CAPINTEC, INC. CAPTUS® 4000e

Figure 17-5

3. From Figure 17-14, select the DICOM Worklist button. Figure 17-15 DICOM Patient
Worklist Screen will appear.

November 16 Figure 17-6 DICOM Patient Worklist Screen 17-15

DICOM

CAPINTEC, INC. CAPTUS® 4000e

4. Query Worklist Provider provides an interface for the user to perform a query of the
Worklist provider on the hospital PACs or RIS system. Select the Query Worklist
Provider button. Figure 17-16 will appear.

Figure 17-7

Start Date/End Date
The Start/End Date queries the Worklist for specific date or a range of dates. The
arrow button is used to expand or collapse from a specific date or a range of dates.
The drop arrow button exposes a monthly calendar to help select the dates.

Warning: Querying a Worklist Provider without a specific data may return a large set
of Worklist entries.

Modality
Modality queries the Worklist Provider for a specific DICOM Modality. NM is Nuclear
Medicine. PT is PET. MR is Magnetic Resonance.

Scheduled Station AE Title
Schedule Station AE Title queries the Worklist Provider for entries scheduled for a
specific AE title. Not all Worklist Providers or departments schedule for specific AE
title. The schedule AE title can be different than the local AE title and remote AE title.

Schedule Performing Physicians Name
Schedule Performing Physicians Name queries the Worklist for the specific
performing physician. The provider should support Wildcard search of “*, ?”. Check
the Worklist providers DICOM conformance statement for details.

17-16 DICOM November 16

CAPINTEC, INC. CAPTUS® 4000e

Accession Number, Patient Id and Patient Name
These provide the user with the means to search for a specific name, ID or Accession
number in the Worklist. DICOM requires that all Worklist Providers to support queries
by Patient Name, Patient ID and Accession number but not all providers have this
functionality. The provider should support “*,?” “*,?”. Check the Worklist providers
DICOM conformance statement for details.

5. Select the Query button. Once the Worklist is queried, a list of patients available is
displayed similar to that shown in Figure 17-17.

Figure 17-8

Import Patient
The patient that is scheduled can be selected by checking the box and highlighting. Select
the Import Patient button.

The imported patient will be displayed in the Thyroid Uptake Patient Directory list (Figure
17-14) with a Patient Status of “Data Partially Entered”. Refer to CHAPTER 7: THYROID
UPTAKE; SECTION: EDITING PATIENT INFORMATION for instructions on editing patient
information. Complete the data entry for the nuclide used.

November 16 DICOM 17-17

CAPINTEC, INC. CAPTUS® 4000e

Figure 17-9

VIEWING THE DICOM REPORT
Once the study is completed, the DICOM report can be viewed.

17-18 Figure 17-10 November 16

DICOM

CAPINTEC, INC. CAPTUS® 4000e

From Figure 17-19, select the Report button. Figure 17-20 will appear.

Figure 17-20

Select the DICOM Report button. The DICOM Report window will appear displaying the
information that will be transferred to the DICOM Server. Refer to Figure 17-21 and Figure

17-22.

November 16 Figure 17-11 DICOM Report Window – Page 1 17-19

DICOM

CAPINTEC, INC. CAPTUS® 4000e
Select the Next Page button to see the rest of the report.

Figure 17-12 DICOM Report Window – Page 2

EXPORTING A STUDY
Once the study is completed it is ready to export to the DICOM server.
From the DICOM Report window, select the Export button. When the transfer is completed
successfully, Figure 17-23 will appear.

Figure 17-13

Select the OK button. The Report has been sent to the DICOM Server and you will be
returned to Figure 17-20.
Select the Exit button to close the Report window. Figure 17-19 will appear.

17-20 DICOM November 16

CAPINTEC, INC. CAPTUS® 4000e

Select the Exit button to exit the Thyroid Uptake screen. The CAPTUS® 4000 Main screen
will be displayed.

VERIFYING STUDY
Check the PAC’s Server to confirm that the report was sent over.

Figure 17-14

November 16 DICOM 17-21

CAPINTEC, INC. CAPTUS® 4000e

This page intentionally left blank.

