staff screen.

Utility Screen
From the Main screen, touch the UTILITY button. Figure 4-20 Utility Screen with Login
Button will appear.

Figure 4-20 Utility Screen with Login Button

The currently logged-in user’s name will be displayed in the lower right portion of the Utility
screen next to the Login button.

To login as a different user, touch the LOGIN button. Figure 4-21 Login Screen with Guest
button will appear.

October 15 SYSTEM SETUP 4-23

CAPINTEC, INC. CAPTUS® 700t

Figure 4-21 Login Screen with Guest button

To abort changing users, touch the CANCEL button. Figure 4-20 Utility Screen with Login
Button will re-appear.

Touch the User Name: field box. An alphanumeric keypad will appear. Input the desired
user’s name (case sensitive) and touch the ACCEPT button. Figure 4-21 Login Screen with
Guest button will re-appear with the User Name: field box populated with the entered user
name.

To cancel inputting the user name and return to Figure 4-21 Login Screen with Guest button,
touch the CANCEL button.

Touch the Password: field box. An alphanumeric keypad will appear. Input the desired
user’s password (case sensitive) and touch the ACCEPT button. Figure 4-21 Login Screen
with Guest button will re-appear with the Password: field box populated with asterisks.

To cancel inputting the password and return to Figure 4-21 Login Screen with Guest button,
touch the CANCEL button.

Once both the User Name: and Password: field boxes are populated, touch the ACCEPT
button. Figure 4-20 Utility Screen with Login Button will re-appear with the new user’s name
displayed next to the Login button.

4-24 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

Setup Staff Screen
From the Main screen, touch the SETUP button. Figure 4-7 Setup Screen will appear.

From Figure 4-7 Setup Screen, touch the STAFF button. Figure 4-8 Setup Staff Screen will
appear.

To login as a different user, touch the LOGIN button. Figure 4-22 Login Screen will appear.

Figure 4-22 Login Screen

To abort changing users, touch the CANCEL button. Figure 4-8 Setup Staff Screen will re-
appear.

Touch the User Name: field box. An alphanumeric keypad will appear. Input the desired
user’s name (case sensitive) and touch the ACCEPT button. Figure 4-22 Login Screen will
re-appear with the User Name: field box populated with the entered user name.

To cancel inputting the user name and return to Figure 4-22 Login Screen, touch the
CANCEL button.

Touch the Password: field box. An alphanumeric keypad will appear. Input the desired
user’s password (case sensitive) and touch the ACCEPT button. Figure 4-22 Login Screen
will re-appear with the Password: field box populated with asterisks.

October 15 SYSTEM SETUP 4-25

CAPINTEC, INC. CAPTUS® 700t

To cancel inputting the password and return to Figure 4-22 Login Screen, touch the
CANCEL button.

Once both the User Name: and Password: field boxes are populated, touch the ACCEPT
button. Figure 4-8 Setup Staff Screen will re-appear and the new user will be logged-in.

Guest Login
When the Security Status is set to Traditional, a user can log in as a Guest.

To log in as a Guest, touch the UTILITY button on the Main screen. Figure 4-20 Utility
Screen with Login Button will appear.

Touch the LOGIN button. Figure 4-21 Login Screen with Guest button will appear.

Touch the Login as Guest button. Figure 4-20 Utility Screen with Login Button will re-appear
with “Guest” displayed next to the Login button.

The user will now be logged in as a Guest and can perform all functions but cannot access
any of the Protected Modules.

4-26 SYSTEM SETUP October 15

CAPINTEC, INC. CAPTUS® 700t

CHAPTER 5

SYSTEM INITIALIZATION

SET DATE AND TIME.....................................................................................5-1
CHOOSING Ci or Bq ......................................................................................5-4
CHOOSING DATE FORMAT ..........................................................................5-5
PRINTING .......................................................................................................5-7

Printer Power Requirements .....................................................................5-7
Printer Selection.........................................................................................5-8
Okidata Menu Set-Up .................................................................................5-9
Epson LX-300+II Menu Set-Up ..................................................................5-11
USB PC DRIVER............................................................................................5-17
SCREEN ADJUSTMENTS .............................................................................5-18
Sleep Timeout............................................................................................5-18
Sleep Brightness .......................................................................................5-19
Brightness .................................................................................................5-19
Screen Calibration.....................................................................................5-20
SPEAKER VOLUME ......................................................................................5-21

SET DATE AND TIME
The system date and/or time is set from the Main screen.

