CRC®
Remote Display Unit
Manual Stock No. 9250-0113 Rev. D – April 2012
OWNER’S MANUAL
Capintec, Inc.
Sales and Marketing
and Customer Support
6 Arrow Road
Ramsey, NJ 07446
Phone (800) ASK-4CRC
Fax (201) 825-1336
Capintec, Inc.
Only Authorized Service Center
620 Alpha Drive
Pittsburgh, PA 15238
Phone (800) 227-6832
Fax (412) 963-0610
Copyright© 2005 Capintec, Inc.
®CRC is a registered trademark of Capintec, Inc.
ALL RIGHTS RESERVED
CAPINTEC, INC CRC® Remote Display
TABLE OF CONTENTS
SAFETY ................................................................................................................................1-1
SYMBOL DEFINITIONS ....................................................................................................1-1
LABEL DEFINITION ..........................................................................................................1-2
DC In (9V 1.5A)..............................................................................................................1-2
Contrast .........................................................................................................................1-2
Off/On ............................................................................................................................1-2
Data ...............................................................................................................................1-2
ISP .................................................................................................................................1-2
CAUTIONS AND NOTES ..................................................................................................1-3
GENERAL SAFETY TIPS..................................................................................................1-3
FUNCTIONAL & TECHNICAL DESCRIPTION ....................................................................2-1
FUNCTIONAL DESRIPTION.............................................................................................2-1
TECHNICAL DESCRIPTION.............................................................................................2-1
Environment Requirements............................................................................................2-1
Operational ................................................................................................................2-1
Storage ......................................................................................................................2-1
Power Requirements......................................................................................................2-2
AC Power Supply ......................................................................................................2-2
Remote Display Power Requirements.......................................................................2-2
Dimensions ....................................................................................................................2-2
SYSTEM SETUP...................................................................................................................3-1
GENERAL..........................................................................................................................3-1
RECEIVING CONDITION EXAMINATION ........................................................................3-1
UNPACKING AND INSTALLATION ..................................................................................3-1
ASSEMBLY .......................................................................................................................3-2
TURN ON PROCEDURES ................................................................................................3-2
OPERATION .........................................................................................................................4-1
GENERAL..........................................................................................................................4-1
SETUP ...............................................................................................................................4-1
KEYBOARD .......................................................................................................................4-1
Layout ............................................................................................................................4-1
Key Usage .....................................................................................................................4-2
MODE Key.................................................................................................................4-2
SELECT Key .............................................................................................................4-2
ACCEPT Key.............................................................................................................4-2
START Key................................................................................................................4-2
MEASURING ACTIVITY....................................................................................................4-2
MEASURING BACKGROUND ..........................................................................................4-5
ZERO TEST.......................................................................................................................4-8
VOLTAGE TEST..............................................................................................................4-10
July 07 TABLE OF CONTENTS TOC-1
CAPINTEC, INC CRC® Remote Display
CLEANING AND MAINTENANCE .......................................................................................5-1
GENERAL ......................................................................................................................... 5-1
CLEANING INSTRUCTIONS ............................................................................................ 5-1
PREVENTATIVE MAINTENANCE .................................................................................... 5-1
SERVICING....................................................................................................................... 5-2
DISPOSAL ........................................................................................................................ 5-2
TROUBLESHOOTING ...................................................................................................... 5-3
Nothing on appears on the display ................................................................................5-3
ACCESSORIES AND REPLACEMENT PARTS ............................................................... 5-3
INDEX
WARRANTY
TOC-2 TABLE OF CONTENTS July 07
CAPINTEC, INC CRC® Remote Display
CHAPTER 1
SAFETY
These warnings and instructions for use form an integral part of the CRC® Remote Display
and must therefore be kept available for consultation at all times. Precise compliance with
the instructions is an essential condition for normal use, correct application and thus safety of
the user.
SYMBOL DEFINITIONS
Attention: consult owner’s manual for further information
DC Voltage
“ON” (power)
“OFF” (power)
Date of manufacture
0413 CE Mark
Waste in Electrical and Electronic Equipment (WEEE) –
This symbol indicates that the waste of electrical and
electronic equipment must not be disposed as unsorted
municipal waste and must be collected separately.
Environmentally Friendly Use Period (EFUP) – 20 years
from the date of manufacture – Toxic or hazardous
substances or elements contained in the unit will not leak
or mutate under normal operating conditions resulting in
any environmental pollution, bodily injury or damage to
assets.
