The words you are searching are inside this book. To get more targeted content, please make full-text search by clicking here.

Water treatment guideline for hemodialysis therapy_ บรรยrายสมาคมพยาบาลโรคไต-combined

Discover the best professional documents and content resources in AnyFlip Document Base.
Search
Published by supaporn charoenkai, 2024-04-10 11:54:41

Water treatment guideline for hemodialysis therapy_ บรรยrายสมาคมพยาบาลโรคไต-combined

Water treatment guideline for hemodialysis therapy_ บรรยrายสมาคมพยาบาลโรคไต-combined

Schematic flow diagram of on-line ultrapure fluid preparation and monitoring technique Ultrafilter = logarithmic reduction for bacteria > 7 logarithmic reduction for endotoxin > 4 Endotoxin only -R2A,TGEA -Membrane Filtration -ET -R2A,TGEA,TSA -MF, spread plate -ET -Action level Option


ISO 23500 :2019


2.2.2.4 กรณีมีการนํา endotoxin-retentive filter ที่มีการรับรองคุณภาพโดย บริษัทผู้ผลิตมาใช้กรองนํ้ ายาไตเทียมให้มีคุณภาพบริสุทธิ์ สูง (ultrapure dialysis fluid) อาจยกเว้นการส่งตรวจหาปริมาณแบคทีเรียและ endotoxin ในนํ้ ายาไตเทียม จากเครื่องไตเทียมนั้นได้ โดยต้องปฏิบัติตามคําแนะนําและเงื่อนไขการรับรอง คุณภาพของบริษัทผู้ผลิตในการใช้ endotoxin-retentive filter ดังกล่าว ทั ้ งนี ้ ยังคง ต้องส่งนํ้าบริสุทธิ์ตรวจเป็ นประจําตามข้อ 2.2.1 ผลที่ได้ต้องอยู่ในเกณฑ์มาตรฐาน ของนํ้าบริสุทธิ์สําหรับฟอกเลือดด้วยเครื่องไตเทียม (แบคทีเรียในนํ้าบริสุทธิ์น้อย กว่า 100 CFU/mL endotoxin ในนํ้าบริสุทธิ์น้อยกว่า 0.25 EU/mL ดู คําอธิบาย ผนวก ก ข้อ 2) แนวทางปฏิบัติสมาคมโรคไต 2564 การตรวจสอบค ุณภาพนํ้ายาไตเทียม


HD station HD station Pre-treatment Particle filter Softener Carbon 5µ filter Raw water RO Pump RO Chlorine feed? Dialyzer reuse 1 2 3 HD station In case of applying endotoxin filters on dialysis fluid pathway Bacteria <100 CFU/ml Endotoxin <0.25 EU/ml *If use TSA -*Action level


2.2.2.5 กรณีมีการทํา on-line hemodiafiltration หรือ on-line hemofiltration และมี การนํา endotoxin-retentive filter ที่มีการรับรองคุณภาพโดยบริษัทผู้ผลิตมาใช้ กรองนํ ้ ายาไตเทียมบริสุทธิ์สูงเพื่อผลิตเป็ นสารละลายทดแทน (substitution fluid) โดยปฏิบัติตามคําแนะนําและเงื่อนไขการรับรองคุณภาพของบริษัทผู้ผลิตในการใช้ endotoxin-retentive filter ดังกล่าว ไม่จําเป็ นต้องส่งตัวอยางสารละลายทดแทน่ ตรวจหาปริมาณแบคทีเรียและ endotoxin ทั ้ งนี ้ ยังคงต้องส่งนํ้าบริสุทธิ์ตรวจเป็ น ประจําตามข้อ 2.2.1และผลที่ได้ต้องอย่ในเกณฑ์มาตรฐานนํ้าบริส ู ุทธิ์สําหรับฟอก เลือด โดยมีปริมาณแบคทีเรียน้อยกว่า 100 CFU/mL และ endotoxin น้อยกว่า 0.25 EU/mL (ดู คําอธิบาย ผนวก ก ข้อ 2) แนวทางปฏิบัติสมาคมโรคไต 2564 การตรวจสอบค ุณภาพนํ้ายาไตเทียม


