The words you are searching are inside this book. To get more targeted content, please make full-text search by clicking here.
Discover the best professional documents and content resources in AnyFlip Document Base.
Search
Published by capintec, 2018-02-19 20:13:15

CAPTUS 600 Manual

CAPTUS 600 Manual

CAPINTEC, INC. CAPTUS® 600

CHAPTER 20

CLEANING AND MAINTENANCE

INTRODUCTION

This chapter provides the information necessary for the user to perform the basic
maintenance of instrument cleaning, fuse replacement, and general preventative
maintenance. There are no internal adjustments or calibration settings that may be done by
the user without voiding the warranty.

REFER ALL SERVICING TO A QUALIFIED SERVICE REPRESENTATIVE!

It is recommended that periodic (every five years) re-calibration of the CAPTUS® 600 be
performed by Capintec’s Authorized Service Center to guarantee the instrument’s high
reliability is maintained. Contact the Capintec’s Authorized Service Center in Pittsburgh for
servicing or re-calibration at 1-800-227-6832.

CLEANING INSTRUCTIONS

CAUTION

DISCONNECT THE POWER BEFORE CLEANING.

TO AVOID ELECTRICAL SHOCK OR DAMAGING OF THE CAPTUS® 600, NEVER GET
WATER OR LIQUIDS INSIDE THE DETECTOR HOUSINGS OR THE READOUT
ENCLOSURE.

DO NOT USE AN AEROSOL DISPENSOR TO SPRAY THE EQUIPMENT WITH ANY
CLEANING SOLUTION OR LIQUID.

TO AVOID DAMAGING THE CASE OR DISPLAY SCREEN, DO NOT USE AROMATIC
HYDROCARBONS, CHLORINATED SOLVENTS, OR METHANOL-BASED CLEANING

SOLUTIONS.

DO NOT USE ANY FORM OF SCRAPER TO CLEAN THE FRONT FACE OF THE
DETECTOR, AS THE NaI CRYSTAL MAY BE DAMAGED.

• FLOOR STAND: Use a soft cloth that is lightly dampened with water, window
cleaner, or a mild detergent. Wipe the surfaces as necessary to remove dust or
smudges.

November 07 CLEANING AND MAINTENANCE 20-1

CAPINTEC, INC. CAPTUS® 600

• DETECTORS: Use a soft cloth that is lightly dampened with water, window cleaner,
or a mild detergent. Wipe the front surface and or the well as necessary to remove
smudges of surface contamination.

Note: If contamination is suspected, dispose of the cleaning cloth in an approved
container for contaminated materials.

• POINTER ASSEMBLY: Use a soft cloth that is dampened with propanol alcohol to
sterilize or clean the black pointer and assembly.

• READOUT: Wipe down with a damp cloth; do not use solvents or aerosol cleaners.

• PRINTER: Wipe down with a damp cloth; do not use solvents or aerosol cleaners.
Wipe the outside surfaces of the printer case to remove the dust, smudges, and
stains. The interior of the printer does not require cleaning.

PREVENTATIVE MAINTENANCE

The Quality Assurance Tests described in CHAPTER 6: ACCEPTANCE & QUALITY
ASSURANCE TESTS, SECTION: QUALITY ASSURANCE TESTS, should be performed
periodically as stated.

Tests must be performed in an environment where the temperature is stable within a range
of +50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing. The unit should be powered-up for at least one-half hour prior to performing any
measurements. No other precautions need to be observed.

CAUTION
If these environmental requirements are not followed,

the instrument may display erroneous readings.

Note: Never put a sharp instrument like a pen into the Well or Collimator. If the aluminum
coating of the NaI Detector is scratched, allowing light and moisture to penetrate, the
detector will be ruined. The rod source is plastic and can be safely placed against the
surface of the detectors when handled carefully.

If the unit fails to pass Auto Calibration Test or any other test, the user should not attempt to
perform any adjustments to the system. In this event, please contact Capintec’s Authorized
Service Center for further assistance.