July 07 SAFETY 1-1
CAPINTEC, INC CRC® Remote Display
LABEL DEFINITION
Located on one side of the Remote Display is a label (Figure 1-1).
Figure 1-1
DC In (9V 1.5A)
This receptacle is where the 9 Volt power supply (100~240VAC, 50~60Hz, 1.2A) is
plugged in.
Contrast
These controls adjust the contrast of the LCD Displays. The “U” contrast adjustment
controls the 16×1 Upper display. The “L” contrast adjustment controls the 20×2 Lower
display.
Off/On
This toggle switch applies input power to the unit.
Data
SPI Bus Interface port, which is attached to the CRC®-Ultra Chamber Unit.
ISP
In-System-Programming port, which is attached to an RS-232 port on a computer.
1-2 SAFETY July 07
CAPINTEC, INC CRC® Remote Display
CAUTIONS AND NOTES
CAUTION
There are no internal adjustments in the Remote Display, which may be done by the user
within the conditions of the warranty. Refer all servicing to qualified personnel.
GENERAL SAFETY TIPS
• Unplug the equipment before cleaning it. Use only a damp cloth; do not use solvents
or aerosol cleaners.
• To protect the equipment from overheating, do not use the equipment directly in front
of a radiator or heat register.
• Do not use the equipment near water, or spill liquids of any kind into the equipment.
• Be sure that your power source matches the rating listed on the unit.
• To avoid damaging the power cord, do not place anything on it or place it where it will
be walked on. If the cord becomes damaged, replace it immediately.
• Do not try to service this equipment yourself. Refer servicing to qualified service
personnel.
July 07 SAFETY 1-3
CAPINTEC, INC. CRC® Remote Display
This page intentionally left blank.
CAPINTEC, INC CRC® Remote Display
CHAPTER 2
FUNCTIONAL & TECHNICAL
DESCRIPTION
FUNCTIONAL DESRIPTION
The Remote Display Unit is controlled by the Dose Calibrator. Its function is to remotely
display information that is present on the CRC® Dose Calibrator display. The CRC® Remote
Display Unit communicates via a RS-485 interface and can be located as much as 50 feet
from the Chamber.
TECHNICAL DESCRIPTION
Environment Requirements
Operational
The instrument should be located where the level of the background radiation is as
low and as constant as possible.
The instrument should be located where the temperature is stable within a range of
+50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing to warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of
27 – 31 inches of mercury.
Storage
The instrument should be stored where the temperature is stable and the range is
from +39°F to +115°F (+4°C to +43°C) and the maximum relative humidity is 90%
non-condensing to warrant maximum reliability.
The instrument should be stored where the barometric pressure is within a range of
15 – 33 inches of mercury.
CAUTION
If these environmental requirements are not followed, the instrument may display erroneous
readings.
July 07 FUNCTIONAL & TECHNICAL DESCRIPTION 2-1
CAPINTEC, INC CRC® Remote Display
Power Requirements
CAUTION
If the input voltage to the following items is not within the stated limits, the unit may not
function correctly or may be damaged.
AC Power Supply
100~240VAC, 50-60Hz, 1.2A
Remote Display Power Requirements
+9.0 VDC, 1.5 Amps
Dimensions
Height............ 15.6cm ..... (6.2in)
Width ............. 19.7cm ..... (7.8in)
Depth............. 5.6cm ....... (2.2in)
Weight ........... 0.7kg ........ (1.6lb)
2-2 FUNCTIONAL & TECHNICAL DESCRIPTION July 07
CAPINTEC, INC CRC® Remote Display
CHAPTER 3
SYSTEM SETUP
GENERAL
Initial installation and checkout procedures are described in this section.
RECEIVING CONDITION EXAMINATION
Be sure to verify that the shipping carton is received in good condition, i.e., no damage
should be visible and the box should be dry and clean.
Should the instrument be received in a damaged condition, save the shipping container and
the packing material and request an immediate inspection by the carrier.
Capintec, Inc. is not responsible for the damage which occurs during shipment but will make
every effort to help obtain restitution from the carrier.
UNPACKING AND INSTALLATION
The instrument is packed and shipped as a complete unit. All the accessories are contained
in the carton.
• Remove all outer packing material and tapes. The shipping and packing material should
be saved for future use.
• The following equipment should be found upon unpacking:
Remote Display Unit
Power Supply
RS-485 Connection Cable
July 07 SYSTEM SETUP 3-1
CAPINTEC, INC CRC® Remote Display
ASSEMBLY
Verify that the power is off on the CRC® Dose Calibrator.