Schematic flow diagram of on-line ultrapure fluid preparation. Ultrafilter = logarithmic reduction for bacteria > 7 logarithmic reduction for endotoxin > 4 Sample points in case of applying endotoxin filters on dialysis fluid pathway Option -R2A,TGEA,TSA -MF, spread plate -ET -Action level


HD station HD station Pre-treatment Particle filter Softener Carbon 5µ filter Raw water RO Pump RO Dialyzer reuse 1 2 3 HD station Online HDF Endotoxin & Bacteria culture q 1 month In case of doing online HDF and validated ETRF application Concentrate mixing Bacteria <100 CFU/ml Endotoxin <0.25 EU/ml *If use TSA -*Action level


2.2.2.5 ..................................... ทางเลือกกรณีที่หน่วยต้องการตรวจสอบคุณภาพและไม่มีปัญหาในการเกบ ็ ตัวอยางส่ ่งตรวจ ในการติดตั ้ งเครื่องไตเทียมครั้งแรกเพื่อทํา on-line hemodiafiltrationควรตรวจคุณภาพนํ ้ ายาไตเทียมบริสุทธิ์ สูงก่อนผานเข้ ่า endotoxin-retentive filter ตัวสุดท้ายที่ใช้ผลิตสารละลายทดแทนโดยการตรวจหา ปริมาณแบคทีเรียและ endotoxin การลงเพาะเชื ้อต้องใช้วิธี membrane filtration ตามที่ระบุไว้ในข้อ 2.4 ผลที่ได้ต้องอยูในเกณฑ์ของนํ ่ ้ ายาไตเทียมบริสุทธิ ์ สูง และ/ หรืออาจส่งตัวอยางสารละลายทดแทนตรวจหาปริมาณ ่endotoxin แนวทางปฏิบัติสมาคมโรคไต 2564 การตรวจสอบค ุณภาพนํ้ายาไตเทียม


Schematic flow diagram of on-line ultrapure fluid preparation. Ultrafilter = logarithmic reduction for bacteria > 7 logarithmic reduction for endotoxin > 4 Sample points in case of applying endotoxin filters on dialysis fluid pathway (Alternative measures for on-site validation) Endotoxin only -R2A,TGEA -Membrane Filtration -ET -R2A,TGEA,TSA -MF, spread plate -ET -Action level


จุดเก็บตัวอย่าง เดือนที่ 1 เดือนที่ 2 เดือนที่ 3 เดือนที่ 4 Bacteria c/s Endotoxin Bacteria c/s Endotoxin Bacteria c/s Endotoxin Bacteria c/s Endotoxin Dialysis water 1.จุดสุดท้ายของแต่ละวงท่อ 1.1 วงท่อ standard HD √ √ √ √ √ √ 1.2 วงท่อ online HDF √ √ √ √ √ √ √ √ 2.อ่างล้างตัวกรอง (หมุนเวียน) √ √ √ √ √ √ 3.จุดผสมนํ้ ายาไตเทียมเข้มข้น √ √ √ √ √ √ กรณีมีการทําonline HDF √ √ √ √ √ √ √ √ Dialysis fluid 1.วงท่อ standard HD (หมุนเวียน) √ √ √ √ √ √ ถ้ามี endotoxin filter + + + + + + 2.วงท่อ online HDF(หมุนเวียน) √ √ √ √ √ √ √ √ 2.1 เครื่อง online HDF + + + + + + + + 2.2 เครื่อง standard HD √ √ √ √ √ √ √ √ ถ้ามี endotoxin filter + + + + + + + + ตารางแสดงความถีของการส่งตรวจคุณภาพนํ่ าเป็ นประจํา ้