CAPINTEC, INC. CAPTUS® 4000e

INDEX

A Viewing, 14-5
Absorbed Dose ALERT, 9-1 Auto Calibration, 5-5

Adding Patients, 9-11 About, 5-10
Counting Parameters Frame, 9-5 Auto Range, 13-3
Counting Time, 9-5
Creating an Archive, 9-21 B
Data Acquisistion, 9-20 Battery Replacement
Date of Intake, 9-4
Delete Incident, 9-8 Computer, 16-9
Deleting Patient Information, 9-15 Bioassay, 11-1
Edit Incident, 9-9
Editing Patient Information, 9-14 Adding Staff Members, 11-9
Incident / Site ID, 9-4 Cross Contamination, 11-9
Main Screen, 9-2 Deleting Staff Member, 11-11
Measurements, 9-16 Editing Staff Member Information, 11-10
New Incident, 9-6 Measurements, 11-12
Opening an Archive, 9-22 Setup, 11-2
Patient Reports, 9-23, 9-26 Viewing and Printing Reports, 11-15
Probe Distance, 9-5 BioAssay, 2-4
Repetitions Frame, 9-5 Blood Tests, 2-3
Screening Level, 9-5 Cr51 Blood Volume, 2-3, 10-30
Select Incident, 9-6 I-125 Blood Volume, 2-3, 10-18
Select Isotope, 9-5 RBC Survival, 2-3, 10-43
Setup Incident Parameters, 9-4 button
Accession Number, 17-17 Abort, 5-6, 5-9, 12-7
Accessories, 16-12 Absorbed Dose ALERT, 9-2
License Information Activity, 8-28
DICOM, 17-1 Add, 6-3, 7-5, 8-6, 9-11, 10-5, 10-21,
Activity, 10-6, 10-23, 10-35, 10-46
Add User, 4-8 10-33, 10-45, 11-9, 15-3, 15-8
Not a Staff Member, 4-9 Add Storage Provider, 17-10
Staff Member, 4-8 Add to Group, 8-8
Adding Storage Provider, 17-10 Add User, 4-7, 4-8
Adding Worklist Provider, 17-6 Add Worklist Provider, 17-6
adjustments, 1-3 Add/Edit Comment, 11-15, 11-17
Admin Log On, 4-6 Additional Fields, 15-4, 15-9
Administered to Standard Activity Ratio, Administer Dose, 7-19
10-20, 10-21, 10-32, 10-33 Administer Now, 7-24
Administered Volume, 10-32, 10-33 Archives, 14-1
AE Title, 17-5 Auto Calibration, 5-5
Application Entity, 17-5 Background, 8-14, 9-18
Archive, 14-1 Bioassay, 11-1
Creating (Backup), 14-3 Calibration, 13-14
Opening, 14-4 Change Group Name, 8-8
Printing, 14-5 Change Password, 4-7
Change Setup Values, 9-11, 9-13
Check All, 8-12, 14-6
Check None, 8-12, 14-5
Chi-Square Test, 5-16