CAUTION: Changing the date or time while a timed test is in progress (for example,
a 24 hour Thyroid Uptake Test) might affect the results or introduce
unexpected error. Care should be taken to evaluate any potential impact
for ongoing tests before changing the date or time.

To change the date and/or time, touch the displayed date/time in the upper-right part of the
screen. A numeric keypad will appear to allowing the user to input a 3-digit password as
shown in Figure 5-1 Enter Password Screen.

October 15 SYSTEM INITIALIZATION 5-1

CAPINTEC, INC. CAPTUS® 700t

Figure 5-1 Enter Password Screen

Input the password (last 3 digits of the Readout serial number) by touching the appropriate
numbers on the keypad. Touch the ACCEPT button to accept the password. Figure 5-2
Enter Current Time Screen will appear.

5-2 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 5-2 Enter Current Time Screen

The screen displays the currently set date and time. Above and below each of the fields are
+ and – buttons, respectively. To adjust the displayed value for a specific field, touch the +
button to increase the displayed value or the – button to decrease the displayed value.

For the year, +10 and -10 buttons are provided to quickly change the tens column of the
years.

Touch the ACCEPT button to accept the new date/time or the CANCEL button to cancel any
changes. The Main screen will re-appear.

October 15 SYSTEM INITIALIZATION 5-3

CAPINTEC, INC. CAPTUS® 700t

CHOOSING CI OR BQ
Measurements can be displayed in Curies or Becquerels.

To select Ci or Bq, touch the SETUP button located on the Main screen. Figure 5-3 Setup
Screen will appear.

Figure 5-3 Setup Screen

In the Activity Unit: section at the top of the Setup Screen, touch the radio button next to the
preferred unit for measurements:

• Ci Only .......... the system will be set to display measurements in Curies only.
• Bq Only ......... the system will be set to display measurements in Becquerels only.

Verify that the desired radio button is selected.

Press the HOME or BACK button to save the changes and return to the Main screen.

5-4 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

CHOOSING DATE FORMAT
The date can be entered, displayed and printed in 3 formats:

• Month Day Year (mm/dd/yyyy)

• Day Month Year (dd/mm/yyyy)

• Year Month Day (yyyy/mm/dd)

In the Activity Unit: section at the top of Figure 5-3 Setup Screen, the current date format is
displayed in the upper right hand corner. The default date format is Month Day Year. This
format will be shown throughout the manual.

To change the date format, touch the Date Format: field box. Figure 5-4 Date Format
Selection Screen will appear.

Figure 5-4 Date Format Selection Screen

Touch the desired date format button. The Setup Screen will re-appear showing the selected
date format. If Day Month Year has been selected, Figure 5-5 Setup Screen after Day Month
Year Selected will be displayed. If Year Month Day has been selected, Figure 5-6 Setup
Screen after Year Month Day Selected will be displayed.

October 15 SYSTEM INITIALIZATION 5-5

CAPINTEC, INC. CAPTUS® 700t

Figure 5-5 Setup Screen after Day Month Year Selected

Figure 5-6 Setup Screen after Year Month Day Selected

5-6 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

PRINTING

CAUTION: If any printer other than one of the models supplied by Capintec is used,
Electromagnetic Interference (EMI) may be introduced into other devices
located in the same general area as the CAPTUS® 700t.

The system does not need a printer to function properly. When a printer is set up to work with
the system, a PRINT button will appear on all screens in which results can be printed from. If
the PRINT button is not displayed, then the system is either not set up to work with a printer
or printing is not permitted from that particular screen.

The CAPTUS® 700t may be used with the following printers:
• USB printers
o HP DeskJet 6000/8000 Series Inkjet (or equivalent) printer
• RS-232 printers
o Epson TM-295II Slip printer
o Epson TM-U220D Roll printer
o Okidata Microline 320
o Epson LX-300+II

Other printer models may be used. Please contact the factory for information on which
printers can be utilized.