Attach the Remote Display Cable (15 pin male) to the Remote connector on the Chamber
(labeled “REMOTE”) and the other end (15 pin female) to Remote Display Unit (labeled
“DATA”).
Attach the Power Cable to the 9VDC receptacle located on the Remote Display Unit.
TURN ON PROCEDURES
1. Verify that the power is off on the CRC® Dose Calibrator.
2. Be sure the power switch is off. (Push “POWER” toggle switch located on the unit to
the position marked “O”.)
3. Verify that the power supply output jack is in the 9VDC input jack.
4. Plug in the power supply cord into a grounded three-wire outlet of the specified
voltage rating.
5. Turn on the Remote Display Unit using the Power switch by pushing it to the “I”
position. The two displays on the unit should appear as shown in Figure 3-1.
Note: The screen will display the revision level of the installed software.
CAPINTEC_
CRC
Remote Display R1.1
Figure 3-1
6. Turn on the power to the CRC® Dose Calibrator.
3-2 SYSTEM SETUP July 07
CAPINTEC, INC CRC® Remote Display
CHAPTER 4
OPERATION
GENERAL
The Remote Display Unit is controlled by the CRC® Dose Calibrator. Its function is to
remotely display information that is present on the CRC® Dose Calibrator display. The CRC®
Remote Display Unit communicates via a RS-485 interface and can be located as much as
50 feet from the Chamber.
SETUP
The setup of the Remote Display Unit is as follows:
1. Verify that the Remote Display Unit has been setup as indicated in CHAPTER 3:
SYSTEM SETUP; SECTION: ASSEMBLY.
2. Verify that the unit is powered on, as indicated in CHAPTER 3: SYSTEM SETUP;
SECTION: TURN ON PROCEDURES.
KEYBOARD
Layout
The CRC® Remote Display keyboard is shown below.
Figure 4-1 Keyboard
April 12 OPERATION 4-1
CAPINTEC, INC CRC® Remote Display
Key Usage
General usage of keys is briefly described. Specific key usage will be given in the
appropriate sections.
Note: When the CRC® Dose Calibrator Readout is in a Chamber Menu, the keys on the
Remote Display are inactive.
MODE Key
When this key is pressed the unit displays the Mode Menu. One of the following will
be displayed on the 16×1 Upper LCD:
Read Background
Zero Test
Voltage Test
Read Activity
SELECT Key
Pressing this key allows the user to cycle through the 4 mode selections.
ACCEPT Key
From the Mode Menu, pressing this key will select the displayed mode.
From one of the 3 test modes, pressing this key will accept the displayed reading.
START Key
Pressing this key starts the selected test. The key has no function in the Mode Menu.
MEASURING ACTIVITY
The Remote Display will default to the Activity mode on power up and after accepting the
results from any test. Refer to Figure 4-2 Activity Display Screens.
23.56mCi_
Tc99m 12:26
Activity R
Figure 4-2 Activity Display Screens
If the Remote Display is in the Mode Menu, press the SELECT key to cycle thru the menu
selections until “READ ACTIVITY?” appears on the Upper LCD. Refer to Figure 4-3.
4-2 OPERATION April 12
CAPINTEC, INC CRC® Remote Display
READ ACTIVITY?_
SELECT to change
ACCEPT to begin_
Figure 4-3 Activity Mode Screens
Press the ACCEPT key to select READ ACTIVITY Mode. The following screens will appear
as shown in Figure 4-4 Activity Start Screens:
_
Read Activity
Press START key_
Figure 4-4 Activity Start Screens
Press the START key to begin Activity Measurement. Figure 4-5 Activity Measurement
(Tc99m) Screens will appear:
The Upper LCD shows the activity of the selected isotope being measured.
23.56mCi_
Tc99m 12:26
Activity R
Figure 4-5 Activity Measurement (Tc99m) Screens
April 12 OPERATION 4-3
CAPINTEC, INC CRC® Remote Display
The first line of the Lower LCD shows the nuclide presently selected and the current time.
The second line shows that the Remote Display is in the Activity mode and will also show
what chamber type is being read (“R” is for a 12 atmosphere chamber, “P” is for a 5
atmosphere (PET) chamber).
While performing an Activity measurement, the isotope being measured can be changed by
pressing the SELECT key.
Figure 4-6 Isotope Selection Mode Screens will appear. The Lower LCD will display the next
nuclide in the list (of the 8 nuclides in the Readout Unit’s memory).