ETRF Validation : Performance & Conditions • Properties : LRV for bacteria & endotoxin • Lifetime : time of replacement/number to be used • Leakage ? : vision & leak test (pressure test) • Disinfection : type of disinfectants • Quality of inlet water • Storage location : temperature, humidity, sunlight, nearby chemical storage


Five major sites of contamination B Canaud. Blood Purif 2000;18:200–213 Endotoxin retentive filters


แนวคิดของการอบฆ่ าเชื ้อคือต้องทําให้บ่อยเพียงพอที่จะ ควบคุมไม่ให้มีปริมาณแบคทีเรียและendotoxin สูงกวา ่ action level ได้ตลอดเวลา ไม่ใช่ทําเพียงแค่ แกปัญหาเมื่อตร ้ วจพบ ปริมาณเชื ้ อจุลชีพและ endotoxin มากกวาเกณฑ์ ่ แนวทางปฏิบัติสมาคมโรคไต 2564 การอบฆ่าเชื้อในระบบจ่ายนํ้าบริส ุทธิ์


6.3.2 ต้องมีการอบฆ่าเชื ้ อในระบบจ่ายนํ ้ าบริสุทธิ ์ และถังเกบนํ ็ ้ าบริสุทธิ ์ อยาง่ น้อยทุก 3 เดือน และต้องพิจารณาทําก่อนกาหนด หากพบจํานวนแบคทีเรี ํ ยหรือ ปริมาณ endotoxin ในระบบจ่ายนํ ้ าบริสุทธิ ์ มากกวา ่100 CFU/mL หรือ 0.25 EU/mL ตามลําดับ ในการอบฆ่าเชื ้ อแต่ละครั ้ งควรพิจารณาอบฆ่าเชื ้ อในชุด RO (แผนกรอง ่RO ด้านขาออกที่ให้ผลผลิตนํ้ าบริสุทธิ ์ ) และในสายส่งนํ ้ าบริสุทธิ ์เข้า เครื่องไตเทียมพร้อมกนไป ั แนวทางปฏิบัติสมาคมโรคไต 2564 การอบฆ่าเชื้อในระบบจ่ายนํ้าบริส ุทธิ์


6.3.3ถ้ามีการทํา on-line hemodiafiltrationควรเพิ่ มความถี่ในวงรอบการอบฆ่าเชื ้ อ เพื่อคงคุณภาพนํ้ าบริสุทธิ ์ให้อยูในเกณฑ์มาตรฐานได้ตลอดเวลา ่ ทั ้ งนี ้ ขึ้ นอยูก่บผลั การตรวจคุณภาพนํ ้ าบริสุทธิ ์ ของหน่วย ถ้าผลการตรวจไม่ได้ตามเกณฑ์ควร พิจารณาหยุดการทํา on-line hemodiafiltration(ดูคําอธิบายในผนวก ก ข้อ 6) 6.3.4กรณีเตรียมนํ้ ายาไตเทียมเข้มข้นใช้เองในหน่วย ต้องอบฆ่าเชื ้ อในระบบเตรียม นํ ้ ายาไตเทียมเข้มข้นชนิดไบคาร์บอเนตทุกสัปดาห์กรณีมีระบบจ่ายนํ ้ ายาไตเทียม ส่วนกลาง (central dialysis fluid delivery system, CDDS) ต้องอบฆ่าเชื ้ อในระบบ ทุกวัน แนวทางปฏิบัติสมาคมโรคไต 2564 การอบฆ่าเชื้อในระบบจ่ายนํ้าบริส ุทธิ์