The Quality Assurance tests should be immediately performed if:
• The equipment has been subjected to extreme physical stress,
• Liquids enter the readout unit, and/or Probe and Well detectors, or
• Any cable shows signed of damage

20-2 CLEANING AND MAINTENANCE November 07

CAPINTEC, INC. CAPTUS® 600

Collimator Block Inspection

Once a year, the Collimator Mounting Block (1) should be inspected for the following:
• excessive play between the Spring-Arm shaft (2) and the Collimator Mounting Block,
• excessive play between the Collimator pin (3) and the Collimator Mounting Block.

If there is excessive play or the knob(s) cannot be tightened, contact Capintec’s Authorized
Service Center for a information on replacement items, instructions or on-site service.

Figure 20-1

WARNING
FAILURE TO FOLLOW THESE INSTRUCTIONS COULD RESULT IN SERIOUS INJURY

OR PRODUCT MALFUNCTION.

SERVICING

The system is covered by a one year limited warranty, under normal conditions of use.

Other than the Readout Unit and Isolation Transformer’s fuses (see Fuse Servicing in this
chapter), there are no user serviceable parts contained in the system.

Every five years, the system should be returned to Capintec’s Authorized Service Center for
a complete verification.

CAPINTEC, Inc.
620 Alpha Drive
Pittsburgh, PA 15238
Phone (412) 963-1988, 1-800-227-6832
Fax (412) 963-0610

November 07 CLEANING AND MAINTENANCE 20-3

CAPINTEC, INC. CAPTUS® 600

DISPOSAL

The following items should be taken into consideration before disposing. These items should
be disposed of in accordance with local and national regulations. Please contact Capintec,
Inc. or an authorized disposal company to decommission your equipment.

4

5

3
54

1
2

1

6

5
1

No. Recycling/Material Code 4
1 External Electrical Cables
2 Lithium Battery Figure 20-2
3 LCD Display Important Information
4 Printed Circuit Boards
Contained on the HPC/CPU printed circuit
5 Lead board inside of the readout unit.
6 Printer LCD with gas discharge backlight
HPC/CPU, A/D Board, Power Supply, LCD
Controller, Dynode Board, Printer
Lead Shielding inside Collimator, around
Well Counter, on Probe Detector Base
Contact printer manufacturer

20-4 CLEANING AND MAINTENANCE November 07

CAPINTEC, INC. CAPTUS® 600

FUSE SERVICING

CAUTION

For continued protection, replace only with
same type and rating of fuse(s)

Power Line Fuses

The power line fuses are located on the front side of the Isolation Transformer Unit directly
below the main power cord connector. The type and rating of these 2 fuses are listed on the
label that is located on the top of the Isolation Transformer Unit.

To change these fuses:

1. Turn off the 2 power switches that are located on the back of the Readout Unit and on
the right side of the Printer.

2. Unplug the power cord that is connected to the front of the Isolation Transformer Unit.

3. Use a small screwdriver to pry open the tab of the fuse drawer that is located directly
below the power cord connector. The drawer will drop down to expose the ends of the
2 fuses. (Refer to Figure 20-3)

4. Remove the blown fuse(s) from the holder and replace it (them) with the fuses
specified on the top label.

5. Rotate the fuse drawer to the upright position and push the drawer back into the fuse
holder.

Figure 20-3

6. Reconnect the power cord to the Isolation Transformer Connector and turn on the 2
power switches (Readout Unit and Printer).

7. Verify the CAPTUS® 600 System is functioning correctly by performing the Daily
Tests.

November 07 CLEANING AND MAINTENANCE 20-5

CAPINTEC, INC. CAPTUS® 600

Readout Unit Fuses

Two fuses are located in the power entry module next to the power cord connector on the
back panel of the Readout Unit. The type and rating of these fuses are listed on the label
located on the back panel of the Readout Unit.

To change these fuses:
1. Turn off the CAPTUS® 600 power switch and unplug the line cord from the power
entry module.