_
Change to In111
ACCEPT to read_
Figure 4-6 Isotope Selection Mode Screens
Further presses of the SELECT key will cycle through the remaining isotopes in the list.
Pressing the ACCEPT key will select the displayed isotope. The example shown below is for
Ga67.
15.16mCi_
Ga67 12:28
Activity R
Figure 4-7 Activity Measurement (Ga67) Screens
4-4 OPERATION April 12
CAPINTEC, INC CRC® Remote Display
MEASURING BACKGROUND
Background Measurement Mode is selected via the Mode Menu. To display the Mode Menu
press the MODE key. Press the SELECT key until Figure 4-8 Background Mode Screens is
displayed.
READ BACKGROUND?
SELECT to change
ACCEPT to begin_
Figure 4-8 Background Mode Screens
Press the ACCEPT key to select BACKGROUND Mode. Figure 4-9 Background Start
Screens will appear.
_
Background
Press START key_
Figure 4-9 Background Start Screens
Press the START key to begin Background Measurement. Figure 4-10 Background
Measurement Screens will appear.
April 12 OPERATION 4-5
CAPINTEC, INC CRC® Remote Display
Please Wait_
Background 12:26
Figure 4-10 Background Measurement Screens
When a measurement is available and the background reading is less then 27µCi (1MBq),
Figure 4-11 Background Measurement (OK) Screens will appear showing the measurement
value and “OK”.
1.44µCi OK
Background 12:26
ACCEPT to save_
Figure 4-11 Background Measurement (OK) Screens
Press the ACCEPT key to save the Background measurement being displayed.
When the background reading is between 10µCi (0.4MBq) and 530µCi (20MBq), then the
measurement value and “HIGH” .will be displayed as shown in Figure 4-12 Background
Measurement (HIGH) Screens.
4-6 OPERATION April 12
CAPINTEC, INC CRC® Remote Display
30.44µCi HIGH
Background 12:26
ACCEPT to end_
Figure 4-12 Background Measurement (HIGH) Screens
Pressing the ACCEPT Key when “HIGH” is displayed, will end the measurement and not
save the background value.
When the background reading is greater than 530µCi (20MBq), then “TOO HIGH” .will be
displayed as shown in Figure 4-13 Background Measurement (TOO HIGH) Screens.
TOO HIGH_
Background 12:26
ACCEPT to end_
Figure 4-13 Background Measurement (TOO HIGH) Screens
Pressing the ACCEPT key when this message is displayed, will end the measurement and
not save the background value.
April 12 OPERATION 4-7
CAPINTEC, INC CRC® Remote Display
ZERO TEST
Zero Test Mode is selected via the Mode Menu. To display the Mode Menu press the MODE
key. Press the SELECT key until Figure 4-14 Zero Test Mode Screens is displayed.
ZERO TEST?_
SELECT to change
ACCEPT to begin_
Figure 4-14 Zero Test Mode Screens
Press the ACCEPT key to select ZERO TEST Mode. Figure 4-15 Zero Test Mode Source
Verification/Start Screens will appear.
Remove sources_
Zero
Press START key_
Figure 4-15 Zero Test Mode Source Verification/Start Screens
Verify all sources have been removed from the vicinity of the Chamber.
Press the START key to begin the Zero Test measurement. Figure 4-16 Zero Test Mode
Measurement Screens will appear.
4-8 OPERATION April 12
CAPINTEC, INC CRC® Remote Display
Please Wait_
Zero 12:36
Figure 4-16 Zero Test Mode Measurement Screens
When a measurement is available, Figure 4-17 Zero Test Measurement (OK) Screens will
appear showing the measurement value and “OK”.
-0.05 mV OK
Zero 12:37
Press ACCEPT to save
Figure 4-17 Zero Test Measurement (OK) Screens
Press the ACCEPT key to save the Zero measurement being displayed.
If the measured value has drifted since the last measurement (±0.3mV), Figure 4-18 Zero
Test Measurement (DRIFT) Screens will appear showing the measurement value and the
“DRIFT” message. Check to make sure that no sources are in the area. If any sources are
found, remove them and repeat the measurement.
0.31 mV DRIFT
Zero 12:38
Press ACCEPT to end_
Figure 4-18 Zero Test Measurement (DRIFT) Screens
April 12 OPERATION 4-9
CAPINTEC, INC CRC® Remote Display
Pressing the ACCEPT key when this message is displayed will end the measurement.