November 16 INDEX i-1

CAPINTEC, INC. CAPTUS® 4000e

Clear All ROIs, 13-9 Load Previous, 13-10, 13-11, 13-13
Clear Spectrum, 13-2, 13-17 Location Comment, 8-6, 8-10
Clear This ROI, 13-9 Manual Calibration, 5-8, 13-17
Comment, 8-17, 9-19 MCA, 13-2
Constancy Test, 5-11 MDA Test, 5-20
Convert, 10-23, 10-34 Measure Efficiency, 6-2, 6-5, 11-3
Count, 8-13, 8-16, 9-14, 9-15, 9-17, 11- New, 8-6
New Incident, 9-4, 9-6
12, 12-5, 13-2, 13-17 Next, 6-6, 8-17, 11-4, 15-6, 15-7
Count Patient Dose, 7-18, 7-19 Next Page, 17-20
Count Selected Patients, 9-17, 10-2, Next Patient, 9-20
Open Archive, 9-22, 14-4
10-10, 10-26, 10-38, 10-48, 15-12 Open Previous Results, 12-8, 12-9
Counting Setup, 13-7 Print, 5-22, 8-17, 9-19, 9-24, 10-53, 10-
Cr51 Blood Vol Test, 10-30
Create Archive, 9-21, 14-3 54, 11-7, 12-7, 13-14, 14-7, 15-15
Cs137 Efficiency, 5-5, 5-12 Print / View History, 5-10, 5-15, 5-19,
Custom Protocol, 15-2
Delete, 5-26, 6-5, 7-9, 7-16, 8-11, 9-16, 5-22
Print All, 11-17
10-9, 10-25, 10-37, 10-48, 11-11, 15- Print All Listed, 14-8
7, 15-11 Print Latest Results, 11-15, 11-17
Delete All, 7-10, 8-11, 9-16, 10-9, 10- Print Protocols, 7-9, 7-10
26, 10-37, 10-48, 11-11, 15-7, 15-11 Print Selected, 8-23, 9-24
Delete Incident, 9-7, 9-8 Print Selection, 14-8
Delete Last Test, 5-25 Print Staff Member, 11-17
Delete User, 4-12 Print/View, 8-17, 8-18
DICOM, 17-2 Print/View History, 5-24
DICOM Report, 5-3, 8-3, 17-19 Protocols, 7-5, 15-3
DICOM Worklist, 17-15 Quality Assurance, 5-2
Display Graph, 7-27, 10-53 Query, 17-17
Echo, 17-7, 17-9, 17-11, 17-13 Query Worklist Provider, 17-16
Edit, 6-4, 7-8, 8-9, 9-14, 10-8, 10-24, Red Blood Cell Survival, 10-43
10-36, 10-47, 11-10, 15-10 Re-Measure, 3-11, 6-9, 8-15, 9-20, 10-
Edit Incident, 9-4, 9-10
Enter Comment, 7-27, 10-17, 10-53 14, 15-14
Enter Efficiency, 6-2, 6-10, 11-7 Remove from Group, 8-8
Enter ID, 13-7 Report, 4-7, 4-11, 7-25, 9-23, 10-15,
Excel File, 15-8
Excel Report, 15-15 10-52, 11-15
Exit, 12-8, 14-2 Restore Original Values, 7-18
Exit Archive, 9-23 Return to Wipe, 8-5
Export, 9-27, 17-20 Revalidate User, 4-13
Export Selected, 8-23, 9-28 ROI From Isotope, 13-9
I 125 Blood Test, 10-18 ROIs and Detector, 15-5
Import Patient, 17-17 Save, 13-18, 17-5, 17-7, 17-9, 17-11,
Incident Setup, 9-4, 9-6
Insert File, 3-12, 10-17, 10-54 17-13
Invalidate User, 4-12, 4-13 Save Results, 12-8, 13-10, 13-12, 13-
Isotope Library, 8-3, 8-4, 12-4
Isotopes, 6-1 13
Lab Tests, 10-2 Save to File, 3-12, 5-23, 5-25, 7-26, 9-
Load From File, 12-3, 12-5
25, 10-17, 10-52, 10-54, 11-16, 12-4,
14-7
Schilling Test, 10-3
Search, 8-19, 8-20

