The words you are searching are inside this book. To get more targeted content, please make full-text search by clicking here.
Discover the best professional documents and content resources in AnyFlip Document Base.
Search
Published by , 2019-09-06 19:00:12

attachment 1 (1)

attachment 1 (1)

g GE Medical Systems LUNAR
726 Heartland Trail
Product Data Madison, WI 53717
Rev 0 888-795-8627
January 2002 Fax: 608-826-7102

Internet – gemedical.com

LUNAR PRODIGY Series

Site Specification Guide

PRODIGY PRODIGY
263cm 201cm
103.5" 79.5
8'-0"
8'-0"

DD

9'-0" 7'-6"

Full Compact

Site planning is an important measuring 8.0’ x 7.5’ and is indicated Page 1 of 2
consideration in any medical facility when exam space is at a premium.
since space is always at a premium. Room size shown is the suggested
Optimal site planning maximizes the minimum physical space requirement.
performance and use of space without
compromising patient and operator Please note that some states may
comfort. require additional distance between
the computer console and the scan
PRODIGY has been designed to fit table. Check with your local and state
efficiently into any clinical operating agencies for any additional
environment. Shown above are space requirements.
requirements for PRODIGY. The
PRODIGY is shown installed in a
typical 8.0’ x 9.0’ examination room.
The compact system fits in a room

LUNAR Prodigy Series

g GE Medical Systems LUNAR
726 Heartland Trail
Product Data Madison, WI 53717
Rev 0 888-795-8627
January 2002 Fax: 608-826-7102

Internet – gemedical.com

LUNAR PRODIGY Series

Site Components and Specifications

Full-Size Table Environmental Requirements TeleDensitometry and Networking
Specifications • Operating Temperature Range • Access to LAN or dedicated analog

Dimensions and Weight 18-27° C (65-81° F) phone line required.
• Length: 263 cm (103.5 in) • Operating Humidity Range • Connection to non-isolated LAN or
• Width: 109.3cm (43.5 in)
• Height: 128.3 cm (50.5 in) 20-80% non-condensing phone line requires a distance of
• Weight: 272.16 kg (600 lbs) • Power Rating 1.83 M between the operator's
console and the scan table to
Compact-Size Table 600vA comply with IEC, UL and CSA safety
Specifications requirments.
• Heat Output
Dimensions and Weight Scanner 150 BTU/hr (idle) 1500 Additional detailed information
• Length: 201 cm (79.5 in) BTU/hr (energized) host PC 500 is available in the LUNAR
• Width: 109.3 cm (43.5 in) BTU/hr Prodigy Safety Information and
• Height: 128.3 cm (50.5 in) Technical specifications manual
• Weight: 254 kg (559 lbs) Radiation (p/n LNR8933US)

Common Specifications • Scatter and leakage: LASER RADIATION
<0.6 mR/hr (6 µSv/hr) @ 1 m (39 in) DO NOT STARE INTO BEAM
Operator Space POWER OUTPUT: 1.00mW MAX.
• Length: 78.5 cm (31 in) • The estimated total annual exposure WAVELENGTH: 670nm
• Width: 63.3 cm (25 in) to the operator is calculated as CLASS II LASER PRODUCT
• Height: 48.1 cm (19 in) follows: This equipment complies with
21 CFR 1040.10 and 1040.11
Electrical Power 720 studies/yr x 1 min/Study x 1 hr/
• Dedicated 20A (isolated ground) 60 min. x 0.6mR/hr=7.2 mR/yr

120 VAC single duplex outlet • Assumptions:
(placed at the operator's console. - User is 1 meter from table center
See room layout drawings) - 60 patients /month
• Power must meet - 1 study = 1 spine and 1 femur scan
IEEE 519-1992 standards:
110/115 VAC +/-10%, THD <5%* GE Medical Systems LUNAR
recommends consulting with your
GE Lunar recommends consulting Stae Radiation regulatory agency to
a licensed electrician to insure that insure that you comply with local
your site meets the system power ordinances.
requirements.
Computer Workstation**

• computer
• high-resolution monitor
• keyboard
• HP Professional printer
• mouse

LUNAR and PRODIGY are trademarks of GE Lunar Corporation.
Windows is a trademark of Microsoft Corporation.

* Some sites may require power conditioning equipment. Contact GE LUNAR for details and specifications.

** The computer workstation is normally connected to the PRODIGY isolated power output connection. If non-isolated devices are connected, i.e.
telephone or modem lines, the computer must be plugged into a normal 110 VAC wall connection and a minimum separation distance of 1.83m
(6 ft) is required between the scanner and the computer to comply with IEC, UL, and CSA requirements.

*** Not available on compact tables.

Specifications may change to reflect changes in technology.

SL119E 1/02

LUNAR Prodigy Series Page 2 of 2


Click to View FlipBook Version