Printer Power Requirements

Included Printer (Recommended)
HP DeskJet 6000/8000 Series Inkjet (or equivalent) Printer (optional)
100-240Vac, 50/60Hz, 1.5A

Optional Printers
Epson TM-295II Slip Printer (optional)
Input: 100-240Vac, 50/60Hz, 1.2A; Output: +24Vdc, 0.8A

Epson TM-U220D Roll Printer (optional)
Input: 100-240Vac, 50/60Hz, 1.0A max; Output: 24Vdc, 1.5A

Okidata Microline 320 (optional)
120Vac, 60Hz, 0.9A or
230/240Vac, 50/60Hz, 0.5A

Epson LX-300+II printer (optional)
99-132Vac, 50/60Hz, 0.6A (maximum 1.4A depending on the character type)
198-264Vac, 50/60Hz, 0.3A (maximum 1.4A depending on the character type)

October 15 SYSTEM INITIALIZATION 5-7

CAPINTEC, INC. CAPTUS® 700t

Printer Selection
If a printer was included with the CAPTUS® 700t, the system will already be set to use the
included printer. If a printer is being added or the system did not include a printer at the time
of purchase, then the CAPTUS® 700t must be set to use the printer in order to provide
printed data.

To view and/or change the selected printer, touch the SETUP button located on the Main
screen. Figure 5-3 Setup Screen will appear.

In the Printer section on the Setup Screen, the following options are available:

• NONE........................if no printer is attached to the system: the calibrator will
continue to function properly but without the ability to print.

• USB/HP.....................USB printer port, HP Officejet 6000 series Inkjet (or equivalent)
printer: prints all data on regular paper.

• 232/SLIP ...................RS-232 printer port, Epson TM-295II Slip Printer: prints
measurements on tickets or regular paper and all other data on
regular paper.

• 232/ROLL..................RS-232 printer port, Epson TM-U220D Roll Printer: prints all
data on roll paper.

• 232/OKI-TICKET .......RS-232 printer port, Okidata Microline 320 printer: prints
measurements on tickets or regular paper and all other data on
regular paper.

• 232/OKI-LINE ............RS-232 printer port, Okidata Microline 320 printer: prints
measurements in a single line on regular paper and all other
data on regular paper.

• 232/LX300-TICKET ...RS-232 printer port, Epson LX-300+II printer: prints
measurements on tickets or regular paper and all other data on
regular paper.

• 232/LX300-LINE........RS-232 printer port, Epson LX-300+II printer: prints
measurements in a single line on regular paper and all other
data on regular paper.

To select the desired printer, touch the radio button next to the preferred printer type. Verify
that the desired radio button is selected.

Press the HOME or BACK button to save the changes and return to the Main screen.

5-8 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

Okidata Menu Set-Up
If an Okidata Microline 320 printer was ordered with the CAPTUS® 700t system, the Okidata
printer should be set up to function properly when received.

Should you experience printer difficulties, such as not printing, improper print spacing,
printing the wrong characters, etc., make sure that the printer Setup is exactly as indicated in
Figure 5-7 Listing of the Turbo OKIDATA 320 Printer Setup Data. If any differences are
found, they must be corrected.

The buttons referred to are all located on the front panel of the printer. The Paper Lever is on
the right side of the top surface of the printer near the paper roller.

To print and correct the Setup:

1. Make sure that the form paper is properly loaded in the printer.

2. Press the SEL button (to take printer off-line).
3. Press SHIFT + MENU (SEL) to enter the Setup Menu.
4. Press the PRINT button and the entire Setup listing will print.
5. If there are no differences between what just printed and Figure 5-7 Listing of the

Turbo OKIDATA 320 Printer Setup Data go to step 10.

6. Press the GROUP button. The first line of the first group of lines of the Setup will
print.

7. If all lines in the group just printed match the corresponding lines in Figure 5-7 Listing
of the Turbo OKIDATA 320 Printer Setup Data, repeat step 6 for the next group.

8. If any line in the group just printed does not match the corresponding line in Figure
5-7 Listing of the Turbo OKIDATA 320 Printer Setup Data, it must be corrected. For
example, to change the Baud Rate to 4800:

a. Press the GROUP button until Serial I/F is printed.
b. Press the ITEM button until Baud Rate is printed.
c. Press the SET button until 4800 BPS is printed.
9. If any differences remain, return to step 6 for the next group.

10. Press the SHIFT + MENU (SEL) buttons to return to SEL (on-line).

Note: The Setup printout shown in Figure 5-7 Listing of the Turbo OKIDATA 320 Printer
Setup Data is for example only. Your printer Setup may differ slightly from that shown.
Verify that the settings, which are the same between your Setup and that shown,
concur.