2. Insert the tip of a small flat bladed screwdriver into the notch of the power entry
module just to the left of the switch. Twist the screwdriver to open the fuse cover
door. (Refer to Figure 20-4).

3. To remove the first of the two fuse carriers, insert the tip of the screwdriver behind the
arrow and pull it out. Repeat this process for the second fuse.

4. Remove the blown fuse(s) from the carrier and replace it (them) with the fuses
specified on the back label.

5. Re-insert the fuse carriers with the arrows pointing up (Refer to Figure 20-4).

Figure 20-4
6. Close the fuse cover door and snap it into place.
7. Replace the line cord and turn on the power switch for the CAPTUS® 600.
8. Verify the CAPTUS® 600 System is functioning correctly by performing the Daily

Tests.

Printer Fuse
The printer fuse is not accessible from the outside of the printer case and must be replaced
by a qualified service representative.

20-6 CLEANING AND MAINTENANCE November 07

CAPINTEC, INC. CAPTUS® 600

ACCESSORIES AND REPLACEMENT PARTS

The following accessories and replacement parts are available from Capintec. Call Capintec
at 1-800-227-6832 for answers to your questions or to place an order.

• Printer Paper: Fan fold paper, 1000 sheets per box....................... 9282-0007
• Printer Ribbon: Okidata 184 printer ................................................ 2001-0048
• Printer Cable: Shielded RS232 serial, 6 ½ feet long ...................... 8400-1217
• Printer: Okidata Microline 184 Turbo.............................................. 5430-0017
• Well liners: 5 per package .............................................................. 5120-0176
• Detector cable, coaxial 10 feet long8 .............................................. 8400-2236
• Flat Field Detector: 2M2/2 type ...................................................... 1200-0028
• Drilled Well Detector: 2MW2/2 type................................................ 1200-0029
• Detector Base Assembly: PM-600/2000......................................... 5420-2126
• Well Detector: 1” Shielding ............................................................. 5420-2033
• Well Detector: 2” Shielding ............................................................. 5430-0069
• Rod Source: Cs137 (0.5 µCi) ......................................................... 0975-137R
• Power cable: hospital grade, 15 feet long ...................................... 8125-0024
• Fuse: T ½ Amp. 250 Volts. (Readout Unit) 2 Req. ........................ 2110-0048
• Fuse: T 3.15AH, 250 Volts. (Power Line- 115Volts). 2 Req. .......... 2110-0077
• Fuse: T 2.5AH, 250 Volts. (Power Line- 230Volts). 2 Req. ............ 2110-0079
• CAPTUS® 600 Owners Manual ....................................................... 9250-0072
• CAPTUS® 600 Schematics.............................................................. 5131-5007
• Application Note, Re-Calibration ..................................................... 5412-5051

SHIPPING

If for any reason the CAPTUS® 600 must be returned to Capintec, the shipping carton must
contain the following or equivalent labeling as shown in Figure 20-5.

Label stipulating the maximum environmental conditions for safe storage and shipment.

November 07 Figure 20-5 20-7

CLEANING AND MAINTENANCE



CAPINTEC, INC. CAPTUS® 600

A INDEX

Acceptance Tests, 6-1 frequency, 6-1
Activity Conversion Factors, 5-12 cc, 19-6
Additional Nuclides, 12-7 Cesium-137 standard, 5-7