6.3.6 หน ่วยไตเทียมอาจต้องกาหนดความถี่ในการอบฆ ํ ่ าเชื ้ อบ ่ อย กวาที่ก ่ าหนดในแนวทางปฏิบัตินี ํ ้เพื่อให้ได้ประสิทธิผลในการป้องกนั และควบคุมไม่ ให้ปริมาณแบคทีเรียและ endotoxin สูงกวา ่ action level ในตลอดช่วงเวลาที่อยูในวงรอบการอบฆ ่ ่ าเชื ้ อ แนวทางปฏิบัติสมาคมโรคไต 2564 การอบฆ่าเชื้อในระบบจ่ายนํ้าบริส ุทธิ์


นํ ้ ายาฆ่าเชื ้ อ ความเข้มข้น อุณหภูมิ ระยะเวลา ความคงค้าง Formaldehyde 4% 25 ̊ c 24 hr < 3 ppm Peracetic acid (PAA) ตามบริษัทผู้ผลิต กาหนดํ (~ 0.02-0.5%) 25 ̊c ตามบริษัทผู้ผลิต กาหนด ํ (0.5-11 hr) < 1 ppm Ozone* 0.2-0.5 mg/L 10 min < 0.1 ppm Sodium hypochlorite 0.5-1% 30 min Hot water* >80 ̊ c 30 min ตารางที่4 แสดงตัวอยางวิธีการต ่ ่างๆในการอบฆ่าเชื้อ ความเข้มข้นของนํ้ ายากาจัดเชื ํ้อและระยะเวลาที่ใช้ *วัดที่ปลาย loop


Outline • Standard Quality of dialysis water & dialysis fluid • Monitoring & maintenance • Actions • Portable water-purification system for HD • In-center dialysis concentrate preparation


แนวทางปฏิบัติ สมาคมโรคไต 2564 Actions


Process bacteria culture/ endotoxin level <action level (<50CFU/mL / <0.125 EU/mL) bacteria culture /endotoxin level >action level แต่ยังไม่สูงกว่าเกณฑ์ (50-100CFU/mL / 0.125 -0.25 EU/mL) bacteria culture /endotoxin level สูงกว่าเกณฑ์ (>100CFU/mL / >0.25EU/mL) ผ้ป่ วยมีอาการแสดงู ทางคลินิกที่สัมพันธ์ กับสิ่งปนเปื้ อนใน ระบบนํ้าบริสุทธิ์ 1. การรายงานผู้รับผิดชอบและการทบทวนวิเคราะห์หาสาเหตุ 1.1 ดําเนินการทํา HD / monitor ตามปกติ √ √ 1.2 รายงานหัวหน้าพยาบาล √ √ √ 1.3 รายงานแพทย์หัวหน้าศูนย์ไตเทียม √ √ √ 1.4 พิจารณาควรหยุดการให้บริการฟอกเลือดหรือไม่ √ * √ ** 1.5 ประเมินข้อมูลย้อนหลังของผลการตรวจ bacteria culture และ ผล endotoxin ของนํ้ า RO และ dialysis fluidย้อนหลังอยางน้อย ่3 เดือน √ √ √ √*** 1.6 ทําการทวนสอบวิธีการเก็บนํ้ าบริสุทธิ์ส่งตรวจ หาก สงสัยการปนเปื้อนเชื้อจากวิธีการเก็บ ให้รีบทําการส่ง ตรวจซํ้ า (ก่อนวงรอบปกติ) และเก็บตัวอยางจากจุด่ อื่นเพิ่ มเติมจากจุดที่เคยทําเป็ นประจํา เช่น ต้นทาง ระบบจ่ายนํ้ าบริสุทธิ์ ถังเก็บนํ้ าบริสุทธิ์ ผลผลิตของ ชุด RO dialysis fluidจากเครื่องฟอกเลือดเครื่องอื่น เป็ นต้นเพื่อเปรียบเทียบผลและวิเคราะห์หาตําแหน่งที่ เกิดการปนเปื้อน √ √ 1.7 ทําการทวนสอบวิธีการอบฆ่าเชื้อระบบนํ้ าคุณสมบัติ ของนํ้ ายา (cleaning/ disinfection) ที่ใช้ และระยะเวลา อบฆ่าเชื้อ √ √ √ 2. การอบฆ่าเชื้อ