October 15 SYSTEM INITIALIZATION 5-9

CAPINTEC, INC. CAPTUS® 700t

Printer Control Emulation Mode IBM PPR

Font Print Mode Utility
Font DRAFT Mode HSD
Font Pitch 10 CPI
Font Proportional Spacing No
Font Style Normal
Font Size Single

Symbol Sets Character Set Set I
Symbol Sets Language Set American
Symbol Sets Zero Character Slashed
Symbol Sets Code Page USA

Rear Feed Line Spacing 6 LPI
Rear Feed Form Tear-Off Off
Read Feed Skip Over Perforation No
Rear Feed Page Length 11”

Bottom Feed Line Spacing 6 LPI
Bottom Feed Form Tear-Off Off
Bottom Feed Skip Over Perforation No
Bottom Feed Page Length 11”

Top Feed Line Spacing 6 LPI
Top Feed Bottom Margin Valid
Top Feed Page Length 11”
Top Feed Wait Time 1 sec
Top Feed Page Length Control by Actual Page Length

Set-Up Graphics Uni-directional

Set-Up Receive Buffer Size 64K

Set-Up Paper Out Override No

Set-Up Print Registration 0

||||||||||||||||||||||||||||||||||||||||||||| <S-E-T--- |||||| S-H-I-F-T-+>SET |||||||||||||||||||

Set-Up Operator Panel Functions Full Operation
Set-Up Reset Inhibit Yes
Set-Up Print Suppress Effective No
Set-Up Auto LF No
Set-Up Auto CR No
Set-Up SI Select Pitch (10CPI) 17.1 CPI
Set-Up SI Select Pitch (12CPI) 12 CPI
Set-Up Time Out Print Valid
Set-Up Auto Select No
Set-Up Centering Position DEFAULT
Set-Up ESC SI Pitch 17.1 CPI

Parallel I/F I-Prime Buffer Print
Parallel I/F Pin 18 +5v
Parallel I/F Bi-Direction Enable

Serial I/F Parity None
Serial I/F Serial Data 7/8 Bits 8 bits
Serial I/F Protocol X-ON/X-OFF
Serial I/F Diagnostic Test No
Serial I/F Busy Line SSD-
Serial I/F Baud Rate 4800 BPS
Serial I/F DSR Signal Invalid
Serial I/F DTR Signal Ready on Power Up
Serial I/F Busy Time 200 ms

Figure 5-7 Listing of the Turbo OKIDATA 320 Printer Setup Data

5-10 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

Epson LX-300+II Menu Set-Up
If an Epson LX-300+II printer was ordered with the CAPTUS® 700t system, the printer should
be set-up to function properly when received.

Should you experience printer difficulties, such as not printing, improper print spacing,
printing the wrong characters, etc., make sure that the printer Setup is exactly as shown in
Figure 5-9 Epson LX-300+II Printer Current Settings. If any differences are found, they must
be corrected.

Follow the steps below to enter Default Setting mode. If single sheets are used, a new sheet
of paper will need to be loaded each time the printer ejects a printed page.

The buttons referred to are all located on the front panel of the printer.

1. Make sure that the paper is properly loaded and the printer is off.

2. Hold down the TEAR OFF (FONT) button and then turn on the printer.

The printer enters Default Setting mode and prints instructions (as shown in Figure
5-8 Epson LX-300+II Printer Language Selection Instructions) for selecting the
language. The current language is indicated by which control panel lights are on. (For
English, only the Pause light is on.)

3. If a different language is desired, press the LF/FF button until the lights show the
preferred language.

4. Press the TEAR OFF (FONT) button to print a list of all the current settings as shown
in Figure 5-9 Epson LX-300+II Printer Current Settings.

October 15 SYSTEM INITIALIZATION 5-11

CAPINTEC, INC. CAPTUS® 700t

Figure 5-8 Epson LX-300+II Printer Language Selection Instructions

5-12 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 5-9 Epson LX-300+II Printer Current Settings

5. If there are no differences between what just printed and Figure 5-9 Epson LX-300+II
Printer Current Settings, turn off the printer to exit Default Setting mode.
If changes are needed, proceed to step 6.

6. Press the TEAR OFF (FONT) button. The printer prints instructions for changing the
settings as shown in Figure 5-10a Epson LX-300+II Printer Changing Settings
Instructions.