i-2 INDEX November 16

CAPINTEC, INC. CAPTUS® 4000e

Select All, 8-22 Captus Screens, 3-7
Select Incident, 9-4, 9-7 Add Comments Screen, 3-12
Set / View Test Parameters, 12-2 Measurement Screen, 3-11
Set Gain, 13-15 Typical Test Screen, 3-9
Settings, 17-4
Setup, 5-3, 5-9, 7-16, 10-4, 10-19, 10- CAPTUS Screens
Typical Input Screen, 3-10
31, 10-44, 11-2
Setup Defaults, 8-4 Cautions and Notes, 1-3
Setup Groups, 8-7 Change Password, 4-10
Setup QA Tests, 5-3 Channel Data Display, 13-2
Sign Off, 4-6 Chi-Square Test, 5-15
Start Procedure, 5-6, 5-9, 5-13, 6-7, 6- Cleaning, 16-2
Cleaning and Maintenance, 16-1
9, 10-11, 10-27, 10-39, 10-49, 11-5, Cleaning Instructions, 16-2
11-13, 15-13 Collimator Block Inspection, 16-6
Stop, 13-2 Computer Battery Replacement, 16-9
Stop and Save, 3-11, 8-15, 9-20, 10-15, Constancy Test, 5-11
15-14 Counting Methods
Summary Reports, 8-19, 8-24
System Parameters, 5-5, 5-23 Decay Correct Administered Dose, 7-3
Test By Location, 8-26 Measure Same Reference Dose, 7-4
Tests By Wipe Type, 8-27 Counting Parameters, 12-4
Thyroid Uptake, 7-2 Counting Parameters Frame, 9-5
Time Activity Analysis, 12-1 Counting Time, 9-5
Unit Conversion, 10-22, 10-34 Cr51 Blood Volume Test, 10-30
Unselect All, 8-22 Adding Patients, 10-33
Update, 4-12 Editing Patient Information, 10-35
View, 15-7 Measurements, 10-38
View Selection, 14-7 Patient Reports, 10-42
Wipe Test, 8-1 Results, 10-42
Zoom, 13-3 Setup Counting Parameters, 10-31
Cross Contamination, 11-9
C Cross Contamination ratio, 11-7
Calibration Cross Contamination ratios, 11-8
Cursor Readout, 13-2
Auto, 5-5 Custom Protocols, 15-1
Manual, 13-17 Adding, 15-3
Calibration Activity, 11-4 Adding Patients, 15-8
Calibration Date, 10-46, 11-4 Creating a Workbook, 15-16
Calibration Time, 10-46, 11-4 Data Acquisition, 15-13
Cap-DICOM, 17-1 Deleting, 15-7
Configuring, 17-1 Deleting Patient Information, 15-10
Export Study, 17-20 Editing Patient Information, 15-10
Import Patient, 17-17 Excel Workbook, 15-16
Import Patients, 17-14 Making Measurements, 15-12
Query Worklist, 17-14 Patient Data, 15-8
Settings, 17-4 Protocols, 15-3
Storage Provider, 17-10 Reports, 15-15
Verify Study, 17-21 Selecting Workbook, 15-7
Viewing Report, 17-18 Setting Up the Workbook, 15-18
Worklist Provider, 17-6 Starting, 15-2
Worklist/DICOM connection, 17-3 Viewing, 15-7

November 16 INDEX i-3

CAPINTEC, INC. CAPTUS® 4000e

Viewing Patients from Workbook, 15-24 emergency screening, 9-2
EMI, 1-4, 1-5, 4-2, 4-3
D End Channel, 13-9
Data Environment Requirements, 2-8, 4-3

displaying by date, 4-13 Operational, 2-8, 4-3
Data Acquisition, 9-20, 10-14 Storage, 2-9, 4-3
Data mode, 12-7 Exiting, 4-14
Date of Intake, 9-4 Export Raw Spectrum, 13-13
Decay Corr Only, 12-7 Export to Excel, 13-12
Default Values, 10-44 Exporting a Study, 17-20
Delete Incident, 9-8
Delete User, 4-12 F
Deleting Storage Provider, 17-13 Functional & Technical Description, 2-1
Deleting Worklist Provider, 17-9 Functional description, 2-2
Detectors Fuse Servicing, 16-9

Description, 2-4 Computer, 16-11
Probe, 2-8 Power Line Fuses, 16-9
Well, 2-8 Printer, 16-11
DICOM, 17-1
Configuring, 17-1 G
Export Study, 17-20 Gain, 13-15
Import Patient, 17-17 General Information, 10-7, 10-23, 10-35,
Import Patients, 17-14
Query Worklist, 17-14 10-46
Settings, 17-4 General Operating Instructions, 3-1
Storage Provider, 17-10 General Setup, 4-1
Verify Study, 17-21 Graph, 12-7
Viewing Report, 17-18
Worklist Provider, 17-6 H
Worklist/DICOM connection, 17-3 High Voltage, 13-15
Disinfecting, 16-2 Horizontal Zoom, 13-3, 13-4
Disinfecting Instructions, 16-4 Host Name, 17-7, 17-10
Display Patient, 17-5
Displaying Data by Date, 4-13 I
Displaying Selected User, 4-12 I-125 Blood Volume Test, 10-18
Disposal, 16-5
Dosage Information, 10-6, 10-23, 10-34, Adding Patients, 10-21
10-46 Deleting Patient Information, 10-25
Dose Administration, 7-4 Editing Patient Information, 10-24
Measurements, 10-26
E Patient Reports, 10-30
Edit Incident, 9-9 Results, 10-30
Editing Storage Provider, 17-12 Setup Counting Parameters, 10-19
Editing Worklist Provider, 17-8 ID, 12-4
Efficiency Import Patient, 17-17
Incident
Entering, 6-9 Delete, 9-8
Isotope, 8-30 Edit, 9-9
Measuring, 6-5 New, 9-6
Electromagnetic Interference, 1-4, 1-5, 4- Select, 9-6
2, 4-3 Incident / Site ID, 9-4
Inspection