Trouble Shooting


Summary: user must do about water treatment system for HD • Monitoring fluid quality • Raw water chlorine (& conductivity) testing -> at least weekly • Post-carbon water chlorine test -> daily, before start HD operation • Post-softener water hardness test -> daily, at the end of unit operation • RO water conductivity (or resistivity) -> daily • RO percent rejection -> daily • RO water & dialysis fluid - bacteria test -> monthly • RO water & dialysis fluid - endotoxin test -> every three months - in case of doing online HDF -> every month • Chemical analysis of RO water (AAMI) -> yearly • Monitoring and maintenance of equipments • Record parameter of equipment status per flow sheet checklist -> daily • Replacement of equipment -> as recommended or as need • Disinfection water treatment system -> at least every 3 months • Actions and revalidation -> rapidly - depend on severity of problems


Outline • Standard Quality of dialysis water & dialysis fluid • Monitoring & maintenance • Actions • Portable water-purification system for HD • In-center dialysis concentrate preparation


เครื่องผลิตนํ้ าบริสุทธิ ์ชนิดเคลื่อนที่มีจุดประสงค์ใช้สําหรับการฟอกเลือด ผู้ป่ วยเฉพาะรายที่ข้างเตียงผู้ป่ วย เนื่องจากเป็ นผู้ป่ วยที่มีสภาวะร ่างกายไม่คงที่ มีความเสี่ยงในการเคลื่อนย้ายและมีความเสี่ยงต่อการฟอกเลือด นํ้ าบริสุทธิ ์ที่ใช้ จึงต้องมีคุณภาพตามมาตรฐาน แนวทางปฏิบัติสมาคมโรคไต 2564 เครื่องผลิตนํ้าบริส ุทธิ์ สําหรับการฟอกเลือดด้วยเครื่องไตเทียม ชนิดเคลื่อนที่ได้ (Portable RO)


เครื่องผลิตนํ้าบริส ุทธิ์ สําหรับการฟอกเลือดด้วยเครื่องไตเทียม ชนิดเคลื่อนที่ได้ (Portable RO) Conditions • Mobility • Compact size • Equipments -as less as possible • Intermittently or non-routinely used Requirements • Produce standard quality of water & enough amount for HD • Need special design & maintenance • ใช้สําหรับการฟอกเลือด ผู้ป่ วยเฉพาะรายที่ข้างเตียง ผ้ป่ วย ู • ผ้ป่ วยมีสภาวะร่างกายไูม่ คงที่ มีความเสี่ยงในการ เคลื่อนย้าย • ผ้ป่ วยมีความเสี่ยงตู่อการ ฟอกเลือด


พater Treatment System for Acute Dialysis Setting (Portable RO) Use good quality of raw water Pre-treatment - Double carbon bed Prevent bacterial growth Use frequently –run daily if possible Adequate disinfection Monitor fluid quality Early equipment replacement –short lifetime HD machine - endotoxin filter - disinfection before use


9.1 คุณภาพนํ ้ าดิบที่นํามาใช้เตรียมนํ้ าบริสุทธิ ์ชนิดเคลื่อนที่ได้ ควรมีคุณภาพเทียบเท่า มาตรฐานนํ ้ าดื่ม 9.2 ระบบกรองนํ ้ าเบื ้องต้น (Pre-treatment) 9.2.2ควรมีชุดลดความกระด้าง (softener) เพื่อป้องกนไม ั ่ให้เกิดตะกรันอุดตันในแผนกรอง ่ RO อาจพิจารณาไม่ใช้ชุดลดความกระด้างได้ถ้านํ้ าดิบมีค่า hardness น้อยกวา ่ 10 grain แต่ ต้องมีวิธีการบํารุงรักษาชุด RO หรือปรับวงรอบการเปลี่ยนแผนกรอง ่RO อยา่งเหมาะสม แนวทางปฏิบัติสมาคมโรคไต 2564 เครื่องผลิตนํ้าบริส ุทธิ์ สําหรับการฟอกเลือดด้วยเครื่องไตเทียม ชนิดเคลื่อนที่ได้ (Portable RO)