October 15 SYSTEM INITIALIZATION 5-13

CAPINTEC, INC. CAPTUS® 700t

Figure 5-10a Epson LX-300+II Printer Changing Settings Instructions

5-14 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 5-10b Epson LX-300+II Printer Changing Settings Instructions

October 15 SYSTEM INITIALIZATION 5-15

CAPINTEC, INC. CAPTUS® 700t

Figure 5-10c Epson LX-300+II Printer Changing Settings Instructions

7. Press the LF/FF button to scroll through the options for the selected setting until the
desired setting is printed. Then press the TEAR OFF (FONT) button to select the next
setting to be changed.

8. When all changes are complete, turn off the printer to exit Default Setting mode.

The new settings will remain in effect until they are changed again or the printer is reset to
the default settings.

Note: The Setup printout shown in Figure 5-9 Epson LX-300+II Printer Current Settings is
for example only. Your printer Setup may differ slightly from that shown. Verify that
the settings, which are the same between your Setup and that shown, concur.

5-16 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

USB PC DRIVER
This selection is provided for adjusting the USB port settings when connecting the CAPTUS®
700t to a Nuclear Management software system. The USB PC Communications protocol can
either be Legacy (HID (Human Interface Device)) or CDC (Communications Device Class).

By default, the USB PC Driver is set to “Legacy”. The default will work Capintec’s
Communications software. This setting may need to be changed if the CAPTUS® 700t is
connected to some other Nuclear Management software system.

To change the USB PC Driver protocol, from Figure 5-3 Setup Screen, touch the box
containing LEGACY. Figure 5-11 USB PC Driver Select Screen will appear.

Figure 5-11 USB PC Driver Select Screen

Touch the CDC button to select the CDC protocol. Figure 5-3 Setup Screen will appear with
CDC displayed in the USB PC Driver box.

To select Legacy, from Figure 5-3 Setup Screen, touch the box containing CDC and then
touch the LEGACY button on Figure 5-11 USB PC Driver Select Screen.

October 15 SYSTEM INITIALIZATION 5-17

CAPINTEC, INC. CAPTUS® 700t

SCREEN ADJUSTMENTS
The screen’s timeout (Sleep Timeout), the screen’s brightness level when it is in timeout
mode (Sleep Brightness) and the screen’s brightness level when it is in operational mode
(Brightness) can be adjusted.

To make screen adjustments, touch the SETUP button located on the Main screen. Figure
5-12 Setup Screen will appear.

Figure 5-12 Setup Screen

Sleep Timeout
Sleep Timeout will blank the screen after the set time period. The Sleep Timeout can be set
to anywhere between OFF and 60 minutes in one minute increments.

The timeout period can be adjusted using the following methods:
• Touch and hold the Sleep Timeout slider and drag it along the slider bar to the left to
decrease the time it takes for the screen to blank-out. Drag the slider to the right to
increase the time it takes for the screen to blank-out.
• Touch anywhere along the slider bar and the slider will move to that location.

The selected timeout period (in minutes) is displayed to the right of the slider bar.

5-18 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

When done making screen adjustments, press the HOME or BACK button to save the
changes and return to the Main screen.

Sleep Brightness
Sleep Brightness adjusts the brightness (intensity) of the screen when the screen is in Sleep
Timeout mode. The Sleep Brightness level can be set to anywhere between 10 and 100 in
increments of 1.

The level can be adjusted using the following methods:

• Touch and hold the Sleep Brightness slider and drag it along the slider bar to the left
to decrease the brightness level. Drag the slider to the right to increase the brightness
level.

• Touch anywhere along the slider bar and the slider will move to that location.

The selected Sleep Brightness level is displayed to the right of the slider bar.

When done making screen adjustments, press the HOME or BACK button to save the
changes and return to the Main screen.

Brightness
The Screen Brightness (intensity) adjusts the brightness (intensity) of the screen when in
operational mode. The Brightness can be changed to conform to lighting conditions and can
be set to anywhere between 10 and 100 in increments of 1.

The intensity can be adjusted using the following methods:

• Touch and hold the Brightness slider and drag it along the slider bar to the left to
decrease the brightness level. Drag the slider to the right to increase the brightness
level.

• Touch anywhere along the slider bar and the slider will move to that location.

The selected Screen Brightness level is displayed to the right of the slider bar.

When done making screen adjustments, press the HOME or BACK button to save the
changes and return to the Main screen.