If the measured value is out of range, Figure 4-19 Zero Test Measurement (OUT OF
RANGE) Screens will appear. Check to make sure that no sources are in the area. If any
sources are found, remove them and repeat the measurement. If no sources were found,
contact Capintec’s only Authorized Service Center in Pittsburgh at 1-800-227-6832.
Out of Range_
Zero 12:38
Press ACCEPT to end_
Figure 4-19 Zero Test Measurement (OUT OF RANGE) Screens
Pressing the ACCEPT key when this message is displayed will end the measurement.
VOLTAGE TEST
Voltage Test Mode is selected via the Mode Menu. To display the Mode Menu press the
MODE key. Press the SELECT key until Figure 4-20 Voltage Test Mode Screens is
displayed.
VOLTAGE TEST?_
SELECT to change
ACCEPT to begin_
Figure 4-20 Voltage Test Mode Screens
Press the ACCEPT key to select VOLTAGE TEST Mode. Figure 4-21 Voltage Test Start
Screens will appear.
4 - 10 OPERATION April 12
CAPINTEC, INC CRC® Remote Display
_
Chamber Voltage
Press START key_
Figure 4-21 Voltage Test Start Screens
Press the START key to begin the Voltage Test measurement. Figure 4-22 Voltage Test
Measurement Screens will appear.
Please Wait_
Chamber Volts 12:36
Figure 4-22 Voltage Test Measurement Screens
When a measurement is available, Figure 4-23 Voltage Test Measurement (OK) Screens
will appear showing the Chamber Voltage and “OK” .
151.4V OK
Chamber Volts 12:36
Press ACCEPT to save
Figure 4-23 Voltage Test Measurement (OK) Screens
Press the ACCEPT key to accept the Voltage Test measurement and end the test.
April 12 OPERATION 4 - 11
CAPINTEC, INC CRC® Remote Display
The measurement is compared with the value input at the factory. If the results are out of
range, the message “FAIL” appears as shown in Figure 4-24 Voltage Test Measurement
(FAIL) Screens. If this occurs, contact Capintec’s only Authorized Service Center at 1-800-
227-6832.
140.2V FAIL
Chamber Volts 12:36
Press ACCEPT to end_
Figure 4-24 Voltage Test Measurement (FAIL) Screens
Press the ACCEPT key to accept the Votlage Test measurement and end the test.
4 - 12 OPERATION April 12
CAPINTEC, INC CRC® Remote Display
CHAPTER 5
CLEANING AND MAINTENANCE
GENERAL
This chapter provides the information necessary for the user to perform the basic
maintenance of instrument cleaning and general preventative maintenance. There are no
internal adjustments or calibration settings that may be done by the user within the conditions
of the warranty.
REFER ALL SERVICING TO A QUALIFIED SERVICE REPRESENTATIVE!
Contact the Capintec’s only Authorized Service Center in Pittsburgh for servicing at 1-800-
227-6832.
CLEANING INSTRUCTIONS
WARNING !
TO AVOID ELECTRICAL SHOCK OR DAMAGING OF THE REMOTE DISPLAY, NEVER
GET WATER OR LIQUIDS INSIDE THE ENCLOSURE.
DO NOT USE AN AEROSOL DISPENSER TO SPRAY THE EQUIPMENT WITH ANY
CLEANING SOLUTION OR LIQUID.
TO AVOID DAMAGING THE CASE OR DISPLAY SCREEN, DO NOT USE AROMATIC
HYDROCARBONS, CHLORINATED SOLVENTS, OR METHANOL-BASED CLEANING
SOLUTIONS.
If the Remote Display Unit requires cleaning, wipe down with a damp cloth; do not use
solvents or aerosol cleaners.
PREVENTATIVE MAINTENANCE
Other than maintaining the appearance of the outer case no other preventive maintenance is
required.
April 12 CLEANING AND MAINTENANCE 5-1
CAPINTEC, INC CRC® Remote Display
SERVICING
The system is covered by a one year limited warranty, under normal conditions of use.
There are no user serviceable parts contained in the system.
Should the Remote Display need servicing, it should be returned to Capintec’s only
Authorized Service Center.
CAPINTEC, Inc.
620 Alpha Drive
Pittsburgh, PA 15238
Phone (412) 963-1988, 1-800-227-6832
Fax (412) 963-0610
DISPOSAL
The following items should be taken into consideration before disposing. These items should
be disposed of in accordance with local and national regulations. Please contact Capintec,
Inc. or an authorized disposal company to decommission your equipment.