input limits, 3-12 input limits, 3-11
adjustments, 1-5 CF

Display Contrast, 3-11 input, 5-14
High Voltage, 3-10 measure, 5-14, 5-15
Offset Voltage, 3-10 Chi-Square Test, 6-2, 6-5
Alphanumeric Entries, 3-7 Collimator Block Inspection, 20-3
ARROW Keys, 3-6 Contrast, 3-4
Auto Calibration Conversion Factors, 5-11
printing, 6-8 generic, 12-6
test, 6-5 setting, 5-12
Counting Methods, 3-9, 7-1
B Counting Time Base, 7-1, 7-3
CPM, 5-11
Backspace Key, 3-6 Curies / Becquerel setup, 5-4
Becquerel / Curies setup, 5-4 Cursor, 3-7
Bioassay mode, 13-7
ROI, 19-7
measurements, 17-5
printing report, 17-7 D
Protocol, 17-4
Protocol input limits, 3-12 Daily Tests, 6-1
Report, 17-7 Data Entry Input Limits, 3-11
Staff Data, 2-10, 17-1 Date
Staff Member
entry format, 5-1
adding, 17-2 input, 3-8
deleting, 17-4 Setting system, 5-2
editing, 17-3 Dead Time, 3-9, 7-1, 19-5
blank entries, 3-8 Decimal Point, 3-7
Blood Volume Detector, 2-2
input limits, 3-12 Diagnostics, 2-5, 6-12
test, 2-10 Printout, 6-13
Blood Volume (Cr51), 16-1 Display, 2-5
Blood Volume (I125), 16-6 Display Contrast adjustment, 3-11
BSP Key, 3-6 Disposal, 20-4
Dose Hematocrit, 16-2
C DPM, 5-11

Calibration Source, 5-7 E
data, 2-5
Electromagnetic Interference, 1-5
Calibration Test, 6-2 EMC, 2-6
EMI, 1-5

November 07 INDEX i-1

CAPINTEC, INC. CAPTUS® 600

energy resolution, 6-4 Isolation Transformer, 2-4
ENTER Key, 3-7
Entering Nuclide Name, 5-13 K
Environment Requirements, 2-3
error message Keyboard, 3-6

measurement error, 5-16 L
ETL Listed, 2-6
Lab Tests
F Blood Tests, 2-10
introduction, 16-1
Facility Name, 5-3 Blood Volume (Cr51)
fine gain, 6-1 calculation, 16-3
Full Width at Half Maximum, 6-7, 19-7 Dose Hematocrit, 16-2
FUNCTION Keys, 3-5 measurements, 16-2
Functional Description, 2-7 Patient Hematocrit, 16-2
FWHM, 6-7 result screen, 16-3
Blood Volume (I125)
G calculation, 16-7
hematocrit, 16-6
Graphics Screen, 13-6 measurements, 16-7
result screen, 16-7
H General Procedure, 14-1
Measurement Procedure, 14-4
Hematocrit Patient Demographics, 14-1, 14-3
Dose, 16-2 RBC Survival
Patient, 16-2 calculation, 16-12
Data screen, 16-11
High Voltage, 2-1, 3-10 hematocrit, 16-11
adjustment, 3-10 result screen, 16-12
Schilling Test
HOME Key, 3-4 calculation, 15-3
Hospital Name, 5-3 measurements, 15-1
Result screen, 15-3
I Setup, 14-1
Urine Test, 2-9, 14-3
input CF, 5-14 introduction, 15-1
input limits, 3-11
Labels
Additional Nuclide, 3-12 Caution, 1-2
Bioassay Protocols, 3-12 High Voltage, 1-4
Blood Volume, 3-12 Warning, 1-2
Cs137 Standard, 3-11
Patient Demographics, 3-12 limits
RBC Survival Measurement, 3-12 Additional Nuclide, 3-12
Staff Data, 3-13 Bioassay Protocols, 3-12
Thyroid Uptake Measurements, 3-11 Blood Volume, 3-12
Thyroid Uptake Protocols, 3-11 Cs137 Standard, 3-11
Wipe, 3-12 data entry input, 3-11
Input Out of Range, 3-8 Patient Demographics, 3-12
invalid values, 3-8