9.2.3 ต้องมีชุดกรองคาร์บอน (carbon filter) เพื่อดักจับสารคลอรีนไม่ให้หลุด ลอดเข้าไปสัมผัสกบแผันกรอง ่RO ชุดกรองคาร์บอนควรมีการจัดวางไว้ 2 ชุด แบบอนุกรม โดยใช้ผงคาร์บอนชนิด granular activated carbon มีค่า EBCT รวมกนอยั างน้อย ่10 นาที อาจใช้ dense carbon block ทดแทน granular activated carbon ได้ ก่อนการฟอกเลือดให้กบผู้ป่ วยต้องมีการตรวจปริมาณคลอรีนใน ั นํ้ า ตัวอยางจากจุดเก่ บตัวอย ็ างระหว่ างคาร์บอน ่2 ชุดนี้ กรณีไม่มีจุดเกบตัวอย ็ างให้ ่ เกบตัวอย ็ างจากนํ ่ ้ าบริสุทธิ์ที่ผลิตได้แทน ค่าคลอรีนที่ได้ต้องน้อยกวา ่ 0.1 mg/L ถ้ามีการใช้งานนานกวา ่ 8 ชัวโมง ควรมีการตรวจคลอรีนซํ่ ้ า แนวทางปฏิบัติสมาคมโรคไต 2564 เครื่องผลิตนํ้าบริส ุทธิ์ สําหรับการฟอกเลือดด้วยเครื่องไตเทียม ชนิดเคลื่อนที่ได้ (Portable RO)


ISO 23500-1/2019 Annex G: Special considerations for acute haemodialysis


9.4 เครื่องไตเทียมที่นํามาต่อใช้เพื่อการฟอกเลือด ต้องมี Endotoxin retentive filter ใช้กรองนํ้ ายาไตเทียมก่ อนผานเข้าตัวกรองเลือด และควรอบฆ ่ ่ าเชื ้อเครื่อง ไตเทียมก่ อนฟอกเลือดให้กบผู้ป่ วยทุกครั ั้ ง 9.5.2 ส ่งตัวอยางนํ ่ ้ าบริสุทธิ ์เพาะเชื ้ อทุก 1 เดือน และตรวจหาปริมาณ endotoxin ทุก 3 เดือน แนวทางปฏิบัติสมาคมโรคไต 2564 เครื่องผลิตนํ้าบริส ุทธิ์ สําหรับการฟอกเลือดด้วยเครื่องไตเทียม ชนิดเคลื่อนที่ได้ (Portable RO)


9.5.4 ต้องมีการอบฆ่าเชื ้ อในระบบผลิตนํ้ าบริสุทธิ ์ชนิดเคลื่อนที่ได้ตามที่บริษัทผู้ผลิตกาหนด ํ แต่ต้องอยางน้อยทุก ่1 เดือน และพิจารณาทําก่อนกาหนดถ้าไม ํ ่มีการใช้งานติดต่อกนหลายั วัน หรือเมื่อผลการตรวจสอบคุณภาพนํ้ าบริสุทธิ ์ไม่ได้ตามเกณฑ์มาตรฐาน การอบฆ่าเชื ้อให้ รวมถึงสายส่งนํ ้ าบริสุทธิ ์เข้าเครื่องไตเทียมด้วย 9.5.5ควรเปิ ดให้เครื่องผลิตนํ้ าบริสุทธิ ์ชนิดเคลื่อนที่ได้ทํางานทุกวัน อยางน้อยวันละ ่ 15 นาที เพื่อลดโอกาสการก่อตัวของแบคทีเรีย แนวทางปฏิบัติสมาคมโรคไต 2564 เครื่องผลิตนํ้าบริส ุทธิ์ สําหรับการฟอกเลือดด้วยเครื่องไตเทียม ชนิดเคลื่อนที่ได้ (Portable RO)