1
2
No. Recycling/Material Code Important Information
1 External Electrical Cable Control Board, Keyboard, LCD’s
2 Printed Circuit Boards
5-2 CLEANING AND MAINTENANCE April 12
CAPINTEC, INC CRC® Remote Display
TROUBLESHOOTING
Some problems may be very easy to diagnose and correct in the field with little or no
equipment. If a problem should occur, check here before you call for service. You may be
able to save a considerable amount of time and money.
Nothing on appears on the display
Make sure the CRC® Dose Calibrator is plugged into a live outlet and is turned on.
Make sure the Remote Display Unit is plugged into a live outlet and is turned on.
Verify that the Connecting Cable is properly attached and secured.
ACCESSORIES AND REPLACEMENT PARTS
The following accessories and replacement parts are available from Capintec. Contact
Capintec at 1-800-227-6832 for answers to your questions or to place an order.
Power Supply – IN: 100~240VAC, 50-60Hz, 1.2A, OUT: 9VDC, 5A........... 9100-0223
Connection Cable ........................................................................................ 8400-2441
Remote Display Unit.................................................................................... 5130-2211
Additional copies of Owner’s Manual .......................................................... 9250-0113
Note: Circuit diagrams, component parts lists, descriptions and calibration instructions are
available to appropriately qualified personnel.
April 12 CLEANING AND MAINTENANCE 5-3
CAPINTEC, INC. CRC® Remote Display
This page intentionally left blank.
CAPINTEC, INC CRC Remote Display
A INDEX
ACCEPT key, 4-2
Activity START key, 4-2
Measurement, 4-2 L
Assembly, 3-2 Label Definition, 1-2
B M
Background maintenance, 5-1
Measurement
HIGH, 4-6
Measurement, 4-5 Activity, 4-2
TOO HIGH, 4-7 Background, 4-5
Bias Test Voltage Test, 4-10
FAIL, 4-12 Zero Test, 4-8
MODE key, 4-2
C
Cleaning, 1-3, 5-1 R
Receiving Condition Examination, 3-1
D Requirements
disposal, 5-2
Environment, 2-1
I
Installation, 3-1, 4-1 S
safety, 1-1
K SELECT key, 4-2
key servicing, 1-3, 5-2
START key, 4-2
ACCEPT, 4-2
MODE, 4-2 U
SELECT, 4-2 Unpacking, 3-1
START, 4-2
Keyboard V
ACCEPT key, 4-2 Voltage Test
Key Usage, 4-2
Layout, 4-1 Measurement, 4-10
MODE key, 4-2 OK, 4-11
SELECT key, 4-2
Z
Zero Test
Measurement, 4-8
April 12 INDEX i-1
CAPINTEC, INC. CRC® Remote Display
This page intentionally left blank.
CAPINTEC, INC. CRC® Remote Display
NEW PRODUCT WARRANTY
Conditions and Limitations
CAPINTEC warrants our products to be free from defects in material and
workmanship for a period of 12 months after delivery to the original buyer
(unless indicated otherwise). Our obligation under this warranty is limited
to servicing or adjusting products returned to our factory for that purpose
promptly upon any claim under this warranty, provided our examination
discloses to our satisfaction that the affected parts were originally
defective and have not been subjected to misuse, abuse, mishandling or
improper operation. This warranty extends to every part of the product
except batteries, fuses, ink, magnetic storage media, toner, transistors,
tubes or any other normally consumable item. In no event shall we liable
for transportation, installation, adjustment or any other expenses which
may arise in connection with such products or their servicing or
adjustment.
This warranty is expressly in lieu of any other express or implied
warranties, and we make no warranty that the products sold herein are
merchantable or are suitable for any particular purpose. The benefits of
this warranty shall extend only to the buyer and to none other. We shall
have no liability of any nature arising out of the use or application of the
product in conformity with this warranty. If the product shall fail to
perform in conformity with the warranty made herein, we shall be liable
solely for the repair and replacement of the product as provided
hereinabove, and shall have no other liability in respect of such failure or
performance, or any consequences therefrom, including, without
limitation, business impairment or loss, personal injury or property
damage.
No agent, employee, or representative of ours has any authority to bind
us to any representation or warranty concerning the product sold herein,
and unless a representation or warranty made by an agent, employee or
representative is specifically included herein, it shall not be enforceable
by the buyer. No waiver, alteration or modification of the foregoing
conditions shall be valid, unless made in writing and signed by an
executive officer of our corporation.
CAPINTEC, INC. CRC® Remote Display
This page intentionally left blank.