October 15 SYSTEM INITIALIZATION 5-19

CAPINTEC, INC. CAPTUS® 700t

Screen Calibration

The touch screen was calibrated at the factory. However, over time, the calibration of the
touch screen may drift slightly. In order to use the touch screen correctly, it needs to be
calibrated properly. Without the right calibration, the touches will be off center and may
inadvertently select the wrong button.

The CAPTUS® 700t has a built-in touch screen calibration utility. To perform a screen
calibration, touch the SCREEN CALIB button.

Figure 5-13 Touchscreen Calibration Screen will appear displaying a target in the upper left
corner. Touch the center of the target. After the target is touched it will disappear and
another target will appear in the lower right corner of the screen. Touch the center of this
target. After this target is touched it will disappear and another target will appear in the center
of the screen. Touch the center of this target. After the target is touched, Figure 5-12 Setup
Screen will re-appear. The touch screen calibration procedure is complete.

Figure 5-13 Touchscreen Calibration Screen

5-20 SYSTEM INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

SPEAKER VOLUME
The volume of the system speaker can be varied. The TEST button located to the right of the
Volume slider bar provides a continuous audible tone at the set volume level so that the
volume of the speaker can be heard while adjusting the slider. Touch the TEST button to
hear the tone. Touch the TEST button again to turn off the tone.

The speaker volume can be adjusted using the following methods:

• Touch and hold the Volume slider and drag it along the slider bar to the left to
decrease the volume level. Drag the slider to the right to increase the volume level.

• Touch anywhere along the slider bar and the slider will move to that location.

When done making volume adjustments press the HOME or BACK button to save the
changes and return to the Main screen.

October 15 SYSTEM INITIALIZATION 5-21

CAPINTEC, INC. CAPTUS® 700t

This page intentionally left blank.

CAPINTEC, INC. CAPTUS® 700t

CHAPTER 6

DETECTOR INITIALIZATION

GENERAL .......................................................................................................6-1
CHANGING DETECTOR.................................................................................6-5
TEST SOURCE DATA ....................................................................................6-6

Serial Number (S/N) Field ..........................................................................6-7
Calibrated On Field ....................................................................................6-8
Calibrated Activity Field.............................................................................6-9
EFFICIENCY DATA........................................................................................6-10
Wipe Full Spectrum Efficiency .................................................................6-11
Nuclide Efficiency .....................................................................................6-13
USER NUCLIDES ..........................................................................................6-24
Adding a User Nuclide ..............................................................................6-24
Editing a User Nuclide ..............................................................................6-29
Deleting a User Nuclide ............................................................................6-30

GENERAL
This section describes initialization and parameter setup for the Probe Detector and optional
Well Counter. This includes inputting the Cs137 Test Source information, setting nuclide
efficiencies and adding User Nuclides.

October 15 Figure 6-1 Main Screen 6-1

DETECTOR INITIALIZATION

CAPINTEC, INC. CAPTUS® 700t

To access the Detector setup, touch the SETUP button located in the lower right portion of
Figure 6-1 Main Screen. Figure 6-2 Setup Screen will appear.

Figure 6-2 Setup Screen

From the Setup screen, touch the ADVANCED DETECTOR button. A numeric keypad will
appear to allowing the user to input a 3-digit password as shown in Figure 6-3 Enter
Password: Screen.

6-2 DETECTOR INITIALIZATION October 15

CAPINTEC, INC. CAPTUS® 700t

Figure 6-3 Enter Password: Screen

Input the password (the last 3 digits of the Readout serial number) by touching the
appropriate numbers on the keypad. Touch the ACCEPT button to accept the password.
Figure 6-4 Advanced Detector Setup Screen will appear.

October 15 DETECTOR INITIALIZATION 6-3

CAPINTEC, INC. CAPTUS® 700t

Figure 6-4 Advanced Detector Setup Screen

The selected Detector will be displayed in the lower left-hand corner of the screen.

To exit Figure 6-4 Advanced Detector Setup Screen,
• touch the BACK button – Figure 6-2 Setup Screen will appear or
• touch the HOME button – Figure 6-1 Main Screen will appear.

Note: The BIO ASSAY, THYROID UPTAKE and THYROID UPTAKE NORMAL VALUES
buttons will only be displayed if a Probe or Drilled Probe Detector is included with the
system. The WIPE TESTS and RBC SURVIVAL NORMAL VALUES buttons will only
be displayed if the Well Detector or Drilled Probe is included with the system.

6-4 DETECTOR INITIALIZATION October 15