i-4 INDEX November 16

CAPINTEC, INC. CAPTUS® 4000e

Collimator Block, 16-6 MCA, 13-1
Spring-Arm, 16-7 Acquisition, 13-2
Instructions for Sample, 15-7 Changing Detectors, 13-3
Integral, 13-9 Count Data, 13-2
Intended Use, 2-1 Counting Setup, 13-7
Invalidate User, 4-13 Customizing Settings, 13-14
Investigation Level, 11-3, 11-8 Export, 13-12
Isotope Efficiency, 8-30 Export Raw Spectrum, 13-13
Isotope Library, 6-2 Gain, 13-15
Printing, 6-2 High Voltage, 13-15
Viewing, 6-2 Manual Calibration, 13-17
Isotope Used in Bioassay, 11-3 Printing Data, 13-14
Isotopes, 6-1 Recall Counting Setup, 13-14
Adding, 6-3 Recall Settings, 13-13
Deleting, 6-5 Recall Spectrum, 13-11
Editing, 6-4 Regions of Interest, 13-8
Measuring Efficiency, 6-5 Restore Count Data, 13-10
Restore Current Settings, 13-14
L Restore Settings, 13-10
Lab Tests, 2-3, 10-1 ROI, 13-8
Save Count Data, 13-10
Cr51 Blood Volume Test, 2-3, 10-30 Save Counting Setup, 13-13
I-125 Blood Volume Test, 2-3, 10-18 Save Settings, 13-10, 13-13
RBC Survival Test, 10-43 Save Spectrum, 13-10
Schilling Test, 10-3 Spectrum ID, 13-7
Urine, 2-3 Spectrum Scale, 13-3
Labels Threshold, 13-16
Warning and Caution, 1-2 Zero Offset, 13-16
Latest Results, 11-1, 11-15 Zooming, 13-3
License Information, 17-1
Live Counting Time, 12-4 MDA Test, 5-19
Live Time, 13-7 Measurement Name, 15-6
Load Previous Counting Setup, 13-14 Measurement Protocols, 7-3
Load Previous Settings, 13-13
Load Previous Spectrum, 13-11 Adding, 7-5
Log On, 4-5 Counting Methods, 7-3
Admin, 4-6 Deleting, 7-9
User, 4-7 Dose Administration, 7-4
Log On Report, 4-11 Editing, 7-7
Delete User, 4-12 Printing Report, 7-10
Displaying Data by Date, 4-13 Measurement Type, 15-6
Displaying Selected User, 4-12 Measurements
Exiting, 4-14 Quality Assurance Tests, 4-16, 5-1
Invalidate User, 4-13 Setup, 4-17
Printing, 4-13 Measuring Efficiency, 6-5
Revalidate User, 4-13 Modality, 17-7, 17-16
Lot Number, 10-6, 10-23, 10-35, 10-46 Mutli-Channel Analyzer, 2-5

M N
Maintenance, 16-4 New Incident, 9-6
Manual Calibration, 13-17 Normal Range, 10-20, 10-32
Normal Range for 50% Survival, 10-44

November 16 INDEX i-5

CAPINTEC, INC. CAPTUS® 4000e

Normal Range of Excretion, 10-5 Q
Number of Measurements, 15-4 Quality Assurance, 5-1