i-2 INDEX November 07

CAPINTEC, INC. CAPTUS® 600

RBC Survival Measurement, 3-12 Nuclides
Staff Data, 3-13 Additional, 12-7
Thyroid Uptake Measurements, 3-11
Thyroid Uptake Protocols, 3-11 O
Wipe, 3-12
liners, 1-5 Offset Voltage adjustment, 3-10
Live Time, 3-9, 7-1 One Capsule Method, 2-8
log-on password, 3-1
P
M
password
MCA log-on, 3-1
Background, 19-2 setup, 3-8
function keys, 19-4
results screen, 19-3 Patient
Calibration, 19-2 Demographics, 8-2
cutoff channel, 18-4 entering, 14-3
features, 18-1 input limits, 3-12
Measurement, 19-4 Hematocrit, 16-2
automatic peak searching program, print summary, 8-5
19-7 selecting, 8-4
function keys, 19-6
Measurements, 19-1 Power up, 3-1
Measuring a Source, 19-4 Pre-Dose Measurement, 2-7
Nuclide Activity, 19-8 PRINT Key, 3-5
Region of Interest., 19-7 Printer, 2-4
ROI, 19-7 Probe
Setups, 18-1
adding, 18-6 calibration, 6-7
deleting, 18-6 calibration source, 5-8
printing, 18-6 Detector, 2-6
ROI, 18-5 Screen, 3-3
PROBE Key, 3-2
MDA Test, 6-9 Prompt, 3-6
measure CF, 5-14, 5-15 Protocol, 7-4
measurement error message, 5-16 adding, 7-6
Measurement Errors, 3-10 deleting, 7-11
Minimum Detectable Activity, 6-9 editing, 7-11
mode printing, 7-11
selecting, 8-3
cursor, 13-7
scale, 13-7 Q
Monthly Tests, 6-1
Multi-Capsules, 2-9 Quality Assurance Tests, 6-1
Multichannel Analyzer, 2-2, see MCA Daily tests, 6-1
Multichannel Pulse Height Analyzer, 2-5 Monthly tests, 6-1
Quarterly tests, 6-2
N
Quarterly Tests, 6-2
Nuclear Data, 2-4

November 07 INDEX i-3

CAPINTEC, INC. CAPTUS® 600

R Counting Methods, 7-1
General Instructions, 8-1
RBC Survival Test, 2-10 Measurements
Measurement input limits, 3-12
input limits, 3-11
Readout Unit, 2-4 Measuring Neck and Thigh, 8-7
Real Time, 3-9, 7-1 Nuclides, 2-7
recalibration, 1-6 One Capsule Method, 2-8
Red Blood Cell Survival. See RBC
Administered Dose, 9-6
Survival Test Entering Capsule Data, 9-2
Regulatory Listings, 2-6 General Procedure, 9-1
Requirements Measureing Capsule(s), 9-3
Pre-Dose Meausrement, 9-1
Environment, 2-3, 4-1 Viewing Results, 9-9
Power, 4-2 Patient Demographics, 8-2
ROI cursor, 19-7 Pre-Dose Measurement, 8-9
Print
S Patient Summary, 8-5
Results, 8-11
S Key, 13-7 Protocols, 2-7, 7-4
Safety, 1-1 adding, 7-6
Scale Mode, 13-7 Calculation Method, 7-9
Schilling Test, 2-9, see Lab Tests choosing nuclide, 7-8
screen deleting, 7-11
editing, 7-11
contrast, 3-4 input limits, 3-11
Dimmer, 3-4 Maximum number of, 7-6
Probe, 3-3 Normal Values, 7-10
Well, 3-3 Number of Uptakes, 7-8
Servicing, 20-3 printing, 7-11
setup password, 3-8 repetitions, 7-8
SOFT keys, 3-5 selecting, 8-3
SPACE Key, 3-7 Region of Interest, 7-3
Staff Data input limits, 3-13 Selecting Patient, 8-4
Standard Capsule Method, 2-8 to Measure, 8-6
System Setup, 7-1
Functions, 3-2 Stand, 2-2
Initialization, 5-1 Standard
Power Requirements, 2-3 adding, 11-2
Setup, 4-1 deleting, 11-3
editing, 11-3
T measuring, 11-4
printing, 11-4
Technical Description, 2-1 Standard Capsule Method, 2-8
Technical Specifications, 2-2 Administered Dose, 11-7
Test, 2-5 General Procedure, 11-1
Pre-Dose Measurement, 11-7
of probe, 6-3, 6-10 Selecting Standard, 11-6
of well, 6-3, 6-10 Viewing Results, 11-9
TEST Key, 3-2, 6-2, 6-9 Two Capsules Method, 2-8
Thyroid Uptake Administered Dose, 10-1
Accessing Studies, 8-1
INDEX November 07
i-4