Outline • Standard Quality of dialysis water & dialysis fluid • Monitoring & maintenance • Actions • Portable water-purification system for HD • In-center dialysis concentrate preparation


7. ส่วนประกอบและการบํารุงรักษาระบบผลิตและจ่ายนํ ้ ายาไตเทียมเข้มข้นใช้เอง ภายในหน่วยไตเทียม 7.1 อุปกรณ์และเครื่องมือทุกชนิดที่ใช้ในกระบวนการผลิต และการจัดเกบนํ ็ ้ ายา ไตเทียมเข้มข้น ต้องทําจากวัสดุที่ไม่ทําปฏิกิริยาเคมี หรือกายภาพกบสารเคมีในนํ ั้ ายาไต เทียมเข้มข้น 7.2 ถังผสมนํ ้ ายาไตเทียมเข้มข้นต้องมีลักษณะเหมาะสม มีการกรองอากาศภายใน ถัง มีใบพัดหมุนวนตลอดเวลาระหวางการผสมนํ ่ ้ ายา และ ต้องได้รับการทําความสะอาด หลังใช้งานทุกครั้ ง 7.3 ใช้วัตถุดิบที่มีฉลากรับรองคุณภาพ ตามเภสัชตํารับ หรือบริษัทผ้ผลิต ในการู ผลิตนํ้ายาไตเทียมเข้มข้น เกณฑ์มาตรฐาน ตรต. องค์ประกอบที่7 ระบบผลิตนําบริสุทธิ ้ และนํ ์ ายาไตเทียมเข้มข้ ้ นมาตรฐาน


7. ส่วนประกอบและการบํารุงรักษาระบบผลิตและจ่ายนํ ้ ายาไตเทียมเข้มข้นใช้เอง ภายในหน่วยไตเทียม 7.4 กรณีที่มีการผลิตนํ้ ายาไตเทียมเข้มข้นเกบในภาชนะไว้ก ็ ่อนนํามาใช้ ต้องบรรจุ ในภาชนะสะอาด ไม่มีสารเคมีตกค้าง ภาชนะปิ ดสนิท และมีฉลากระบุ ส่วนผสมของ นํ้ายา และวันที่ผลิตอย่างชัดเจน สําหรับภาชนะที่จะใช้บรรจุนํ้ ายาไตเทียมไบ คาร์บอเนตที่นํากลับมาใช้ซํ้ า ต้องได้รับการทําความสะอาด และผานกระบ่วนการฆ่าเชื ้ อ ก่อนบรรจุนํ ้ ายาไตเทียมไบคาร์บอเนตใหม่ทุกครั ้ ง 7.5 ต้องทําการอบฆ่าเชื้อโรคระบบท่อจ่ายนํ้ายาไตเทียมเข้มข้นชนิดไบคาร์บอเนต ไปยังเครื่องไตเทียมทุ กสัปดาห์ 7.6 ต้องมีการส่มตรวจระดับความเข้มข้นของเกลือแร่ และกลุ ู โคสในนํ้ายาไตเทียม ทุ ก 3 เดือน หรือเมื่อเปลี่ยนแหล่งที่มาของวัตถุ ดิบ เกณฑ์มาตรฐาน ตรต. องค์ประกอบที่7 ระบบผลิตนําบริสุทธิ ้ และนํ ์ ายาไตเทียมเข้มข้ ้ นมาตรฐาน