O Printing History, 5-24
Operator Schedule, 5-2
Setting Up Parameters, 5-2
Profile, 2-2 Tests, 4-16
Training, 2-2 Viewing History, 5-24
Query Worklist, 17-14
P
Parameters R
Raw and Decay Corr, 12-7
Chi-Square Test, 5-4 Raw Only, 12-7
Constancy Test, 5-4 RBC Survival Test, 2-3, 10-43
MDA Test, 5-4
Setting Up QA, 5-2 Adding Patients, 10-45
Password Deleting Patient Information, 10-47
change, 4-10 Editing Patient Information, 10-46
Patient Directory, 9-15, 10-2, 10-9, 10- Measurements, 10-48
10, 10-15, 10-18, 10-25, 10-26, 10-30, Patient Reports, 10-52
10-37, 10-38, 10-42, 10-43, 10-47, 10- Results, 10-51
48, 10-51, 10-52, 15-2, 15-10, 15-12 Setup Counting Parameters, 10-44
Patient HCT, 10-23, 10-35 Real Time, 13-7
Patient Information, 7-10, 10-22, 10-34, recalibration, 1-5
10-45 Regulatory Listings, 2-10
Adding, 7-10 Repetitions, 10-44
Deleting, 7-15 Repetitions Frame, 9-5
Editing, 7-13 Replacement Parts, 16-12
Viewing, 7-13 Report
Patient Status, 10-2, 10-10, 10-14, 10-15, Log On, 4-11
10-18, 10-26, 10-30, 10-38, 10-42, 10- Printing, 4-13
43, 10-48, 10-51 Requirements
Peak Counts, 13-7 Environment, 2-8, 4-3
Pointing Device Cursor, 3-2 Power, 2-9, 4-4
Port, 17-7, 17-11 Restore Current Settings, 13-14
Pre- Dose Measurement, 10-5 Revalidate User, 4-13
Pre-Defined Channel Ranges, 13-4 ROI Information, 12-4
Pre-Dose Measurement, 10-5 ROI Number, 13-8
Preventive Maintenance, 16-4
Printing QA History, 5-24 S
Printing Report, 4-13 Safety, 1-1
Probe Distance, 9-5 Safety Tips, 1-6
Profile Sample HCT, 10-35
Operator, 2-2 Sample Preparation, 10-6, 10-23, 10-35
Protocol Name, 15-4 Sample Volume, 10-20, 10-21
Protocols Save Counting Setup, 13-13
Adding, 7-5 Save Settings, 13-13
Deleting, 7-9 Schedule Performing Physicians Name,
Editing, 7-7
Printing Report, 7-10 17-16
Scheduled Station AE Title, 17-16
Schilling Test, 2-3, 10-3

i-6 INDEX November 16

CAPINTEC, INC. CAPTUS® 4000e

Adding Patients, 10-5 Software, 2-7
Data Acquisistion, 10-14 Stand, 2-6
Deleting Patient Information, 10-8 Well Detector, 2-8
Editing Patient Information, 10-7 System Parameters, 4-15
Measurements, 10-10 Viewing, 5-23
Patient Reports, 10-15 System Power Requirements, 2-9, 4-4
Results, 10-15
Setup Counting Parameters, 10-4 T
Screening Level, 9-5 Table, 12-7
Select Incident, 9-6 Technical Description, 2-4
Select Isotope, 9-5, 12-4
Service List, 17-7, 17-11 Detector, 2-4
Servicing, 16-6 Multi-Channel Analyzer, 2-5
Setup Technical Specifications, 2-6
General, 4-1 Environment Requirements, 2-8, 4-3
Measurement, 4-17 Regulatory Listings, 2-10
System Parameters, 4-15 System Hardware, 2-6
Turn On Procedure, 4-4 System Power Requirements, 2-9, 4-4
Shipping, 16-13 Text Color, 17-5
Shortcut Keys, 3-13 Threshold, 13-16
All Screens, 3-13 Thyroid Uptake, 2-2, 7-1
Input Screens, 3-13 Administering Dose, 7-24
Sign In, 4-6 Changing Setup, 7-16
Sign Off, 4-6 Counting Patients, 7-20
spillover, 11-9 Making Measurements, 7-18
Spring-Arm Inspection, 16-7 Patient Reports, 7-25
Staff Directory, 11-16 Quick Setup Guide, 7-3
Staff ID, 11-9, 11-10 Results, 7-24
Staff Member Directory, 11-1, 11-11, 11- Viewing Setup, 7-16
12, 11-15 Time Activity Analysis, 12-1
Standard Dilution, 10-4, 10-5, 10-6 Measurements, 12-5
Start Channel, 13-9 Results, Saving, 12-8
Start Date/End Date, 17-16 Setting up Parameters, 12-2
Startup Tips window, 4-14 Viewing Saved Results, 12-9
Storage Provider, 17-10 Time Between Measurements, 12-4
Adding, 17-10 Total Urine Volume, 10-6
Editing, 17-12 Training
Study Operator, 2-2
Exporting, 17-20
Verify, 17-21 U
Study Date, 17-5 Urine Lab Tests
Summary Reports, 8-24
Symbol Definitions, 1-1 Schilling Test, 2-3
System Hardware Urine Sample Aliquot, 10-4, 10-5, 10-6
Computer, 2-6, 2-7 User
Detector Inputs, 2-7
Isolation Transformer, 2-6 Delete, 4-12
MCA Interface PCB, 2-7 Displaying Selected, 4-12
Printer, 2-6 Invalidate, 4-13
Probe Detector, 2-8 Revalidate, 4-13
User Log On, 4-7
User Selectable Range, 13-5