CAPINTEC, INC. CAPTUS® 600

General Procedure, 10-1 Activity for Generic Conversion Factor,
Measuring Reference Capsule(s), 10- 13-7

3 Activity of a Nuclide, 13-7
Pre-Dose Measurement, 10-1 areas, 12-3
Viewing Results, 10-7 Background, 12-5, 13-1
view patient information, 8-5
Time too high, 13-2
input, 3-8 Count Times, 12-4
Setting system, 5-2 Descriptions, 12-2
Two Capsules Method, 2-8 General, 12-5, 13-3
Graphics Screen, 13-6
U input limits, 3-12
Location
Units
Becquerel, 5-4 adding, 12-3
Curie, 5-4 deleting, 12-4
switching, 5-4 editing, 12-4
measuring, 13-3
uptake measurement, 9-8, 10-6, 11-9 energy range, 13-4
User Key select area, 13-3
Receiving, 12-5
Assignment, 12-9 Region of Interest, 12-5
UTIL Key, 3-2, 3-5, 5-4, 5-7, 5-9, 6-6, 6-12 report, 13-5, 13-7
Utilities, 2-5 results, 13-5
interpreting, 13-9
V Results Activity Screen, 13-5
sample report, 13-5
Voltage Sealed Sources, 12-5
High, 1-4, 1-5, 2-1, 3-10 Setup, 12-1
Trigger Levels, 12-4
W Unrestricted Areas, 12-5
Work Areas, 12-5
Warm up, 3-1
Well X

calibration, 6-7 X-Ray/Gamma Ratio, 17-5
calibration source, 5-8
Detector, 2-6 Y
Screen, 3-3
WELL Key, 3-2, 14-2 YES or NO Answers, 3-9
Wipe Test, 2-9, 13-1

November 07 INDEX i-5



NEW PRODUCT WARRANTY

Conditions and Limitations

CAPINTEC warrants our products to be free from defects in material and
workmanship for a period of 12 months after delivery to the original buyer
(unless indicated otherwise). Our obligation under this warranty is limited to
servicing or adjusting products returned to our factory for that purpose
promptly upon any claim under this warranty, provided our examination
discloses to our satisfaction that the affected parts were originally defective
and have not been subjected to misuse, abuse, mishandling or improper
operation. This warranty extends to every part of the product except
batteries, fuses, ink, magnetic storage media, toner, transistors, tubes or any
other normally consumable item. In no event shall we liable for
transportation, installation, adjustment or any other expenses which may
arise in connection with such products or their servicing or adjustment.

This warranty is expressly in lieu of any other express or implied warranties,
and we make no warranty that the products sold herein are merchantable or
are suitable for any particular purpose. The benefits of this warranty shall
extend only to the buyer and to none other. We shall have no liability of any
nature arising out of the use or application of the product in conformity with
this warranty. If the product shall fail to perform in conformity with the
warranty made herein, we shall be liable solely for the repair and
replacement of the product as provided hereinabove, and shall have no
other liability in respect of such failure or performance, or any consequences
therefrom, including, without limitation, business impairment or loss,
personal injury or property damage.

No agent, employee, or representative of ours has any authority to bind us
to any representation or warranty concerning the product sold herein, and
unless a representation or warranty made by an agent, employee or
representative is specifically included herein, it shall not be enforceable by
the buyer. No waiver, alteration or modification of the foregoing conditions
shall be valid, unless made in writing and signed by an executive officer of
our corporation.






Click to View FlipBook Version