- สัญญาการบํารุงรักษาระบบนํ ้ าบริสุทธิ ์และเครื่องไตเทียมที่ยังมีอายุสัญญา - ผลการตรวจคุณภาพนํ ้ าบริสุทธิ ์ (dialysis water) ทางเคมี (AAMI) ของปี ล่าสุด - ผลการตรวจปริมาณแบคทีเรียและ endotoxin ในนํ้าบริสุทธิ์และนํ้ายาไตเทียม - บันทึกการเปลี่ยน endotoxin filters ตามวาระที่บริษัทผ้ผลิตกําหนดู - ตารางการตรวจระบบนํ ้ าบริสุทธิ ์ ประจําวัน - การตรวจปริมาณคลอรีนหลังถังคาร์บอน และในนํ้ าดิบ - คุณภาพระบบ RO - ตารางการบํารุงรักษาระบบนํ ้ าบริสุทธิ ์ - การอบฆ่าเชื ้ อ - การเปลี่ยนอุปกรณ์ตามวาระ - ผลการสุมตรวจค่า electrolytes ใน dialysis fluid _กรณีหน่วยผลิตนํ้ ายาไตเทียม เข้มข้นใช้เอง (ทุก 3 เดือน) เอกสารที่เกี่ยวข้องในการตรวจของตรต.


1. ตารางแสดงแผนการส่งเพาะเชื้อน้ำบริสุทธิ์ และ น้ํายาไตเทียม และ แผนผังเส้นทางเดินน้ำ RO ภายในหน่วยไตเทียม ที่ มีการระบุตำแหน่งของเครื่องไตเทียม และอ่างล้างตัวกรอง ให้ หมายเลขกำกับเครื่องไตเทียมที่ใช้สื่อสารในหน่วยไตเทียม (ตรงกับใบรายงานผลเพาะเชื้อน้ํา บริสุทธิ์และ น้ํายาไตเทียม) **** กรณีที่มีวงจรทางเดินน้ำ มากกว่า 1 loop ให้ระบุหมายเลขของ loop และกรณีที่มีการทํา Hemodiafiltration ให้ระบุตำแหน่งเครื่องที่ใช้ทำ HDF ด้วย 2. ผลเพาะเชื้อน้ําบริสุทธิ์ และ น้ํายาไตเทียม และมีการตรวจวัด endotoxin ตามเกณฑ์ มาตรฐาน ตรต. 5-6 เดือน ย้อนหลัง โดยให้เดือนแรกของผลที่ส่งมา ยังเป็นผลการตรวจที่อยู่ ในช่วงของการรับรองที่กำลังจะหมดอายุนี้ **** การส่งเพาะเชื้อน้ำบริสุทธิ์และ น้ํายาไตเทียม ต้องใช้วิธีการมาตรฐานสำหรับการเพาะเชื้อน้ำ ที่ระบุในแนวทางปฏิบัติเรื่อง การเตรียมน้ำบริสุทธิ์เพื่อการฟอกเลือดด้วยเครื่องไตเทียม พ.ศ. 2564 3. ผลการทดสอบการปนเปื้อนของสารเคมีของปีล่าสุด 4. สัญญาการบํารุงรักษาระบบผลิตน้ําบริสุทธิ์ ที่ยังมีอายุสัญญา 5. สัญญาการบํารุงรักษาเครื่องไตเทียม ที่มีระบุ serial number ของเครื่อง และยังมีอายุสัญญา 6. รายการตรวจสอบประจําวัน ของการทํางานของ RO


microSiemens/cm (µS/cm) แก้คําผิด


Take home messages • In modern era of hemodialysis technology which high-flux & super high-flux dialyzer and on-line producing substitution fluid are used, the quality of water treatment system is crucial and should be validated. • We can not prove the quality of dialysis fluid at real-time used and the sterility of substitution fluid is beyond the lab sensitivity to verify. • We have to use validation process to make sure that the quality of dialysis fluid are within the required range at all time. • Using equipments according to validation process of manufacturer, adequate monitoring & maintenance of water treatment system, rapid corrective actions and revalidation after correction are the principal task to achieve high fluid quality for HD and ensuring patient safety.


Click to View FlipBook Version