November 16 INDEX i-7

CAPINTEC, INC. CAPTUS® 4000e

V Measurements, 8-12
Verify Study, 17-21 Printing Results, 8-23
Vertical Zoom, 13-3 Printing Summary Reports, 8-29
Vertictal Zoom, 13-3 Results, 8-17
Viewing QA History, 5-24 Search, 8-19
Selecting Isotopes, 8-4
W Selecting Results, 8-22
warm-up period, 2-9 Setting up Defaults, 8-3
Weight, 10-34 Sorting Data, 8-21
Windows Operating Conventions Summary Reports, 8-24
Viewing and Printing Results, 8-18
Buttons, 3-6 Viewing Data Selectively, 8-21
Windows Operating Conventions, 3-1 Workbook
Creating a Report, 15-23
Command Button, 3-3 Entering Formulas, 15-22
Common Controls, 3-1 Patient Selection, 15-20
Frame, 3-3 Protocol Selection, 15-19
List Controls, 3-6 Worklist
Menu Bars, 3-6 Query, 17-14
Scroll Bars, 3-5 Worklist Provider, 17-6
Spin Control, 3-6 Adding, 17-6
Text Box, 3-4 Deleting, 17-9
Windows Controls, 3-2 Editing, 17-8
Windows Operating Conventions Worklist Storage
Keyboard Control, 3-7 Deleting, 17-13
Wipe Tests, 2-3, 8-1 Worklist Threshold, 17-7
Adding Locations to Preset Groups, 8-6 Worklist/DICOM connection, 17-3
Adding Locations to User Defined
Z
Groups, 8-7 Zero Offset, 13-16
Deleting Locations, 8-10 Zoom
Editing Location Information, 8-8
Exiting Search, 8-23 Horizontal, 13-3, 13-4
Exiting Summary Reports, 8-30 Vertical, 13-3
Exporting Results, 8-22
Isotope Efficiency, 8-30

i-8 INDEX November 16

CAPINTEC, INC. CAPTUS® 4000e

NEW PRODUCT WARRANTY

Conditions and Limitations

CAPINTEC warrants our products to be free from defects in material and
workmanship for a period of 24 months after delivery to the original buyer
(unless indicated otherwise). Our obligation under this warranty is limited
to servicing or adjusting products returned to our factory for that purpose
promptly upon any claim under this warranty, provided our examination
discloses to our satisfaction that the affected parts were originally
defective and have not been subjected to misuse, abuse, mishandling or
improper operation. This warranty extends to every part of the product
except batteries, fuses, ink, magnetic storage media, toner, transistors,
tubes or any other normally consumable item. In no event shall we liable
for transportation, installation, adjustment or any other expenses which
may arise in connection with such products or their servicing or
adjustment.

This warranty is expressly in lieu of any other express or implied
warranties, and we make no warranty that the products sold herein are
merchantable or are suitable for any particular purpose. The benefits of
this warranty shall extend only to the buyer and to none other. We shall
have no liability of any nature arising out of the use or application of the
product in conformity with this warranty. If the product shall fail to
perform in conformity with the warranty made herein, we shall be liable
solely for the repair and replacement of the product as provided
hereinabove, and shall have no other liability in respect of such failure or
performance, or any consequences therefrom, including, without
limitation, business impairment or loss, personal injury or property
damage.

No agent, employee, or representative of ours has any authority to bind
us to any representation or warranty concerning the product sold herein,
and unless a representation or warranty made by an agent, employee or
representative is specifically included herein, it shall not be enforceable
by the buyer. No waiver, alteration or modification of the foregoing
conditions shall be valid, unless made in writing and signed by an
executive officer of our corporation.

CAPINTEC, INC. CAPTUS® 4000e

This page intentionally left blank.